SPECIFICATIONS

Document Sample
SPECIFICATIONS Powered By Docstoc
					SPECIFICATIONS:

1. Incinerator:

   1. Should be electrically operated and specifically made for biomedical wastes
   2. Design should be such that it can be installed indoor with a room space of 112 sq ft at
       7th floor
   3. Should meet the CPCB emission standard
   4. Should consist of dual chamber for efficient burning
   5. Should have a separate ash collection chamber
   6. Capacity should be at least 10kg per hour
   7. Should be provided with emission chimney, height should be as per CPCB norms
   8. Should be provided with dilution process system
   9. Should be provided with foul odor treatment system
   10. Should be provided with gravity settler
   11. Insulation should be with glass wool
   12. Burning chambers should be lined with fire bricks
   13. Loading should preferably be from top
   14. Should be provided with ash collection tray
   15. The burning chamber should have some tilting device for easy dropping of ash to the
       ash collection chamber
   16. Should be provided with a control panel consisting of solid state digital temperature
       controller cum indicator
   17. The supplier should take the responsibility to procure NOC from the WBPCB for
       installation of the Incinerator.
   18. Power: 220-240VAC, 50-60Hz.
   19. Accessories needed to run the instrument smoothly.
   20. Five years maintenance including the guaranty period.
   21. Kolkata based after sale service with proven track record.




                                              1
2. Horizontal Sterilizer:

1. Should be rectangular, horizontal, double chambered
2. Should be robust, rugged
3. The inner chamber should be made of stainless steel
4. The outer chamber or jacket should be made of rolled steel
5. Should have a single door made of rolled steel and lined with SS on both sides
6. The door should have self locking device
7. The door should have radial locking system
8. The door should be fitted with heat resistant rubber gasket
9. The outer jacket preferably be lined with asbestos or glass wool on the outer side to
   prevent heat loss
10. The boiler should be made of thick stainless steel and should be located just under the
   sterilizer
11. Heating coils should be fitted in both the sides of the boiler for uniform heating
12. Boiler should be provided with water gauge glass, pressure gauge and safety valve, water
   inlet and outlet valves, dial thermometer
13. Boiler should be provided with „low water level cut off‟ device
14. The sterilizer should have multi-port control valve for setting it in different mode e.g.,
   sterilizing or fast exhaust mode
15. Should be provided with quick exhaust valve, ejector valve, vacuum drier, steam trap,
   steam exhaust line, screen for draining inner chamber, vacuum breaker for the jacket
16. Should be provided with pressure gauges for inner chamber and jacket
17. Should have a composite switch box comprising of pressure control switch, main switch,
   pilot lamps, relays, etc, necessary for operation of the sterilizer
18. Should be provided with Trolley and SS loading carriage
19. Should have a robust stand for holding the whole unit
20. Should be electrically operated (total load 18kw)
21. Should be provided with SS trays for loading materials as accessory
21. Power:      220-240VAC, 50-60Hz.
22. Accessories needed to run the instrument smoothly.




                                               2
23. Five years maintenance including the guaranty period.

24. Kolkata based after sale service with proven track record.



3. Low Speed Cold Centrifuge:

   1. Microprocessor controlled table top refrigerated centrifuge with adjustable
       speed from 300 to 15,000 rpm and max. RCF Value 24,500g or higher.
   2. CFC- free refrigerant. Digital display (LCD) for speed, time, sample/chamber
       temp. , RCF.
   3. Maintenance -free frequency -controlled induction drive. Acceleration &
       deceleration profile: 9 / 10.
   4. Temp. range : – 10 deg. C to + 40 deg. C.
   5. Pre cooling function to achieve 4 deg. C should be available with direct button.
   6. Max timer range: 9 h , 99 min + continuous
   7. Drive system: Direct, brushless induction drive.
   8. Imbalance Detection System: Smart Spin technology for advanced rotor
       management system to identify imbalance during run due to liquid spillage or
       tube breakage.
   9. It should have Auto Lock rotor system so that it can be installed and removed
       without any tools in less than 5 Sec. With just a push of a button for quick and
       easy change of rotors.
   10. Dual timing mode for at start and at speed.
   11. Dual temperature mode for Chamber temperature and Sample temperature.
   12. Enhanced security with password protection of programs and lid openings.
   13. It should be certified for quality & safety by International Certification Agency,
       like UL, CSA certified, CE marked, IVD compliant etc.
   14. It should have an option for automatic lid opening at the end of the run.
   15. It should have a minimum of 5 direct recall program keys and capability for up to
       99 programs.
   16. The angle rotors must be manufactured from a highly corrosion and fatigue
       resistant Carbon Fiber material. The fixed angle carbon fiber rotors should be
       warranted for at least 12 to 15 years.
   17. The buckets and rotors sealing lids must be certificate for bio-containment by a
       3rd party Lab of worldwide recognition.
   18. Power: Suitable for 230V, single phase, AC, 50 Hz operation.
   19. Accessories: a) 4 x 400 ml swing out rotor with max. speed 5,000 rpm at 4 deg.
       C with 400 ml autoclavable round bucket with 15 ml & 50 ml round and conical
       tube adaptors to be quoted.
       b ) 6 x 100 ml Fiberlite Fixed angle rotor with 24,500 x g or higher at 4 deg.
       C with 50 ml and 15 ml round and conical tube adaptors to be quoted.
        c) 30 x 2 ml Microliter sealed aluminium fixed angle rotor with 25,800 x g with
       adaptors for 0.2 ml to 0.5 ml .
        d) Swing out Microtitre plate rotor with 4,000 rpm. or higher should be quoted.



                                            3
   20. Suitable servo voltage stabilizer with high voltage low voltage cut off
      circuit, auto reset with delay timer and spike eliminator etc. also should be
      quoted.
   21. User list & some performance certificates from users in & around Kolkata
       should be provided.
   22. Prompt & efficient after-sales service should be available from Kolkata.
   23. Five years maintenance including the guaranty period.


4. Low Temperature Immersion Probe:

   1.   Temperature range ambient to –78 to1000 C.
   2.   Operable at 220V/50Hz.
   3.   Digital read out of set and actual temperature.
   4.   Quick achieve of set temperature.
   5.   Standard flexible probe.
   6.   Self luming electric control system.
   7.   Fully insulated bath with support rod boss head of suitable size.
   8.   Power: 220-240VAC, 50-60Hz.
   9. The system should be provided with UPS to run the instrument smoothly.
   10. Accessories needed to run the instrument smoothly.
   11. Five years maintenance including the guaranty period.
   12. Kolkata based after sale service with proven track record.


5. High Speed Cold Centrifuge:

   1. Microprocessor controlled table top refrigerated centrifuge with adjustable speed and
        with quick run facility.
   2. Temp. range from -90C to +400C,
   3. CFC-free refrigerant.
   4. Digital display (LED) for speed, time, temp., RCF, acceleration & deceleration
        profiles & program.
   5. Maintenance- free induction drive Imbalance detection: Electronic imbalance
        detection to detect imbalance, even if it occur due to tube breakage or liquid spillage
        during the run.
   6. It should have tested in accordance with international standard for safety and
        performance, and document should be provided.
   7. Maximum noise level should be less than 55 db(A).



                                             4
   8. Power: 220- 230VAc, single phase, 50-60 Hz operation.
   9. Accessories:
       (i) 6x250 ml Fixed angle rotor with maximum RCF higher than 18,500 x g with 50 ml
       & 15 ml conical tube adaptors.
       (ii) 4 x 750 ml swing-out rotor with max. RCF higher than 4,500 x g with 750 ml
       round bucket, cap & 50 ml conical & round tube adaptor to be quoted. Also, adaptor
       for tissue culture flask for 75 ml & 25 ml to be quoted.
       (iii) 12 - 24 x 1.5/2 ml rotor with max. speed 30,000 rpm & max. RCF 60,000 x g to
       be quoted.
       (iv) 6 - 8 x 50 ml Fixed angle rotor with RCF higher than 40,000 x g.
Optionally to be quoted:
(i) Microplate Rotor with max . RCF higher than 5, 000 x g.
(ii) 4 x 250 ml swinging bucket rotor with max. RCF higher than 6,950 x g.
(iii)14 x 50 ml Fixed angle rotor with max. RCF higher than 17,000 x g.
10. Stabilizer: Suitable capacity servo voltage stabilizer with high voltage, low voltage cut-
off, auto reset with delay timer & spike eliminator should be quoted.
12. Accessories needed to run the instrument smoothly.
11. Five years maintenance including the guaranty period.
14. Kolkata based after sale service with proven track record.

6. CO2 Incubator

1. One chamber CO2 Incubator with work chamber volume approx. 150 liters.
2. Air jacketed heating system (not direct heating system) with temp. Control by
    microprocessor.
3. Temp. Range from ambient +3 o C to + 55 o C
4. Temp. Deviation with respect to time is + 0.1K.
5. CO2 control by Microprocessor based & by heat resistant Thermal Conductivity
   Detector to withstand heat of decontamination.
6. CO2 range 0 to 20% or better and CO2 accuracy + 0.1% by volume.
7. System should have Interactive Control with Touch screen for rapid access of
    important information for each critical parameter in the Incubator.
8. Fast Recovery time for Temp., CO2 and Humidity should be available for good culture
    growth. Temp. recovery time at 37C : less than 10 min. CO2 recovery time for 5%
    CO2 : less than 5 min. Humidity recovery at 95% rH ( Humidity loss during door
    opening 30 seconds ): not more than 20 minutes.




                                             5
9. System should have built-in automatic decontamination facility to remove bacteria,
    fungi, spores, mycoplasma etc. without removing sensor, fan or any other fitting. The
    system should employ automatic decontamination routine at moist heat of 90 o C
    or dry heat of 180 deg C.
10. Validation certificate of the decontamination routine from any accredited agency such
    as FDA, CAMR etc should be enclosed.
11. At least 3 nos. stainless steel perforated shelves should be supplied.
12. It should be supplied with access port to allow any cable, plug or tubing to be
    easily inserted into or out of the chamber.
13 Interior chamber should be made of stainless steel with electro polish finish to
    have highest quality of inner surface with rounded corners on all sides for easy
    cleaning. The shelves and fan impeller also should be made of stainless steel and
    should not have nuts or bolts for shelf supports to reduce the scope of growth of
    contamination. No plastic should be inside to avoid VOC.
14. Built-in audible and visual water level alarm should be available when the water
    reservoir needs to be refilled to ensure a constant high level of humidity and to
    prevent cultures from drying out.
15. Independent electronic over temp. Protection with separate sensor should be available
    for sample protection.
16. The system should have large digital display for both temp. & CO2 simultaneously.
17. System should have one main door and one glass doors .
18. Suitable for 230V, single phase 50Hz operation.
19. Suitable servo voltage stabilizer with high voltage low voltage cut off circuit, auto
    reset with delay timer and spike eliminator etc. & Double stage CO2 pressure
    regulator with stainless steel diaphragm & 18Kg CO2 Cylinder filled with CO2 gas
    99.5% or better purity required for operation of CO2 Incubator should also be
    quoted.
20. Warranty 2 years.
21. Optional item: Oxygen control accessory with 3 nos. Interior gas tight split doors
    should be quoted.
22. User list & some performance certificates from users in & around Kolkata should be
    provided prompt & efficient after-sales service should be available from Kolkata.
23. Accessories needed to run the instrument smoothly.
24. Five years maintenance including the guaranty period.


7. Electronic Balance:

1.     Microprocessor based with full tare facilities.
2.     Weighing Range:      0.0 to 40gm/120gm
3.     Readability:         0.01mg/0.1mg
4.     Repeatability:       0.02mg (10gm)/0.1mg (100gm)
5.     Linearity:           0.2mg
6.     Weighing Pan:        Stainless Steel
7.     Eccentric load Dev. At ½ max. capacity:       0.3mg (50gms)
8.     Fully automatic temperature controlled internal calibration



                                           6
9.       Backlitt Graphic Display with touch screen operation.
10.      Tare Range:           0.0 to 40gm/120gm
11.      Power: 220-230VAc, 50-60Hz.
12.      The Balance should be supplied with following features:
         (i)    Feed through weighing below the balance.
         (ii)   Display various units of weight.
         (iii) Facility for determination of densities of solid, liquid and pasty substances
13.      ISO 9001/ ISO 14001 Certificate should be provided for quality and
         environmental standard.
14.      Accessories needed to run the instrument smoothly.
15.      The system should be provided with UPS to run the instrument smoothly.
16.      Five years maintenance including the guaranty period.
17.      Kolkata based after sale service with proven track record.


8. Inverted Microscope:

      1. Inverted Phase contrast Microscope with Infinity & Harmonic Corrected (HC)
          Optics
      2. Transmitted light stand with Long Life LED illumination with service life of
          approx. 50,000 hrs. and uniform color temperature. Should have integrated auto-
          shut off facility.
      3. Magnification from 50X to 400X or higher.
      4. Objectives 5x, 10x Phase, Long distance 20x Phase and Long distance 40x Phase
          contrast corrected objective with numerical aperture 0.7 or higher. All the
          objectives should be working without any oil.
      5. Fixed stage coaxial double knob for coarse and fine focus on both sides, a 4-fold
          revolving nosepiece with object guide to accept all the different inserts
      6. The coaxial control drive should be ergonomic, low-lying & extremely accurate
          and sensitive
      7. Binocular Phototube with fixed photo tube, with 30° viewing angle, with inter
          pupillary adjustment, with constant focus and beam splitter positions vis/phot:
          50/50%, fixed
      8. Condenser with a free working distance of 40 mm and a numerical aperture of
          0.45 for Bright field & Phase Contrast applications
      9. Should have facility for sample observation through eyepiece, operating and
          viewing the camera display and macro positioning of the stage simultaneously, i.e
          without changing the sitting positions
      10. Scientific color digital camera for on-screen microscope image display,
      11. PC, Printer & 1.0 KVA UPS to be quoted along with the system
      12. Kolkata based after sales service with proven track record

      Optional items to be quoted separately:




                                              7
      1. Fluorescence attachment with 50W Mercury with filter system for Blue
         excitation (BP450/490), green excitation (BP515-560) & UV excitation
         (BP340-380) with spare mercury lamp
      2. Scientific Color Cool Digital camera with resolution atleast 2592 x 1944
         pixels, 5 Mpixel, 1/2" CCD sensor, pixel size 2.78 um x 2.78 um, colour filter
         RGB (Bayer), color depth 36 bits, exposure time 1 msec- 600 sec. with c-
         Mount


9. Bright Field Microscope with Photographic System

   1. Microscope should have Infinity & Harmonic Corrected (HC) Optical system
   2. Microscope Stand should be with Long Life LED Illumination, full white light, no
       blue filter to be required LED illumination of atleast 6000K temp. & 100,000 hrs.
       at full intensity - providing cool white light with a lifetime of atleast 50 years for
       cost saving
   3. Fixed 5-fold revolving nosepiece for bright field objectives (PLAN Achromat
       Objectives 10x/0.25, 20x/0.40, 40x/0.65, 100x/1.25 Oil with wide-field eyepiece
       pair 10x/20), with stage carrier and centrable condenser holder, right- and left-
       hand operation (preferable) and clamp screw, stabilized power supply 90- 250V,
       commutable 50/60 Hz; with Koehler kit with variable field diaphragm.
   4. Simultaneous focus and stage control facility
   5. Stand with height adjustment focus knobs with coarse and fine focus
   6. Right hand mechanical stage Stand with Koehler Illumination, with non-
       extending rack and slide holder. ErgoStage for right- or left-hand operation.
   7. Binocular Phototube with 30° viewing angle, with constant focus and beam
       splitter positions vis/phot: 50/50%, fixed.
   8. Color coded Abbe Condenser
   9. Scientific color digital camera for on-screen microscope image display,
   10. PC, Printer & 1.0 KVA UPS to be quoted along with the system
   11. Accessories needed to run the instrument smoothly.
   12. Kolkata based after sales service with proven track record


10. Liquid Nitrogen Tank:

          1. Wide mouth ultra cold storage for biological specimens in Liquid
             Nitrogen.
          2. Advanced insulating system with very low evaporation rate to provide a
             maximum –1900 C throughout the flask with as little as 2 inches of Liquid
             Nitrogen.
          3. Secured locking hasp to prevent unauthorized access.
          4. Stain steel racks to accommodate freezing vials.
          5. Digital display of temperature.
          6. Alarm system to aware the level of Liquid Nitrogen.
          7. Storage capacity: a) 50L flask, b) 100L flask



                                             8
          8. Four wheeled roller base.
          9. Power: 220-230VAc, 50-60Hz.
          10. Accessories needed to run the instrument smoothly.
          11. Five years maintenance including the guaranty period.
          12. Kolkata based after sale service with proven track record.


11. High Voltage Power Pac:

A. Sequencing Power Pac:
1.    Programmable high voltage power supply for 5000V/500mA/400W with constant
      volt/ current/ Power operating mode with automatic crossover.
2.    Built in timer range: 0 – 999min.
3.    Four output terminal.
4.    Auto power recovery feature after power failure is required.
5.    The system should have “No Load detection” capability.
6.    Power: 220-240VAC, 50-60Hz.
7.    Accessories needed to run the instrument smoothly.
8.    Kolkata based after sale service with proven track record.

B. Universal Power Pac:
1.     Programmable power supply for 500V/2.5A/500W with constant volt/ current/
       Power operating mode with automatic crossover.
2.     Built in timer range: 0 – 999min.
3.     Four output terminal.
4.     Auto power recovery feature after power failure is required.
5.     The system should have “No Load detection” capability.
6.     Power: 220-240VAC, 50-60Hz.
7.     Accessories needed to run the instrument smoothly.
8.     Kolkata based after sale service with proven track record.

C. High Current Power Pac:
1.    Programmable power supply for 250V/3.0A/300W with constant volt/ current/
      Power operating mode with automatic crossover.
2.    Built in timer range: 0 – 999min.
3.    Four output terminal.
4.    Auto power recovery feature after power failure is required.
5.    The system should have “No Load detection” capability.
6.    Power: 220-240VAC, 50-60Hz.
7.    Accessories needed to run the instrument smoothly.
8.    Kolkata based after sale service with proven track record.




                                           9
12. Microtome:

1.     Low maintenance & backlash free micrometer feed system with cross roller
           bearings for vertical stroke.
2.     Precision cylinder guide system for horizontal precision movement.
3.     Selection thickness setting from 0.5 to 60µm,
                                      from 0 to 2µm in 0.5µm steps,
                                              2 to 10µm in steps of 1µm,
                                              10 to 20µm in steps of 2µm,
                                              20 to 60µm in steps of 5µm.
4.     Specimen retraction: 220µm
5.     Mechanical trimming function at two steps (10 & 50µm)
6.     Vertical specimen stroke length:       60mm (+/- 1 mm).
7.     Horizontal specimen feed: Approx. 28mm.
8.     Hand wheel lockable in any position using two independent locking systems.
9.     Specimen orientation facility: 8 deg in X, Y, Z direction.
10.    Disposable blade holder system with lateral displacement facility.
11.    Knife 16cm profile-C with knife holder.
12.    Accessories needed to run the instrument smoothly.
13.    Kolkata based after sale service with proven track record.
14.    Five years maintenance including the guaranty period.


13. Freeze Drier:

       1.  Digital display of condenser temperature.
       2.  Ultimate temperature of condenser (Trap_ -1100C.
       3.  Condenser capacity 2.0Kg/24hrs.
       4.  Stainless steel condenser AISI 316:4 litres.
       5.  Built in start delay for the compressor (Step starting).
       6.  Microprocessor conrolled with alarms.
       7.  Acrylic base for use on top on condenser of Freeze Dryer. It should includes
           connection for vacuum pump & vacuum release. It should have provision for
           connecting acrylic chamber with lid & with 8 ports fitted with rubber valves
           for connecting RB flask of different sizes.
       8. Manifold tree with 4 ports with cocks for freeze drying.
       9. Adaptor for various sizes RB flask (50 to 1 liter).
       10. Ampoule manifold with 16 ports for connecting glass ampoules with sealing
           torch.
       11. Direct driven high vacuum pump – with anti suck back device. Capacity 30
           lit/min, ultimate vacuum 0.001 mBar with connecting nozzle & hose.
       12. Provision for connecting Vacuum Centrifuge with 120x1.5/2ml rotor (to be
           quoted optionally).
       13. Preservation of temperature sensitive products.
       14. Freezing, drying, vacuum influence for freeze-drying or lyophilization cycle
           of different volume of liquid.



                                          10
        15. Attachment for drying of samples of different volumes under full vacuum or
            partial pressure.
        16. Facility for monitoring of temperature and vacuum.
        17. CFC free.
        18. User friendly.
        19. Power:      220-240VAC, 50-60Hz.
        20. Suitable capacity servo voltage stabilizer with high voltage, low voltage cut-
            off, auto reset with delay timer & spike eliminator should be quoted.
        21. Accessories needed to run the instrument smoothly.
        22. Kolkata based after sale service with proven track record.
        23. Five years maintenance including the guaranty period.


14. Ultrasonic Cell Disrupter:

   1.  Standard probes ½″ titanium disrupter probe for volumes 10 to 250ml included.
   2.  Heavy duty generator.
   3.  Microprocessor controlled.
   4.  Auto tuning.
   5.  Temperature monitoring and control.
   6.  Compact size- Lightweight, portable.
   7.  Probe tip.
   8.  Cup horn /Probes- 3/16″ for (2 to 25ml)
                           3/32″ for (0.2to0.50ml)
                          1/2″ for (10 to 250 ml)
                          1/2″ replacement probe tip
                          3/4″ for (25 to 500ml)
                          3/4″ for (50 to1000ml)
   9. Sound enclosure,
   10. Rosette cooling cell and micro cup horn.
   11. Digital wattmeter and LCD digital display.
   12. Quick disconnect cables.
   13. Load monitoring thumb activated pulsing button.
   14. Power: 220-240VAC, 50-60Hz.
   15. Accessories needed to run the instrument smoothly.
   16. Kolkata based after sale service with proven track record.
   17. Five years maintenance includes the guaranty period.


15. Environmental Shaker:

   1.      Size: a) Internal: 890 X 630 X 420 mm, b) External: 1200 X 750 X 1000 mm
           c) Platform: 750 X 480 mm
   2.      Temperature: 50C to 450 C
   3.      CFC free refrigeration
   4.      Temperature uniformity: ± 10 C, Temperature accuracy : ± 0.20 C



                                            11
          Timer: Wait on time, Wait off time (Max. 99hr Min. 1hr)
   5.     Programmable Digital display with membrane Touch type key pad
   6.     Run time upto 99 hr
   7.     Shaking: a) Motion: Orbital, b) Frequency: 10 - 250 r.p.m., c) Timer:
          Programmable.
   8.     Driving System: Brushless D.C. Motor with feed back control.
   9.     Material: a) Internal & External: Stainless steel, b) Platform: Anodized
          Aluminium plate, c) Lid: With viewing glass window and supported by Gas
          Spring, d) Insulation: Non-flammable polystyrene
   10.    Custom logical safe control system
   11.    Hydraulic over temperature limit.
   12.    The instrument should have following provisions: a) Univeral platform, b)
          clamp for 50 ml flask, c) clamp for 100 ml flask, d) clamp for 250 ml flask, e)
          clamp for 500 ml flask, f) clamp for 1000 ml flask, g) clamp for 2000 ml
          flask.
   13.    Power:      220-240VAC, 50-60Hz.
   14.    Accessories needed to run the instrument smoothly.
   15.    The system should be provided with UPS to run the instrument smoothly.
   16.    Kolkata based after sale service with proven track record.
   17.    Five years maintenance includes the guaranty period.


16. Photocopier:

   1. Digital Photocopier cum printer model – 1 Tray (250 Sheets) & Bypass Tray (50
   Sheets).
   2. Speed:- 20 pages per minute B/W (A4 Size)
   3. Duty cycle:- Per month up to 20,000 copies
   4. RAM :- 64 MB RAM
   5. Resolution:- Printing & Copy 600 x 600
   6. Paper Size:- A3, A4, B4, B5.
   7. Connectivity:- USB 2.0
   8. Print Drivers:- Windows 2000/XP/2003 Server, Vista, Linux.
   9. Basic Copy Features:- Auto Reduction & Enlargement, N-Up, ID
   10. Card Copy, Auto Fit.
   11. Media handling:- 60 to 110 gsm.
   12. Media Type:- Plain paper, Labels, Transparencies.
   13. Accessories needed to run the instrument smoothly.
   14. Power: 220-240VAC, 50-60Hz.
   15. Stabilizer:- 1.5 KVA.
   16. Kolkata based after sale service with proven track record.
   17. Five years maintenance includes the guaranty period.


17. Mini Electrophoresis Set:




                                          12
1. Mini protein electrophoresis cell for running hand cast or pre-cast PAGE and SDS-PAGE
   gels.
2. Mini Trans blot electrophoresis transfer module for western blotting.
3. Electro-eluter elution of sample bands or entire lanes for protein recovery.
4. Accessories needed to run the instrument smoothly.
5. Kolkata based after sale service with proven track record.
6. Five years maintenance includes the guaranty period.




                                     13

				