Affinity constants from SPR equilibrium analysis by dfsiopmhy6

VIEWS: 48 PAGES: 2

									                                                            AUTOLAB APPLICATION NOTE

   Affinity constants from SPR equilibrium                  To test the binding capacity of the surface the analyte    can be determined. It is always important to check
                   analysis                                 in a concentration of approximately 10 times the           the quality of the fits!
                                                            estimated binding constant is injected and the SPR         Req as function of the free analyte concentration can
Two ways to assay affinity constants for biomolecular       signal at equilibrium assayed. Preferably this would       be plotted (Fig. 2) and non-linearly fitted to a one-site
interactions are described. Firstly, the affinity for an    be in the range of 60 to 100 mº. If it is too low the      binding isotherm (Eq. 1).
immobilized ligand (KC), and secondly, affinity of the      assay will be less accurate. If it is too high, the
interaction in solution (KS), using SPR competition         interaction can be slow, especially with low analyte               [analyte] free            
experiments. Although affinity constants can be             concentration; the correction for analyte depletion         Req =                            .Rmax (1)
                                                            (see below) can be large; and the effects of mass                  [analyte] + K             
derived from kinetic information, here the equilibrium
                                                            transport will be more evident.
                                                                                                                                        free  C          
signal is used, to avoid complications of mass
transport limitation and the use of binding models that       In the actual experiment solutions of analyte are
                                                            made in a concentration range around the estimated         With non-linear fitting programs (not our KE program)
may not be appropriate for the system under                                                                            readily available now, a non-linear fit is preferred over
investigation.1 The cuvette design of the Autolab SPR       KD value. Solutions of analyte are made in buffer e.g.
                                                            HBS buffer pH 7.4. The highest concentration should        a linear fit e.g. from a Scatchard plot. The fit yields
is pre-eminently suitable to obtain the signal at                                                                      values for KC and Rmax, the value of Req at saturation
equilibrium.                                                be approximately 5-10 times the estimated KD value,
                                                            the lowest concentration at least equal to KD, but         of binding.
The species (a protein or other macromolecule) that
is added in solution and binds to the immobilized           preferably lower. However, at low concentrations it
                                                            can take a long time before equilibrium is reached!        Depletion correction
ligand is defined as the analyte.                                                                                      In the cuvette-based ESPRIT instrument the
                                                            For an interaction plot at least 5 data-points are
                                                            needed. Sensorgrams of the analyte interactions are        concentration of free analyte decreases due to
Affinity of an analyte to an immobilized                                                                               binding to the surface, for this a correction can be
                                                            recorded (Fig. 1). From the sensorgrams the SPR
ligand on the SPR sensor chip.                                                                                         made according to Eq. 2.
                                                                                                                                                2
                                                                                           signal at equilibrium is
                                                                                           determined. This is                                                Req .S.10−9
After initialization of the SPR chip, the initial step is                                  easily done with the        [analyte free = [analyte total −
                                                                                                                              ]               ]                                 (2)
coupling of one of the reactants to the SPR chip (see                                      Autolab           kinetic                                      122.MWanalyte.Vbulk
Application Note 31). The conditions can vary much                                         software. Make an
and depend amongst others on isoelectric point and                                         overlay of the blank        S of the standard sample cell is 2.6 mm2, Vbulk under
the type of sensor chip. The following is based on                                         corrected curves of         standard conditions is 35 µL. MW is the molecular
EDC/NHS as the coupling strategy. To allow                                                 interest, position the      weight of the analyte. The effect of depletion
complete access to the analyte-binding site, for small                                     baseline at the start on    correction is shown in
ligands insertion of a spacer between the binding                                          zero for all curves, and    Fig. 2: without correction
epitope and the sensor dextran matrix is advisable.                                        fit a part of the           (dotted     line,      open
When a stable surface is needed covalent coupling is                                       association        phase    symbols) KC in this
preferred over non-covalent immobilization like             including a part of the equilibrium signal, with the       example is found to be
biotin/avidin coupling. Following the standard              monophasic association model. The equilibrium              8.0 nM, and with
EDC/NHS procedure, as a rule of thumb one could             signal (Req) is given by the fit parameters E+R(0). If     correction          (closed
start with 1-2 mM ligand reacting during 5-10 min with      equilibrium is not reached, e.g. for a low                 symbols, solid line) 5.5
the activated sensor surface (see Application note          concentration (see Fig. 1), using this method also Req     nM.     The       depletion
31).


AUTOLAB ELECTROCHEMICAL INSTRUMENTS ARE DEVELOPED AND PRODUCED BY ECO CHEMIE B.V. IN THE NETHERLANDS                                   appl032-1
www.ecochemie.nl.
                                                          AUTOLAB APPLICATION NOTE

correction can be readily calculated in a spread                                               A − A 2 − 4[analyte] total .[ligand] total
                                                                                                                                            extremely flexible proteins if their conformational
sheet.                                                    [analyte] free = [analyte] total −                                                freedom is limited by binding to the sensor surface.
                                                                                                                   2
                                                                                                                                            Deviations between KS and KC are also observed if
Affinity of an analyte to a ligand in solution.                                                                                             not the proper value of the analyte concentration is
                                                          in which A = [ K S ] + [analyte]total + [ligand ]total                            used.
From SPR competition experiments the equilibrium                                                                                              The big advantage of this approach is that series of
constant of a molecular interaction in solution (KS)      Substitution of Eq. 4 into Eq. 1 gives the complete                               ligands can be investigated using the same sensor
can be derived. KS is defined by Eq. 3.                   expression, which can be introduced into a non-linear                             surface, without preparing individual surfaces for
                                                          fitting program. It should be noted that this                                     every ligand.
       [analyte] free .[ligand ] free, solution           expression contains KC, therefore the affinity to the
KS =                                              (3)     sensor surface must be determined (section A of this                              References
       [analyte − ligand ]complex, solution               Note). In the fit the numerical values of [analyte]total
                                                          and KC is introduced, and the fit yields values for KS                            (1)    Schuck, P.; Minton, A. P. Kinetic analysis of
                                The experiments are       and Rmax.                                                                         biosensor data: elementary tests for self-consistency.
                                performed          with      A few remarks concerning the application of this                               Trends Biochem Sci 1996, 21, 458-460.
                                samples containing a      approach:
                                constant concentration    If binding constants are not needed, and only the                                 (2)      de Mol, N. J.; Plomp, E.; Fischer, M. J. E.;
                                of analyte in the         relative affinity of ligands has to be established, from                          Ruijtenbeek, R. Kinetic analysis of the mass transport
                                presence       of     a   the inhibition curves IC50-values can be derived                                  limited interaction between the tyrosine kinase lck
                                concentration range of    without fitting to the KS model.                                                  SH2 domain and a phosphorylated peptide studied
                                ligand. Req determined    The fixed analyte concentration in the competition                                by a new cuvette-based surface plasmon resonance
                                from the sensorgrams      experiments is best chosen as 1 to 4 times KC.                                    instrument. Anal Biochem 2000, 279, 61-70.
                                (see section A of this    In the fit all concentrations should be in the same unit
                                Note) is plotted as a     (M, µM or nM). KS will be returned in that same                                   (3)    de Mol, N. J.; Gillies, M. B.; Fischer, M. J. E.
                                function of the ligand    concentration unit.                                                               Experimental and calculated shift in pK(a) upon
concentration in solution (Fig. 3). In the presence of    The returned value of Rmax is the value at complete                               binding of phosphotyrosine peptide to the SH2
competing ligand Req is determined by the affinity to     saturation of all binding sites and should be                                     domain of p56(lck). Bioorg Med Chem 2002, 10,
the immobilized ligand and the amount and affinity of     compatible with the value from the KC experiment.                                 1477-1482.
the ligand in solution. The amount of sensor-bound        Also in these experiments loading of the chip should
analyte is again described by Eq. 1, but now the          not be too high, as correction for depletion is not
amount of free analyte is diminished by binding to the    feasible here. For lower MW analytes one has to note
ligand in solution. Based on this a modification of Eq.   especially this point (compare Eq. 2)
1 can be derived which fits the data of Fig. 3 3. Here
the expression of [analyte]free is given for the          In general for a bimolecular interaction the values for
competition conditions (Eq. 4).                           KC and KS are found to be similar. Deviations can be
                                                          found for dimer interactions (e.g. GST-fusion
(4)                                                       proteins), high MW analytes (MW >100 KDa) due to
                                                          lower partition in the dextran layer of the sensor, and


AUTOLAB ELECTROCHEMICAL INSTRUMENTS ARE DEVELOPED AND PRODUCED BY ECO CHEMIE B.V. IN THE NETHERLANDS                                                      appl032-2
www.ecochemie.nl.

								
To top