					         Equipment Specifications for Gynecological Operation Theatre Table


         UNSPSC Code:
         ECRI Code:
1 Description of Function

  Sl   Name                                                          Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder
 1.1 Operating tables provide an elevated surface that
     supports the patient’s body during surgical procedures,
     stabilizing the patient’s position and providing optimal
     exposure of the surgical field.
2 Operational Requirements

  Sl   Name                                                          Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder
 2.1 High quality suitable for gynecological surgery Width
     3.5 feet, length 6- 6.5 ft with stirrups for the support of
     legs and breakable on all table positioning Mattress
     should be x-ray translucent. Should have all the
     accessories Electro- Hydraulic O.T. Table with
     central controls.
3 Technical Specifications

  Sl   Name                                                          Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder
 3.1 Operation Table should be electrically operated with
     facilities for remote control along with manual
     backup and have the following specifications.

       1• O.T. table with four section table top with divided
       foot section.
       2• All table positioning i.e. Height, Back section, lateral
       tilt, trendelenburg and Reverse - trendelenburg.
       Upward Movement should be achieved by pressing
       down on the foot pedal and reverse movement by lifting
       the pedal as well as by remote control.
       3• The manual position selector must be located on the
       head end of the table and should have the provision to
       shift the position selector to the foot section also
       depending on the surgery.
       4• Gas springs should facilitate the easy adjustment of
       the head and foot section.
       5• The casing on the frame and centre supporting
       column should be made up of Hygienic stainless steel
       of high grade rust proof quality.(epoxy coated steel)
       6.Height 700 -1050 mm, trendelenburg/reverse
       trendelenburg = -25 to +25 degree, back section = -15
       to + 70 degree, side tilt = +/- 15degree, leg section = -
       90 to 0 degree, gas spring assisted and detachable. head
       rest = -45 to + 25 degree, manually, gas spring assisted,
       detachable. Facility perineal unit.
       7.Mattress of antistatic PU integral
       8. Remote control and additional controls on table
       base for emergency.
       9. DC power back should be inbuilt.
       10. Maneuverability with basal frame with multiple
       mounting facility.


4 System Configuration Accessories, spares and consumables

  Sl   Name                                                        Technical Bidders
                                                                   Specs     Deviation
                                                                   quoted    if any
                                                                   by bidder
 4.1 Should be supplied with following Accessories.
     A• Padded Arm rest with straps pair with clamps
     B• Side support pair with clamps.
     C• Shoulder supports pair with clamps
     D• Knee crutches a pair with clamps.
     E• Infusion rod with clamp
     F.Drain tray
5 Environmental factors

  Sl   Name                                                        Technical Bidders
                                                                   Specs     Deviation
                                                                   quoted    if any
                                                                   by bidder
 5.1 Shall meet IEC-60601-1-2 :2001(Or Equivalent BIS)
       General Requirements of Safety for Electromagnetic
       Compatibility. or should comply with 89/366/EEC;
       EMC-directive.
 5.2 The unit shall be capable of operating continuously in
     ambient temperature of 20-30 deg C and relative
     humidity of 15-90%
 5.3 The unit shall be capable of being stored continuously
     in ambient temperature of 0-50deg C and relative
     humidity of 15-90%
6 Power Supply

  Sl   Name                                                     Technical Bidders
                                                                Specs     Deviation
                                                                quoted    if any
                                                                by bidder
 6.1 Power input to be 220-240VAC, 50Hz fitted with
     Indian plug
 6.2 UPS of suitable rating with voltage regulation and spike
     protection for 60 minutes back up.
7 Standards, Safety and Training

  Sl   Name                                                     Technical Bidders
                                                                Specs     Deviation
                                                                quoted    if any
                                                                by bidder
 7.1 Should be FDA, CE, UL or BIS approved product
 7.2 Manufacturer should have ISO certification for quality
     standards.
 7.3 Comprehensive training for lab staff and support
     services till familiarity with the system.
 7.4 Comprehensive warranty for 2 years and 5 years CMC
     after warranty including UPS.
 7.5 Electrical safety conforms to standards for electrical
     safety IEC 60601-1 (OR EQUIVALENT
     international/national standard)General requirement for
     Electrical safety of Medical Equipment.
8 Documentation

  Sl   Name                                                     Technical Bidders
                                                               Specs     Deviation
                                                               quoted    if any
                                                               by bidder
8.1 User/Technical/Maintenance manuals to be supplied in
    English.
8.2 List of important spare parts and accessories with their
    part number and costing.
8.3 Compliance Report to be submitted in a tabulated and
    point wise manner clearly mentioning the page/para
    number of original catalogue/data sheet. Any point ,if
    not substantiated with authenticated catalogue/manual,
    will not be considered.
8.4 Certificate of calibration and inspection.
8.5 List of Equipments available for providing calibration
    and routine Preventive Maintenance Support. as per
    manufacturer documentation in service/technical
    manual.
8.6 Log book with instructions for daily, weekly, monthly
    and quarterly maintenance checklist. The job
    description of the hospital technician and company
    service engineer should be clearly spelt out.

				