					                    NOTICE INVITING QUOTATION (NIQ)

Dated: 7.12.2007

The department of Textile Technology invites quotations for purchase of the
following items with specifications and terms and conditions mentioned below; The
vendors may quote one or both items:

1. Differential Scanning Calorimeter (DSC) with modulation (stand alone unit)

Features                  Specification               Remarks
Useable Temperature range Better than -80 oC (min) To Specify range with
of DSC                    700 oC (max)                intracooler

Modulation:                 Yes; sinusoidal                Specify range and types
                            temperature modulation         of modulations
                            superimposed with basic
                            temperature ramping
Cell architecture           Robust measurement cell        User report on this aspect
                            architecture with protection   would be critically
                            of electronics from sample     evaluated
                            vapors and moisture; and
                            easy clean-up of cell;
Temperature Precision        ±0.05 oC or better
Temperature                 ±0.1 oC
reproducibility
IntraCooling arrangement    Both (a) Refrigerated          Specify cooling rate using
with gas drying             cooling system to -80 oC       both arrangements, faster
                            (or better), and               and accurate cooling
                            (b) easily detachable          controls would be
                            quench cooling accessory       preferred
                            using user usable cooling
                            liquid such liq nitrogen
Heating rate                 Min 0.1 oC/min to max.
                            better than 150 oC/min
Sensor                                                     Specify the type used and
                                                           the position of sensors;
                                                           sensors closer to sample
                                                           pans, accurate, and with
                                                           faster response time
                                                           would be preferred
Sensitivity                 0.3 W or better
Baseline reproducibility    ± 15W or better               Curvature of baseline
                                                           should be 10 mW or
                                                           better over 300 oC range;
                                                           easy to recalibrate
                                                           baseline
Calorimetric precision and ±0.1% or better
reproducibility (on Indium)
Sample type                 Solid/powder / Liquid          Anti spilling
Gas Purging                Inert gas purging
Indium height /Width Ratio 15 or better @ 1.0 mg at 10 Specify parameters of test
(mW/oC)                    o
                            C/min under Nitrogen         if different and its effect
                                                         on ratio
Dual mass flow controllers For precise purging, gas      Specify flow rates and
                           mixing and switching          precision;
Sample pan crimper         Hermetically sealed and       Crimper for all types must
                           standard crimping; and        be included
                           other suitable crimping for
                           overflowing materials
Sample pans                Hermetic: specify the
                           quantity
                           Standard: specify the
                           quantity
                           pressure cell: 2Nos.          Define range and quality
                           minimum
Software:                  Modulated DSC should be
                           completely controlled with
                           a user friendly interface;
                           complete analysis and real
                           time separation of kinetic
                           and thermodynamic
                           components of heat flow;
                           plotting and multiple curve
                           analysis with legend
                           marking; full control of all
                           accessories and mass flow
                           controllers; easy export of
                           data; All possible modules
                           of software must be quoted
                           with features of analysis
Controls                   Unit should be controllable
                           independently preferably
                           from a touch panel and
                           from computer software;
                           instrument should be easy
                           to calibrate and baseline
                           should be easy to stabilize
Warranty/Guarantee         Cost of comprehensive 5       Also quote yearly cost for
                           years warranty/guarantee      beyond 5 years for service
                           must be quoted separately; and comprehensive
                           but is must                   warranty.
Service Support and        The company must have         Define response time and
application support        service and application       service network
                           engineers trained and
                           available full time in India;
                           preferably near Delhi
Quality and performance    The decision for technical
                           shortlisting would depend
                           both on specification given
                             above and
                             quality/performance
                             feedback from the current
                             users and information
                             provided by the vendor(s)
Accessories and options      All accessories available
                             for the model must be
                             specified and mentioned
                             with the quote



Thermogravimetric Analyser (TGA); stand alone unit

Features                     Specification
TGA compatibility            with multiple furnace
                             types, including but not
                             limited to the following:
                             • Standard, low mass
                             ceramic furnace
                               • Quartz-lined evolved-
                               gas analysis furnace

Temperature Range         ambient to 1000 C or Specify heating times
                          better
Isothermal    Temperature ±1°C
Accuracy
Isothermal    Temperature ±0.2°C or better
Precision
Heating Rate Range        Min 0.01 oC/min to max.
                          better than 90°C/min in
                          0.01°C/min increments (for
                          standard furnace)

                             Quote heating rates for
                             EGA furnace

Furnace Cooling              Forced Air 1000°C to           Faster cooling would be
                             100°C in < 15 min              preferred
Weighing Capacity            1 gram
Sensitivity                  0.1 g or better
Weighing Precision           ± 0.01% or better
Possibility   of      online With Mass Spectrometer or      Quote connectors as
Interfacing                  FTIR;                          options (if not standard)
mass flow controllers        dual digital; Flow rate must
                             be programmable within
                             operating software and
                             allow     automated     gas-
                             switching      during     an
                            experiment
Balance cooling:            Integrated preferred
Purging                     Furnace design to purge
                            gas flow to minimize
                            buoyancy effects
Calibration                 Automatic                       Specify method
Modes of operation          Fully        Programmable:
                            Isothermal,          dynamic
                            heating     rate,    constant
                            heating rate, or step type
                            isothermal
Spares:                     Quote quantities of spares
                            to be provided with
                            equipment and cost of
                            additional quantity
Software                    Full operation and control
                            of      instrument       with
                            computer; online display
                            and complete analysis of
                            the data
Warranty/Guarantee          Cost of comprehensive 5         Also quote yearly cost for
                            years must be quoted            beyond 5 years.
                            separately; but is must.
Service Support and         The company must have           Define response time and
application support         service and application         service network
                            engineers trained and
                            available full time in India;
                            preferably near Delhi
Quality and performance     The decision for technical
                            shortlisting would depend
                            both on specification given
                            above and
                            quality/performance
                            feedback from the current
                            users and information
                            provided by the vendor(s)
Accessories and options     All accessories available
                            for the ,model must be
                            specified and mentioned
                            with the quote
Must quote Options:         Modulation                      Quote separately if not
                            Evolved      gas     analysis   part of the item
                            furnace


Terms and Conditions

   1. (a) Quotation should be directly from Original manufacturer or authorized
      sales agent. (b) Relevant agency certificates, validity of agency certificate
      from Foreign Principals (in case of foreign manufacturer), (c) proprietary
   certificate (if any, for any component or instrument quoted) must be submitted
   along with quotation.

2. (a) Prices should be CIF Delhi inclusive of all taxes, delivery charges,
   installation and training charges; IIT Delhi is exempt from central excise and
   custom duty, Delivery Period should be specifically mentioned and should be
   as small as possible (b) guarantee or warranty conditions must be clearly
   specified, (c) validity of quotation must be till end of March 2008, (d) Mode of
   payment through LC only (any kind of advance payment clause should be
   avoided), (e) Details about installation, commissioning of & training on the
   equipment be mentioned, (f) Supplier must submit TIN number/PAN number
   as applicable.

3. Only those manufacturers who have manufactured and supplied the above
   mentioned equipment to more than one nationally reputed Indian research
   institutions or laboratories need to submit quotations with complete address,
   Tel. No. and name of the user. Quotations without proper and complete
   references related to functioning of the quoted equipments will not be
   entertained.

4. Technical bids (with certificate of compliance and deviations from the
   specifications) and financial bids should be submitted separately in sealed
   covers by 28 December 2007 by 12 Noon Indian standard time to the
   following address:

           Attn of: Dr. Ashwini K. Agrawal
           Room No. TX- 206 or TX- 216
           Dept. of Textile Technology
           Indian Institute of Technology, Delhi
           Hauz Khas, New Delhi 110016, India

				