									                   FETAL MONITOR (Antepartum)            (a)

1. Should have facility to monitor two fetal heart rates externally using Ultrasound
    for transducer
2. Facility to monitor uterine activity externally by tocotransducer
3. Upgradable upto Triplet monitoring
4. Cross Channel Verification when both US transducer picking same signal(For
    Twins)
5. Twins offset facility (for Twins)
6. Beat to Beat sensitivity should be minimum 25 BPM through US transducers
7. Should have Minimum 6.5 Inch Color TFT High Resolution Display.
8. Should Have Touch Screen Facility
9. Should have at least One Hour Trace Storage and Facility for taking a fast
    Printout
10. FMP Parameter (Fetal Movement Profile)
11. Quantitative Fetal Movement Analysis with Percentage movement Calculation
    and Averaging Facility
12. Remote Event Marker Facility
13. Bradycardia and Tachycardia Alarm with Time Duration Facility
14. NST Timer facility
15. Green Grid Z fold thermal paper printing by integrated Thermal Printer
16. Three Speed recording
17. Interface to Fetal Telemetry System
18. Attachable to OB Trace View
19. Water Proof transducers with LED Indicators
20. Should have automatic Channel Management so that the fetal monitor can detect
    any Transducer in any socket.
21. Should have intelligent Technical and Physiological Alarms
22. Wall Mount facility should be available
23. Upgradation key for on-site upgradation

    Optional
    Trolly
    Should be upgradable to Maternal ECG and NIBP (Optional)
                     FETAL MONITOR (Intrapartum)            (b)

  The Fetal Monitor should have :-
   Facility to Monitor up to three fetal heart rates externally using ultrasound
     transducers.
   Should have facility to Monitor Uterine activity externally by TOCO Transducer.
   DECG via combined Toco/ECG “Toco+” transducer (also can measure MECG)
   Should have facility to monitor intrauterine pressure
   Upgradable upto Triplet monitoring
   Twins Monitoring should be Standard.
   Cross Channel Verification when both US transducer picking same signal(Fro
     Twins)
   Twins offset facility (for Twins)
   Beat to Beat sensitivity should be minimum 25 BPM through US transducers
   Should have Minimum 6 Inch or more Color TFT High Resolution Display.
   Should Have Touch Screen Facility for easy operations
   Should have at least One Hour Trace Storage and Facilty for taking a fast Printout
   FMP Parameter (Fetal Movement Profile)
   Quantitative Fetal Movement Analysis with Percentage movement Calculation
     and Averaging Facility
   Remote Event Marker Facility
   Bradycardia and Tachycardia Alarm with Time Duration Facility
   NST Timer facility
   Green Grid Z fold thermal paper printing by integrated Thermal Printer
   Three Speed recording
   Interface to Fetal Telemetry System
   Attachable to Central Monitoring
   Water Proof transducers with LED Indicators
   Should have automatic Channel Management so that the fetal monitor can detect
     any Transducer in any socket.
   Should have intelligent Technical and Physiological Alarms
   Wall Mount facility should be available

Optional
   Trolley
   Maternal heart rate with ECG waveform display (Optional)
   6 second maternal ECG strip, with a choice of real-time recording or “snapshot”
      together with the fetal trace (Optional)
   Should be upgradable to Maternal NIBP and pulse oximetry (Optional)
                                 Ultrasound Machine

a) Portable B/W

1. Should be portable and light weight approx 10kg
2. The system shall be capable of providing the following imaging and operating modes.
    B mode, M mode, B+B mode, B+M mode
    Cine storage - min 32 frames
    Image storage - min 16 images
    Full measurement and Analysis capabilities for Ob/Gyn applications

3.  Should have minimum 16 channel digital beamformer
4.  Should have facility for TGC slide pots
5.  System should have multiple foci at least 4
6.  Should have at least 7 inches display
7.  Should have alphanumeric keyboard
8.  Should have general measurement & calculations i.e. distance, circumference, area,
    volume and heart rate.
9. Should have facility of display annotation
10. Should have multifrequency probes
11. Should have capability of dual probe connectivity
12. Facility to be attached with PC should be there
13. System should be quoted with the following:
    a) Convex Probe
    b) Transvaginal probe
    c) Biopsy of transvaginal probe
    d) B/W Thermal printer
    e) Trolley for mounting the machine
    f) Suitable online UPS for backup upto 1 hour.
14. Zoom: Picture- in-picture zoom facility
15. Facility for optimization of image on the press of a key
b) Portable Ultrasound Machine with Colour Flow

Should be portable light weight approximately 5 kg
Reqd Probes:-            Wide-band,Multifrequency
                         Convex Array (2-5 MHz)
                         Endovaginal / Endorectal (4-10 MHz)
                         Linear Array (MHz)
                         Phased Array (1. 4-3.6 MHz)

Scanning Modes:-       Doppler modes Color, Power, PWD,
                       Directional Power
                       Triplex Mode
Software/Features:-    Tissue Harmonics facility
                       Auto Doppler, virtual convex
                       Auto Optimization, Spilit screen with Real time 2D and Color flow.
Scanning depth:-       Max Dept h 30 cm

Storage:-              Image Storage , on-board capacity, no.of stored images/loops
                       40 GB >30,000 images on board hard drive
Battery back up:-      Should have battery backup for min 1 hour.
CD R/W:-               CDRW , included
                       Cine Image Storage , multiple loops
                       Calculation package
                       Report package , option
                       Cardiac Scanning , Phased Array
Monitor/ Screen size:- Monitor Display 10. 4" High Res Monitor ( high cont rast)

Zooming:-              Pan/ Zoom
OPTIONS:-              3D Capability option
                       Connectivity DICOM option
                       Sony thermal printer, Color deskjet printer
                       Trolly
c) Traditional Low End Black & White

   1. Should be premium black & white ultrasound system
   2. The system shall be capable of providing the following imaging and operating
       modes
        B mode, M mode, B+B mode, B+M mode
        Cine storage - min 64 frames
        Image storage - min 10 images
        Full measurement and analysis capabilities for Ob/Gyn applications
   3. Should have minimum 48 channel digital beamformer
   4. Should have facility for TGC Slide pots
   5. System should have multiple foci at least 4
   6. Should have at least 12 inches display
   7. Should have alphanumeric keyboard
   8. Should have general measurement & calculations i.e. distance, circumference,
       area, volume and heart rate.
   9. Should have facility of display annotation
   10. Should have multifrequency probes
   11. Should have capability of dual probe connectivity
   12. System should be quoted with the following:
       a) Convex Probe
       b) Transvaginal probe
       c) Biopsy of transvaginal probe
       d) B/W Thermal printer
       e) Trolley for mounting the machine
       f) Suitable online UPS for backup upto 1 hour.
   13. Facility for connecting with PC by direct or additional attachment
d) Traditional with Colour Flow

1. Should have capability for the user to adjust keyboard height for increased operator
    comfort
2. The system shall be capable of providing the following imaging and operating modes
     2D, M mode, color M- mode, Rotation of M mode cursor
     Power Doppler with directional mode
     Ultrasound/Color Angio mode
     Fully Steerable Pulsed & continuous wave Doppler
     On screen Cine Storage & Image recall
     Digital image storage and patient Archive with true scanner frame rates
     Full Measurement and analysis capabilities
     Imaging frequencies from 2 MHz to 12 MHz
     Review of stored ultrasound images
     User adjustable B Colorization maps, gain settings, color Doppler baseline, angle
         correction and other important parameters with live/frozen/arc hived images/loops
3. System should have Virtual Convex in Linear transducer
4. System should have support facility to compound large organs real time in single
    field of view
5. System should support 3D applications in all applications and transducers including
    intracavity probes, should provide ability to capture 3D data. The 3D image obtained
    should be rotated to any position, 360 degree angle.
6. 3D of grayscale, CFM, grayscale + CFM should be possible.
7. System integrated keyboard for easy patient data, annotation and report entries
8. Should have a display of single, dual or quad images side by side
9. System should have an programmable architecture with data processing of phase,
    amplitude and frequency, raw data digital reply for cine/single loops allowing the
    adjustment of all major parameters and measurements.
10. Should have a built in digital archival system for image storage and archival with
    reporting facilities. The images should be stored and analyzed, extensive post
    processing, re- measurement, analysis, generation of new reports in CD should be
    available
11. Digital Cine reply, allowing store and reply ultrasound images including 2D color
    Doppler. The Cine reply should allow the user to change gain, contrast, sweep speed,
    base line etc. image parameters.
12. Should have a hi- fi Pan Zoom capability with live/frozen/stored images. Should have
    capability of zooming the archived Cine loops.
13. Should be directly compatible with Color inkjet printers.
14. System should be quoted with the following:
         a) Convex probe
         b) Transvaginal probe
         c) Linear probe
         d) Biopsy of transvaginal probe
         e) B/W Thermal printer
         f) Color inkjet printer
         g) Suitable online UPS for backup upto 1 hour
                                   Color Doppler System (4D)


1    The system should be a state of the art, high end digital system with whole body applications
     including obstretic, gynecology, pediatric, urology, pediatric, and small parts and vascular
     application . The system shoud have both adult and pediatric cardiac facilities with continuous
     wave doppler imaging .
2    Specify the No of probe connectors
3    The system should have a floadting key board with Backlit alphanumeric display
4    The system shoud have a non interlaced high resolution Monitor with tilt and swivel facitlity.

5    The system should have Integrated recording keys for remote control storage and printing options

     The system should have more than 8000 digital processing channels. Higher channels is desirable.
6
     Pls mention the no of processing channels
     The system should have windows Operating System. Pls specify the operating system used in the
7
     machine.

8    The system should have minimum 80 user defined preset per tranducer, More presets is desirable.

     The system should have advanced calculation package for all mentioned application - obs , gyn , tv,
9
     tr, small parts , doppler and cardiac( adult and pediatric )
     The system should have the below mentioned modes - B mode, CF mode, PW mode, Power angio,
10
     M mode.
11   All transducers should have tissue harmonics imaging as standard.
12   The system should have Real time triplex imaging on all transducers.
13   The system should have Panaromic View imaging
14   The system should have trapezoidal imaging for all linear probes.
15   The system should have Volume 3d and 4d Imaging on all transducer
     System should have real time compounding with all probes. It should have facility for Real time
16
     compounding with Color doppler mode too.
     The system should have one touch optimisation function for adjusting doppler function while doing
17
     doppler scans .
     System should have post processing facility on Freezon mode for Pulse and continuous wave
18
     doppler imaging .

19   System should have facility for multiple frequency selection for fat patients at one touch of a button

20   System should have a maximum depth of 30 cm .
21   Dynamic range of 120Db or more .
22   Real time 4d should be available on all tranducers
23   4d Frame rates should be more than 20 frames/sec
24   System should have real time zoom and zoom facility on frozen image too.
25   System should have a facility for Real time Calculation for Spectral doppler .
26   The system should have extended loop capability of upto 3 minutes or 2000 frames at acquisition rates
27   System should have mimimum 40 GB harddisk facilty for image storage .
     System should have an integrated image management system for printing and storing images for the
28
     offline analysis
29   System should have Dicom facility to connect it to the hospital server/printer
30   System should have inbuilt CD/DVD drive for copying images directly on CD
31   The system should store real time loops in Bmode and Color mode .
32   The system should have direct printing facility on both thermal and color injet printer
33   System should have broad band, high frequency tranducers with 5 frequecny selection option
34   Multifrequency convex transducer of 2-5Mhz for abdomen , obst and gyn application
35   Multifrequency linear tranducer of 5-12 Mhz for small parts and vascular imaging
36   Biopsy Needle guides with kit
37   Multifrequency Volume transducer from 2- 5Mhz for obstretic and abdominal 4d imaging
38   Multifrequency Volume transducer from 5-9 Mhz for transvaginal and transrectal 4d imaging
39   Multifrequency Volume transducer from 5-12 Mhz for small parts , and breast 4d imaging
                               Video Colposcope

   The video colposcope must have magnification from minimum 1x to maximum
    25x
   Display of Magnification Indicator, stop watch on monitor
   CCD resolution must be 750 lines and number of pixels 8,00,000
   The image sensor must be 1/6” super HAD color CCD with high speed digital
    signal processing
   Super bright LED light source, color temperature 6000 0 K, life more than 1000hrs
   The system shall have floor level indicator on the stand of video colposcope as
    well as to have Anti- vibration image acquisition control for proper alignment and
    capture best images
   Should be able to control essential parameters like automatic as well as manual
    focusing, zooming, light source, E Flip, Mirror image, image freeze and electronic
    green filter directly on the hand- held unit
   The system shall have SVHS and video output for simultaneous connection to
    medical grade monitor, 14” TFT/LCD screen as well as to PC
   Compatible patient data and video management software with associated
    hardware and personal computer with 40 GB hard disk, Uninterruptible power
    supply and color inkjet printer to facilitate to take colposcopy images with the
    colposcopy report on hard copy and to store still images and procedure on CD.
   Mobile Trolley for personal computer
                              Ope rating Hysteroscopy Instrume nt

S No.
        Specification
 1.     Endoscopic Video Camera System (Came ra Head & Came ra Control Unit)
        Digital Single Chip Medical-Video-Camera – Color system PAL, NTSC with integrated Image
        Processing Module with
        Digital contrast enhancement & digital anti-grid filter for use with fiber scopes
        Should have integrated Parfocal Optical Zoom lens system, f=25-50mm (2x)
        Should have the additional facility of Electronic 2x zoom, selectable in 4 steps
        Integrated still image function
        Digital Video output DV-For digital recording directly on the lap-top or any other digital recording
        device with fire wire port
        Should have the Auto Exposure System for consistent illumination level
        High horizontal image resolution of more than 450 lines
        Automatic white balance with memory function for two settings.
        Character generator
        Composite & S-VHS compatibility
        2 programmable function keys on the camera head for control of camera functions or video
        printer/recorder functions and other peripheral units
        Camera head fully soakable for disinfection
 2      Light Source (with BNC connecting cable, main cords)
        Xenon Light Source Auto LP including xenon short ARC Lamp 300 watts, BNC video cable 1.5 m
        long, power cable 50/60 Hz, 100-240 VAC having following specifications:
        Power Supply: 100-240 VAC
        Frequency : 50/60 Hz
        Colour Temperature: 6000 K Constant approx.
        With continuously adjustable light intensity from 0 to 100%
 3      Xenon Bulbs (spare)- Xenon short ARC Lamp 300 watts for the above light source
 4      Fiber Optic Cable : Heat resistant 4.8mm, 250cm
 5      Monitor: Medical grade color monitor minimum 14” with REG input.
        Multinorm monitor for PAL, NTSC
        High horizontal resolution of more than 450 TV lines
        Outstanding imaging quality
        S-VHS, RGB compatibility
        Built in speaker system
 7      Electronic Suction and Irrigation Pump: Complete set along with accessories
        Set operating voltage:100 – 240 VAC, 50/60 Hz consisting of : integrated Power Cord, connecting
        cable, 3 tubing sets irrigation HYS, for single use, sterile VACU safe Pack. Precise presetting of
        suction and irrigation parameters via touch keys, Adjacent display scales for set values and actual
        value.
 8      Electro Surgical Unit: Power supply 220-240 Volt AC, 50/60 Hz, with special features of:
        Coagulation degree preselectable in several stages
        Soft coagulation, Standard coagulation, Spray coagulation, Voltage controlled cutting
        Auto stop function, Auto start function, HF functions can be activated either via footswitch on the
        handle.
        Complete set along with accessories including Neutral Electrode, Connecting cord for Neutral
        Electrode, High frequency cord monopolar, High frequency cord bipolar, Two paddle foot switch,
        Adaptors
 9
        Mobile Trolley
 10     Telescope: 2mm diameter, 0o angle, 26 cm length, autoclavable
 11     Telescope: forward, Oblique, 4mm diameter, 300 angle, 30cm long, autoclavable
12   Telescope: 4mm diameter, 120 angle, 30cm length, autoclavable
     Telescope with contact hysteroscope mode 4mm, 300 , 30cm, magnification up to 60x
13   Diagnostic Sheath
     2.7mm diameter with one stopcock and one Luer- lock adaptor
14   Single flow operating sheath
     3.4mm diameter channel for operating instrument (5Fr) with one stopcock and one Luer lock
     adaptor
15   Continuous flow ope rating sheath
     4.1mm diameter, with one stopcock and one Leur- lock adaptor
16   Diagnostic Sheath : 5.1mm diameter
17   Single flow operating sheath with channel for semirigid 5Fr. Instruments with 1 stopcock
     and 1 LUER- lock adaptor, size 5.4mm
18   Continuous flow ope rating sheath with 1 stopcock and 1 LUER- lock adaptor, size 6mm
19   Ope rating Instruments: Grasping forceps – double action jaws, 5Fr, 34cm
20                            Scissors – blunt, 34cm, 5Fr
21                            Scissors – pointed, 34cm, 5Fr
22                            Biopsy spoon forceps, double action jws, 5Fr, 34cm
23                            Myoma fixation screw, 5Fr, 34cm
24                            Unipolar needle electrode 5Fr, 32cm
25                            Unipolar ball electrode 5Fr, 53cm
26
                                     Polypectomy loop, Unipolar, 5Fr. 34cm
27                            Vaporization electrodes – Cutting loop, unipolar
28                            Vaporization electrodes – roller electrode 5mm, 3mm, unipolar
29
                                      Bipolar Vaporization electrode 5Fr., 36cm
30   Resectoscope Sheath: 8mm diameter, 2 Luer lock adaptors, oblique beak with rotatable
     inner sheath, ceramic insulation
31
     Working element of resectoscope
32   Obturator - Standard obturator
33   Resectoscope electrodes: Cutting loop, angled and straight
34                            Coagulating electrode –ball 3mm & 5mm
35                            Coagulating electrode – barrel 3mm, 5mm
36                            Coagulating electrode – needle
37                            Coagulating electrode – micro knife
38
     Uterine grasping forceps and cannula
39
     Brushes for instrument cleaning
40   Pressure cuffs with - Silicon tubing set, rubber foot pump, manometer
41
     Formaline chamber
42
     Protection tubes for electrode
43
     Protection tubes for telescope
                    SURGICAL OPERATIVE MICROSCOPE


1.    Compact Microscope body with 1:6 zoom ratio – Apochromatic optics
      magnification range from 1.9x to 18.2x continuously variable. Newest
      apochromatic design delivers extraordinary image quality improved resolution
      and incredible color fidelity.

2.    12.5x wide field eyepieces with locking diopters.

3.    Field of view should be 112m to 11m continuously variable.

4.    Spot illumination from 65 m – 11 m continuously variable.

5.    Variable objective lens from 200 mm to 415 mm of continuously variable focal
      length without changing front lens.

6.    180° inclinable binocular tube with focal length f=200mm.

7.    Pair of handles for easy control of zoom , focus & Illumination /
      Brightness controls

8.    180 w xenon built in illumination with Backup of 180W Xenon.         The sturdy
      floor stand having magnetic brakes & cluches for all six axes (3 for microscope
      and 3 for stand).

9.    1 CCD camera attachment .

10.   Face to Face attachment for Assistant surgeon with 22.5 degree dovetail adapters .
                                    Infant Warmer

Should have following features
1. PID based microprocessor technology with internal accuracy of 0.01 0 C
2. Skin manual controlled
3. Photo glow skin, set and actual temperature display on LED
4. Ceramic infrared heater
5. Easy access of an infant from all sides due to clear acrylic collapsible side panel
6. On board temperature calibration facility with feather touch key
7. Entire warmer height up down facility through remote control using linear activators
8. Baby tray up and down (trendelenburg facility through remote control using linear
    activators
9. 3600 rotating monitoring tray
10. Swivelling 3 drawer facility
11. Facility for auto reset the system during hang up due to power fluctuation
12. Swivel heater source for taking X-Ray
13. Halogen examination lamp (12 Volts 50 Watts)
14. % of heater output display
15. Apgar Timer
16. Stop Watch
17. Baby attending alarm
18. Electronic soft-touch keyboard with key pad lock facility
19. Audio & Visual alarm
20. Mute auto programmable
21. Safety alarm message with status LED
22. Safety high skin temperature cut off at 38 0
23. Power failure alarm and probe fail alarm
24. Separate acrylic X-Ray guide for positioning the cassettes
25. Instrument tray
26. IV pole
27. Pole for infusion pump
28. 4 castor
29. Unit should be CE/FDA certified and company should be ISO 9001
Optional:
         Should be provided with complete Resuscitation kit including Silicon Manual
            Resuscitation bag larayngoscope Miller 0 & 1, disposable ET tubes, 3 Nos and
            electrically operated slow suction mounted on the unit
         Unit should have facility for undersurface phototherapy with two blue and two
            white tubes
         Unit should have optional facility for bubbling CPAP with built in compressor
            and Air/O 2 mixer
         Unit should have optional facility of SpO2 monitoring facility with
            plethysmograph and digital conversion of signal at probe level giving good
            low perfusion monitoring and motion artificial correction.
                                Neonatal Phototherapy

                             (Single Surface Mobile CFL)

Single surface CFL phototherapy should be provided with Overhead source unit portable
undersurface
1. Overhead unit should be provided with an irradiance of more than 10-12 w/am/cm2
   with two white light and 4 special blue tubes (18 Watts)
2. Unit should be mounted on a canopy with adjustable height and angle adjustment
   facility.
3. Unit should work on 220V AC
4. Unit should be mounted on castor wheels with easy mobility.
5. Unit should have Head up/down and I/V stand facility
6. Unit should be provided with digital flux meter to indicate flux w/nm/cm2 with 420-
   480nm filter and should also provide flux/hr and total flux. With automatic alarm
   indicating tube change.
7. Unit should have IEC-301 safety test report

Optional:

      Unit should be mounted on Infant care trolley with transparent acrylic collapsible
       side facility (compatible with radiant heat warmer) and shelf.
      Undersurface unit should be provided with 2 special blue CFL light, 18 Watts and
       two white CFL lights
                                        ICU Monitor

Multi Channel Bedside Monitor with facility of ECG, Pulse Oximeter and non-invasive
BP monitoring.

Display of dual trace ECG & plethysmograph on CRT

Display of Sp02 %, Pulse Rate, Heart Rate, Systolic/Diastolic BP meas urements

Facility of three lead selection with lead fault indicator

Auto manual & STAT modes for BP measurement.

Non volatile alarm limits setting storage

Protected against diathermy and defibrillation.

Built in battery back up.
                      Boyles Trolley (Work Station)

Technical Specifications:
   A. Anesthesia System

   1. Compact Breathing system with minimal number of parts and tube connections to
      reduce the potential for leaks and misconnects
   2. Bag/vent switch to turn the ventilator on/off.
   3. The breathing system components should be easy to disassemble and fully
      autoclavable and latex- free.
   4. Selectatec type mounting for 2 vaporizers.
   5. Integrated circle absorber system with analog pressure gauge and integrated
      Adjustable Pressure Limiting Valve.
   6. The machine should have auxiliary outlet for fresh gas.
   7. Reusable and autoclavable soda lime canister.

   B. Ventilator

   1. Volume Controlled with electronic PEEP. Ventilator should be integrated for
       safety & minimization of leakage.
   2. Exhaled Volume Monitoring facilities
   3. Direct access to ventilator parameter settings
   4. Smart alarms to direct user to specific problems and affected parameters
   5. Pressure bar graph for visual reference on a breath-by-breath
   6. Latex- free autoclavable bellows
   7. Alarms for
           a.         Low/High Tidal volume
           b.         Low/High Minute volume
           c.         Apnea alarm
           d.         Low airway pressure
           e.         (Plimit) range: upto 99 cm H2 O
           f.         Sustained airway pressure
   8. Modes of ventilation: Volume Control mode
   9. Ventilation Parameters
           a.         Tidal volume range upto 1500 ml
           b.         Rate upto 65 breaths per minute
           c.         Inspiratory/expiratory ratio: 2:1 to 1:6
           d.         Inspiratory pause adjust: 5% to 60% or inspiratory time
           e.         Positive End Expiratory Pressure (PEEP): Range: OFF, 4 to 30 cm
              H2 O (increments of 1 cm H2 O)
   10. Battery backup

   C. Low Flow Delivery

      a.             Automatic Fresh gas flow compensation
      b.             Minimum O 2 flow of 50 ml,
      c.             Dual cascaded flow tubes for O 2 and N 2 O
  D. Safety
     a.     Chain-Link type mechanical hypoxic guard to provide minimum 25% O 2
            concentration across all O2 /N 2 O mixtures.
     b.     Oxygen Failure warning should be there
     c.     Nitrous Oxide should be cut-off in case of failure of O2 supply pressure

  E. Vaporizers

      a.    Temperature, pressure and flow compensated spill-proof vaporizers for
            Halothane and Isoflurane

  F. Multichannel monitor with minimum 10” color screen. Suitable for adult
     patients.
     Should have standard parameters –
     ECG, SPO2, NIBP, Temperature, Respiration and IBP
     Graphic & Tabular trends for 24 Hrs
     Arrhythmia detection & Alarm
     ST Segment Analysis
     Network capability
     Capnography
     Suitable mount for mounting the monitor on Anaesthesia Machine
     Should have inbuilt rechargeable battery backup.

  G. Anesthesia machine, circle absorber, ventilator, Multichannel Monitor &
     vaporize r should be from same company for easy serviceability

All the above features should be highlighted & numbered in the original
brochure/catalogue of the item (to be attached with the offer).
High Pressure High Vaccum Horizontal Rectangular Steam Ste rilizer Fully Automatic (a)

   (a) Rectangular in shape constructed of heavy duty stainless steel (SS-304/316) door and
       jacket of stainless steel.
   (b) The pipes and fittings of stainless steel, the valves ball type – sterlizer mounted on
       MS sturdy tubular type stand made of SS 304
   (c) All other construction detail as per IS 3829
   (d) Designed as per international standard
   (e) Chamber of autoclave tested for hydraulic test pressure performance as per BS & IS
       standard
   (f) Working pressure of autoclave from 1.2 to 2.1 Kg/Cm2 and vaccum up to 26” Hg
       with built in vaccum pump arrangement
   (g) Door gasket design expansion type – silicon and joint less
   (h) Accurate in sterilization time and temperature
   (i) Door interlock
   (j) Over heat protection
   (k) Electronic timer for controlling the sterilizing time
   (l) Excess pressure safety valve
   (m) Self diagnosis fault message
   (n) Safety against electric overload
   (o) Visual and audio alarm for easy operation
   (p) Electrically operated
   (q) Size       Width          Length          Height
                  600mm          900mm           1500mm
   (r) Product must be (ISI mark) and CE certified
        High-Pressure Cylindrical Ste rilizer –Vertical     (b)


Bearing ISI mark-IS 3829 III (ISI) and CE (mark)
The chamber of heavy gauge SS chamber is fitted with two safety valve of SS,
pressure gauge- water outlet value and air in let value, SS steam condenser, tube
with stainless steel eject value
        Size: Diameter             Depth     Load
              400mm                600mm     6KW
                                    Big Boilers

                           (Bowl – Sterilizer CE & ISI)

Used in clinics & CSSD to disinfect instruments with hydraulic lid, which closes
the sterilizer smoothly.
Lid can be lifted with pedal type device
Thermostatically controlled – supplied with instrument tray and hanger system to
dip instrument fully inside water
Size:    Width             Depth         Height           Load
         600mm             600mm         600mm            6 KW
                       Ventouse Extractor

a) Hand held unit for delivery of fetus
b) Traditional unit with suction for delivery of fetus
                                Medical Gas Pipe Line System
Minimum qualifications & must fulfill criteria and conditions
                                                   Part-1
Tenderer who is quoting the bid should be an ISO 9001: 2000 certified company for quality assurance. Copy of the
valid ISO certificate must be enclosed with their tender bid.

Tenderer should submit letter of the Authority of their Foreign Principal / M anufacturer for the quoted imported
products. Only sole and exclusive distributor/agent can bid the tender on behalf of their principals/foreign
manufacturer.

Tenderer should have factory trained persons from the Foreign Principal / M anufacturer for the quoted imported
products. Copy of the certificate in this regard must be submitted with the t ender.

Firms should have sound financial background to undertake the project and should have a bank solvency of minimum
one hundred lakhs or above and also have annual turn over of above seven hundred lakhs for the last financial year
ending 2006-2007. CA certificate and bankers certificate both should be submitted in this respect and failing which bid
will be straight way rejected.

Firms should submit copy of CE certificate with CE no. for the quoted products. Without CE no. products will be
treated as no compliance to the specifications and conditions.

Firm should not quote any optional items. If any firm quotes any optional items, they will disqualified. Firm must quote
strictly as per the desired specifications and items mentioned in BOQ.

                                PART-II
TECHNIC AL S PECIFICATIONS FOR MED ICAL GAS PIPE LIN E S YS TEM

OXYGEN S UPPLY S YS TEM
Gas Manifold for Oxygen :The central oxygen manifold supply facility shall comprise of two cylinder banks
extendable type of size 6 + 6 bulk oxygen cylinders complete with copper tail pipe with bull nose fittings suitable for
cylinder valves and cylinder support system. The tail pipes shall be fitted to individual non-return valves of the
cylinder manifold for easy removal of cylinders without disturbance to system operation. M anifold should also have 12
brass non-return valves for individual oxygen cylinder. Threaded connections for each manifold block with non-return
valve shall be distantly separate for Oxygen and Nitrous Oxide so as to eliminate possibility of inadvertent
interchange. Each manifold shall be provided with a terminal header and a NPT connection for automatic gas change
over system (Automatic gas control panel).

Fully Automatic Control Panel for Oxygen : Should fully meets and complies with BS and HTM -2022, C11
standards. Should be duly CE marked with CE no. to be specified. An Automatic control panel (ACP) shall be
provided. The panel shall consist the following but not limited to the following : - Automatic manifold control panel
should provides a duty or standby gas supply from two cylinder banks with automatic changeover the manifold
indicates local status and has a remote indicator facility. Should have minimum 500 L/min output flow available.
Automatic manifold control panel should be designed to supply the M GPS with a constant pressure supply via a control
panel form two equal banks of cylinders and the changeover from the „duty‟to the „standby‟ bank of cylinders should
be automatic. Automatic changeover from the duty to the standby bank should occur at a cylinder pressure that will
ensure the maximum usage of the contents of the duty bank. If the normal operation of the changeover control depends
on an electrical supply the design should ensure that, in the event of an electrical supply failure, there should not be any
disruption to the flow of gas into the pipe line supply and on restoration of the supply, the original running bank shall
return on line. The unit should have separate pressure regulating valves for each bank of cylinders and the control
system should be designed for ease of maintenance. Cylinders can be changed, or the pressure regulating valve
removed for overhaul without loss of continuity of the gas supply.

Emergency Oxygen System : Emergency oxygen cylinder manifold should be of two cylinders emergency manifold
and at all two cylinders at on end with a high pressure high flow regulator with double gauges, two pig tail pipes and
two non return valves manifold with Top & middle frame for Oxygen to connected with the hospital pipe line system
with valve.

Oxygen Generation S ystem :M icroprocessor based Oxygen Generator plant to produce 94%+/-1% of medical grade
oxygen minimum 15 m3 continuously through single unit. Oxygen so produced should be supplied through pipeline at
4.1 bar pressure. System should be PSA based. It should be supplied with Oxygen Analyzer with minimum shelf life of
system. Compressed air system compatible for oxygen generator module and compressed air 4bar for OT and OCU
areas. Compressed air system with single air compressors of non-lubricated type along with respective accessories. It
should be skid mounted, fitted with 40 HP 3 phase TEFC 440volts electrical motor, along with desiccant heat -less type
air dryer, after cooler with moisture separator and auto drain valve. water recycling pumps (1 No.), Cooling Tower (1
No.), One No. of air receiver, and one no. of oxygen storage tank of suitable capacity fitted with valves, pressure
gauges should be supplied. Tenderer should provide training to hospital staff for a period of minimum fifteen days for
operation and maintenance of the system. The supplied air compressors should be capable enough to provide 4bar
compressed air to various areas like OT, ICU, Recovery through gas terminal points.

Nitrous Oxide S ystem : The central N2O manifold supply facility shall comprise of two cylinder banks extendable
type of size 2 + 2 bulk N2O cylinders complete with 2nos. copper tail pipe with bull nose suitable for cylinder valves
and cylinder support system. The tail pipes shall be fitted to individual non-return valves of the cylinder manifold for
easy removal of cylinders without disturbance to system operation. M anifold should also have two brass non-return
valves for individual N2O cylinder. Threaded connections for each manifold block with non-return valve shall be
distantly separate for Nitrous Oxide so as to eliminate possibility of inadvertent interchange. Each manifold shall be
provided with a terminal header and a NPT connection for automatic gas change over system (Automatic gas control
panel).

Fully Automatic Nitrous Oxide Control Panel : Should fully meets and complies with BS and HTM -2022, C11
standards. Should be duly CE marked with CE no. to be specified. An Automatic control panel (ACP) shall be
provided. The panel shall consist the following but not limited to the following : - Automatic manifold control panel
should provides a duty or standby gas supply from two cylinder banks with automatic changeover the manifold
indicates local status and has a remote indicator facility. Should have minimum 500 L/min output flow available.
Automatic manifold control panel should be designed to supply the M GPS with a constant pressure supply via a control
panel form two equal banks of cylinders and the changeover from the „duty‟to the „standby‟ bank of cylinders should
be automatic. Automatic changeover from the duty to the standby bank should occur at a cylinder pressure that will
ensure the maximum usage of the contents of the duty bank. If the normal operation of the changeover control depends
on an electrical supply the design should ensure that, in the event of an electrical supply failure, there should not be any
disruption to the flow of gas into the pipe line supply and on restoration of the supply, t he original running bank shall
return on line. The unit should have separate pressure regulating valves for each bank of cylinders and the control
system should be designed for ease of maintenance. Cylinders can be changed, or the pressure regulating valve
removed for overhaul without loss of continuity of the gas supply.

Emergency Nitrous Oxide S ystem : Emergency N2O manifold system should be two cylinder emergency manifold at
all two cylinders at on end with a high pressure high flow regulator with double gauges, two pig tail pipes and two
non return valves manifold with Top & middle frame for N2O to connected with the hospital pipe line system with
valve.

Vacuum (suction ) Supply S ystem :- Tenderer should provide two vacuum pumps, air cooled typ e, 2 Nos. base frame
mounted, air cooled type , each pump having piston displacement –60cfm (1695 LPM ), free air delivery to suction
approximately 90% of P.D., maximum working pressure 29” of Hg or 730 mm of H g, single stage, two cylinder, fitted
with M .I. channel frame complete with V-belt drive, belt guard etc. 2nos. 5HP 3phase TEFC 440volts electric motor ,
Silencers at vacuum pumps discharge end – 2 Nos. Vacuum Filter with drain valve – 2 Nos., one 2000 liters water
capacity vacuum receiver . One vacuu m pressure gauge. Two Y stanners.

Air Filtration System : One set of four – stage filter (one as working and one standby) Should be provided as
follows: The filters should be made of die cast aluminum housing with epoxy powder paint on the outside and
anodized surface treatment inside to prevent corrosion and ensure extra long life. The filters Should have maximum
contaminant removal efficiency with minimum pressure drop. Total 4 stages of filters Should be used (stage 1 & 2
Should be installed before the desiccant dryer and balance two stages after.

Copper Tube /Piping
All copper tubes should be medical grade manufactured to BS EN 13348:2001 latest standards. It should have Hard
ness to R250 half hard. It should also be certified that pipes are suitable for the particulars service and cross
connections anti-confusion tests has been carried out by the contractor. Pipe sizes should be used as per BS standards:
42mm OD X 1.2 mm thk
28mm OD X 0.9 mm thk
22mm OD X 0.9 mm thk
15mm OD X 0.9 mm thk
12mm OD X 0.7 mm thk

Fittings :Fittings shall be wrought copper, brass or bronze conforming to BS: 864 part 2 and suitable for a steam
working pressure of 17 bar and especially made for brazed socket type of connections.
Installation: M edical gas pipe shall be installed in utmost cleanliness and in a manner to prevent any mechanical &
chemical damage, exposure to excessive heat, splashing, dripping or permanent contact with oil, greases etc. and
proximity to electrical installations subject to spark etc. Only pipes, fittings and valves which have been degreased and
brought in polythene sealed bags shall be used on site. Pipe fixing brackets shall be non-ferrous or non-deteriorating
plastic suitable for the diameter of the pipe. All screws used shall be chromium plated brass. Wherever pipes run in
service ducts or voids contractor shall ensure adequate ventilation for each area and prevention of concentration or
gases in event of any leakage.
M aster Alarm Panel : Should fully meets and complies with BS and HTM -2022, C11 standards. Should be duly CE
marked with CE no. to be specified. M aster Alarm Panel (M AP) shall monitor the central gases, vacuum and
compressed air and work or indicate abnormal conditions as per specified herein. M AP shall be located in t he gas
manifold room in the basement of the APC Building. Alarm should be designed to monitor piped gas pressure (high
and low pressure on up to six services), via pressure switches, in theatres, intensive care units, recoveries, private
rooms and wards etc. It should also act as a slave to another alarm within 250 meters, using a three core screened
cable (four cores if signals are to be returned to the master unit).Local area alarms should provide indication of the
condition of gas (Normal , High or Low Pressure) at the point of use, by monitoring the internal pressure of the
pipeline. The method of monitoring should be by individual pressure switches for high pressure, low pressure, and low
vacuum and each switch should be fitted with an end of line monitoring resistance, matched to the alarm panel, to
enable the alarm panel to detect any faults on the system wiring or signal transmission. Alarm panel should have
battery back up will be provided within the system to enable the alarm panel to function normally in the event of mains
power failure. The internal battery must be used to keep alarm panels operational in the event of mains power failure.
Should be designed to monitor high and low pressure from local pressure switches.

Area Alarm Panels : Should fully meets and complies with BS and HTM -2022, C11 standards. Should be duly CE
marked with CE no. to be specified. Area alarm should be equipped with a local sensor module, which is installed
inside the box. Each sensor should be connected to a gas pipe with a gas specific DISS check unit and be clearly
marked and identified with a color-coded label for the gas specified. Area alarm should be designed to monitor piped
gas pressure (high and low pressure on up to six services), via pressure switches, in theatres, intensive care units,
recoveries, private rooms and wards etc. It should also act as a slave to another alarm within 250 meters, using a three
core screened cable (four cores if signals are to be returned to the master unit).Local ar ea alarms should provide
indication of the condition of gas (Normal , High or Low Pressure) at the point of use, by monitoring the internal
pressure of the pipeline. The method of monitoring should be by individual pressure switches for high pressure, low
pressure, and low vacuum and each switch should be fitted with an end of line monitoring resistance, matched to the
alarm panel, to enable the alarm panel to detect any faults on the system wiring or signal transmission. The alarm panel
should have the ability to display up to six services each with normal & two fault conditions. Alarm panel should have
battery back up will be provided within the system to enable the alarm panel to function normally in the event of mains
power failure. The internal battery must be used to keep alarm panels operational in the event of mains power failure.
Should be designed to monitor high and low pressure from local pressure switches. Should be available in surface or
flush format.

Bed Head Horizontal Panel : The panel should integrate the mechanical, electronic at the patient bed side with an
aim to facilitate easier access to doctors, enhance the comfort of the patient and of aesthetics of the ward. The bed head
panel (horizontal) should be made from high strength extruded aluminum profiles with powder coating with a choice of
color to suit the interior décor of the hospital. It should be custom built to house gas and electrical outlets. Bed heads
panels should be supplied with powder coating on aluminum with a choice of color to suit the interior décor of the
hospital. Each bed head panel should be 5 feet long and should have following:-
                     Provision for Two Oxygen outlets.
                     Provision for One Vacuum outlet
                     Provision for One MA4 Air 4bar outlets
                     6 Nos. 5/15amp -3 Pin Electrical Switch Sockets.
                     One utility basket
                     One Infusion pump rail (stainless steel) mount pole with adaptor
                     One “IV” tree (stainless steel) rail mount with adaptor
                     One M onitor Tray with rail mount clamps

Terminal Units ( Gas Outlet Points )
Should fully meets and complies with BS and HTM -2022, C11 standards. It should be duly CE marked and CE no. to
be specified on it. The terminal outlet of all the services (e.g.O 2, N2O, CA 4, & Vacuum) should conform to
International Standard and the features required are : Should have Integral check valve Integral check valve – allows
removal of the housing and socket assemblies for maintenance without closing down the entire pipeline and each outlet
should be individually tested. It should be hundred metal construction. Full metal to metal seal on maintenance check
valve ensures no degradation over time. It should be of all hundred percent metal and must incorporate a sheerplance
that ensures a fail – safe condition after accidental damage or bed jacking (causing no da mage to first fix and enabling
easy replacement without isolation). Construction of the terminal unit should be of machined brass and die – cast
chrome collar with stainless steel rolling pins. Each of the gas specific components must have the gas service engraved
onto it, to ensure safety and compliance with standard. Each wall mounted outlet point must be enclosed in a white
ABS decorative mounting box with an ABS fascia and both the mounting box and the fascia shall have rounded corners
to avoid the possibility of injury. The box should be supplied with a flush mounting bezel as a plaster finish. Should
have safety features like p ositive action of rolling pin latch mechanism which hold the probe securely, anti rotational
locking bar and the gas indexing pin are cast into the socket assembly and cannot come loose or be removed and gas
specific indexed – eliminates the risk of connecting a socket assembly of one gas to the terminal block of another,
either during installation or maintenance.
Oxygen Flow Meter with Humidifier Bottle
It should be CE marked and CE no. to be specified . Fully quality assurance medical devices. It should be pressure
compensated prevents back pressure build up on flow indicator, durable polycarbonate flow tube with cover. It s hould
be made up of anodized aluminum body and control knob. Flow meter should have twin graduated scale which must
provides precision control permanent scale graduations. Flow meter should be placed in the vertical position. To use
it, gradually turn on the knob and adjust the position on the ball in accordance with the desired flow rate. The flow
meters should be of 1-15 LPM range for oxygen and with inlet pressure 50-60psi. Humidifier bottle should be
unbreakable, reusable to disinfectants and complements. It should be autoclavable at 121degree centigrade and made of
polycarbonate material, 250 cc capacity.

Theatre Vacuum Unit
It should be CE marked and CE no. to be specified. Fully quality assurance medical devices. It should be mounted on a
trolley . It must consists of the following :-1no. Suction Regulator and 2nos. 4000ml polysulphone collection jar and
both to be mounted on a trolley. Suction Regulator : The unit should include one suction regulator. Suction regulator
should be supplied with a safety jar, including and antibacterial filter and an anti overflow safety device. Should have
wide membrane continuous suction controller with reliability & fine adjustment. Should have plastic body and cover,
preventing risk of corrosion. Should have v acuum levels : 0-1000mbar/hPa. Should have vacuum gauge fitted with a
protective bumper device. Should have on/off knob allowing for the quick restoration of a readjusted vacuum level.
M ust have central adjustment knob with a color coded for 0-1000 mbar/hPa. Should have polycarbonate 150cc safety
jar, autoclavable at 121º C, unbreakable, fitted with an anti overflow safety device and equipped with a plastic
antibacterial filter. Suction regulator must have a unique serial number stamped on the body of each suction regulator ,
thereby allowing for identification and trace ability. The unit should include two polysulphone collection jar of 4000ml
capacity and should be unbreakable and autoclavable upto 134º C with screw –fit lid Collection jar must be fitted with
an extremely simple anti overflow safety device, thereby ensuring easy maintenace. Should be totally transparent, they
ensure perfect sucked liquid visibility. Each suction jar must have a batch number at the bottom be for identification
purposes, a batch number is marked at the bottom of each jar.

WARD VAC UUM UN IT
It should be CE marked and CE no. to be specified. Fully quality assurance medical devices. Ward Suction Unit
consists of two parts : a. Suction Regulator and b. Polysulphone Jar of 2litres with lid a)Suction Regulator :Suction
regulator should be supplied with a safety jar, including and antibacterial filter and an anti overflow safety device.
Should have wide membrane continuous suction controller with reliability & fine adjustment. Should have plastic
body and cover, preventing risk of corrosion. Should have vacuum levels : 0-1000mbar/hPa. Should have vacuum
gauge fitted with a protective bumper device. Should have on/off knob allowing for the quick restoration of a
readjusted vacuum level. M ust have central adjustment knob with a color coded for 0-1000 mbar/hPa. Should have
polycarbonate 150cc safety jar, autoclavable at 121º C, unbreakable, fitted with an anti overflow safety device and
equipped with a plastic antibacterial filter. Suction regulator must have a unique serial number stamped on the body of
each suction regulator , thereby allowing for identification and trace ability. b) Polysulphone Collection Jar : Collection
jar must be of 2litrers capacity and should be unbreakable and autoclavable upto 134º C, Jar capacities must be
minimum two liter with screw –fit lid Collection jar must be fitted with an extremely simple anti overflow safety
device, thereby ensuring easy maintenance. Should be totally transparent, they ensure p erfect sucked liquid visibility.
Each suction jar must have a batch number at the bottom be for identification purposes, a batch number is marked at
the bottom of each jar. Polycarbonate jar will not be accepted and only polysulphone jar will considered.

Low Flow Vacuum Unit
It should be duly CE marked with CE no. to be specified on it. Suction Unit consists of two parts : a. Suction Regulator
and b. Polysulphone Jar of 1litres with lid a)Suction Regulator :Suction regulator should be supplied with a safety jar,
including and antibacterial filter and an anti overflow safety device. Should have wide membrane continuous suction
controller with reliability & fine adjustment. Should have plastic body and cover, preventing risk of corrosion. Should
have vacuum levels : 0-250mbar/hPa. Should have vacuum gauge fitted with a protective bumper device. Should have
on/off knob allowing for the quick restoration of a readjusted vacuum level. M ust have central adjustment knob with a
color coded for 0-1000 mbar/hPa. Should have polycarbonate 150cc safety jar, autoclavable at 121º C, unbreakable,
fitted with an anti overflow safety device and equipped with a plastic antibacterial filter. Suction regulator must have a
unique serial number stamped on the body of each suction regulator , thereby allowing for identification and trace
ability. b) Polysulphone Collection Jar : Collection jar must be of 1litrers capacity and should be unbreakable and
autoclavable upto 134º C, Jar capacities must be minimum two liter with screw –fit lid Collection jar must be fitted
with an extremely simple anti overflow safety device, thereby ensuring easy maintenance. Should be totally
transparent, they ensure perfect Sucked liquid visibility. Each suction jar must have a batch number at the bottom be
for identification purposes, a batch number is marked at the bottom of each jar.

High pressure antistatic rubber tube for O2, N2O, Air & Vacuum : D uly color coded for individual services i.e.
white for Oxygen, Blue for N2O, Black for Air and Yellow for Vacuum, antistatic rubber tube. Should be as per
medical devices directive and CE marked with CE no. to be specified.

Electrical Control Panel inside Plant Room : Common electrical control panel for Air-Compressors and Vacuum
Pumps will be cubicle type, complete with M ain Incomer, Bus Bar arrangement, Voltmeter with VSS, phase indicating
lights, phase loss/phase Reversal indicating light, individual M CB for all starters, individual Ammeters for all motors,
Single Phasing preventer, Contactors, overload relays, control Circuit M CB, Start / Stop push buttons, Auto/M anual
switches, Pump “ON” and pump “TRIP” indicating lights, Sequencing Relays (Separate for Air compressor and
vacuum pumps) with overriding features. The panel should be cubicle type with side cubicles for bus bars for separate
sub-circuit to load breaking isolating switch. There should be individual compartment which should contain the
control circuit of each compressor separately with associated equipment for that particular compressor or vacuum
pumps. Steel bulk heads should be provided to minimize the risk of the spread of fire from one compartment to the next
. Warning notices indicating the presence of medium voltage should be attached to each compartment as necessary to
conform with IEE regulations. The load breaking isolating switch preferably to be incorporated in the control assembly
and interlocked with its cover. Auxiliary apparatus and associates circuits should be protected by M CCB mounted
within the enclosure. Each motor should have its own STARTER which should be rated for frequent duty in
accordance with BS:587 and fitted with thermal overload protection. The starter should be of automatic type so that
once switched on, the motor will automatically restart after an interruption of the supply. All 3 phase motors should
have single phasing preventors. Each star / delta should be mechanically and electrically interlocked to prevent
simultaneous star and delta connections. An industrial grade ammeter to BS:89 should be connected to the yellow
phase connection to each main motor. A manual / auto switch should be provided for each compressor assembly so
that each set may be run continuously manually.

ELECTRICAL WORK inside Plant Room : ospital will supply 3phase cable with distribution board and at one point
inside the plant room. Electric able with fixtures an fittings for manifold room will be done by the hospital. All switch-
gear and motor control centre e.g. switches statement etc. volt and ampere meters. Contactors shall be of reputed make
e.g. L&T, Siemens, GEC or English Electric. All switch gear motor etc. shall be of the same make for
interchangeability. All electrical equipment shall be earthed in an approved manner as per I.E.E. rules and acceptable
to the local authority. Earthing station shall be provided by the contractor. No medical gases pipe shall be used for
electrical earthing. Entire installation shall be done taking care to follow all safety regulations for electrical installation
of piped medical gases system. Two main supply of the required KW shall be provided up to the electrical control
panel. The wiring after the control panel has to be provided by the supplier as per IEE regulations. Following material
must be inside the plant room.
1.         Cables ,
2.         G.I Earth Wire
3.         Saddling
3.         Thumbling
4.         Gland
5.         Control Cable
               PART-III BOQ FOR MEDIC AL GAS PIPE LIN E S YS TEM
 Sr.      Description                                                                                                         Qty
No.
1.        M edical grade Copper Tubes/Pipes (Should be strictly as per tender technical specifications )
          Outer Dia           Thickness
          15 mm                0.9 mm                                                                                         1200mtrs
          22 mm                0.9 mm                                                                                         900mtrs
          28 mm                0.9 mm                                                                                         250mtrs
          42 mm                1.2 mm                                                                                         145mtrs

2.       6 +6 size manifold with 12nos.tail pipe and 12nos. non return valves                                               1set
         (Should be strictly as per tender technical specifications )

3.       Two cylinder emergency oxygen manifold system with 2nos.tail pipe and 2nos. non return valves (Should              1set
         be strictly as per tender technical specifications )

4.       2x2 size manifold System for N2O with 2nos.tail pipe and 2nos. non return valves                                   1set
         (Should be strictly as per tender technical specifications )

5.       Two cylinder emergency N2O manifold system with two tail pipe and two non return valves (Should be                 1set
         strictly as per tender technical specifications )

6.       Central Vacuum Supply System (Should be strictly as per tender technical specifications )                          1set

7.       Electrical Control Panel (Should be strictly as per tender technical specifications )                              1set

8.       4 Stage Air Filtration System (Should be strictly as per tender technical specifications )                         1set

9.       M edical Gas Area Line Pressure Alarm (Should be strictly as per tender technical specifications )
         3Gas +1 Vacuum                                                                                                     3nos.
         2Gas +1 Vacuum                                                                                                     1no.
         1Gas +1Vacuum                                                                                                      4nos.
10.        Bed Head Horizontal Wall Panel for ICU (Should be strictly as per tender specifications )                     4nos.

11.        Terminal Units (Gas Outlets Points with S.S. Probe) Oxygen-50nos. ,Vacuum-5nos., N2O-7nos.,                   12nos.
           Compressed M edical Air 4Bar-11nos. (Should be strictly as per tender technical specifications )

12.        O2 Fully Automatic Control Panel (Should be strictly as per tender technical specifications )                 1no.

13.        N2O Fully Automatic Control Panel (Should be strictly as per tender technical specifications )                1no.

14.        High Pressure Antistatic Rubber Color Coded Tube                                                              160mtrs
           Oxygen-White color ,Vacuum-Yellow color,N2O-Blue color, Air-Black color
           (Should be strictly as per tender technical specifications )

15.        Oxygen flow meter with humidifier bottle (Should be strictly as per tender technical specifications )         4nos.

16.        Ward Vacuum Unit (Should be strictly as per tender technical specifications )                                 27nos

17.        Low flow Ward Vacuum Unit (Should be strictly as per tender technical specifications )                        2nos

18.        High Suction Theatre Suction Unit (Should be strictly as p er tender technical specifications )               11nos.

19.        M aster Alarm (Should be strictly as per tender technical specifications )                                    1no.

20.        Bidder should provide all electric wiring and cable inside the gas manifold and plant room.                   1set
           ***** 3phase Electric supply from one source which will be provided by the hospital

21.        Low Pressure silicon rubber tube (Should be strictly as per tender technical specifications )                 120meters

22.        Oxygen Generation System (Should be strictly as per tender technical specifications )                         1no.

23.        Area Valve Service Unit (Should be strictly as per tender technical specifications )
           Oxygen Single Service AVSU                                                                                    11nos.
           N2O Single Service AVSU                                                                                       04nos.
           MA4 Air Single Service AVSU                                                                                   5nos.
           Vacuum Single Service A VSU                                                                                   11nos.



             List of Outlets for MGPS at Gyanecology Deptt of KGMC-Lucknow


            FLOO
 S.No.      R                   LOCATION (Area)                     O2      N2O         CA 4   Vac   F/ M W/V      T/V          L/V   BHP
      1        GF      Plant Room                                    1        1          1      1      1      1
      2        GF      Endoscopy OT 1 Table (Wing A)                 2        1          1      2      1      1    1             1
      3        GF      Major Gyne OT 2 Tab le (W ing A)              4        2          2      4      1      1    2             1
      4        GF      Recovery Roo m 2 Bed (Wing A)                 2                   2      2      2      2
      5        FF      Emergency OT 2 Tab le (W ing A)               4        2          2      4      1      1    2             1
      6        FF      Labor Roo m 3Table ( W ing A)                 3                          6      3           3             3
      7        FF      Labor Roo m 4Table ( W ing A)                 4                          8      4           4             4
      8        FF      Eclims ia Roo m 4 Table (Wing A)              4                          4      4      4
      9        FF      ICU 9 Beds ( Wing B)                          11                  2      9      9      9                        4
      10       FF      Post Op VIP Room 3 Bed (Wing B)               3                          3      3      3
      11       FF      Endoscopy OT 1 Table (Wing B)                 2        1          1      2      1      1    1             1
      12       FF      Neontal Ward                                  10                        10      10                       10
                       Total                                         50       7          11    55      4      23   13           21     4
                              Diesel Generating Set
A. Diesel Engine
Kriloskar, Jaleson or equivalent Oil Engine Ltd,. Make diesel Engine Water colled with
canopee Air/water cooled Horizontal totally enclosed, electric start, compression ignition,
4 stroke, multicylinder designed to run continuously at 1500 rpm complete with

Radiator colling (applicable to WC Engine only)
Radiator fan with Guard (Applicable to WC Engine only)
Flywheel with starter ring
Oil bath/ Dry type air cleaner as applicable.
Exhaust gas driven Turbocharger (only in TC engines)
Fuel Filter
Water temperature gauge (applicable to W.C. Engine only)
Lub oil temperature gauge
12V/24V electric start system complete with starter, dynamo/alternator, cutout, charging
ammeter.
Engine control panel consisting of ON/OFF key switch, lub oil pressure gauge, ammeter
and cutout
Engine maintenance manual with spare parts list
High Head Temp. and low lub Oil pressure protection
Electronic Governor from 250 KVA and above.

B. Alternator 50 KVA
Self excited self regulated, three phase, 4 wire, 415 Volt, 50 c/s, 1500 RPM Screen
protected, drip proof conforming to IS-4722/BS-2613 of Kirloskar/Crompton /Stamford
make

C. Control Panel
   With AMF control Panel (automains failure)
                       O.T. Instruments (General) for Gynaecology

S No.                                 Instruments
 1       Foerster Sponge holding forceps SERR, STR, 25cm

 2       Sims anterior vaginal wall retractor 26cm

 3       Sims vaginal speculum Comp 80mm x 35/40mm

 4       Cuscos Vaginal speculum

 5       Cervical speculum

 6       TI Cannula

 7       Millin Bladder retractor

 8       Bowl lifter

 9       Green Armytage placenta & ovum forceps 21cm

 10      Hegar UT dilator set 4mm – 17mm in metal box

 11      Teale vulsellum forceps 3 x 3T, CVD 23cm

 12      Blake uterine curette sharp & blunt 27cm

 13      Seanger ovum forceps STR 27cm

 14      Scissors

      a) Surgical scissors SH/BL STR 14.5cm

      b) Mayo dissection scissors CVD 23cm TC insert

      c) Mayo dissecting scissor STR 17cm

      d) Mayo dissecting scissor CVD 17cm

      e) Metzanbaum dissecting scissors BL/BL CVD 23cm

 15      Needle holders

      a) Jacobson‟s CVD 18cm

      b) Crilwood needle holder STR 18cm TC insert

      c) Crilwood needle holder CVD 18cm TC insert

      d) Mayo Heqar needle folder 20cm
16    Spencer Wells artery forceps STR 20cm

17    Spencer Wells artery forceps CVD 20cm

18    Oschner Kocher artery forceps 1x2T STR 30cm

19    Kocher haemostatic forceps 1/2 T STR 24cm

20    Kocher haemostatic forceps 1/2 T CVD 24cm

21    Allis tissue forceps 5 x 6 T 19cm

22    Allis tissue forceps 5 x 6 T 15cm

23    Somer uterine grasping forceps 23cm

24    Doyen myoma instrument

25    Schubert‟s cervical biopsy forceps

26    Wrigley‟s forceps 28cm

27    Deaver‟s retractor 75mm 30cm

28    Deaver‟s retractor 25mm 30cm

29    Haigh Ferguson obstetric forceps

30    Tuboplasty set

31    Dressing forceps standard 1x2 T STR 20cm

32    Tissue forceps 1x2 T STR 25cm

33    Tissue forceps 1x2 T STR 20cm

34    Mosquito artery forceps STR

35    Mosquito artery forceps CVD

36    Micro mosquito artery forceps STR

37    Micro mosquito artery forceps CVD

38    Dressing Drum Stainless Steel:
      Size 9x5, size9x9,11x5,11x9.5,14x5,14x9.5,15x12

39.   Stainless Steel Container Should be side filter type & fitted with hinges:
      Size:600x300x260mm,600x300x210mm,600x300x160mm,460x300x260mm,
      460x300x210mm,460x300x160mm,300x300x260mm,300x300x210mm,
      300x300x160mm, Textile Filter
                                    Birthing Bed
Approx. overall size : 2070mmL x 850mmWx550mm to 850mmH with out mattress.
Three section top.     Backrest, Trendelenburg/ Reverse Trendelenburg (6 0 approx)
adjustment on screw mechanism. Height adjustment on hydraulic pump. Leg section can
be telescoped under the backrest for lithotomy position. Removable stainless steel head
and leg bows with laminated panels. Removable stainless steel side railings. Foot
supports for high/low chair positions. Middle and leg section top removable for cleaning.
Pair of upholstered aluminium lithotomy crutches mounted on stainless steel rods.
Stainless steel tray. Three section mattress. Four imported castors, 125mm dia., made
from non-rusting composite material, two with brake. All mild steel components should
be thoroughly pre-treated chemically to remove rust and foreign matter like Grease, Oil
etc. by dip tank process pre-treatment system. The treated Metal Surface should be
coated with Epoxy Polyester Powder and oven baked at 180 degree to 200 degree
Centigrade to avoid contamination of the clean metal surface from dust particles.



Alternative – As above with fully motorized bed with battery backup with built in battery
charge.
                                    Eclampsia Bed

Stretcher size : 1830mm Lx 555W (approx.). Trolley should be fitted of X-Ray
permeable removable two section stretcher, backrest raised on ratchet. Trolley should
have gas spring assisted trendelenburg / reverse trendelenburg positions. Trolley should
have height adjustment facility by motorized actuators with a height adjustment from
560mm x 810mm (approx.). Trolley should be fitted on four 125mm dia imported swivel
castors , two with brake. Trolley should be fitted with corner buffers, sliding x-ray
cassette, accessories tray, oxygen cylider holder, stainless steel telescopic saline rod and
with swing away stainless steel side rails. Quote trolley with sectional mattress. All mild
steel components should be thoroughly pre-treated chemically to remove rust and foreign
matter like Grease, Oil etc. by dip tank process pre-treatment system. The treated Metal
Surface should be coated with Epoxy Polyester Powder and oven baked at 180 degree to
200 degree Centigrade to avoid contamination of the clean metal surface from dust
particles.
                                 Post Operative Bed


Approx. Bed Frame Size 2050mm L x 910mmW.                   Height minimum 490mm and
maximum 710mm. Bed should have trendelenburg/ reverse trendelenburg facility. Bed
should consist of four section top of minimum thickness of 18G CRCA mild steel sheet
of which backrest and kneerest should be pivoted on the main frame and should be
adjustable on screw mechanism and second section should be fixed. Leg section should
be interconnected to knee section and it should be adjusted in relation to the knee section
with the help of ratchet mechanism. All four sections should be fabricated with round
tube and flat sections suitably reinforced. Top section should have uniformly perforated
holes for effective ventilation. Main frame of the bed should be of rectangular tubular
welded construction, fitted with aluminum brackets at each corner. Bows should consist
of outer tube in one piece double bend to achieve bow shape (inverted U-shape) and
should be fitted with laminated panels with the help of Stainless Steel (304 quality)
brackets. Chariot base frame of the bed should have rectangular tube construction.
Seprate screw mechanism should be used to achieve kneerest adjustment and
trendelenberg and reverse trendelenberg position. The bed should be fitted with 125mm
dia non rusting swivel castors. All mild steel components should be thoroughly pre-
treated chemically to remove rust and foreign matter like Grease, Oil etc. by dip tank
process pre-treatment system. The treated Metal Surface should be coated with Epoxy
Polyester Powder and oven baked at 180 degree to 200 degree Centigrade to avoid
contamination of the clean metal surface from dust particles.


               (1)     Side Railings
               (2)     Stainless Steel Heavy duty I.V.Rod
               (3)     Mattress (Sectional)
               (4)     Urine Bag Holder
               (5)     Chart Holder


Alternatively : Quote fully motorized bed with rust free castors.
                                     Patient Bed

(A)    Patient Bed – Two Section


Aprrox. Bed Frame size 1980mm L x 910mmW.                  Height 600mm.     Two section
perforated top. Backrest adjustment by crank mechanism. Stainless Steel tubular head
and foot bows with laminated panels. Provision for I.V. Rod at four locations. All mild
steel components should be thoroughly pre-treated chemically to remove rust and foreign
matter like Grease, Oil etc. by dip tank process pre-treatment system. The treated Metal
Surface should be coated with Epoxy Polyester Powder and oven baked at 180 degree to
200 degree Centigrade to avoid contamination of the clean metal surface from dust
particles. Bed should be quoted with sectional mattress and following accessories.


               (1)    Stainless Steel telescopic I.V.Rod
               (2)    Sectional Mattress
               (3)    Side Railings

(B)    Patient Bed – Four Section

Approx. Bed Frame size 1980mm L x 910mmW.                  Height 600mm.    Four section
perforated top. Two separate screws for backrest and kneerest ad justments. Stainless
Steel tubular head and foot bows with laminated panels.     Provision for I.V. Rod at four
locations. All mild steel components should be thoroughly pre-treated chemically to
remove rust and foreign matter like Grease, Oil etc. by dip tank process pre-treatment
system. The treated Metal Surface should be coated with Epoxy Polyester Powder and
oven baked at 180 degree to 200 degree Centigrade to avoid contamination of the clean
metal surface from dust particles.    Bed should be quoted with following optional
accessories.


               (1)    Stainless Steel I.V. Rod
               (2)    Sectional Mattress
               (3)    Side Railings
                                        O.T. Table
   i.   Height, Trendelenburg / Reverse Trendelenburg and lateral tilt operated
        electrically with handset control.
  ii.   Overall approx. size without accessories : 1830mm L x 560mm W.
 iii.   Height adjustment on electromechanical actuator from 760mm to 960mm.
        (approx)
 iv.    Trendelenburg / Reverse Trendelburg 25 o and lateral tilt 15o on elctromechanical
        actuator.
  v.    X-Ray permeable laminated board top in three sections with easily removable leg
        section.
 vi.    Stepless Backrest adjustment 65o Up/25o Down on gas spring.
vii.    Heavy Duty base covered with impact resistant, non rusting, moulded top.
viii.   Fitted with Non-rusting castor wheels with provision to brake the tab le.
 ix.    Table is fitted with rechargeable batteries.
  x.    Should be supplied with Three section foam pad, Armrest, Lithotomy crutches,
        Anesthesia Screen, Body strap.
 xi.    All mild steel components should be thoroughly pre-treated chemically to remove
        rust and foreign matter like Grease, Oil etc. by dip tank process pre-treatment
        system.     The treated Metal Surface should be coated with Epoxy Polyester
        Powder and oven baked at 180 degree to 200 degree Centigrade to avoid
        contamination of the clean metal surface from dust particles.
                                      ICU BED

Approx. Bed Frame Size 2050mm L x 910mmW.                  Height minimum 490mm and
maximum 710mm. Bed should have trendelenburg / reverse trendelenburg facility with
the help of screw mechanism.. Bed should consist of four section top of minimum
thickness of 18G CRCA mild steel sheet of which backrest and kneerest should be
pivoted on the main frame and should be adjustable with the help of screw mechanism.
Backrest should be adjustable on screw mechanism and second section should be fixed.
Leg section should be interconnected to knee section and it should be adjusted in relation
to the knee section with the help of ratchet mechanism. All four sections should be
fabricated with round tube and flat sections suitably reinforced. Top section should have
uniformly perforated holes for effective ventilation. Main frame of the bed should be of
rectangular tubular welded construction, fitted with aluminium brackets at each corner.
Stainless Steel bows should consist of outer tube in one piece double bend to achieve
bow shape (inverted U-shape) and should be fitted with laminated panels with the help of
Stainless Steel (304 quality) brackets. Chariot base frame of the bed should have
rectangular tube construction. The bed should be fitted with 125mm dia non rusting
swivel castors. All mild steel components should be thoroughly pre-treated chemically
to remove rust and foreign matter like Grease, Oil etc. by dip tank process pre-treatment
system. The treated Metal Surface should be coated with Epoxy Polyester Powder and
oven baked at 180 degree to 200 degree Centigrade to avoid contamination of the clean
metal surface from dust particles.     Bed should be quoted with following optional
accessories.


               (1)    Full Length Side Railings
               (2)    X-Ray Permeable Backrest
               (3)    Stainless Steel Heavy duty I.V.Rod
               (4)    Mattress (Sectional)
                                  Patient Trolley

Approx. Size 2030mm L x 560mm W x 820mm H. Mild steel tubular frame work
mounted on four swivel castors of 200mm dia. Removable stretcher top made of
aluminium sheet supported on tubular frame with mattress of 2”.         All mild steel
components should be thoroughly pre-treated chemically to remove rust and foreign
matter like Grease, Oil etc. by dip tank process pre-treatment system. The treated Metal
Surface should be coated with Epoxy Polyester Powder and oven baked at 180 degree to
200 degree Centigrade to avoid contamination of the clean metal surface from dust
particles.
                                     Wheel Chair

Approx. Overall size : 670mm W x 1120mm D x 920mm H. Welded frame construction
of round tubes. Two solid rubber tyred bicycle wheels with brakes & self propelling
stainless steel hoops. Two swivel castors, approx. 200mm dia infront. Minimum Frame
size of round steel 22.2 x 18 G tubes and 19.05 x 18 G tubes. Mild steel tubular
construction fitted with cushion seat and back. Wheel chair is fitted with 26” dia rim of
bicycle wheel fitted on specially developed and heat treated axle with solid tyre in the
rear.   In the front 200mm dia castor wheels are fitted.     In front of castor wheels,
aluminium foot paddles are provided on adjustable brackets. Two handles are provided
with the hand grips. Brakes are provided on rear wheel to hold the chair to stop on 5
degree ramp. All mild steel components should be thoroughly pre-treated chemically to
remove rust and foreign matter like Grease, Oil etc. by dip tank process pre-treatment
system. The treated Metal Surface should be coated with Epoxy Polyester Powder and
oven baked at 180 degree to 200 degree Centigrade to avoid contamination of the clean
metal surface from dust particles.
                               Bedside Patient Cabinet

Approx. Length - 400mm. Approx. Width - 400mm. Approx. Height -820mm. Body
consisting of two sides and back should be made from one piece. Top should be fitted
with stainless steel sheet with raised edges on three sides.    One drawer should be
provided below the top. Drawer should be followed by open storage space which should
be followed by closed door cabinet. Door of the cabinet box should be pivoted at top and
bottom. Base of the locker should be fitted with non rusting castors of minimum 50mm
dia. All mild steel components should be thoroughly pre-treated chemically to remove
rust and foreign matter like Grease, Oil etc. by dip tank process pre-treatment system.
The treated Metal Surface should be coated with Epoxy Polyester Powder and oven
baked at 180 degree to 200 degree Centigrade to avoid contamination of the clean metal
surface from dust particles.
                               Patient's Cardiac Table
Approx. Size : Top 810mm L x 352mm W . Top should be laminated in two sections.
Laminated top section should be fitted on mild steel square tubular telescopic stem with
geared S.S. handle for height adjustment from approx. 760mm to 1050mm. Base frame
should be of mild steel rectangular tubular base frame mounted on four castors of 50mm
dia. All mild steel components should be thoroughly pre-treated chemically to remove
rust and foreign matter like Grease, Oil etc. by dip tank process pre-treatment system.
The treated Metal Surface should be coated with Epoxy Polyester Powder and oven
baked at 180 degree to 200 degree Centigrade to avoid contamination of the clean metal
surface from dust particles.
                                      Foot Step
                              (Single Step/ Double Step)

Approx. Height : 230mm. Approx. Step size 510mmL x 300mmW. Mild Steel tubular
frame fitted with aluminium tread flats. Legs fitted with rubber feet. All mild steel
components should be thoroughly pre-treated chemically to remove rust and foreign
matter like Grease, Oil etc. by dip tank process pre-treatment system. The treated Metal
Surface should be coated with Epoxy Polyester Powder and oven baked at 180 degree to
200 degree Centigrade to avoid contamination of the clean metal surface from dust
particles.
                                   Mayo’s Trolley

Adjustable height of Stainless Steel tray from 760mm to 1270mm (approx). Tray size
560mmL x 400mmW. Stainless Steel tubular frame mounted on four swivel, twin wheel
non rusting castors, 50mm dia. All stainless steel components should be of 304 quality.
                                   Revolving Stool

Revolving Stool should have tubular tripod base with stainless steel revolving top.
Minimum height adjustable of 480mm to 670mm by screw mechanism. Legs should be
fitted with rubber feet. All stainless steel used should be of 304 grade quality. All mild
steel components should be thoroughly pre-treated chemically to remove rust and foreign
matter like Grease, Oil etc. by dip tank process pre-treatment system. The treated Metal
Surface should be coated with Epoxy Polyester Powder and oven baked at 180 degree to
200 degree Centigrade to avoid contamination of the clean metal surface from dust
particles.
                               Soiled Linen Trolley


Approx. Size : 910MM H X 510MM dia. Trolley should be made of Stainless Steel 304
quality tubular frame round in shape. The trolley should be fitted with 100mm dia rolling
castors for free movements.    Trolley should be supplied with removable high quality
washable two canvas bag.
                                Examination Couch

Approx. Size 1890mm L x 560mm W x 840mm H. Couch should have gas spring
assisted adjustable backrest with upholstered top. Upper section with three drawers.
Lower section has cabinet with three separate doors, B.P. holder tray near headrest. Body
of the couch should be made of CRCA sheets. All mild steel compone nts should be
thoroughly pre-treated chemically to remove rust and foreign matter like Grease, Oil etc.
by dip tank process pre-treatment system. The treated Metal Surface should be coated
with Epoxy Polyester Powder and oven baked at 180 degree to 200 de gree Centigrade to
avoid contamination of the clean metal surface from dust particles.


Please quote as above with „U‟ cut with lithotomy rods.
                                      Saline Stand

Saline stand should have strong Mild Steel tubular construction mounted on five pronged
base fitted with non rusting swivel castors of 50mm dia. Stainless steel saline rod should
have double hooks. Stainless steel I.V. rod should be height adjustable from minimum
1620mm to 2340mm.         All mild steel components should be thoroughly pre-treated
chemically to remove rust and foreign matter like Grease, Oil etc. by dip tank process
pre-treatment system. The treated Metal Surface should be coated with Epoxy Polyester
Powder and oven baked at 180 degree to 200 degree Centigrade to avoid contamination
of the clean metal surface from dust particles.
                  Bio Medical Waste Segregation Container

Biomedical waste segregation container made of M.S. sheet of 18 gauge, paddle lifter lid,
drum capacity of 25 liter, 40 liter and 50 liter power coated with hazardous symbol
complete. As per CPCB norms. Inside with Rubber coating. Each set consist of Red,
Yellow & Blue containers

Optional:
   - Frame to hold the 3 bins firmly.
   - Frame to hold the 3 bins firmly with four wheels
   - Waste collection bags min. 125 micron thick, size for 25 liter capacity container,
       set of Red, Yellow and Blue color as per Environmental protection Act
       specification
                             Plasma TV

Plasma Display Panel     :   Drive Method – AC Type, 42”, 16:9 Aspect Ratio

Contrast Ratio           :   4000:1

Screen Size              :   920 X 518mm 42”(1,056mm)

No. of Pixels            :   408.960 (852[W]X 480[H])

Audio Output             :   16 W (8W+8W)

Operating Conditions     :   Temperature 00 C-400 C;
                             Humidity:20%-80% (Non Condensation)

Color System             :   NTSC, PAL, PAL60, SECAM, Modified NTSC

Terminals AV Slot Type   :   Video In/Out (BNC):1.0Vp-p
                             (75 ohms or High-Impedance)
                             S-Video (Mini DIN 4-pin):Y:1.0 Vp-p (75 ohms)
                                     C:0.286 Vp-p (75 ohms)
                             Audio In L/R (RCA Pin Jack X 2):0.5 Vrms (High
impedance)

PC                       :   Mini D-Sub 15 pin
                             fH 15-110 Khz:48-120 kHz
                             R.G.B. : 0.7 Vp-p (75 ohms)
                             Y:1.0 Vp-p (75 ohms; including sync)
                             PB,PR/CB,CR:± 0.7 Vp-p(75 ohms); HD,VD/1.0-
                             5.0 Vp-p (High impedance)
                             Audio In (M3 jack): 0.5 Vrms (High impedance)

Service Life             :   60000 hrs or higher

Additional Feature       :   Remote system monitoring
                             Fan less operation- for quiet performance
                             3d color management system this finer
                             Level of control produces more expensive images
                             Dual picture mode, 4x digital zoom, multi function
                             Slots for future expendability of inputs
                      Motorised Remote Control Screen

Electrically Operated 220V AC Single Phase Motor Mounted inside Roller
Chamber
Internal Thermal Overload Projection
Preset Limit Switch
Automatically Stop in UP-Down Position
Torque Limit Control to Protect Motor from Damage
Dual Remote Control

         1.    Cordless Remote Control – for Up, Down & Stop
         2.    Wired Control Switch for Up, Down & Stop
Size 72 X 96 inches.
                      Overhead Projector

Should be Portable
High Grade Vari-Focus Three Element Projection Lens.
Fersnal Lens DIN A4 size (285X285mm) Hard Coated and Scratch Resistance
Twin Halogen Lamp 24V/250W with Instant Lamp Change.
Power Supply 240V AC
Carry Case
                                   Portable X-Ray

Allengers-100           100mA-100KVP, Full Wave Rectified X-Ray Generator with
                        Stationary Anode X-Ray Tube as per ISI specification
Output                  8KW as per ISI (IS:7620-Part-I)
Rating- Radiography     Above 75mA to 100mA
KVP Range               45 – 100 KVP in steps of 5 KVP step
Timer                   Solid State Electronic Timer with timing range from 0.04 to 8 Sec. In
                        24 steps
Control                 Attractive and Ergonomic designed control panel has digital display
                        of mAs, KVP and Radiographic mA. Independent Voltmeter to
                        indicate line voltage. mA meter, Voltage Compensator fine and
                        coarse to compensate Voltage is provided. Booster transformer is
                        provided in control panel for stabilization of filament voltage. Tech.
                        Selector for selection of Rad mA, KVP selector from 45 to 100 &
                        Time Selector switch for selection of Radiographic time. Ready & X-
                        Ray switches, Bucky selector Switch provided. Machine is made
                        ON/OFF with push to ON/OFF Type switches.
Overload Protection &   Automatic trip off overload circuit breaker. Overload, X-Ray & line
safety Features         indicators provided. Electronic overload protection with simultaneous
                        protection from high input voltage/KVP/mA/Time provided. Fuses
                        for all circuits provided in the machine
Tube Head               Self-contained Tube Head containing Stationary anode X-Ray Tube,
                        Compact Heavy Duty Full Wave Rectified H.T. Transformer,
                        Filament transformer High Voltage Silicon Rectifiers etc. all oil
                        immersed and hermetically sealed Tube head can be tilted 360 degree
                        with full flexibility for use in operation theatres and in ward for
                        bedside radiography
Accessories              a) Manual Light Beam-cum-shutter diaphragm
                         b) Hand Switch with retractable cord
                         c) Aluminum filter
                         d) Lead Apron
Stand (CM)               Mobile Stand with 4-wheel design, which ensures easy mobility and
                         steering. Counter balanced articulated tube carriage allowing smooth
                         movements of the Tube Head in vertical and Horizontal plane, Lead
                         lined cassette storage box. Double steel rope for counter balancing
                         gives additional safety as per ISI-7620. Large castor nylon wheels
                         for easy mobility. The stand is designed for maximum
                         manoeuvrability of the unit.
Power Requirement        10 KVA, 440 Volts, 3 Phase, AC supply, Max. Allowable Line
Portable Grid            Regulation + 10%

  Should be passed by BARC & Radiation regulatory bodies
  Spare parts availability
                                    ECG Machine
1. Single channel ECG Machine
2. Lead I, II, III, aVR, aVF or V
3. Real time acquisition and printing of selected lead
4. High fidelity thermal array recording
5. Sharp & clear trace
6. Electronic lead selection
7. AC mains/built- in rechargeable battery
8. Machine should be supplied with patient cable, clipon electrodes, recording paper,
   chest electrode (Adult & Pead.) with rubber bulb, jelly & instruction manual warranty
   of machine – 12 months.
      I.V.F. Lab along with all instruments & etc. on turn key basis
                                (S.No.1(a) to 14)

     1.(a)   LAMINAR FLOW (IVF WORK STATION WITH ACCESSORIES)

IVF Workstation for Oocyte denudation and embryo culture. Fully heated ( at 37º
C)stainless steel table. Temperature control achieved by dual Sensor controlled
Digital Microprocessor based Temperature regulator. Unit should also have
stand by control system with lock and key. Workstation should produce class
100 air space and should have differential pressure indicator. Noise level should
be less than 63 db. The workstation should have built in monitor, image grabbing,
storing and copying device connected to microscope with CCD Camera. Unit
should have metal framed dust cover operated by gas springs.

IVF Workstation should have CO2 Incubator built in to unit ( approx. 14 Ltrs
capacity) for temporary storage of gametes and embryos. Unit should have
Falcon 14 ml Test Tube heater, heating independently with it’s own rechargeable
battery. Such four test tube heaters should be housed in base station. Base
station should have definite place on workstation with it’s own electrical plug.
Test Tube heater should have indication of Temp. status and battery status.
Should also have slot for viewing liquid level. The table width should be minimum
1200 mm

Accessories.

   1. CO2 incubator approx 14 Ltrs capacity
   2. TFT Monitor with Image grabbing, copying and storing device.
   3. Standby Temperature control system
   4. Differential pressure indicator
   5. UV Light
   6. Portable Mini Incubator
   7. Warming Block for 4 well Nunc dish 2 Nos
   8. Warming Block for 60 mm culture dish 2 Nos
   9. Warming Block Test Tube 14 cc
   10. Warming Block Test Tube 6 cc
   11. “C” Mount camera adapter
   12. Glass hoodØ 130 mm h 100 mm – 2 Nos
   13. Extra prefilter – 2 Nos
   14. Base station for rechargeable Test Tube heaters with LCD display
   15. Portable Test Tubes ( 14 ml) heaters with rechargeable battery, heater,
       sensor,and LED indicators for battery and Temperature.- 4 Nos
1.(b) LAMINAR LOW (ANDROLOGY WORK STATION WITH ACCESSORIES)

Andrology Workstation for sperm screening , preparation & analysis.. Fully
heated ( at 37º C)stainless steel table Temperature control achieved by dual
Sensor controlled Digital Microprocessor based Temperature regulator. Unit
should also have stand by control system with lock and key. Workstation should
produce class 100 air space and should have differential pressure indicator.
Noise level should be less than 63 db. The workstation should have built in
monitor, image grabbing, storing and copying device connectable to microscope
with CCD Camera. Unit should have metal framed dust cover operated by gas
springs.

Andrology Workstation should have Falcon 14 ml Test Tube heater, heating
independently with it’s own rechargeable battery. Such four test tube heaters
should be housed in base station. Base station should have definite place on
workstation with it’s own electrical plug. Test Tube heater should have indication
of Temp. status and battery status. Should also have slot for viewing liquid level.
The table width should be minimum 1200 mm

Accessories.

   1. TFT Monitor with Image grabbing, copying and storing device.
   2. Standby Temperature control system
   3. Differential pressure indicator
   4. UV Light
   5. Portable Mini Incubator
   6. Warming Block Test Tube 14 cc
   7. Warming Block Test Tube 6 cc
   8. Extra prefilter – 2 Nos
   9. Base station for rechargeable Test Tube heaters with LCD display
   10. Portable Test Tubes ( 14 ml) heaters with rechargeable battery, heater,
       sensor,and LED indicators for battery and Temperature.- 4 Nos
2. CO2 Incubator
CO2 Incubator should have minimum 135 Ltrs usable space ( volume) Incubator
should have seamless chamber with direct heating system and high temperature
decontamination at 121º C . Incubator should have infrared sensor with fully
automatic auto zero system and non removable type during decontamination cycle to
avoid accidental damages and to maintain decontamination integrity. 72 hour data
storage should be logged for all conditions and events every 15 seconds and viewable
on screen in a simple real time graphical manner. Humidity Tray should be directly
heated. Minimum four shelves and eight doors for the incubator. Incubator should
have rapid recovery system to enhance temperature recovery time of embryos or
gammets.


Technical Specifications


Capacity: Minimum 135 Ltrs usable volume
CO2 range: 0.2% TO 2.0 %
CO2 Control :± 0.1 %
Temperature range: 2º C above ambient to 50º
Temperature Control: ± 0.1º C
Shelf – Non perforated 4 Nos
Oxygen control: 1-19%


Accessories


1.   Key lock on outer door
2.   Centronics port
3.   Inline pressure regulator
4.   Rapid Recovery system for 4 well nunc dishes withi 4 dish holders
5.   Rapid Recovery System for 60 mm Falcon petridishes with 4 dish holders
6.   Digital infrared CO2 Analyser.
3. Centrifuge (Temperature Regulated)

The Centrifuge with swing at rotor head for processing semen by swim up or
density gradient technique centrifuge accepts 8 tubes of 15 ml tube, will have G
Force indication, rpm indication, and timer indication. Centrifuge chamber to be
heated at 37º C to maintain sperms at the same temperature throughout.
Centrifuge should have nine different programmes each one should be
independently set for Temperature, Time, G Force and or RPM. Centrifuge
should have brushless AC Motor. Centrifuge should have on screen tube
selection Centrifuge should have programmed acceleration and deceleration.
4. CRYOCAN :

Liquid Nitrogen Containers, made up of High Vacuum Cryogenic container of
Aluminum and with canister.

   1. CAPACITY - 26 Litres,
   2. CAPACITY - 11 Litres,
5. MAKLER COUNTING CHAMBER

The Makler Counting Chamber should be 10 microns deep. 1/10 th of thedepth of
ordinary hemocytometers, It should be constructed from two pieces of optically
flat glass, the upper layer serving as a cover glass, with a 1 sq mm fine grid in
the center subdivided into 100 squares of 0.1 x 0.1 mm each. Spacing should be
firmly secured by four quartz pins.
6. Inverted Microscope, M icromanipulator (ICSI) With Camera,
   Software and CASA System & Other Accessories

This should have Inverted Microscope with on off light, sextuple revolving
nosepiece, magnification changer 1X and 1.6x and 2 step light path selector and
hexagonal driver. It should also have illuminator pillar (Tiltable) with condenser
holder and 4 filter holder including diffusion filter. Should have Lamp house
for100 halogen including long connection cord. Inverted Microscope should be
supplied with 2 way mechanical stage with flexible right hand vertical low drive
controls including stage insert plate. Inverted Microscope should have long
distance work condenser with 4 positive turret, Hoffman modulation system with
20x and 40 x objectives.

Work Station should have Oil hydraulic Joystic set consisting of 2 each of drop
handle oil hydraulic micromanipulators, three axis motorized coarse positioners,
universal joint units and adapters. The unit should also have Microinjector with
metal syringe including glass syringe, injection holder for holding side and
injecting side.

The work station should have glass heating stage with high resolution CCD
Camera, TFT Monitor 17” with image grabber, copying and storing device.

The work station should have Surgical Lazer Shot System with Laser direct via
live video display with crosshair measurement gridoverlay direct calibration and
direct interactive measurement possible

The unit should be supplied with Digital USB Camera. The imaging software
should facilitate digital video recorder, net work broadcast functioning, time lapse
recording and power point slide show. The entire equipment should be housed
in antidust cabinet having positive sealing and openable door with grass springs.
The entire unit should be mounted with ant vibration pad to minimize vibration.

TECHNICAL SPECIFICATIONS
      Coaxial coarse/fine knobs:tension adjustment on the right side
       Fine focus knob graduated
       Stage movement (XY direction) on rack and pinion
      Quadruple revolving noseouece
      Plane stage 120 x 132 nn
      With right hand mechanical stage
      Abbe condensor N.a. 1.25 (oil immersion), with aperture iris diaphragm.
      Blue filter
      Universal power supply
      Built-in 6V 20W illuminator
      Input voltage 100V-240V
      8cc immersion oil
      Dust cover
      Mirror unit (Plano-concave)
      Objectives: 10X,20X, 40X, 100X
      Eyepiece: iCWHK10X(LB eyepiece 10X), F.N. 18mm, (anti fungus)
       Phase contrast attachment
       The System should have image analysis for automatic generation of
        spermiograms and morphological results. Easy handling should be based
        on Windows user interface. The image should be grabbed from video
        camera which mounted on Microscope processing of Video tape is also be
        possible.

       Following should be possible with the CASA system.
   1. Automatic analysis of sperm motility and morphology. Calculation of a
       complete spermiogram in a few seconds, multiple recordings for each
       sample should be possible.
   2. Storage of patients data and sample recorded on template in
       MS Access and SQL-Server
   3. Classification of spermatozoa according to WHO categories
   4. Print of Spermiogram by means of user defined HTML templates
   5. Automatic export of data into MS Excel for further statistical analysis
   6. Storage fo images from morphological analysis
   7. Use of standard hardware to reduces investment need.
   8. Measured parameters should be spermato zoa head ( area, length and
       width in microns, roundness, acrosome area, midpiece area and head
       asymmetry.
   9. Measured parameters in motility should be:
            Number of spermatozoa
            Concentration of Spermatozoa
            VCL curvilinear Velocity
            VSL straight-line velocity
            VAP average path velocity
            ALH amplitude of lateral head displacement
            LIN linearity
            WOB wobble
            BCF Beat cross frequency
The hardware of the system should consist of Pentium IV PC, Windows XP, high
resolution Video Camera, image processing board, phototube and C-mount
adapter , objectives for motility analysis and morphological analysis and
standard Makler chamber.

SCOPE OF SUPPLY
   1.  Inverted Microscope
   2.  Hoffman Modulation system
   3.  Oil hydraulic moterised Micromanipulators
   4.  Injectors
   5.  Glass Heating stage
   6.  High Resolution USB digital camera
   7.  TFT Monitor 17”
   8.  Image grabbing, storing & copying device
   9.  Antidust cabinet with metal framed cover operated by Gas spring with
       Antivibration pads.
   10. Laser shot System with advanced software with digital video recording,
       network broadcast, power point slide, function & time lapse recording
7. Lab. Refrigerator (Drug Refrigerator) – Capacity 150 Ltrs to 180 Ltrs

It should be constructed of double walls, the exterior should be made of sheet
steel while the stainless steel interior should be provided with a number of
compartments having transparent plexiglass doors. The unit should be fitted
with castor wheels for ease of mobility. Internal cabinet lamp should be
illuminated whenever the door is opened. The cabinets should be provided with
roll out type tray for proper storage. It should have a frost free performance
thereby avoiding periodical defrosting.


The controls should comprise of fully automatic digital temperature indicator cum
controller. On/Off switches, indicating lamps and protective fuse. It should also
have incorporated electronic safety circuit which gives audio-visual alarm in case
of either power failure, temperature variation beyond ± 2º C of the set
temperature of 4º C or door open, thereby protecting the precious contents
stored inside the cabinet.
The Drug Storage Cabinets, should incorporate a built in temperature recorder
and servo controlled voltage stabilizer.
8. OVUM ASPIRATION PUMP

The Ovum Aspiration Pump should provide smooth low volume vacuum at pre-
determined negative pressures, allowing simple, low turbulence, oocyte recovery.

    It should have Foot operated switch with Vacuum gauge and Regulation
     manufactured to BS 5724 Class 1 and CE Safety
     Regulation.
    Vacuum should be fully variable from 0-500mm Hg, & should have a High
     vacuum control which produces a pre-set300mm Hg.
    It should have a Fluid trap to prevent accident aspiration of fluids into the
     vacuum unit.
      Suction units It should feature super quiet diaphragm pumps for reliable,
         low maintenance service.


     TECHNICAL DATA
1.   Low Pressure                      : 0-550 mm Hg = 0-666 mbar.
2.   Volume                           : 2x100ml
3.   Connecting for Aspirating tubing : 2 to 4 mm
4.   Power Supply                      : 230 V
5.   Power Frequency                    : 50/60 Hz
6.   Power Point                       : 50 VA
7.   Nominal Current                  : 220mA
8.   Power Line Isolation             : 0.315 AT
9. DEEP FREEZER (Embryo Freezer)

The Embryo Freezer should meet the requirements of Human Embryo Transfer.
It should be a portable freezer with its own internal battery with a thorough put of
up to 100 Straws per cycle. It should be "Solid State" design without moving
components. There should be no solenoid Valves, pumps or motor to freeze.
The straw plate should be precisely controlled and the temp. should be balanced
against the cold liquid nitrogen chamber.

It's features should be as follows :

1. DISPLAY FUNCTIONS : When a Program is running the Screen Display
   should all the necessary Information to the User of Correct Performance.

2. STERILISING : It should have a temperature range of 150C to - 100C. It
   should have a sterilizing program Designed to hold the Straw plate temp. at
   125C for 30 Minutes.

3. PROGRAMS : It should store up to eight programs each containing up to
   Eight Lines. Individual Programs should be password protected to prevent
   unauthorized editing or use. Program options should be for temperature rate
   from 0.1C to 5.0C per minute, timed hold from 1 to 99 minutes, indefinite
   Pause, freefall (max. possible range of temp. change) and program end.

4. OPERATING SECURITY : To ensure best possible embryo survival, the unit
   should hzve various safety features. If the Liquid Nitrogen Reservoir is filled,
   the freezer should be automatically switched on, preventing Unnecessary
   cooling of the straw chamber. A low liquid nitrogen warning system should
   operates. The freezer should not be switched off when running a program
   until the program has finished, or been stopped & the straws removed.

5. PORTABILITY : It should be designed with portable use in mind. The
   controlled rate freezer should  Operate on Power Supplies from 80 to 260 V
   AC or on 12 V DC including its own internal, rechargeable battery. The
   Battery should be run freezing programs up to three hours long without
   need of recharging the handle in the centre of the freezer and the freezer
   should be balanced to allow easy carrying with Only one hand.

   6. RECORD KEEPING : It should include automatic data logging of all freezer
   runs. At the end of each run,A list of the last eight freeze runs should be
   presented with the last one at the top. The user should have the choice of
   printing details and a Graph or sending the log file to a Computer or Terminal.
10.GLASSWARE & DISPOSABLES


 1      “TARSONS “
        DISPOSABLE NONTOXIC PRESTERILISED SEMEN COLLECTION JARS
        TARSONS PRODUCTS
 2      4 WELL DISH
        “Nunc” Cat.No.144444
 3      PAPER NAME STICKERS
 4      FREEZING STRAWS (SUPREME, EMBRYO)
        “Cryo-Bio-System”
        Cat No.00657
 5      ALCOHOL (ETHANOL) 500 ML
 6      TEST TUBE STAND
 7      PARAFILM ROLL
 8      IVF TEST TUBES POLYSTRELE
        Falcon/“Nunc”145470 –7 ml round bottom sterile tube with cap
 9      IVF TEST TUBES POLYSTRELE
        Flacon/“Nunc” 366060 15 ml conical sterile tube with Cap
 10     IVF SEROLOGICAL PIPETTES 1 ML
        Falcon/“Nunc” 159609- 1 ml serological sterile pipette
 11     IVF SEROLOGIAL PIPETTE, 5 ML
        Falcon / “Nunc” 159625 5ml serological sterile pipette
 12     IVF GRADE PETRIDISH 35 MM
        Falcon /“Nunc” 150255 35 x 10 mm sterile petridish
 13     IVF GRADE PETRIDISH 60 ML
        Falcon/“Nunc” 150270 60x15 mm sterile P petridish with lid
 14     IVF GRADE PETRIDISH FOUR WELL DISH
        Falcon /“Nunc” Cat.No.144444 POLYSTERENE STERILE 4 WELL DISH
        WITH LID
 15     OVUM PICK UP NEEDLE & OVUM PICKS UP SET (without valve) Tube 60
        cms Needle 30 cms Gauge 17
        Reproline Medical GmbH-Germany 461230 LF
 16     EMBRYO TRANSFER CATHETER
        “CCD” – Frydman Catheter length 4.5cm. Cat.No.1306045
 17     ICSI PIPETTE INJECTING PIPETTE 35 DEGREE ANGLE
        Reproline Medical GmbH –Germany Cat No.441535
 18     ICSI PIPETTE RI HOLDING PIPETTE 35 degree angel
        Reproline Medical GmbH – Germany Cat No.440735
 19     DENUDING NEEDLES STRIPPER TIPS 150 u
        “Mid Atlantic” Cat.MlX3/IND-150 (Pack of 20)
 20     Denuding needles stripper tips 135 u
        “Mid Atlantic”MLX3-135
 21     DENUDING NEEDLES
        STRIPPER TIPS 175 U
        “Mid Atlantic” MLX3-175
 22     EMBRYO REDUCTION NEEDLE
        “CCD” LU-LOCK NEEDLE
        CAT NO.1403002

        CCD INTERNATIONAL
 23     TARSONS CRYO VIALS FOR EMBRYO FREEZING
     “TARSONS PRODUCTS”
24   CRYO STRAWS FOR EMBRYO FREEZING
     “Cryo-Bio-System”
     Cat No.006578
25   MOUTH ASPIRATION TUBING SET
     Reproline Medical GmbH - Germany – Cat No.441291
26   INSEMINATION CANNULA
     “Shivani Scientific:
     CRYI OREO-SET ( FOR iui) WITH ACCESSORIES FOR IUI ( 2 TUBES, 2
27   PIPETTES, 1 CONTAINER, IUI CANNULA)
     “Surginova” IUI Accessories Set
28   Stripper
29   Labexpert
11. AHU System

       AREA :- 446 SQ.FT
                - 9 FT
                       - 30 ACPH
            - 3500
          - 45± 5
       TEMP :- 23 ± 2

   DETAIL

   1) AIR HANDLING UNIT
     
          – Filter section with filter with VCD
      Mixing Chamber
      Coil Section with 4 row DX coils
     
     
     

       2) CONDENSINT UNIT
       Supply of air cooled condensing unit of absolute capacity given comprising of
       reciprocating compressor complete with 10mt cu piping with nitrile rubber
       insulation of 19mm thick suction line ,M.S stand, expansion valve,dryer,refrigerant
       gas etc

       3) G.I Ducting with 24 swg and 22 s wg
       4) Thermal insulation with 50mm s upply and 25mm return
       5) Electrical panel for AHU to condensing unit
       6) Volume control Dampe r for fresh air

       7) Humidistat for RH




12. Drying Ove n
13. Ph Meter
   Portable pH meter with LCD graphic display screen with Multipoint (1-5) Calibration
   for more linearity.

      pH range - : -2 to +16 pH
      pH resolution : selectable from 0.01 to 0.1 pH
      mV range : ± 2000 mV
      Features mV calibration for accurate redox Measurements.
      Programmable recorder output
      Mains & rechargeable battery operation
      Two way communication with any computer using RS232
14. Wate r Purification System (R.O)
   1.
Sl No. Specifications

   1    Reverse osmosis system

   2    Source of Water should be potable

   3    3 liter purified water per hour

   4    >5M W.Cm Resistivity

   5    <0.2mS/Cm Conductivity

   6    TOC Level Should be < 30ppb,

   7    Water Recovery Up to 15 %

   8    Product water should meets or Exceed Type II water Quality as defined by CAP, NCCLS and
        ISO 3696/ BS 3997.

   9    The System should be supplied with Pre Filtration Kit & Water Softener etc

  10    For all equipments Accessories, spares and available consumables of the equipment should be
        given.

  11    Equipments should be supplied with suitable UPS with one-hour backup

   2.

Sl No. Specifications

  1     Source of Water should be RO feed

  2     Product Flow Rate 1.5 lt/min

  3     Water Resistivity18.2 M Ω .cm (at 250 C)

  4     TOC Level 5 to 10 ppb

  5     Bacteria CFU/ml <1

  6     Particulates >.22µm (p/ml) <1

  7     Product Water should Be Pyrogen free, Nuclease free.

  8      The system should be supplied with Storage Tank of 50 Liter Capacity & other Essential
        Accessories.
                                   Class Room Chair

A.

A Chair having appropriate foldable writing pad approx height 31.5” Ground seat height
19” back 14” depth of seat 17.5” width of seat 19.5” aver all width23.5” over all depth
22”(+100mm) tapestry with cloth or letherite of approved colour The frame of chair
made of 14 to 16 SWG ERW CRC tube elliptical and round duly copper plaited and then
chromium plated of 10 micron (minimum) inter joint locking. Ply made of BWP hot
pressed ergonomically contrued back made of plastic molded, high density foam and
cushioned seat. Having PU arm or unarmed.
Sample must be approved before the bulk supply.

B.

Single ply frame Ply made of BWP hot pressed agronomically contrued , chair made of
elliptical and round 1.1/2” pipe ERW CRC tube of 14 to 16 SWG 3” high density foam
apostered with letherite The frame of chair is copper plaited and then chromium plated of
10 micron (minimum) high density foam and cushioned seat of 42 density apostered with
fabric and letherite of Rs. 350/M dimensions approx all over height 35” ground seat
height 17.5” seat to back height21.5” depth of seat 19.2” width of seat 21” over all width
26” over all depth 31” .
Sample must be approved before the bulk supply.

								