									  HARMANN PHARMACEUTICAL LABS (PVT) LTD LAHORE


SOP FOR DISINTEGRATION TESTING APPARATUS

 TAKE 1000 ml BEAKER AND FILL IT WITH REUIRED SOLVENT UPTO
  800 ml-900ml AND PLACE THEM IN WATER CHAMBER.
 2-POUR THE DISTILLED WATER IN THE WATER CHAMBER OF THE
  APPARATUS UPTO THE SPECIFIED LEVEL IT SHOULD NOT OVER
  FLOW AFTER PLACING BEAKER FILLED WITH REQUIRED SOLVENT.
 3-PUT THE POWER PLUG IN THE SOCKET AND SWITCH IT ON”
  AND THEM INSTRUMENT AND CONFIRM ON POSTION OF THE
  INSTRUMENT BY RED INDICATION DIGITS IN SCREEN.
 4-SWITCH ON THE HEATER AND CONFIRM ITS ON POSTION BY
  RED INDIGATION LITE.
 5- SWITCH ON THE STIR” FOR UNIFORM CIRCULATION OF WATER
  AND HEAT DISTRIBUTION IN OUTER CHAMBER.
 6- PLACE EMPTY BASKET IN BEAKER AND HANG ON THEIR
  RODS.SWITCH ON THE “OSCULATION.
 7- PLACE THE THERMOMETER INSIDE AFTER CERTAIN TIME
  INTERVALS.
 8- AFTER ONE TO TWO OF STEP 7 ON, THE HEATING SYSTEM
  THROUGH MEMBRANE PUSH ON TEMPERATURE OF 37°C ±0.5°C
  WILL BE ACHIEVED THROUGH BUILT IN MICRO PROCESSOR
  CONTROL.
 9- ADJUST THE „rpm” BY PRESSING +10 TO +1 PUSH BUTTONS AS
  REQUIRD
 10- LOAD REQUIRD TIME ON STOPWATCH WITH THE HELP OF
  PUSH BUTTON.
 11- PLACE THE TABLETS / CAPSULES IN BASKET AND CLIP WITH
  ROTATE BAR OR PUT THE TABLETS / CAPSULES IN VESSEL.
 12- LOWER THE UPPER PART OF APPARATOS UP TO STANDARD
  LIMIT, WITH THE HELP OF “↓”PUSH BUTTON.
 13- AT THT AND OF STEP 12 QUICKLY PRESS rpm ON AND TIME
  START ASSEMBLIES WILL ROTATE IN MEDIA AT ADJUSTED rpm TIME
  QUANTING WILL START.

								