					Techniques for the Modern Laboratory




                                 by Dr. Robert Richter


    A practical guide to . . .
    Clean Sample Preparation
      for Trace Metals Analysis
C   LEAN
    HEMISTRY
    Techniques for the Modern Laboratory




                                        by Dr. Robert Richter


           A practical guide to . . .
           Clean Sample Preparation
             for Trace Metal Analysis
© 2003 Milestone Inc.



All rights reserved. No part of this publication may be reproduced,
stored in a retrieval system, or transmitted, in any form or by any
means, electronic, mechanical, photocopying, recording, or otherwise,
without the prior written consent of the publisher.



Printed in the U.S.A.
Edition 1




Teflon®, Dacron®, and Tyvek® are registered trademarks
of E.I. DuPont Corporation.




                                Milestone Press
                                25 Controls Dr..
                                Shelton, CT 06484
                     CONTENTS
Section 1:
Controlling Contamination in Your Laboratory
   Chapter 1:
   The Analytical Blank                            13
   Chapter 2:
   Laboratory Environment                          15
   Chapter 3:
   Materials for Trace Analysis                    21
   Chapter 4:
   Trace Analysis Reagents                         33
   Chapter 5:
   The Analyst: A Source of Contamination          37


Section 2:
Advanced Sample Preparation Techniques
   Chapter 6:
   Closed-Vessel Microwave Digestion               43
   Chapter 7:
   Microwave Evaporation                           57
   Chapter 8:
   Improving Method Detection Limits               65

Section 3:
Preparing Your Laboratory for Trace Analysis
   Chapter 9:
   Laboratory Housekeeping Techniques              77
   Chapter 10:
   Equipment & Supplies for Trace Metal Analysis   81

Appendix A:
References                                         91
                             5
Clean Chemistry: Techniques for the Modern Laboratory




                         6
Introduction
Clean Chemistry: Techniques for the Modern Laboratory




                         8
O
         ver the past several decades, the need for trace and ultra-
         trace elemental analysis has increased. Current drinking
         water regulations require determinations in the low parts
per billion (ppb) range, while semiconductor applications require
parts per trillion (ppt) determinations. There is a growing aware-
ness that improved sample handling and preparation techniques
are required to meet these new standards, and the instruments
and experimental techniques used to achieve them.

   This book focuses on helping the analytical chemist under-
stand and adapt to the latest
trends in trace metal analysis.
The book is divided into three
parts. The first part of the book
focuses on the importance of
the analytical blank, along with          Introduction
techniques for its reduction and
control. The second part deals
with the preparation of the ho-
mogeneous sample solutions for
trace metal analysis. Part three
talks about preparing your lab for trace metal analysis—what
you need and where to get it.




                                 9
Clean Chemistry: Techniques for the Modern Laboratory




                         10
 l   SECTION ONE   l




  Controlling
Contamination in
Your Laboratory
Clean Chemistry: Techniques for the Modern Laboratory




                         12
T
     race elemental methods require the analyst to carry a spe-
     cial sample, known as an analytical blank, through all the
     steps of the analysis. This analytical blank is a measure of
all external sources of elemental contamination, and is used to
make a correction to the measured sample concentration. The
variability of the analytical blank, rather than its absolute value,
is what determines the accuracy of trace metals analysis.

    The overall uncertainty for an analytical measurement
is calculated using the following formula1:

     ������ �   ��������� � �������
    The variability of the analyti-                  Chapter 1
cal blank has little or no effect on
the accuracy of the result when                    The Analytical
the analyte concentration of the
sample is several orders of magni-
                                                       Blank
tude higher than the blank result.
For example, given a sample with
a measured lead concentration
of 500 ± 25 ng and a blank concentration of 10 ± 5 ng, the total
uncertainty would be:

                              ������ � ���� � ��

                                  ������ � ��

    The reported analytical result would be 490 ± 25 ng of lead.
In this example, the blank result has no effect on the accuracy of
the result.
                                       13
           Clean Chemistry: Techniques for the Modern Laboratory



    On the other hand, as the analyte concentration ap-
proaches the blank level, the accuracy of the result can
be compromised by the uncertainty of the analytical
blank measurement. For example, given a sample with
a measured lead concentration of 50 ± 2 ng and a blank
concentration of 10 ± 5 ng, the total uncertainty would be:

                          ������ � ��� � ��

                              ������ � �

    The reported analytical result would be 50 ± 5 ng of
lead. In this example, all of the uncertainty in the analysis
is due to the analytical blank.

    In order to improve the precision and accuracy of
trace metal determinations, steps must be taken to control
and reduce the analytical blank. The contributions to the
analytical blank come from four major sources:

         • Laboratory environment
         • Apparatus and containers used for analysis
         • Reagents
         • Analyst

    The rest of Section One discusses each of these
sources in detail, and offers some strategies for controlling
these contamination sources in your lab.




                                    14
A
        irborne contamination is a major contributor to
        the analytical blank. Dust particles are created in
        the laboratory from the slow weathering of solid
objects, or from chemical processes upon materials. The
main components of these dust particles are Ca, Si, Al,
Fe, Na, Mg, K, Tl, Cu, and Mn.2-4 If left isolated and undis-
turbed, dust particles will have little effect on trace metal
analysis. Unfortunately, this does not usually happen, and
the dust particles become airborne. Airborne particulate
concentrations can go from 0.2
x 106 in the morning to 1.5 x 106
by noon.5 Once airborne, the
heavier particles quickly settle
out, while the lighter ones are         Chapter 2
suspended and carried by the           Laboratory
air currents throughout the
laboratory. Airborne contami-          Environment
nates can also be introduced
to the laboratory from external
sources, through the ventila-
tion system. In Dortmund, Germany, analysts discovered
that their high iron blanks were caused because nearby
industrial plants dispersed 20 tons of ferric oxide, which
contaminated their ventilation system.6

    Particulate contamination is minimized by filtering
the laboratory air with a high efficiency particulate air
filter (HEPA). This filter was developed for the Manhat-
tan Project to remove fissionable particles from the air.7
HEPA filters have an efficiency of 99.97% for particles 0.3
                              15
            Clean Chemistry: Techniques for the Modern Laboratory



                                     µm and larger.            They effec-
                                     tively remove bacteria, pol-
                                     len, fly ash, and dust. Some
                                     common contaminates, such
                                     as tobacco smoke, are not
                                     removed because their parti-
                                     cles are smaller than 0.3 µm.2
                                     Particulate contamination in
                                     a work area is also minimized
                                     by using a laminar air flow
                                     system to provide directional
      HEPA air filter
                                     air flow. Directional air flow
from one end of the work area to the other essentially
creates an air curtain, preventing airborne contamination
from entering, or settling out.

   The best approach to controlling airborne contamina-
tion for trace metal analysis is to construct a clean labora-
tory, or “clean room.” This approach is expensive, and is
used in the production of computer, pharmaceutical, and
aerospace products, and in the production of standard
reference materials.         Clean rooms minimize metal-bear-
ing dust by filtering the air that enters the room through
HEPA filters, using specialized construction materials,
and operating under positive pressure so that the flow
of air is always out of the room. Clean rooms also utilize
antechambers to isolate them from the general laboratory
and provide an area for analysts to change into specialized
clean room attire. In a properly maintained and operat-
ed clean room, sub-parts-per-trillion measurements can be
                                     16
                      Ch. 2: Laboratory Equipment



made.    More information
on clean rooms and clean
room design can be found
in references 8–11.

    A more practical ap-
proach for trace metal
laboratories is the use
of laminar flow hoods.
These units provide areas
of HEPA filtered air for
sample handling, analysis,
and reagent storage. The
                                             Laminar flow hood
standard units are good
for drying containers and apparatus after cleaning, pro-
tecting samples in autosampler trays during analysis, and
preparation of calibration and quality control standards. A
standard unit should not be used to handle or store toxic
or hazardous materials, because all the fumes that are
generated are blown out at the analyst. Exhausted laminar
hoods are available for the handling of hazardous and
corrosive substances. These units are constructed from
nonmetallic components and the HEPA filtered air flows
vertically and is drawn though a perforated work area into
the laboratory ventilation system.

        Clean conditions can also be obtained with a
filtered air enclosure. These enclosures consist of a plexi-
glass box that is purged with filtered nitrogen or air to
remove air particles. The exhausted air is vented to a con-

                                  17
          Clean Chemistry: Techniques for the Modern Laboratory



ventional hood. These units are an inexpensive alternative
to laminar flow hoods, and are good for drying and storage
of containers and apparatus after cleaning, and storage of
samples and reagents.

       Airborne contamination can also be controlled by
minimizing the generation, transportation, and deposi-
tion of atmospheric particles. The major source for trace
metal particulates in the laboratory is the degradation of
metals, paints, cements, plastics, and other construction
materials. Unnecessary shelving, partitions, and furniture
should be removed from the laboratory, because dust and
debris can accumulate on them. Metal furniture should be
replaced with wood. Stainless steel door handles, hinges,
and plumbing should be replaced with plastic equiva-
lents or coated with pigment-free epoxy paint. Bench
tops should be coated with epoxy paint, and for added
protection, covered with Teflon, or polyethylene sheeting
(contact paper). Ceiling panels should be replaced with
ones that have a plastic laminate on each side to prevent
particle formation. Low fiber emitting and dissolvable con-
tent tissues should be used for wiping operations. Floors,
benches, and apparatus should be wiped down with D.I.
water regularly.     Bottles, containers, samples, reagents,
and equipment should be kept isolated from laboratory air
using laminar flow hoods, plastic snap-top Tupperware
boxes, or polyethylene bags.3,4,8-13

       When an electrostatic charge is present, charged
atmospheric particles are attracted to oppositely charged
                                   18
                          Ch. 2: Laboratory Equipment




   Ha s, Silvinum
   Sili on bber



   My styre ane




   Ste d ber




              k in
               air
        lar ne




   Wo Ruber




              m
             u



             h




   Mi an H
             t




   Air an S
   Bra d, Pla




   Silk inu
   Tefl on R



   Po ret
       con




   Go on




   Hu n
   Alu er
      lyu




         m




         ss
         ol
   Po l




        el




        m


        m
       rd




      ca
         o




       lo
         y

      ly



         s
       c




   Pap
      y
        l




   G la
   Vin




   Wo




   Hu
   Sili




   Ny
   Ra
  Most Negative (-)                                     Most Positive (+)

Figure 1. Materials’ tendency to generate a static charge.

surfaces. Electrostatic charges are usually generated by
friction between, and/or separation of, two dissimilar
materials, at least one of which is a nonconductor or a
poor conductor of electricity. The accumulated charge
(static) resides on the surface of, rather than within, the
charged nonconductive object. The highest static charges
accumulate under low humidity, on insulating surfaces
that have a low moisture content.4 Teflon has a tendency
to generate a negative charge, while the components of
airborne contamination become positively charged, espe-
cially at low humidity (Figure 1). The opening of a Teflon
bottle produces a negative charge on the neck of the
bottle, so any positively charged airborne contamination
can immediately become attached to the cap or neck of
the bottle.12 Electrostatic charges on airborne contamina-
tion can be minimized by keeping humidity levels above
50%. The surface charges can be removed by wiping with
a lint-free cloth, lightly wetted with high-purity ethanol or
water. Commercially manufactured static eliminators are
also available.
                                      19
Clean Chemistry: Techniques for the Modern Laboratory




                         20
B
     efore a sample is analyzed for trace metals, it has been
     collected, stored, processed, and prepared. During this
     sequence of events, the sample comes in contact with
many different laboratory tools, containers, and apparatus,
which can deposit trace metal contamination into the sample.
Standard laboratory mortars are made of alumina or glass, and
will therefore contaminate the sample with Al, Si, and Fe. Mills
and blenders are made of stainless steel, and have tungsten
carbide blades, which will contaminate the sample with W, Ni,
Cr, Co, and Fe. Sieves are usually
made of stainless steel with cop-
per wire mesh, which will con-
tribute Cu, Fe, Co, Cr, and Ni. The
ashless filter paper often used to       Chapter 3
filter samples contains trace met-      Materials for
al contamination on the order of
1ppm, and the filtration assem-        Trace Analysis
blies can also impart contamina-
tion to the sample.     For trace
metal analysis it is best to avoid
these practices when possible. Solid samples should be homog-
enized by digesting several samples and combing the digestates,
or by digesting a large sample and aliquoting. Aqueous samples
should be centrifuged in plastic tubes (see below) to remove
particulate material.


Borosilicate Glass
    The most common material used for laboratory contain-
ers and apparatus is borosilicate glass. It is resistant to most
                                21
           Clean Chemistry: Techniques for the Modern Laboratory



acids, but should not be used with HF or boiling H3PO4.
Alkaline solutions should not be heated or stored in bo-
rosilicate glass, because they will gradually solubilize the
glass according to the following equation4:

          2X NaOH + (SiO2)x         ]XNa SiO  2     3
                                                        + XH20

    Borosilicate glass is not a good material for trace
analysis, because it contains high levels of trace metals
(Table 1), and has the potential to absorb analytes from
the sample according to the following equation4,14:

            GlassSiOH + M+         ] GlassSiOM + H           +




Quartz
    An alternative to borosilicate glass is quartz.                Like
borosilicate glass, it is resistant to most acids, but should
not be used with HF, boiling H3PO4, or alkaline solutions.
Quartz is composed almost entirely of SiO2, and its trace
metal concentration depends on the type of quartz and the
method of production. Naturally occurring quartz labora-
tory components are made by electric (Type I) or flame
melting (Type II). Type II quartz has a lower trace metal
concentration because some of the metals are volatilized
in the flame. Synthetic quartz laboratory components are
made by the vapor phase hydrolysis (Type III) or oxida-
tion and electrical fusion (Type IV) of SiCl4. Both of these
methods produce quartz with low trace metal contamina-
tion.3,4,15 The typical trace metal impurity levels for the
various types of quartz are shown in Table 1.

                                    22
            Table 1: Trace element concentrations (µg/g) in borosilicate glass and various types of quartz.a

         Element     Borosilicate Glass      Quartz (Type I)       Quartz (Type II)        Quartz (Type III)
         Al          Major                   74                    68                      < 0.25
         B           Major                   4                     0.3                     0.1
         Ca          1,000                   16                    0.4                     < 0.1
         Cr                                  0.1                   NDb                     0.03
         Cu                                  1                     1                       <1
         Fe          3,000                   7                     1.5                     < 0.2




23
         K           3,000                   6                     <1                      0.1
         Li                                  7                     1                       ND
         Mg          600                     4                     ND                      ND
                                                                                                               Ch. 3: Materials for Trace Analysis




         Mn          1,000                   1                     0.2                     < 0.02
         Na          Major                   9                     5                       < 0.1
         Sb          2.9                     0.3                   0.1                     0.1
     a
         Adapted from references 3 and 15.
     b
         ND = Not Detected.
            Clean Chemistry: Techniques for the Modern Laboratory




Synthetic Polymers
       The low levels of trace metal contamination make
quartz an ideal material for trace metal analysis, but the
cost and availability of common laboratory containers and
apparatus in quartz limit its use in trace metal analysis.
Synthetic polymeric materials are now being employed
as materials for containers and apparatus for trace metal
analysis. The trace metal impurities of these materials
are comparable to those of quartz, but will vary based on
the manufacturing environment, type of molding, molding
components, and polymerization process. A dirty manu-
facturing environment will lead to the incorporation of
airborne particles. Molded components can contain high
levels of the metals (Ni, Al, Mn, Cu, Fe) used to make the
mold.12,16 The most common materials used for trace metal
analysis are polyethylene, polypropylene, and fluorinated
polymers.


Polyethylene
   There are two types of polyethylene used in trace
metal analysis, conventional (low density) and linear (high
density). Low density polyethylene (LDPE) is produced
by high pressure polymerization of ethylene. High density
polyethylene is produced at low pressures, catalyzed by
transition metal oxides ([Al]R3, TiCl4, ZrCl3, VCl3, CrCl3).
Polyethylene is resistant toward concentrated HCl and HF,
but is oxidized by dilute HNO3 and aqua regia. Prolonged
storage of dilute solutions of HNO3 causes the material to
turn brown or yellow. The maximum temperature for LDPE
                                     24
                        Ch. 3: Materials for Trace Analysis




                             Table 2:
       Trace element concentrations (µg/g) in some polymersa

    Element   LDPE      HDPE        PP           PFA          FEP   PTFE
    Al        0.5       30          55
    Ca                  800
    K         > 5,000   > 600                                 0.2   0.23
    Na        1.3       1.5         4.8          0.1          0.4   0.16
    Sb        0.005     0.2         0.6
    Ti                  5           60
    Mn                  0.01        0.2                       0.6
    Zn                  520
a
    Adapted from reference 17.


is 80°C and 110°C for HDPE. The use of LDPE is preferable
to HDPE because it has less trace metal contamination.
(Table 2).


Polypropylene
      Polypropylene is produced catalytically (Al, Ti) from
propylene, and, like HDPE, has elevated levels of some
trace metal contaminates (Table 2). Polypropylene is less
resistant to concentrated HCL and becomes yellow or
brown with prolonged exposure. Its resistance to dilute
HNO3 and aqua regia is similar to that of polypropylene.
Polypropylene is harder and more rigid than polyethylene
and is stable up to 135°C. Polypropylene is well suited for
open vessel digestion containers, and applications that
require sterilization.

                                        25
            Clean Chemistry: Techniques for the Modern Laboratory




Fluorinated Polymers (Teflon®)
    Fluorinated polymers (Teflon®) are more expensive
than polyethylene or polypropylene, but have lower
trace metal impurities (Table 2), and greater chemical
resistance. Fluorinated polymers are attacked only by
molten alkali metals and fluorinated organic compounds
at elevated temperatures. The greater resistance is due
to the high-energy C-F bonds and the protection of the
carbon backbone by the fluorine atoms. The three most
common fluorinated polymers used in trace metal analysis
are poly(tetrafluroethylene)(PTFE), perfluoroalkoxy-fluro-
carbon (PFA), fluorinated ethylene propylene (FEP).


Cleaning Methods
    Picking a material with low trace metal impurities
doesn’t guarantee low blank levels. Trace metal impuri-
ties can be leached from the container and apparatus, by
the sample and reagents used for trace metal analysis.
Comprehensive cleaning procedures must be adopted to
ensure the lowest blank levels. There have been a variety
of cleaning methods reported in the analytical literature,
and the method chosen will vary with the chemical behav-
ior of the element of interest. A good general procedure for
cleaning all types of containers is sequential leaching with
hydrochloric and nitric acids (Tables 3 and 4). For poly-
ethylene, a 48-hour soaking with 10% nitric acid is effective
for initial and routine cleaning.17, 18



                                     26
                      Ch. 3: Materials for Trace Analysis




                            Table 3:
Impurities (ng/cm2 of surface) leached from plastic containers in
one week with 1:1 HNO3/Water. HDPE and LDPE were leached at
       room temperature while FEP was heated to 80ºC.a

    Element        LDPE                 HDPE                FEP
    Pb             0.7                  2                   2
    Tl             1                    <1                  <1
    Ba             2                    < 0.2               4
    Te             < 0.5                0.2                 0.6
    Sn             < 0.8                1                   1
    Cd             0.2                  0.2                 0.4
    Ag             NDb                  0.2                 <8
    Sr             0.2                  1                   0.2
    Se             3                    0.4                 0.2
    Zn             2                    8                   4
    Cu             2                    0.4                 2
    Ni             0.5                  1.6                 2
    Fe             3                    3                   20
    Cr             0.8                  0.2                 0.8
    Ca             10                   0.6                 80
    K              2                    2                   2
    Mg             0.7                  0.6                 8
    Al             1                    1                   6
    Na             8                    10                  6
    Total          38                   50                  148

a
    Adapted from reference 17.
b
    ND = Not Detected.




                                      27
              Clean Chemistry: Techniques for the Modern Laboratory




                            Table 4:
Impurities (ng/cm2 of surface) leached from plastic containers in
 one week with 1:1 HCl/Water. HDPE and LDPE were leached at
       room temperature while FEP was heated to 80ºC.a

    Element         LDPE                 HDPE                 FEP
    Pb              18                   0.6                  2
    Tl              3                    < 0.6                <1
    Ba              0.3                  1                    2
    Te              0.7                  NDb                  2
    Sn              < 0.8                <1                   1
    Cd              0.2                  0.2                  0.6
    Ag              ND                   ND                   <6
    Sr              0.2                  0.2                  <1
    Se              < 0.3                0.4                  0.8
    Zn              1.0                  9                    4
    Cu              0.7                  1                    6
    Ni              0.3                  0.8                  0.8
    Fe              1.0                  1                    16
    Cr              0.3                  0.8                  4
    Ca              0.8                  60                   2
    K               0.7                  1                    1.6
    Mg              0.7                  0.4                  1.0
    Al              10                   4                    4
    Na              42                   6                    2
    Total           38                   50                   148

a
    Adapted from reference 17.
b
    ND = Not Detected.




                                       28
                       Ch. 3: Materials for Trace Analysis



    Steam cleaning with nitric or hydrochloric acid is also
a very effective cleaning method for containers and appa-
ratus (Table 5).19,20 In this method the container is placed
over a PTFE-coated glass rod. Acid in a lower reservoir
is heated, and purified acid vapor travels up through the
glass rod and condenses on the container, removing sur-
face contamination (Figure 2). This method of cleaning is a
popular alternative to the traditional soaking methods for
the following reasons:

  1. The trace metal contamination found in the re-
  agent grade acid remains in the lower reservoir and
  does not come in contact with the component to be
  cleaned.

  2. The clean component does not remain in contact
  with the cleaning acid after the surface contamina-
  tion is removed.

                                    ������ �����




             �������




    ����������
                                   ���� ���������
                                                             Figure 2. Acid
                                                             steam cleaning
                              ��������� ������� ������
                                                             system.

                                         29
            Clean Chemistry: Techniques for the Modern Laboratory



   3. The critical surfaces of the clean component are
   dry when the cleaning process is complete.                       This
   eliminates the need for rinsing and air drying.

   4. The cleaning process takes place in a sealed con-
   tainer, which minimizes airborne contamination and
   provides a clean environment for the components to
   be stored until they are needed.




   A Closer Look

A machine that performs fully automated acid
steam cleaning, as described in the text above.
Purified acid vapor condenses on the containers
placed in the machine, and the sealed environment
protects them until they are needed. (See page 82
for more information.)




                                     30
      Table 5: Comparison of high-temperature acid leaching cleaning vs. acid steam cleaning. Trace metal
     contamination (pg/g) in 5% HNO3 blanks prepared after cleaning are listed below. The acid leaching was
       performed at 180ºC with mixture of HCL and HNO3. The steam cleaning performed with HNO3 only.a
                      TFM Teflon Vessel                                        Quartz Vessel
                                                                                 b
    Element     Acid Leachingb          Steam Cleaningb          Acid Leaching         Steam Cleaningb
    Al          287 ± 46                258 ± 24                 398 ± 28              327 ± 18
    Mg          289 ± 22                232 ± 15                 441 ± 56              347 ± 26
    Na          ≤	121                   ≤	121                    1190 ± 350            608 ± 67
    Fe          ≤	474                   ≤	474                    ≤	474                   ≤	474
    Ni          ≤	55                    ≤	55                     ≤	55                    ≤	55
    Co          ≤	56                    ≤	56                     ≤	56                    ≤	56
    Cu          144 ± 39                117 ± 12                 170 ± 15                109 ± 9
                                                                                                                    Ch. 3: Materials for Trace Analysis




    Cr          ≤	85                    ≤	85                     176 ± 57                ≤	85
    Cd          ≤	72                    ≤	72                     ≤	72                    ≤	72
    Tl          ≤	261                   ≤	261                    ≤	261                   ≤	261
    Pb          ≤	57                    ≤	57                     ≤	57                    ≤	57
    Zn          995 ± 80                ≤	876                    1,640 ± 1,000           1,005 ± 124
a                                                          b
    Adapted from reference 20.                                 Error expressed as one standard deviation (n = 3).
Clean Chemistry: Techniques for the Modern Laboratory




                         32
T
     he instrumentation used for trace analysis (ICP/OES or
     ICP/MS) requires homogeneous solutions for calibration
     and analysis. Calibration solutions are prepared by dilu-
tion with water, and samples are treated with mineral acids. The
purity of the reagents is important because the amount of re-
agent used is usually several orders of magnitude larger than the
original sample size. Trace metal contamination in the reagents
must be low enough to measure accurately the analyte con-
centration in the sample. For example, to measure 10 ng of lead
in a sample that was prepared
with 25 mL of reagents, the lead
contamination in the reagents
must be less than 0.2 ng.

    No single purification method
                                          Chapter 4
is capable of removing all impuri-      Trace Analysis
ties from reagents.   Sub-boiling
distillation has been shown to
                                           Reagents
be the method of choice for acid
purification.   This method uses
infrared heaters to vaporize the
surface liquid. The vaporized liquid is collected on an inclined
water-cooled condenser and drips into the collection container
(Figure 3). Vaporization without boiling is the key element of this
purification process because it prevents aerosolized particles
from depositing on the surface of the condenser and being car-
ried over to the purified acid. High-purity acids produced in this
way have essentially the same concentration as the acids used to
produce them.12,21-23 Table 6 shows the quality of acids that can
be obtained from sub-boiling distillation.
                                33
             Clean Chemistry: Techniques for the Modern Laboratory




     Figure 3. Sub-boiling acid distillation system.


   A Closer Look

A fully automated sub-boiling distillation
system allows chemists to produce their
own high-purity acids, and even re-purify
contaminated acids. (See page 81 for more
information.)




    A good water supply is also essential for trace metal
analysis. The most common methods for achieving this
are sub-boiling distillation (previously discussed), reverse
osmosis, and ion-exchange. Reverse osmosis separates
dissolved material from the water by forcing contami-

                                      34
                       Ch. 4: Trace Analysis Reagents




                              Table 6:
     Trace metal contamination (pg/g) in nitric acid produced by
               sub-boiling distillation in a quartz still.

    Element        Source 1a          Source 2b         Source 3c
    Mg             42                 90                400
    Al             147                700               900
    Ca             157                110               400
    Ti             8                  ND                800
    V              11                 ND                <3
    Cr             5                  60                100
    Mn             2                  ND                7
    Fe             210                350               800
    Co             1                  ND                < 10
    Ni             23                 80                30
    Cu             21                 50                200
    Zn             49                 60                80
    Se             1                  60                ND
    Sr             1                  20                ND
    Ag             2                  6                 ND
    Cd             2                  20                < 30
    Sn             9                  20                ND
    Ba             4                  20                20
    Tl             <1                 60                ND
    Pb             3                  30                40
    B              ND                 ND                200

a
  Reference 24 Microwave evaporation with ICP-MS analysis.
b
  Reference 22 Hot plate evaporation in Class 100 hood with
ID-SSMS analysis.
c
  Reference 23 Inverted Pyrex with filtered air evaporation with
SSMS analysis.

                                    35
              Clean Chemistry: Techniques for the Modern Laboratory



nated water through a membrane against osmotic pres-
sure. The membrane preferentially allows water to pass,
rejecting 90-99% of dissolved ions and particulates. The
reverse osmosis system is generally used as a pre-treat-
ment technique for water before it is further purified by
either sub-boiled distillation (previously described) or
ion-exchange. In the ion-exchange method the contami-
nated water passes through a column of resin. The resin is
composed styrene-divinylbenzene copolymers engineered
to have an affinity for either cations or anions. The resins
exchange hydrogen and hydroxyl ions for the charged
metal contaminates. This results in an exchange of the
trace metal contamination for clean water.                        Sub-boiling
distillation and ion exchange both produce water with low
levels of trace metal contamination (Table 7).




                            Table 7:
       Trace metal contamination (pg/g) in high-purity water.

    Element                Source 1a                   Source 2b
    Mg                     42                          90
    Al                     147                         700
    Ca                     157                         110
    Ti                     8                           ND
    V                      11                          ND
    Cr                     5                           60
    Mn                     2                           ND
a
    Adapted from reference 23.


                                       36
N
        early all trace metals analysis procedures require some
        intervention by the analyst. Serious sample contamina-
        tion can occur as a result of careless manipulation of the
sample. Touching or handling of equipment with bare hands
can cause serious contamination. Dried skin contains 6 µg/g of
Zn and 0.7 µg/g of copper, and sweat contains Na, K, Pb, Ca, and
Mg. Hand lotions and creams contain Al, Zn, Ti, and Mg oxides,
and other trace metal contaminates.      Traces of iron, copper,
gold, silver, platinum, and chromium can be deposited from
watches, rings, and bracelets.
Human hair contains on average
100 µg/g of zinc, 20 µg/g of copper
and iron and 10 µg/g of lead. The         Chapter 5
use of cosmetics by the analyst
can be a significant source of con-
                                        The Analyst:
tamination, because of the metal         A Source of
oxides and other materials which
are added to such products for          Contamination
color and texture (Table 8). Some
hair dyes and shampoos con-
tain selenium and lead. Analysts should avoid the use of these
products when carrying out analyses.

    In order to control analyst contamination the analyst must
first be isolated from the samples.      The most fundamental
precaution is to wear gloves when performing procedures that
require manipulation of the sample. The gloves must be impervi-
ous to skin oils and perspiration and must be powder-free. Clear
polyvinyl chloride or polyethylene gloves are the best for routine
handling of the samples. When working with concentrated acids,
                                37
           Clean Chemistry: Techniques for the Modern Laboratory




                        Table 8:
         Trace metal contamination in cosmetics.
Cosmetic                 Trace Metals Present
Lipstick                 Bi, Zn, Fe, Mg, Ti, Mn
Eye Shadow               Bi, Si, Fe, Mn, Ti, Al, Cr, Mg
Blush                    Si, Fe, Mg, Ti, Ca
Mascara                  Na, Fe, Mg, Ti, Cr, Al
Foundation               Ti, Al, Zn, Fe
Face Powder              Ti, Si, Bi, Fe, Zn, Mg, Ca



nitrile gloves are a good compromise between chemical
resistance and cleanliness, and are often worn in conjunc-
tion with long-cuff vinyl gloves. One should remember
that gloves are only as clean as the last thing touched, and
must be changed on a routine basis to avoid the gloves’
becoming a contamination source. The best results are
achieved when gloves are worn in conjunction with a head
cover, shoe covers, and a laboratory coat.

   Protective garments (lab coats, head and shoe covers)
worn during trace metal analysis should not be cotton or
linen. These materials should be avoided because they
produce a considerable amount of lint, which can con-
taminate the sample. The best materials for laboratory
garments are nylon, Dacron® polyester, and Tyvek®. Gar-
ments made from these materials do not shed fibers, are
lightweight, and are resistant to acids and other reagents.
Garments used in trace metal analysis should also have
all cut edges enclosed, and should be made without com-
                                    38
               Ch. 5: The Analyst—A Source of Contamination



ponents that are susceptible to
corrosion, such as metal buttons
and zippers.

   The analyst can also be re-
sponsible for cross-contamina-
tion from other samples or work
being performed in the labora-
tory. If the analyst is not aware
of the history of the laboratory
containers and apparatus used
for the analysis, contamination
can occur as a result of using a
vessel or container that was ex-
posed to high concentrations of
the analytes of interest. The use
of paper towels in another part
                                               Protective garb—including
of the laboratory can introduce              gloves, head cover, shoe covers,
large quantities of airborne con-             and lab coat—is essential for
tamination into the atmosphere.                        best results.

Careless handling of the reagents
can lead to contamination of the reagents, the sample,
and the calibration standards. Contamination from these
sources is very unpredictable. As an analyst, one must
be constantly aware of one’s actions, and think about how
those actions will affect the blank. Most importantly, the
analyst must avoid those actions that tend to increase the
blank, or whose effects on the blank are unknown.



                                   39
Clean Chemistry: Techniques for the Modern Laboratory




                         40
  l   SECTION TWO   l




    Advanced
Sample Preparation
   Techniques
Clean Chemistry: Techniques for the Modern Laboratory




                         42
T
     he reliability of most methods of analysis depends on
     quantitative conversion of solids to homogeneous solu-
     tions. Conventional wet-sample preparation methods for
the decomposition of solid samples are usually carried out in
vessels containing the sample and a large volume of decomposi-
tion reagent(s), typically 15 to 100 mL. This mixture is heated
for several hours using hot plates, heating mantles, or ovens.
Heating is terminated when the analyst decides that the decom-
position of the sample is sufficiently complete.      This type of
open-vessel digestion has many
drawbacks, which include the use
of large volumes (and multiple
additions) of reagents, potential          Chapter 6
for contamination of the sample
by materials and laboratory en-
                                         Closed-Vessel
vironment, and the exposure of            Microwave
the analyst and the laboratory to
corrosive fumes.                           Digestion
    The   high-pressure    closed-
vessel wet-ashing technique, originally described by Carius, is
a more efficient way to decompose samples for analysis. The
increased pressures allow temperatures beyond the atmospheric
boiling point of the reagent to be reached. While Carius’s method
improves the efficiency of decomposition, it also has several
drawbacks. Analytes can be lost during the opening of the tube
when the contents, under elevated pressure, are released sud-
denly. The analyst is frequently exposed to corrosive reagents, as
well as flying pieces of glass, during the opening of the tube.


                                43
            Clean Chemistry: Techniques for the Modern Laboratory



   A Closer Look

A closed-vessel microwave sample
digestion machine is the most
efficient means of converting solids
to homogeneous solutions. (See pp.
81-82 for more information.)




    Steel-jacketed Teflon lined bombs are now available
to perform similar high-pressure and temperature reac-
tions in thermal ovens. While the higher pressures and
temperatures inside the closed vessels increase reaction
rates, digestions may still require several hours due to the
inefficient, “outside-in” heating mechanism. In addition,
the high pressures involved with both of these conven-
tional closed vessel methods tend to increase the safety
risk of applying these techniques. Closed-vessel micro-
wave decomposition, on the other hand, uses significantly
different technology and fundamentally unique principles
to accomplish sample decomposition.

        Heating by microwave energy is a “cold” in situ
process, producing heat only when there is absorption or
coupling of the microwave energy to the solution or mi-
crowave-absorbing objects. The two primary mechanisms
for the absorption of microwave energy by a solution are
dipole rotation and ionic conductance. In the dipole rota-
tion mechanism, molecular dipoles align with the applied

                                     44
                 Ch. 6: Closed-Vessel Microwave Digestion



electric field. Oscillation of the electric field results in
forced molecular movement of the dipole molecules with
the resulting friction heating the solution. At 2.45 GHz, the
frequency of most laboratory microwave ovens, the di-
poles align, then randomize 5 billion times a second. In the
ionic conduction mechanism, the ionic species present in
solution migrate in one direction or the other according to
the polarity of the electromagnetic field. The accelerated
ions meet resistance to their flow, and heating is a natural
consequence.1     These two unique heating mechanisms
result in rapid heating of solutions, in comparison with
conduction and convection. The heating is so fast that, in
open vessels, vaporization alone can dissipate the excess
energy. This results in solutions’ being able to sustain su-
perheating above their normal boiling points by as much
as 50C for water to 260C for acetonitrile.2,3

         To use closed-vessel microwave decomposition
effectively, one must understand the unique temperature
and pressure relationships involved.                    Gas pressures
inside microwave-closed vessels are not what would be
predicted from the temperature of the liquid phase. The
pressure inside a microwave vessel is dependent upon the
volume of the vessel, and the temperature and composi-
tion of the gas phase. For example, when water is placed
in a high-pressure steel-jacketed Teflon bomb and heated
in a convection oven, an equilibrium vapor pressure is
established.    This vapor pressure is dependent upon
the rate of evaporation and condensation of the water
vapor.    When the temperature is increased, there is a
                                   45
           Clean Chemistry: Techniques for the Modern Laboratory



corresponding increase in the evaporation rate and a de-
crease in the condensation rate, because the vessel walls
heat both the solution and gas phase. The decrease in
condensation rate leaves more water in the vapor phase,
increasing the internal pressure. In contrast, when water
is heated to the same temperature in a microwave closed-
vessel, the internal pressure is significantly less than its
steel-jacketed counterpart. This phenomenon is a direct
result of the microwave heating mechanism used, and the
materials of which the microwave vessel is composed.
The microwave-closed vessel’s liner and outer casing are
microwave transparent and have no insulating capac-
ity. Thus, they remain relatively cool during the heating
process. The cooler the vessel walls, the more efficient
                                            they will be at removing
                                            water molecules from
                                            the vapor phase.           The
                                            increased condensation
                                            rate results in lower
                                            internal      pressures      at
                                            higher         temperatures.
                                            This microwave reflux
                                            action is illustrated in
                                            Figure 1.

                                                   A more complex
                                            example of this phe-

                                            Figure 1. Reflux conditions in-
                                            side a microwave closed vessel.
                                            Adapted from reference 4.

                                    46
                                  Ch. 6: Closed-Vessel Microwave Digestion



nomenon is the closed-vessel microwave heating of nitric
acid. Nitric acid is a polar and partially ionized mixture
which heats rapidly in a microwave field. When heated,
nitric acid partially decomposes into NOx gas. The gas
phase inside a closed vessel becomes a mixture of NOx
gas, nitric acid and water vapor. The pressure of nitric
acid during closed-vessel microwave heating is still lower
than predicted, even after taking into account the partial
pressure of the NOx gas (Figure 2). The decrease in pres-
sure results from the previously described reflux action
and from the loss of the ionic conductance mechanism
of microwave heating in the gas phase. The loss of the
                 70



                 60
                                     Thermal Equilibrium

                                     Microwave Vessel
                 50
Pressure (atm)




                 40



                 30



                 20



                 10



                 0
                      121   130       133    150     165    170     190      192   210   219
                                               Temperature (Celcius)

Figure 2. Internal pressure of nitric acid at different temperatures.

                                                    47
          Clean Chemistry: Techniques for the Modern Laboratory



ionic conductance mechanism in the gas phase means
the NOx gas does not convert microwave energy into heat
efficiently, keeping the pressure increase associated with
the heating of the gas to a minimum.

       This unique temperature and pressure relationship,
found only in closed-vessel microwave heating, becomes
more complex and unpredictable as additional reagents
and samples are added to the solution. Also, the condensa-
tion rate varies with the microwave vessel materials and
geometry, the liquid volume, the duration of heating, and
the system’s ability to dissipate the excess energy. At the
present time there is no conventional method of predicting
the decrease in internal pressure associated with closed-
vessel microwave heating. This phenomenon is one of
the reasons that pressure control is not applicable for
standardizing microwave sample preparation methods.

       The use of closed vessels in microwave decomposi-
tion allows the reagents to be heated above their atmo-
spheric boiling points. The higher temperatures achieved
in the closed system give microwave decomposition a
kinetic advantage over hot plate digestion, as described by
the Arrhenius Equation:
                             dlnk        Ea
                                    =
                              dT        RT 2
   Integration of this equation gives:

                   ln
                         k2
                         k1
                            =
                                 Ea  1
                                       –
                                         1
                              2.303R T1 T2
                                           (          )
                                   48
                Ch. 6: Closed-Vessel Microwave Digestion



   In this expression k1 and k2 are rate constants for the
reaction of interest at T1 and T2 respectively, Ea is the
activation energy, and R is the ideal gas constant. These
equations show that the reaction rate increases exponen-
tially with increasing temperature. This translates into
approximately a 100-fold decrease in the time required to
carry out a digestion at 175 °C when compared to 95 °C di-
gestion.4,6,7 In addition, because the mineral acid converts
the microwave energy into heat almost instantaneously,
rapid heating of the sample is achieved, further decreasing
the reaction times.

       The effectiveness and safety of the microwave
digestion procedure depends on the choice of digestion
reagents. The most common reagents used for microwave
digestion are nitric, hydrochloric, and hydrofluoric acids.
Nitric acid is a strong oxidizing agent and liberates trace
elements as highly soluble nitrate salts. The oxidizing
properties of nitric acid are retained only in concentrated
form and are lost when the acid is diluted below 2M. Nitric
acid is used for the dissolution of metals and organic/bio-
logical materials. Some metals form insoluble oxide films
(Al, Cr, Ti, Nb, Ta) and require the addition of a complex
forming species in order to obtain complete dissolution.
The nitrate ion forms weak complexes with tin, tungsten,
and antimony, and these elements can become hydrolyzed
and precipitate from solution.

       Hydrochloric and hydrofluoric acids are complex-
ing acids. Many metal carbonates, peroxides and alkai hy-
                                  49
           Clean Chemistry: Techniques for the Modern Laboratory



droxides are dissolved by hydrochloric acid. Hydrofluoric
acid is the only acid which will readily dissolve silica-based
materials, and creates strong complexes with elements
that form stable refractory oxides. Alkaline earths, lan-
thanide and actinide elements form insoluble or sparingly
soluble complexes with hydrofluoric acid. These acids are
often used in conjunction with nitric acid to dissolve metal
alloys and noble metals.


Microwave Decomposition and
Clean Chemistry
    Closed-vessel microwave decomposition has several
unique characteristics that have led analysts to reduce
the sources of error and contributions to the analytical
blank that were previously obscured by lengthy sample
preparation procedures.

       As described in Chapter 2, the laboratory air that
comes in contact with the sample can deposit airborne
contaminants into the sample. Open-vessel methods can-
not prevent this type of contamination because the sam-
ples are continuously exposed to laboratory air currents.
The use of closed vessels for microwave decomposition
isolates the sample from laboratory air currents during the
decomposition process, reducing airborne contamination.
By preparing the samples for closed-vessel microwave
dissolution under clean air conditions, we can isolate the
sample, so that it will not come in contact with laboratory
air. Thus, the possibility for contamination by airborne

                                    50
                Ch. 6: Closed-Vessel Microwave Digestion



particles in the laboratory air is eliminated. Similarly, by
performing post sample processing, such as rinsing and
dilution in clean environments, airborne contamination
can be further reduced.

       Reagent contamination contributes to the concen-
tration of the analytes present in the sample. The amount
contributed is a function of the total quantity of reagent
used. For example, using 50 mL of a reagent that contains
a contaminant at the 100 ppb level will contribute 5 µg of
that species as contamination. Unlike open-vessel meth-
ods that require large quantities of reagent due to evapora-
tion losses, closed-vessel microwave methods continually
reflux the reagents and do not allow the reagent vapors to
escape. The reflux conditions inside a microwave-trans-
parent vessel allow much smaller volumes of the reagent
to be used, thus reducing the contamination from the
decomposition reagents.

       The materials used for microwave vessels (fluori-
nated polymers and quartz) are chemically inert to most
dissolution reagents, and provide a non-contaminating en-
vironment for sample preparation. TFM is preferred over
PFA for the construction of microwave vessels because of
its superior chemical and thermal resistance and lower
blank levels (Tables 1 and 2).

    The skill of the analyst is perhaps the most difficult
factor to evaluate. For open-vessel methods, the analyst
must monitor the progress of the dissolution in several

                                  51
              Clean Chemistry: Techniques for the Modern Laboratory




                               Table 1:
     Trace element concentrations (µg/L) of acid blanks prepared
            from identical reagents in PFA and TFM vesselsa

    Element                PFA                        TFM
    Al                     2.7                        ND
    B                      7.5                        1.8
    Ba                     0.9                        DL
    Bi                     0.5                        DL
    Cd                     4.7                        0.1
    Co                     0.8                        DL
    Cr                     0.8                        0.1
    Mo                     0.6                        DL
    Pb                     6.7                        0.03
    Sb                     0.6                        DL
    W                      5.4                        DL
    Zn                     0.6                        0.2
    Zr                     1.0                        DL
a
 Adapted from reference 8.
ND = Not determined.
DL = Concentration below instrument detection limit.


vessels simultaneously. Compounding the problem, the
temperature of each vessel varies with its position on
the hot plate. The analyst must then subjectively decide
whether dissolution is sufficiently complete, and remove
the sample from the heat source. In microwave methods,
the efficiency of the heating, coupled with the direct moni-
toring of the conditions inside the vessel, removes the
subjective judgment of the analyst from the digestion pro-

                                       52
                    Ch. 6: Closed-Vessel Microwave Digestion




                               Table 2:
     Analytical blank values for TFM microwave vessels after fifty
                digestions of environmental samplessa

    Element        µg/L                 Element                µg/L
    As             0.24                 Li                     0.09
    B              0.40                 Mo                     0.26
    Ba             0.04                 Pb                     0.15
    Be             0.03                 Sb                     0.01
    Cd             0.01                 Sc                     0.04
    Ce             0.19                 Se                     0.01
    Co             0.02                 Sr                     0.05
    Cu             0.03                 Ta                     0.01
    Ga             0.02                 Zn                     0.31
a
    Adapted from reference 9.


cess. The automation and standardization of the sample
preparation method through the use of closed-vessel mi-
crowave decomposition serves to reduce the impact of the
analyst’s judgment upon much of the sample preparation
process.

       The combined effect of closed-vessel microwave
decomposition and clean chemistry techniques is shown
in Tables 3 and 4, and Figure 3. The blank results show
significant reduction in the contributions of the outside
environment, reagents, and materials to the analytical
blank level and overall measurement uncertainty. From
Table 4, we can see that this combination of techniques
allows trace metal analysis to be completed with increased

                                      53
              Clean Chemistry: Techniques for the Modern Laboratory



accuracy and precision. Figure 3 (on page 56) shows
that closed-vessel microwave decomposition, coupled
with clean chemistry techniques, is accurate and precise
enough to prepare instrument calibration standards.




                             Table 3:
      Comparison of analytical blank results obtained from hot
     plate and microwave digestion of a certified reference soil.a
                 Concentration expressed as µg/g.

    Analyte     Hot Plateb                       Microwaveb
    As          0.204 ± 0.106                    0.074 ± 0.013
    Cd          0.318 ± 0.122                    0.029 ± 0.019
    Cr          3.35 ± 2.85                      0.104 ± 0.059
    Cu          0.060 ± 0.020                    0.030 ± 0.017
    Pb          0.171 ± 0.076                    0.040 ± 0.019
    Hg          0.037 ± 0.004                    0.017 ± 0.007
    Ni          0.375 ± 0.069                    0.060 ± 0.063
    Se          0.548 ± 0.264                    0.172 ± 0.022
    Tl          0.028 ± 0.020                    0.028 ± 0.020
    V           2.35 ± 0.45                      1.20 ± 0.51
    Zn          2.92 ± 1.43                      1.66 ± 0.93

a
    From reference 10.
b
    Error expressed as 95% confidence interval (n=4).

                                       54
                     Ch. 6: Closed-Vessel Microwave Digestion




                             Table 4:
      Comparison of analytical blank results obtained from hot
     plate and microwave digestion of a certified reference soil.a
                 Concentration expressed as µg/g.

    Analyte   Hot Plateb            Microwaveb              Certified Valuec
    As        12.3 ± 2.27           17.6 ± 0.9              17.7 ± 0.8
    Cd        0.31 ± 0.09           0.41 ± 0.06             0.38 ± 0.01
    Cr        68.8 ± 7.5            123 ± 3                 130 ± 4
    Cu        23.8 ± 2.7            33.5 ± 1.2              34.6 ± 0.7
    Pb        10.8 ± 1.3            17.5 ± 1.1              18.9 ± 0.5
    Hg        0.97 ± 0.14           1.42 ± 0.10             1.40 ± 0.08
    Ni        63.4 ± 3.9            83.2 ± 3.0              88 ± 5
    Se        1.61 ± 0.34           1.54 ± 0.33             1.57 ± 0.08
    Tl        0.29 ± 0.05           0.63 ± 0.02             0.74 ± 0.05
    V         65.6 ± 8.8            119 ± 6                 112 ± 5
    Zn        113 ± 13.5            102 ± 6.1               106 ± 3

a
  From reference 10.
b
  Error expressed as 95% confidence interval (n=4).
c
  95% confidence interval as reported on certificate.

                                       55
              Clean Chemistry: Techniques for the Modern Laboratory




Figure 3. ICP-MS calibration using microwave digested certified reference
materials. The three calibration points represent separate digestions. Adapted
from reference 10.




                                       56
S
         olutions submitted for trace elemental analysis often need
         to be evaporated prior to analysis. The most common rea-
         son for evaporation is to concentrate the sample, because
initial analyte levels are below the instrument detection limits.
The other reason for evaporation is that the sample solution
contains matrix elements that will present problems for analysis.
Traces of hydrofluoric acid will etch the glass components of
ICP and ICP-MS systems, releasing trace element contamina-
tion. Chlorine and fluorine form polyatomic ions that interfere
with the ICP-MS analysis of many
common elements. For example,
40
   Ar35Cl+ interferes with 75As, and
35
     Cl16O+ interferes with 51V.
                                             Chapter 7
           Obtaining    solutions
trace elemental analysis is often
                                    for
                                            Microwave
complicated due to the forma-               Evaporation
tion of volatile species during
the evaporation process. As the
number of solvent molecules de-
creases, the ions begin to recombine, and at dryness the residue
will consist of a mixture of recombined salts. These salts will
have an associated vapor pressure and boiling point that will
vary with the oxidization state and counter anion (Table 5). The
use of traditional heating methods to perform evaporations can
lead to the loss of volatile analytes, because as the evaporation
proceeds, the temperature of the evaporation vessel approaches
the temperature of the heat source. At dryness these tempera-
tures can be in excess of 150°C.


                                    57
               Clean Chemistry: Techniques for the Modern Laboratory




                                Table 5:
                Potentially volatile salts from solution.

    Element                 Volatile Salts              Boiling Point (ºC)
    Arsenic                 AsCl3                       130.2
                            AsF3                        -63
    Antimony                SbF5                        150
                            SbCl5                       79
    Selenium                SeCl4                       170–196a
                            SeF4                        107.8
    Tin                     SnCl4                       115
    Vanadium                VCl4                        152
                            VF5                         111.2
    Chromium                CrF5                        117
a
 SeCl4 sublimes.
Data taken from the following references:

Dean, J.A., ed., Lange’s Handbook of Chemistry, 12th ed., New York:
McGraw-Hill, 1979.

David, R. Linde, ed., CRC Handbook of Chemistry and Physics, 71st
ed., Cleveland: CRC Press, 1990.


          In contrast to traditional evaporation methods, the
unique heating mechanism exclusive to microwave-as-
sisted heating allows for the retention of volatile analytes
during evaporation. As the evaporation proceeds and the
solvent is removed from the system, the matrix volume will
be reduced. As the mass of the sample solution decreases,
the amount of microwave energy absorbed decreases
according to the following equation:
                                        58
                        Ch. 7: Microwave Evaporation



    A Closer Look

Microwave-assisted heat-
ing methods allow for the
retention of volatile analytes
during evaporation. (See pp.
81-82 for more information.)




                               KCp∆Tm
                             Pabs=
                                   t
     In this equation P is the apparent absorbed power in
watts, K is the conversion factor for calorie/s to watts,
Cp is the specific heat, ∆T is change in temperature, m
is the total mass of sample in the microwave, and t is ir-
radiation time. This unique relationship between sample
mass and energy absorption, along with the microwave
vessels being microwave transparent, leads to a decrease
in temperature as the sample approaches dryness. Lower
temperatures at dryness decrease the potential for loss of
volatile species, resulting in more complete recoveries for
volatile analytes11,12 (Figures 4 and 5).

        Another factor that plays a significant role in
evaporation losses is the oxidization state of the analyte
of interest. It has been reported that even mild heating
of an HF solution results in a 20% loss of Se(IV) and a
45% loss of As(III), and a losses of 65%-100% upon dry-
ness. Losses of antimony during evaporation are due to

                                     59
              Clean Chemistry: Techniques for the Modern Laboratory




Figure 4. Evaporation recoveries of select elements from a 9:3 solution of
HNO3/HCl. Initial solution concentration was 500 ppb.


the formation of poorly soluble compounds, and mercury
is due to its being present in its elemental form. These
losses can be prevented by converting these elements to
higher oxidation states i.e. Se(VI), As(V), Sb(V), Hg(II). The
use of traditional decomposition methods can not ensure
a uniform oxidation state, due to matrix interferences
and reaction rate limitations.14 This can be overcome by
coupling closed-vessel microwave decomposition with
microwave evaporation. The elevated temperatures and
pressures decompose the matrix interferences and form
stable complex ions.


                                       60
                         Ch. 7: Microwave Evaporation




Figure 5. Percent recovery of 2.5 ng spikes from 10ml of HCl. Uncertainties
are expressed as 95% confidence intervals with n≥4. Adapted from references
4 and 13.

     For example:

                  Sb2O2(s) + HCL + HNO3                 SbCl6-(aq)

                  AsF3(aq) + HF + HNO3             AsF6-(aq)

     The formation of these stable complex anions leads to
complete retention of these traditionally volatile elements.
(Figures 6 and 7).




                                     61
                                     Clean Chemistry: Techniques for the Modern Laboratory

Concentration (µg/g)




                                                             Element

                       Figure 6. Concentration of analytes in SRM 1566A (Oyster Tissue) following
                       Microwave-Assisted Evaporation of the digestate compared with the certified
                       total concentrations. Uncertainties are expressed as 95% confidence intervals
                       with n≥3. Adapted from references 4 and 11.




                                                              62
                                                Ch. 7: Microwave Evaporation

Concentration (µg/g)




                                                              Element

                       Figure 7. Concentration of analytes in SRM 2710 (Montana Soil) following
                       Microwave-Assisted Evaporation of the digestate compared with the certified
                       total concentrations (* = noncertified concentration). Uncertainties are
                       expressed as 95% confidence intervals with n≥3. Adapted from references 4
                       and 11.




                                                            63
Clean Chemistry: Techniques for the Modern Laboratory




                         64
A
        nalytical chemists are being required to measure lower
        and lower levels of trace metals in samples. This fre-
        quently requires the analyst to make measurements near
the method detection limit, which usually results in decreased
accuracy and precision. The advantages of using closed-vessel
microwave decomposition in combination with clean chemistry
techniques for lowering method detection limits have been dis-
cussed in Chapter 6 of this book, but sometimes these methods
are not enough, and another approach is required to achieve the
desired levels.

       Closed-vessel microwave
digestion techniques require a
minimum volume of 10 mL to               Chapter 8
achieve accurate temperature
monitoring of the reaction con-        Lowering Your
ditions. Modern spectroscopic         Method Detection
techniques        require   samples
submitted for analysis to have             Limits
acid concentration of 10% (v/v)
or less. This requires samples
that were digested with 10 mL of acid to be diluted by a factor
of 100 or more. This problem can be overcome by increasing
the sample size. This works well for samples that do not con-
tain a large amount of organic material. For samples with high
organic content this is usually not an option, because the sec-
ondary gases (CO2 and NOx) produced during the digestion can
cause the vessel to vent when larger sample sizes (greater than
0.5 grams) are used. Recent advances in microwave chemistry
have led to the development of vessel-inside-vessel technol-
                                 65
           Clean Chemistry: Techniques for the Modern Laboratory



ogy as a means to lower method detection limits for highly
organic samples.

       Vessel-inside-vessel technology was developed
through by the collective efforts of Milestone Inc, srl, and
GmbH in the late 1990s. Vessel-inside-vessel technology
uses a smaller secondary vessel inside the primary micro-
wave vessel. The secondary vessel contains the sample
and digestion reagents, and the primary vessel contains
the 10 mL of solution required to achieve accurate tempera-
ture monitoring (Figure 8). This configuration reduces the
amount of acid required for digestion to near stoimetric
quantities, which reduces the dilution factor and increases
the detection limit.

       The use of vessel-inside-vessel technology is also
used for the processing of larger organic sample sizes.
This is accomplished by controlling the reaction kinetics
and lowering the pressure inside the microwave vessel.
Controlling reaction kinetics is especially important when
trying to digest large quantities (0.5 to 1.0 g) of organic ma-
terial, because the potential for auto-catalytic decomposi-
tion increases. When the sample size is small (0.25 g), the
heat released by the oxidization of the organic material
does not cause a significant change in the temperature of
the reaction mixture. As the sample size is increased, the
heat released from the oxidation can cause the reaction
mixture to heat faster that the programmed rate. The rise
in temperature promotes further decomposition, which
results in the microwave vessel’s venting (sometimes at
                                    66
                  Ch. 8: Lowering Your Method Detection Limits




Figure 8 (a and b). Photo and schematic of vessel-inside-vessel technology.


                                      67
                                Clean Chemistry: Techniques for the Modern Laboratory



                   pressures lower than its rating) due to the sudden increase
                   in pressure resulting from the self-sustaining auto-catalytic
                   decomposition (runaway reaction) of the sample (Figure
                   9). The use of vessel-inside-vessel technology helps to
                   control these self-sustaining auto-catalytic reactions by
                   providing a heat sink for the energy liberated during oxidi-
                   zation. This is accomplished by placing water in the outer
                   microwave vessel. The water draws the heat away from
                   the reaction mixture, slowing down the reaction kinetics
                   and preventing a runaway reaction. (Fig 10).

                            The amount of sample that can be safely digested
                   is limited by the amount of pressure generated during the
                   decomposition process. Current microwave vessel tech-
                   nology limits the internal pressure to 100 atm (1450 psi).
Temperature (ºC)




                   Figure 9. Example of runaway reaction during a closed-vessel microwave
                   digestion.

                                                         68
                                      Ch. 8: Lowering Your Method Detection Limits


Temperature (ºC)




                   Figure 10a. Digestion of 5 grams of fresh liver using conventional microwave
                   decomposition. Note the runaway reaction.
Temperature (ºC)




                   Figure 10b. Digestion of 5 grams of fresh liver using vessel-inside-vessel
                   technology. There is no runaway reaction.

                                                          69
           Clean Chemistry: Techniques for the Modern Laboratory



    For most organic samples, this limits the sample size to
0.5 to 0.7 grams. In order to digest organic samples larger
than 0.7 grams, secondary reaction chemistry must be em-
ployed to lower the pressure during microwave digestion.
This is accomplished with vessel-inside-vessel technology
by adding H2O2 to the outer microwave vessel to convert
NOx and CO2 into HNO3 and HCO3 respectively.

       Primary decomposition reaction

       (CH2)x + 2xHNO3            xCO2 + 2xNO + 2xH2O

       Secondary reactions
              ∆
      2H2O2       2H2O + O2

      2NO + O2         2NO2

      2NO2 + H2O          HNO3 + HNO2

              ∆
      HNO2        HNO3 + 2NO + H2O

      CO2 +H2O        H2CO3


    The quantitative effect of this technique is shown in
Figure 11. As you can see, this technique effectively dou-
bles the amount of organic sample that can be digested
using closed-vessel microwave technology. There is also
no transfer of analytes from the inner vessel to the outer
vessel. (Tables 6-8.)



                                    70
                  Ch. 8: Lowering Your Method Detection Limits




               Red line = Temperature      Blue line = Pressure

Figure 11. Quantitative effect of vessel-inside-vessel technology. These
are overlay plots for the microwave digestion of polypropylene pellets.
Line A represents 0.35 grams digested using conventional microwave
digestion. Line B represents 0.35 grams digested with vessel-inside-vessel
technology. Line C represents 0.60 grams digested with vessel-inside-vessel
technology.




                                      71
           Clean Chemistry: Techniques for the Modern Laboratory




                            Table 6:
 Microwave digestion of cell culture media, using vessel-inside-
 vessel technology followed by ICP-AES analysis. Results of six
                       replicate samples.

Element      Average (µg/g)                   % RSD
Ca           454                              5.04
Zn           2.22                             8.24
Fe           9.44                             2.52
Mo           7.03                             5.49
Mg           428                              4.13
K            19,552                           5.66
P            7,527                            3.69
Na           79,461                           2.88




                             Table 7:
  Iron spike recoveries for microwave digestion of cell culture
    media, using vessel-inside-vessel technology followed by
                         ICP-AES analysis.

Spike Amount                          % Recovery
10 µg                                 95.2
25 µg                                 103
50 µg                                 98.8

The average spike recovery is 99.2% with an RSD of 4.3%.




                                    72
                  Ch. 8: Lowering Your Method Detection Limits




                              Table 8:
    Microwave digestion of SRM 1577A (Bovine Liver), using vessel-
        inside-vessel technology followed by ICP-MS analysis.

    Element     Measureda (ng/g)                Certifieda (ng/g)
    Cd          427 ± 50                        440 ± 60
    Co          208 ± 1.9                       210 ± 50
    Mo          3,540 ± 46                      3,500 ± 500
    Pb          135 ± 14                        135 ± 15
    V           104 ± 7                         98.7 ± 1.6
a
    Error expressed as 95% confidence interval.




                                      73
Clean Chemistry: Techniques for the Modern Laboratory




                         74
  l   SECTION THREE   l




     Preparing
Your Laboratory for
  Trace Analysis
Clean Chemistry: Techniques for the Modern Laboratory




                         76
T
       he first two parts of this book provide the theory and
       background for low level trace metal analysis. This chapter
       outlines simple housekeeping and laboratory techniques
that can have an immediate effect on your trace metals determi-
nations.


Laboratory Environment
    As previously discussed, the laboratory environment
can significantly affect low level trace metal analysis. If clean air
facilities are not available for any
reason, then the sample prepara-
tion and analysis areas must be
isolated from the main labora-            Chapter 9
tory. Rooms with access to the
outside via doors or windows,
                                          Laboratory
with high pedestrian traffic, or         Housekeeping
with    exposed   metal   surfaces
should be avoided. Any exposed            Techniques
metal surfaces should be painted
with epoxy paint to prevent metal
contamination. The work surfaces should be sealed with a poly-
urthane or polyacrylic finish (clear heavy-duty contact paper can
also be used) to trap potential contamination. Airborne contami-
nation can be controlled by covering air vents with additional
filtration media and using a portable HEPA filter system that can
be purchased from a home supply store. Floors, benches, and
apparatus should be wiped down with D.I. water regularly with
lint free towels to remove any dust accumulation.


                                 77
          Clean Chemistry: Techniques for the Modern Laboratory




Reagent Handling
   Clean reagents are essential to the success of trace
metal analysis. All reagents and solutions used in trace
metal analysis must be handled with care, in order to
prevent contamination from outside sources. Reagents
should be dedicated to trace metal analysis, and not be
used for other laboratory procedures. Reagents bottles
should be stored in a clean environment to prevent the ac-
cumulation of dust, which can be transferred to the sample
during handling. For calibration solutions, a commercially
available desiccator cabinet (without the dessicant) works
well. For digestion acid bottles, large polyethylene bags
used in layers work well for 100-1,000 mL Teflon bottles.
No foreign objects should ever be introduced into the
original reagent container, and the original reagent bottle
should be opened as little as possible, and for the mini-
mum amount of time. To minimize opening of the original
container, a sub-sample should be transferred into a clean
container for everyday use. Excess reagent should not be
returned to the original bottle under any circumstances.


Preparation Procedures
   Preparation procedures for standards and samples
should be designed to minimize the potential for contami-
nation. Large dilutions of standards or samples should be
prepared using a micropipette with disposable tips. This
approach minimizes the number of dilutions, and prevents
contamination from a glass pipette. When using this ap-
proach, the tips should be rinsed first with DI water to
                                   78
               Ch. 9: Laboratory Housekeeping Techniques



remove any dust, then with a 10% acid solution to remove
any trace metal contamination, and last with two volumes
of the final solution before use.

    Traditional glass volumetric apparatus should also be
avoided, because they can contribute trace metal contam-
ination to the solution. Plastic volumetric apparatus are
available, but disposable graduated polypropylene centri-
fuge tubes work best. The graduation marks are accurate
to 1% RSD or better (dilutions can be done by weight if
greater accuracy is needed). These tubes can be used for
preparation of calibration solutions as well as sample dilu-
tions, and, because they are disposable, carry-over and
cross-contamination problems are eliminated.


Miscellaneous
    As an analyst, one must be constantly aware of one’s
actions, and think about how those actions will affect the
blank. Most importantly, the analyst must avoid those
actions that tend to increase the blank, or whose effects
on the blank are unknown. For example, if you are wearing
gloves to prevent contamination from your hands, but you
touch a dusty or metallic surface, your glove will pickup
trace metal contamination that can be transferred to the
sample. The best approach to trace metal analysis is to
have a routine and follow that routine for every analysis.
If you have a set routine, it will help when a contamination
problem occurs, because you can check each step until
the source of the contamination is found.

                                  79
Clean Chemistry: Techniques for the Modern Laboratory




                         80
T
      his chapter lists some of the laboratory equipment and
      supplies you will need to perform trace metal analysis,
      and provides contact information for many reputable
suppliers of such equipment.


Instrumentation
     Acid Purification
     Milestone has developed a self-contained sub-boiling
distillation system called duoPUR, which allows chemists
to make their own high-purity
acids.     The system consists of
two high-purity quartz distilla-          Chapter 10
tion units.    Each unit contains
two infrared heating elements           Equipment and
that supply a maximum power
of 1,250 W, a water cooled
                                       Supplies for Trace
condenser,      and   a   collection    Metal Analysis
bottle.    The distillation process
is       microprocessor-controlled,
allowing the user to set distillation times and power level.
Distillation rates range from 10 to 200 mL per hour, depending
on the power setting
and the temperature
of the cooling water.
An added benefit of
the system is that you
can re-purify contami-
nated acids instead of
downgrading them.
                                  81
           Clean Chemistry: Techniques for the Modern Laboratory



    Closed-Vessel Microwave Decomposition
                                         Milestone’s Ethos Plus line
                                 of microwave digestion systems
                                 was designed with the trace
                                 metals chemist in mind.             The
                                 system is constructed from cor-
                                 rosion-resistant stainless steel.
                                 The inner chassis is coated with
                                 five layers of electrosprayed
                                 PTFE for added corrosion re-
                                 sistance. The unique software
                                 interface allows precise tem-
perature ramping. You simply create a temperature profile,
and software modulates the power output to follow the
defined heating profile. A variety of digestion rotors and
accessories (evaporation, micro-inserts, etc.) are available
to accommodate all your digestion needs.

                                               Ultra-trace Cleaning
                                             Milestone’s   trace-
                                          CLEAN system is a fully
                                          automated          acid   steam
                                          cleaning system for trace
                                          metal analysis accesso-
                                          ries. This self-contained
                                          system houses an acces-
                                          sories rack and an acid
                                          reservoir.       Once loaded,
                                          the accessories are low-
                                          ered into a sealed cham-
                                    82
               Ch. 10: Equipment and Supplies for Trace Metal Analysis



ber. Nitric acid is repeatedly evaporated and condensed
throughout the chamber, thoroughly cleaning the acces-
sories. The main benefit of the system is that any trace
metal impurities that are present in the acid do not come
in contact with the cleaned accessories.

    Clean Benches and Hoods
    Several companies offer
clean benches and hoods for
trace metal sample prepara-
tion.      Labconco offers the
Purifier   ®
               Trace Metals Work
Station, designed specifically
for the demands of trace met-
als analysis. This work station
is made of non-metallic com-
ponents and provides a class
100 working environment. This
enclosure is well-suited to
applications involving corrosive chemicals, such as acid
digestions, with its optional exhaust fan and PVC duct
work. Several other models are also available.

    Terra Universal Inc. offers everything from free stand-
ing modular clean rooms to portable clean booths. All
units offer HEPA filtration, with some units providing class
1-10 working conditions. Terra also offers static control via
optional ionizers, and static dissipative materials neutral-
ize any static imbalances that may exist in the work area.


                                         83
           Clean Chemistry: Techniques for the Modern Laboratory



                                                      High-Purity Water
                                                      There are three
                                                major        manufactur-
                                                ers     of     high-purity
                                                water         equipment:
                                                Millipore, Barnstead
                                                International,   and
                                                Labconoco. All three
                                                manufacturers offer
     Free-standing modular cleanroom            a variety of systems,
from water softeners and reverse osmosis to high end
polishers capable of producing Type 1 18.2 megohm-cm
water.


Reagents and Standards
   Calibration and Quality Control Standards
   There are three major suppliers of calibration stan-
dards for trace metals analysis. They are SPEX CertiPrep
Inc., Inorganic Ventures, and High Purity Standards. All
three companies offer single, multi-element, and custom
calibration standards for AA, ICP, and ICP-MS analysis.
SPEX offers proficiency testing samples, and Inorganic Ven-
tures and High Purity Standards offer certified reference
materials for quality control procedures.

   Resource Technologies Corporation (RTC) and the
National Instituted of Standards and Technology (NIST)
provide “real world” quality control material for trace
metals analysis. Samples range from soils and sediments

                                    84
             Ch. 10: Equipment and Supplies for Trace Metal Analysis



to glasses and metal alloys. Each sample comes with a
certificate of analysis identifying major and minor trace
element constitutes.

    Reagents
    If you are not going to make your own high-purity acid,
then you will have to buy it. For general and low trace met-
al analysis, high-purity acids can be purchased from J.T.
Baker, Fisher Scientific, and High Purity Standards. Each
company offers double-distilled nitric and hydrochloric
acids in high-purity Teflon bottles. For ultra-trace work,
TAMA Chemicals offers TAMAPURE AA-100, guaranteed
to contain trace metal impurities less than 100 pg/mL,
and TAMAPURE AA-10, guaranteed to contain trace metal
impurities less than 10 pg/mL. All bottles of high-purity
acid come with a certificate of analysis.


Lab Supplies
    Gloves
    There are many different types and brands of gloves
that are suitable for trace metals analysis. I prefer to use
N-Dex Nitrile gloves (available from Fisher Scientific) as
my first layer, and Oak powder-free vinyl gloves as my
top layer. The Oak gloves are available in standard or long
cuff. I use the long cuff because it offers better protec-
tion. The Oak gloves are available from Fisher Scientific or
Terra Universal.




                                       85
           Clean Chemistry: Techniques for the Modern Laboratory



    Plasticware
    As discussed in Section One, plastics are the materials
of choice for trace metals work. Nalgene makes a wide
variety of plastic bottles and accessories for trace metal
analysis. One hard-to-find item that they make is plastic
volumetric flasks. Sizes range from 50mL to 1,000mL, and
flasks are made from either polypropylene or polymethyl-
pentene. Another item that I have found I can’t live with-
out is the free-standing graduated polypropylene tube.
These tubes are available from Fisher Scientific, and are
perfect for sample dilutions and storage.

    Miscellaneous
    A variety of clean chemistry accessories are available
from Terra Universal and Fisher Scientific. Items range
from tacky mats to lint free towels.




                                    86
          Ch. 10: Equipment and Supplies for Trace Metal Analysis




Suppliers’ Contact Information
Milestone Inc.
25 Controls Drive
Shelton, CT 06484
Telephone: 866-995-5100
Fax: 203-925-4241
Website: www.milestonesci.com


Labconco Corporation
8811 Prospect Avenue
Kansas City, Missouri 64132-2696
Phone: 816-333-8811
Fax: 816-363-0130
Website: www.labconco.com


Terra Universial Inc.
700 N. Harbor Blvd.
Anaheim, CA 92801
Phone: 714-526-0100
Fax: 714-992-2179
Website: www.terrauniversal.com


Millipore Corporation
80 Ashby Road
Bedford, MA 01730
Phone: 1-800-MILLIPORE
Website: www.millipore.com




                                    87
          Clean Chemistry: Techniques for the Modern Laboratory



Barnstead International
255 Kerper Blvd.
Dubuque, IA 52001
Phone: 1-800-446-6060
Fax: 536-589-0516
Website: www.barnsteadthermolyne.com


SPEX CertiPrep
203 Norcross Avenue
Metuchen, NJ 08840
Phone: 1-800-LAB-SPEX
Fax: 732-603-9647
Website: www.spexcsp.com


Inorganic Ventures
195 Lehigh Avenue
Suite 4
Lakewood, NJ 08701
Phone: 1-800-569-6799
Fax: 732-901-1903

High Purity Standards
4741 Franchise Street
Charleston SC 29423
Phone: 843-767-7900
Fax: 843-767-7906
Website: www.hps.net




                                   88
          Ch. 10: Equipment and Supplies for Trace Metal Analysis



Resource Technologies Corporation
2931 Soldier Springs Rd.
Laramie, WY 82070
Phone: 1-800-567-5690
Fax: 307-745-7936
Website: www.RT-Corp.com


National Institute of Standards and Technology
Standard Reference Materials Program
Building 202, Room 204
Gaithersburg, MD 20899
Phone: 301-975-6776
Fax: 301-948-3730
Website: www.nist.gov


Mallinckrodt Baker, Inc. (J.T. Baker)
222 Red School Lane
Phillipsburg NJ 08865 U.S.A.
Phone: 1-800-582-2537
Fax: 908-859-6905
Website: www.JTBaker.com


Fisher Scientific
Phone: 1-800-766-7000
Website: www.fishersci.com




                                    89
          Clean Chemistry: Techniques for the Modern Laboratory



Tama Chemicals
Distributed by
Moses Lake Industries
8249 Randolph Rd
Moses Lake, WA 98837
Phone: 509-762-5336
Fax: 509-762-5981


Nalge Nunc International
75 Panorama Creek Drive
Rochester, NY 14625
Phone: 1-800-625-4327
Fax: 585-586-8987
Website: www.nalgenunc.com




                                   90
For Section One
1.     Harris, D.C. Quantitative Chemical Analysis, 2nd Ed., W.H.
Freeman and Company, New York, 1987. p. 38.
2.     J. Ruzicka and J. Stary, Substoichiometry in Radiochemical
Analysis, Pergamon Press, New York, 1968 p. 54-58.
3.     Murphy, T.J. “The Role of the Analytical Blank in Accurate
Trace Analysis”, Proceedings from the Seventh Materials Research
Symposium; U.S. Government Printing Office, Washington D.C.,
1976. p. 509-539.
4.     Howard, A.G. and Statham,
P.J. Inorganic Trace Analysis: Phi-
losophy and Practice, Wiley, New
York, 1993 Murphy, T.J. “The Role
of the Analytical Blank in Accu-
                                          Appendix A
rate Trace Analysis”, Proceedings          References
from the Seventh Materials Re-
search Symposium; U.S. Govern-
ment Printing Office, Washington
D.C., 1976. p. 509-539.
5.      P.W. Morrison, ed. Contamination Control in Electronic
Manufacturing, Van Nostrand Reinhold, New York, 1973, p. 245.
6.      Specker, H.I. Z. Erzbergbau Metallhue 17, 132 (1964).
7.      Gilbert, H. and Palmer, J.H. High Efficiency Particulate Air
Filter Units, TID-7023 USAEC , Washington D.C., August 1961.
8.      Moody, J.R., Analytical Chemistry 52, (1982) p. 1358A-
1376A.
9.      Whyte, W. (Ed.) Cleanroom Design, Wiley, Chichester,
1991.

                                 91
           Clean Chemistry: Techniques for the Modern Laboratory



10.    Boutron, C.F., Fresenius Journal of Analytical Chem-
istry, 337 (1990) p. 482-491.
11.     Austin, P.R., Encyclopedia of Clean Rooms, Bio-Clean-
rooms and Aseptic Areas, 3rd ed., Acorn Industries, 2000.
12.   Zief, M. and Mitchell, J.L. Contamination Control in
Analytical Chemistry, Wiley, New York, 1976.
13.     Patterson, C.C. and Settle, D.M. “The Reduction of
Orders of Magnitude Errors in Lead Analysis of Biological
Materials and Natural Waters by Evaluating and Control-
ling the Extent and Sources of Industrial Lead Contamina-
tion Introduced During sample collecting, Handling and
Analysis”, Proceedings from the Seventh Materials Research
Symposium; U.S. Government Printing Office, Washington
D.C., 1976. p. 321-351.
14.     Sulcek, Z. and P. Povondra. Methods of Decomposi-
tion in Inorganic Analysis, CRC Press, Boca Raton, FL, 1989.
15.     Hetherington, G., Stephenson, G.W., Witerburn, J.A.
Electronic Engineering, May 1969 p. 52
16.     Stevens, M.P. Polymer Chemistry: An Introduction,
Oxford, New York, 1990
17.     Moody, J.R. and Lindstrom, R.M. Analytical Chemis-
try 49, (1977) p. 2264-2267.
18.     Laxen, D.P.H and Harrison, R.M. Analytical Chemistry
53, (1981) p. 345-350.
19.    Tschopel, P.,Kotz, L., Schulz, W., Veber, G. and Toelg
G. Fresenius’ Z. Anal. Chem. 302, (1980) p. 1
20.    Richter, R.C. Spectroscopy, 16(6), (2001) p. 21-24.
21.    Kuehner, E.C., Alvarez, R., Paulsen, P.J., and Murphy,
T.J. Analytical Chemistry, 44, (1977) p. 2050-2056

                                    92
                       Appendix A: References



22.    Moody, J.R and Beary E.S.Talanta, 29, (1982) p.
1003-1010.
23.    Dabeka, R.W, Mykytiuk, A., Berman, S.S., and Rus-
sell, D.S. Analytical Chemistry, 48, (1976) p. 1203-1207.
24.     Richter, R.C., Link, D., Kingston, H.M., Spectroscopy,
15(1), (2000) p. 38.



For Section Two

1.     Mingos, D. M. P.; Baghurst, D. R. Chem. Soc. Rev. 20,
(1991) p. 1-47.
2.     Hoopes, T.; Neas, E.; Majetich, G. Abstr. Pap. Am.
Chem. Soc. 201, (1991) p. 231.
3.     Baghurst, D. R.; Mingos, D. M. P. J. Chem. Soc., Chem
Commun. (1992) p. 674.
4.      Richter, R.C., Link, D.D., Kingston, H.M. Analytical
Chemistry (2001) 73(1) p30A
5.      Journal of Research of the National Bureau of Stan-
dards 30, (1943) p. 110.
6.      Kingston, H. M.; Walter, P. J. Spectroscopy 7 (1992)
p. 22-27.
7.      Kubrakova, I. V.; Formanovskii, A. A.; Kudinova, T.
F.; Kuz’min, N. M. J-Anal-Chem (1999) p. 460-465.
8.      T. Noltner, Spectroscopy 5(4) 1989.
9.      Visini and Rampazzo, Department of Environmental
Science, University of Venice, Private Communication
10.   Richter, R.C. and H.M. Kingston, Department of
Chemistry, Duquesne University Private Communication.


                                93
           Clean Chemistry: Techniques for the Modern Laboratory



11.    Link, D.L. and Kingston, H.M. Analytical Chemistry
(2000) 72(13) p. 2908-2913.
12.    Link, D.L. and Kingston, H.M., Havirilla, G.L, Colletti,
L.P. Analytical Chemistry (2002) 74(5) p. 1165-1170.
13.    Han, Y., Kingston, H.M., Richter, R.C., Pirola, C.,
Analytical Chemistry (2001) 73(6) p. 1106-1111.
14.     Sulcek, Z. and Povondra, P., Methods of Decomposi-
tion in Inorganic Analysis, CRC Press, Boca Raton, FL, 1989.




                                    94
Appendix A: References




         95
          Clean Chemistry: Techniques for the Modern Laboratory




About the Author

   Dr. Robert C. Richter is a research professor at Chi-
cago State University, where he develops and coordinates
student research projects involving microwave-enhanced
chemistry. He has taught analytical chemistry and sample
preparation, for various universities and the American
Chemical Society, since 1992. He has also served as a
consultant on a wide range of corporate and governmental
projects—developing microwave-assisted methods for a
variety of applications, and conducting seminars on clean
chemistry techniques.

    Dr. Richter is also Senior Applications Chemist for
Milestone’s clean chemistry, microwave digestion, and
microwave extraction product lines. In this role, he has
helped academic and commercial laboratories across
North America to adapt themselves for clean chemistry
procedures and standards.