									CT Scanner QC
Presented by:
David E. Hintenlang, Ph.D., DABR Medical Physics Program Director Department of Nuclear and Radiological Engineering Department of Biomedical Engineering American College of Medical Physics 2004 Annual Meeting Scottsdale Arizona
University of Florida

CT Quality Control Motivations
• Appropriate diagnostic benefit vs risk • Image Quality Noise Contrast Spatial resolution • Radiation Risk Dosimetry indices (ie. CTDI)

University of Florida

Presentation Overview
• • • • • • • • Introduction & Background General categories of QC tests Specific QC tests and procedures QC Program recommendations & comparisons Acceptance tests “Annual” QC tests “Daily” QC tests Conclusions & References

University of Florida

Background
Expanded CT capabilities over last decade • Helical and Multislice CT • Decreases scan times - Changes in patient dose • Increased utilization of CT • Complex systems, high utilization, large contribution to patient dose demands a continuous and robust QA program
University of Florida

QC Objectives
Complexity of system offers opportunities for many individual QC tests Level of testing balanced against QC Objectives QC Categories: • Acceptance Testing • Annual Testing • Daily QC • ACR Accreditation
University of Florida

Categories of QC Tests
• • • • Safety Mechanical accuracy Imaging Performance Display System

University of Florida

General QC Tests : Safety
• • • • Electrical Mechanical Laser Radiation

University of Florida

General QC Tests: Mechanical Accuracy
Ensure that aligned patient anatomy is imaged • Alignment light accuracy • Alignment of table and gantry • Table and gantry positioning • Slice localization from scout images

University of Florida

General QC Tests: Image Quality
•Noise and Field Uniformity •CT Number Linearity •Low Contrast Detectability •Spatial Resolution •Display and Hard Copy Image Quality

University of Florida

General QC Tests: Dosimetry
• CTDI • Head & Body • Patient Dosimetry

University of Florida

Alignment Laser Accuracy
Internal & External scan plane lights • Verify correct distance between internal & external scan plane lights • Wrap film around centered phantom • Make pin hole marks along illuminated lines

University of Florida

• Scan 1 slice @ 1mm thickness • Repeat for external alignment laser • Measure distance between line of pierced holes and center of exposed slice

• Misalignment should be less than 2 mm
University of Florida

Table - Gantry Alignment
To ensure that longitudinal axis of table is aligned with isocenter • Measure to locate & mark central axis of table at: head end middle foot end • Hang plumb bob from center of gantry - to table top • Measure distances between plumb bob and central axis of table • Repeat for various gantry angles
University of Florida

Table should be aligned within +/- 5 mm

Table Incrementation Accuracy
Evaluate accuracy/ reproducibility of longitudinal table motion • Tape ruler along table edge • Tape pointer on frame around mid-point of ruler • Load table • Move table 300-500 mm in one direction & return • Repeat for a total of 3-times in each direction Average & Standard Deviation should be less than 3 mm

University of Florida

Table Incrementation Accuracy & Collimation (axial scan)
Evaluate accuracy & reproducibility of longitudinal table motion & width of x-ray beam • Wrap film around centered head phantom • Scan 5 slices for several combinations of slice thickness and gap. • Measure slice thickness and gaps from film • Mark table frame at one end of the phantom • Return to original table position and note any difference from original position

Interslice gap should have < 0.5 mm discrepancy Slice thickness should be accurate to within 1 mm
University of Florida

Table Incrementation, scanned volume, helical pitch accuracy
Evaluate accuracy & reproducibility of longitudinal table motion & helical pitch • Wrap film around centered head phantom • Mark table frame at one end of phantom • Record indicated table position • Obtain volume scans for various slice thicknesses and pitch Measure from film: Total length scanned Spiral slice thickness Spiral gaps (pitch)
University of Florida

kVp Accuracy
• Lock x-ray tube position • Center voltage divider using CT specific filer pack in center of x-ray beam • Check kVp’s at : 80, 100, 140 kVp Tube potential should be within +/- 5%
University of Florida

Half-Value Layer
• Lock x-ray tube position • Set scanning parameters per manufacturer (80 kVp and mAs to obtain 200 mR) • Mount CT chamber parallel to table long axis, centered on isocenter • Vary Al thicknesses Should be at least 2.3 mm Al @ 80 kVp
University of Florida

Exposure Reproduceabiity & Linearity
• Head phantom in head holder on table • Insert CT ion chamber in center of phantom • Perform 4 scans at 80 kVp, 100 mA, 1s • Record exposure for each • Repeat at: 200 mA, 300 mA • Repeat at : 120 kVp, 140 kVp
University of Florida

Radiation Profile Width
Evaluates the accuracy of pre-patient collimator settings • Tape film around centered head phantom • Scan one slice at each slice thickness • Measure width of each slice
Lowest kVp and tube current available space about 2 cm apart
positioned flat on table with 0o gantry angle

Should be within 1 mm of selected thickness
University of Florida

Slice Sensitivity Profile
Evaluate the actual width of the imaged slice • Set standard scanning parameters • Scan phantom insert with “ramps” for each available slice thickness • For each image determine average pixel values of: a) metal ramp and b) acrylic background • Set window width < 10 HU of min width setting • Set window level at average of ramp and background values • Measure width of ramp and multiply by tanq
University of Florida

University of Florida

Image Quality Measures
• • • • • • Noise and Field Uniformity CT Number Linearity Low Contrast Detectability Spatial Resolution Reconstruction times Scout view accuracy

University of Florida

Phantoms & Test Tools
• Testing phantoms available from manufacturers & third parties • Early CT phantoms may not provide sensitivity require to test modern CT systems

University of Florida

Noise & Field Uniformity Axial Scan
Assess noise level under simulated clinical conditions Set standard scanning parameters Position a centered homogenous phantom Scan and select ROI’s Obtain mean pixel value and standard deviation for each ROI Std. Dev. Central ROI < 10 HU Max difference between peripheral and central ROI’s < +/- 5 HU
University of Florida

Noise & Field Uniformity Volume Scan
Assess noise level under simulated clinical conditions • Set standard scanning parameters • Position a centered homogenous phantom • For each voume scanned measure 3 reconstructed slices • center & both ends of volume • Scan and select ROI’s • Obtain mean pixel value and standard deviation for each ROI Std. Dev. Central ROI < 10 HU Max difference between peripheral and central ROI’s < +/- 5 HU

University of Florida

CT Number Linearity
CT numbers should be consistent from system to system • Set standard scanning parameters • Image phantom insert with 5 rods of different materials • Determine mean pixel values of each object & background • Perform for low, mid, & high kVp settings CT number should be within 50 HU for each material

University of Florida

Low Contrast Detectability
Evaluate scanner low contrast object discrimination • Set standard scanning parameters • Scan low-contrast detectability insert • Determine mean value of each hole & surrounding material • Contrast = difference in HU of objects & background • Determine Number of objects visible Should meet manufacturer’s specifications
University of Florida

Spatial Resolution
Evaluate high contrast spatial resolution • Set standard scanning parameters • Scan Spatial Resolution Insert • Determine highest frequency group visually resolved for each image

Should meet manufacturer specifications
University of Florida

Spatial Resolution MTF
Derive the MTF as a measure of limiting spatial resolution • Scan spatial resolution insert • For each group of high contrast resolution patterns determine the mean value & std. deviation. • Determine & plot MTF • Repeat using bone reconstruction filters

Should meet manufacturer’s specifications
University of Florida

Image Reconstruction Time
Evaluate reconstruction times • Set parameters for typical head scan • Scan anthropomorphic head phantom • Measure time required to obtain 1 image slice • Subtract out data collection time • Repeat for anthropomorphic chest phantom and typical body slice Should meet manufacturer specifications
University of Florida

Scout Views
Evaluate scout view production • Position anthropomorphic head phantom • Perform scout view at 120 kVp, 160 mA, 1 mm thickness • Scan length 0  500 mm • Perform scout views for 0, 45 and 90 degree gantry angles • Record time required to perform and print images • Repeat with anthropomorphic chest phantoms
University of Florida

Display and Hard Copy Image Quality
Ensure that information in the video signal is accurately transferred to display media • Reduce room illumination to normal viewing level • Display SMPTE Pattern • Adjust brightness and contrast controls

5% and 95% patches should be visible
University of Florida

Dosimetry Measurement: CTDI
Determine a measure of radiation dose to tissue under different scan conditions • Set standard scanning parameters (i.e. head) • Position CTDI head phantom • Scan with CT ion chamber inserted sequentially in each of 5 positions • Calculate CTDI using peripheral and central values • Repeat with CTDI body phantom • Repeat for various scanning parameters as desired

University of Florida

CT Equipment Quality Control Program
• ACR Accreditation Program for CT • QC program established & implemented under supervision of a qualified medical physicist • Requirements
• Acceptance testing upon installation • Annual medical physicist survey • Continuous Quality Control
University of Florida

Acceptance Testing
• Series of measurements performed to verify that a CT system conforms to required specifications • Specific tests/technical specifications should be included in bid request or purchase agreement • Many more combinations of operating parameters exist than can be practically tested • Must acquire sufficient data to adequately characterize system performance
University of Florida

Considerations for Acceptance Tests
• Determine most important performance specs to test in view of:
• Clinical needs • Scanner capabilities

• Devise a short list of 4-5 “standard” scanning conditions:
• • • • • Standard body Standard head Spine Pediatric Body Pediatric Head

• Pertinent performance tests should be performed for each standard condition to provide a performance baseline • Select a few additional scanner conditions that test performance limits
University of Florida

AAPM Report #39 Recommendations
Essential Tests Localization light accuracy Table Incrementation Radiation Profile Sensitivity Profile Image Noise Field Uniformity High Contrast Resolution Display & Hard Copy Image Quality Radiation Dose Optional Tests
Alignment of table to gantry Table gantry tilt mAs Linearity kVp Accuracy Scan Field Dependence (QCT)

Alternative Tests
Slice localization from Scout image High contrast resolution – MTF Low contrast detectability

University of Florida

Annual ACR Requirements
Localization light accuracy Alignment of table to gantry Table gantry tilt Slice localization from Scout image Table Incrementation Slice thickness -Sensitivity Profile Image Noise Field Uniformity High Contrast Resolution Low Contrast Detectability Artifact Evaluation
University of Florida

CT Number Accuracy & Linearity Display & Hard Copy Image Quality Radiation Dose – CTDI Patient Radiation Dose for select exams Safety Evaluation Visual inspection Audible/visual signals Posting requirements Scattered radiation measurements Other regulatory tests

Monthly to Semi-Annual Tests per AAPM Report # 74
Localization light accuracy Slice localization from Scout image Table Incrementation Slice thickness -Sensitivity Profile Image Noise Field Uniformity High Contrast Resolution Low Contrast Detectability Artifact Evaluation CT Number Accuracy & Linearity Display & Hard Copy Image Quality Radiation Dose – CTDI Safety Evaluation Visual inspection Audible/visual signals Posting requirements Scattered radiation measurements Specified by ACR but not AAPM 74 Patient Radiation Dose for select exams Alignment of table to gantry Table gantry tilt

Dosimetry from Scout Image
University of Florida

Continuous Quality Control
• Required for all CT units • Qualified medical physicist should determine tests, frequency, and responsible individual
• Facility Considerations • CT Usage

• On-site radiological technologist should be identified with responsibility for conducting routine QC
University of Florida

Continuous Quality Control Tests
Localization light accuracy Slice thickness Image Noise Field Uniformity High Contrast Resolution Low Contrast Detectability Artifact Evaluation CT Number Accuracy Display & Hard Copy Image Quality
University of Florida

Daily Tests- AAPM Report #74
CT Number Accuracy Image Noise Field Uniformity Artifact Evaluation All can be obtained from same CT phantom scans Sensitive to a wide variety of scanner problems
University of Florida

i.e. AutoQA LiteTM

Aids to Daily QC
1

• A software program to evaluate and monitor performance of CT scanners • Provides database tool for tracking a limited number of tests: • Image Noise • CT Number accuracy • Field Uniformity • Slice Width • Spatial Resolution
1. The Institute for Radiological Image Sciences, Inc. , MedSoft Technologies, Inc.
University of Florida

AutoQA Lite Overview
• Provides automated analysis of CT phantom images CatPhan series ACR Accreditation phantom Toshiba InnerVision Vendor specific (GE, Siemens, and Philips) • Operates on PC with Dicom 3.0 CT image formats
University of Florida

L

University of Florida

Conclusions
• ACR provides an excellent outline for CT QC Programs • A wide variety of tests exist • Medical Physicist has latitude to define requirements for “continuous” QC testing

University of Florida

CT QC References:
• American College of Radiology Website • AAPM Report No. 39, Specification and acceptance testing of computed tomography scanners, American Institute of Physics, 1993. • AAPM Report No. 74, Quality control in diagnostic radiology, Medical Physics Publishing, 2002. • ImPACT, CT Scanner Acceptance Testing, Version 1.02, 2001 • The ACR CT Accreditation Program and the Medical Physicist, Maynard High, ACMP 2003 Annual Meeting
• www.acmp.org/meetings.html
University of Florida

• www.acr.org (links to “accreditation”)

• www.aapm.org/pubs/reports/

• www.aapm.org/pubs/reports/

• www.impactscan.org/acceptance.htm


								
To top