					PET QUICK REFERENCE - 04/04/07
                       DISCOVERY VCT
gantry/table flr.
anchored?                      YES
minimum floor
thickness                5" conrete floor
                       below gantry (right
gantry cable             side of unit iso
access                       center)
conduit sizes
-OC to Gantry                 3 1/2"
-OC to PDU                    2 1/2"
-PDU to Gantry          3 1/2" and 3 1/2"


PDU & UPS               39" to gantry and
interference limits            OC
gauss field                OC = 1G
proximity limit           Gantry = 1G
                        OC - Gtry = 75'
standard cable          OC - PDU = 75'
lengths                 PDU - Gtry =55'
alternate cable
lengths                        NO
voltage/amps               480V/125A
MDC cat NO.                 E4502AE
UPS                         P5052PS
recommended
ceiling height                 9'-0"
absolute min.
ceiling height                 8'-0"
absolute min. room
size                     25'-6" X 14'-0"
                      5 ft. corr = 80"
 Delivery: min.
                6 ft. corr = 68"
 opening, incl.
                7 ft. corr = 56"
    Dollies
                      8 ft. corr = 44"
PET QUICK REFERENCE - 04/04/07
                       DISCOVERY STE
gantry/table flr.
anchored?                      YES
minimum floor
thickness                5" conrete floor
                       below gantry (right
gantry cable             side of unit iso
access                       center)
conduit sizes
-OC to Gantry                 3 1/2"
-OC to PDU                    2 1/2"
-PDU to Gantry          3 1/2" and 3 1/2"


PDU & UPS               39" to gantry and
interference limits            OC
gauss field                OC = 1G
proximity limit           Gantry = 1G
                        OC - Gtry = 75'
standard cable          OC - PDU = 75'
lengths                 PDU - Gtry =55'
alternate cable
lengths                        NO
voltage/amps                480V/90A
MDC cat NO.                 P5051RJ
UPS                         P5052PS
recommended
ceiling height                 9'-0"
absolute min.
ceiling height                 8'-0"
absolute min. room
size                     24'-6" X 14'-0"
                      5 ft. corr = 80"
 Delivery: min.
                6 ft. corr = 68"
 opening, incl.
                7 ft. corr = 56"
    Dollies
                      8 ft. corr = 44"
PET QUICK REFERENCE - 04/04/07
                        DISCOVERY ST
                            HP60
gantry/table flr.
anchored?                      YES
minimum floor
thickness               5" concrete floor
                       below gantry (right
gantry cable             side of unit iso
access                       center)
conduit sizes
-OC to Gantry                 3 1/2"
-OC to PDU                    2 1/2"
-PDU to Gantry          3 1/2" and 3 1/2"


PDU & UPS               39" to gantry and
interference limits            OC
gauss field                OC = 1G
proximity limit           Gantry = 1G
                        OC - Gtry = 75'
standard cable          OC - PDU = 75'
lengths                 PDU - Gtry =55'
alternate cable
lengths                        NO
                          480V/90A
voltage/amps            P5051RJ (HP60
MDC cat NO.                 ONLY)
                        P5052PS (HP60
UPS                         ONLY)
recommended
ceiling height                 9'-0"
absolute min.
ceiling height                 8'-0"
absolute min. room
size                      14'-0" x 24'-6"
                      5 ft. corr = 80"
 Delivery: min.
                6 ft. corr = 68"
 opening, incl.
                7 ft. corr = 56"
    Dollies
                      8 ft. corr = 44"
PET QUICK REFERENCE - 04/04/07
                    DISCOVERY ST (gold
                          seal)
gantry/table flr.
  anchored?                   YES
minumum floor
   thicknes             5" conrete floor

 gantry cable        Below Gantry (Right
   access           side of unit iso center)

 conduit sizes
 -OC to Gantry                3 1/2"
  -OC to PDU                  2 1/2"
-PDU to Gantry          3 1/2" and 3 1/2"
 PDU & UPS
 interference
     limits          39" to gantry and OC

  gauss field              OC = 1G
proximity limit           Gantry = 1G
                        OC - Gtry = 75'
standard cable          OC - PDU = 75'
    lengths             PDU - Gtry = 55'
alternate cable
    lengths                    NO
                          480V/110A
                           P5050RB
 voltage/amps          (DISCOVERY ST
 MDC cat NO.                ONLY)
                           P5051PS
                       (DISCOVERY ST
      UPS                   ONLY)

recommended
ceiling height                9'-0"
 absolute min.
 ceiling height               8'-0"
 absolute min.
   room size             14'-6" x 25'-0"
  Delivery:         5 ft. corr = 80"
    min.            6 ft. corr = 68"
  opening,          7 ft. corr = 56"
incl. Dollies       8 ft. corr = 44"
PET QUICK REFERENCE - 04/04/07
                    DISCOVERY LS
                     (GOLDSEAL)
gantry/table flr.
anchored?                  YES

                        Must be
                     determined by
                        qualified
                       Structural
minimum floor       Engineering prior
thickness              to delivery
gantry cable        REAR & BELOW
access                 GANTRY
                        NOT
                    RECOMMENDE
conduit sizes             D

PDU & UPS           39" to gantry and
interference limits        OC
gauss field             OC = 1G
proximity limit        Gantry = 1G

                    OC - Gtry = 50'
                     OC - PM = 50'
                    PM - Gtry = 35'
                    Grty-petoc=100'
standard cable      EC-Gtry = 100'
lengths             petoc-EC =100'
                    CT PART ONLY
alternate cable       (CONTACT
lengths              GOLDSEAL)

voltage/amps
MDC cat NO.            480V/110A
w/transformer           P5050AH
UPS                     P5050UP
recommended
ceiling height             9'-0"
absolute min.
ceiling height             8'-0"
absolute min.
sacn room size       23''-0" X 14'-0"
absolute min.
equipment room
size                   6'-0" X 9'-0"
                    5 ft. corr = 80"
Delivery: min.
               6 ft. corr = 68"
opening, incl.
               7 ft. corr = 56"
   Dollies
                    8 ft. corr = 44"
PET QUICK REFERENCE - 04/04/07
                       ADVANCE NXI
                       (GOLDSEAL)
gantry/table flr.
anchored?                    YES
gantry cable
access                      REAR

conduit sizes
-EC to Gantry          (3) 4" & (1) 3"
-OC to EC              (1) 4" & (1) 3"
-OC to Gantry               (1) 4"
PDU & UPS
interference            20" to gantry
limits                   39" to OC

gauss field              OC = 10G
proximity limit         Gantry = 1G
                     OC - Gantry = 100'
standard cable        OC - EC = 100'
lengths              EC - Gantry = 100'
alternate cable
lengths                      NO
voltage/amps              208V/60A
MDC cat NO.               E4502AD



UPS                      E45502JK
recommended
ceiling height               8'-0"

absolute min.
ceiling height               7'-8"

absolute min.
scan room size         20'-0" X 14'-0"
absolute min.
equipment room
size                     6'-0" X 9'-0"
                    5ft. Corr = n/a
   Delivery:
               6ft. Corr = n/a
min. opening,
               7ft. Corr = n/a
 incl. Dollies
                    8 ft. corr = 44"