Routine Measurement of Calcium, Magnesium, Copper, Zinc

Document Sample
scope of work template
							CLIN.CHEM.32/9, 1660-1665 (1986)




RoutineMeasurementof Calcium,Magnesium,Copper,Zinc, and Ironin
                            CoupledPlasmaEmissionSpectroscopy
Urineand Serum by Inductively
David E. Nixon,1 Thomas P. Moyer,’ Peter Johnson,’                     John T. McCall,’    Anders B. Ness,2 Wayne H.
Fjerstad,2 and Mark B. Wehde2

We describe an inductivelycoupled plasma atomicemission                  (1-5). Comprehensive reviews of this technique have been
spectrometer that has been adapted to perform routine,                   published (6, 7). Briefly, the advantages of ICP are:
simultaneous, direct analyses of calcium, magnesium, cop-                    Multi-element
                                                                           #{149}              analysis is truly simultaneous.
per, zinc, and iron in serum or urine without sample digestion               The
                                                                           #{149}dynamic range of linear calibration typically covers
or pretreatment. The system, constructed with inexpensive,               four to five orders of magnitude of concentration.
readily available components, can analyze 1-mL or smaller                  #{149} detection limits and sensitivities are as good as if not
                                                                             The
samples. Results correlate nearly perfectly with those de-               better than thosefor FAAS.
rived by standard atomic absorption techniques (r = 0.98 to                  Solute-vaporization-type
                                                                           #{149}                          interferences do not occur.
                                                                            The obvious advantages, then, of this technique are that
0.997). Using certified serum and urine samples from various
                                                                         these five elements can be determined in one sample simul-
sources, we demonstrate that the instrument yields accurate
                                                                         taneously, with no serial dilutions of specimens, and the
results with a precision better than certified values. The
                                                                         completeanalysis can be performed on one instrument.
instrument is sensitive to one order of magnitude less than
                                                                            In the last 15 years, numerous papers dealing with the
the lower limit of the normal range in serum or urine for all
                                                                         ICP determination       of analytes in biomedical samples have
elements tested, and respondslinearlyto concentrations    two            been published [see reviews by Delves (8), Morrison (9),
orders of magnitude higher than the upper limit of the normal            Mermet and Hubert (10), and Barnes (11)]. Trace elements
range. With the system described here, these five elements               have been determined         in urine, serum, blood, tissue, and
can be assayed with the same or less technical effort than               other biological specimens. Most of this research has been
needed for a single element by atomic absorption.                        published by instrument         manufacturers,  universities, and
                                                                         laboratoriesassociated with the federal government.Except
AddItIonal Keyphrasee: atomic absorption          spectrophotometry      for Schramel et al. (12), we know of no other reported ICP
compared           multi-element analysis                                results for more than control specimens or, at best, a few
                                                                         pooled clinical samples.Unfortunately, an overviewof these
   Historically,      flame (FAAS) and graphite     furnace   atomic     papers yields little methodology that is directly transferable
absorption  spectrophotometry    (GFAAS) have been used by               to the clinical laboratory facing routine analyses of many
many clinical laboratories    to determine  calcium, magne-              urine and serum specimens each day.
sium, copper, zinc, and iron in serum and urine.3 Although                  Here we describethe direct, routine, automated determi-
these techniques provide excellent results, each has limita-             nation of Ca, Mg, Cu, Zn, and Fe in urine or serum by ICP.
tions unique to atomic absorption spectrophotometry:                     We describethe complete analytical facility and stream-
  #{149} one element can be determined at a time.
    Only                                                                 lined sample treatment      and calibration procedures that
    Limited dynamic range forces the dilution of high sam-
  #{149}                                                                 allow the direct determination   of these five elements with
ples.                                                                    no sample digestion or preconcentration.    To establish accu-
  #{149} techniques have limited sensitivity for Cu.
    FAAS                                                                 racy, we analyzed 24-h urine collections and serum speci-
  #{149} is sensitive to solute-vaporization
    FAAS                                      interferences              mens by this technique and by FAAS and GFAAS. Certified
such as the depression of the Ca signal by phosphate.                    urine and serum control specimens were also analyzed and
   Early investigators of inductively coupled plasma (ICP)               the results compared with values obtained by either FAAS
techniques found that many of the limitations     associated             or GFAAS (for copper).
with FAAS techniques were effectively eliminated by ICP
                                                                          Materials and Methods

  Department of’ Laboratory Medicine (H-400) and2 Engineering,            Multi-Element ICP System
Mayo Clinic, Rochester, MN 55905.                                           A schematic diagram of the entire ICP system is shown in
  Nonstandard      abbreviations: FAAS, flame atomic absorption                                      of
                                                                          Figure 1. The components our six-element direct-reading
spectrophotometry;    GFAAS, graphite atomic absorption spec-
trophotometry;ICP, inductively coupled plasma atomic emission             system are identified in Table 1.
spectroscopy.                                                               The 1-rn-focal-length  spectrometer  was designed as a
  Received April 17, 1986; accepted June 6, 1986.                         convertible instrument,  with both a grating drive and an

1660    CLINICAL CHEMISTRY, Vol. 32, No. 9, 1986
                                                                    master program of the IBM-PC controls the preparation          of
                                                                    the ICP for analysis; the compilation of patients’ demo-
                                                                    graphic data from our Laboratory Information         System; the
                                                                    collection and calculation    of data for calibration, control,
                                                                    and sample results; and the management of analysis ifies.
                                                                       The fused quartz plasma torch is based on a design by
                                                                    Scottet al. (1). The preciselyboredand ground quartz tubing
                                                                    was obtained      from Wilmad     Glass Co., Inc., Buena, NJ
                                                                    08310. We restricted the aerosoltip of our torch to 1.2 mm,
                                                                    to direct a narrow stream of argon and sample aerosol into
                                                                    the ICP. The nebulizer used for sample introduction is a
                                                                    modified flxed-crossflow device (Model N058-0368) with
                                                                    spray chamber (Model N058-0368; both from Perkin-Elmer
                                                                    Corp., Norwalk, CT 06856). We replaced the standard
                                                                    molded plastic gas inlet to the nebulizer with a smaller-
                                                                    diameter bored capillary, i.e., a precision-bore    (0.1778 mm)
FIg. 1. Schematic diagramof ICP system                              glass capillary 2.54 mm (o.d.) x 13.5 mm long (Wilmad
                                                                    Glass Co.), attached to a stainless-steel    stop ring 2.21 mm
          Table 1. Components of the System for ICP                 long x 5.23 mm (o.d.) x 2.54 mm (i.d.). This replacement
                                                                    capillary has essentially the same outside dimensions as the
                                                                    original Perkin-Elmer     molded capillary except for overall
Plasma                     Plasma-Therm. Inc., Kresson, NJ          length. We increased the path-length of penetration into the
  High-frequency generator Model HFS-2500E
                                                                    nebulizer body to bring the gas orifice closer to the back of
  Torch                    One-piece quartz construction
                             18 mm I.d. plasma tube                 the sample needle. No modification of the standard Perkin-
                                14 mmi.d.auxiliaryubet              Elmer nebulizer body was needed. The modifications to
Flow meters                      1.2mml.d. aerosol tip              produce this sample introduction       system are shown dia-
                             Linde 201-4335; 201-4334               grammatically in Figure 2. A more detailed discussion of
Monochromater                McPhersonModel216 (1-rn focal          these modifications is presented elsewhere (ms. accepted for
                             length)
  Grating                    1200 grooves/mm;300-nm blaze           publication in Anal Chem).
                             angle                                     The other nebulizer evaluated was a flxed-crossflow de-
  Slits                      Entrance: 25 m x 4 mm                  vice (Model 90-790; Jarrell-Ash, Division of Allied Analyti-
                             Exit: ARL 6910-250 ,..Lm x 4 mm        cal, Waltham, MA 02254).
  Optics                     25 mm x 200 mm focal length,              Instrument-operating    conditions are summarized in Ta-
                             planoconvex                            ble 2. These settings are compromise conditions,attainable
  Filter                     Acton Research Model 395/BR
  Photomultipliers           Hamamatsu   R760 13 mm diam.           with any commercially available ICP. We selected these
Computer                     IBM-PC                                 conditions to optimize the system for Cu analysis. In healthy
Electronics                  Computer interface-Mayo design
                             Multichannel readout-Mayo design                                            Precision capitlary
                                                                                                             gas orifice
                             HV and signal boards-Mayo design            Critical distance                   0.18 mmID
Nebulizers                             M
                             Jarrell-Ash odel90-790                             minimized
                             Perkin-ElmerModelN058-0368
Pump                         Gilson        2;
                                    MinIpuls 10 roller

                                                                                                                         High pressure
exit   slit-frame for multi-element analysis.It canbe operated                                                              gas inlet
as a monochromator,      a spectrograph   for permanent     film
records, and a direct-reading polychromator overa spectrum
of approximately    170 nm. A polychromator     slit-frame,   the
housing for the photomultiplier tubes, and a three-point                                             Solution uptake
self-centering mount were designedand constructed by the                                                  tube
Section of Engineering at Mayo Clinic. The emission wave-           Fig. 2. Schematic diagram of the modification of the Perkin-Elmer
lengths (nm) isolated for each element on the slit-frame are:       nebulizerto allowincorporation ofa precision-bore glass capillary
Zn(I) 213.856, Fe(U) 259.940, MgUI) 293.654, Ca(II) 315.887,
Cu(I) 324.754, and Y(H) 360.073.
                                                                       Table 2. OperatIng CondItIons for the ICP System
   In the polychromator    mode, the electronic amplifier is
placed directly behind the photomultiplier tube8, supplying
the high voltage for eachphotomultiplier and the first stage        Forwardpower                                       1300W
of amplification for the signals. Each amplified signal is          Reflected power                                      <5W
                                                                    Gas flow rates
converted first to a frequency signal and then to a light             Plasma                                           15 Ljmin
pulse for transmission.  This entire circuit is contained inside      Auxiliary                                        Normally off
the spectrometer to avoid being influenced by the high-               Aerosol                                          Varies with nebulizer
frequency magnetic fields emitted by the ICP. Each digi-              Back pressure                                    345 kPaa
tized light signal is reconverted to an electronicfrequency         Samplepump rate                                    2.92 mUmln
signal, then sent to an intermediate microprocessor in the          Observation height                                 11 mm
                                                                    Signalintegration                                  lOs
autosampler.    This processor and bus serve as the link              aModlfj      Perkin-Efmer.
between an IBM-PC computer and the spectrometer.The

                                                                                        CLINICAL CHEMISTRY, Vol. 32, No. 9, 1986 1661
subjects, urinary Cu excretion ranges from 10 to 60 ig/24 h
or approximately 8 to 50 gfL of urine, whereas the concen-               Table 3. ComposItIon of Calibrating Standards and
trations of Ca, Mg, Zn, and Fe in urine are significantly                                     Blanks
higher; thus, the detection limit and sensitivity for copper           Component and composition               Reagent
are more critical and were used to define optimum operating            Blank                       AqueousHNO3, 10 mL/L
parameters.                                                            Urine diluent               AqueousHNO3, 20 mLJL,containing
                                                                                                     100 g of V per liter
Materials                                                              Urine standard,mg/L
                                                                         Ca, 100
   Reagents. Stock solutions of Ca, Mg, Cu, Zn, and Fe were              Mg, 100
purchased as Baker Instra-Analyzed       Atomic Spectral Stan-           Zn, 0.50
dards (J. T. Baker Chemical Co., Phillipsburg, NJ 08865).                Fe, 0.50
Yttrium stock solution was prepared from yttrium nitrate                 Cu, 0.050
purchased from Fisher Scientific Co., Chemical Manufac-                Serumdiluent                Triplydistilledwater
                                                                       Serum standard,mg/L
turing Div., Fair Lawn, NJ 07410. Concentrated        nitric acid        Ca, 122.2
was distilled in glass, then redistilled     in a Teflon-coated,         Mg, 71.7
sub-boiling     still and stored in Teflon bottles. Water was            Zn, 2.85
triply distilled and stored in carboys dedicated to this use.            Fe,   2.18
   Controls. The plasma protein control was Lab-Trol Chem-               Cu,    1.89
istry Control (lot no. LT1O7-1-15; American Dade, Div. of
American Hospital Supply Corp., Miami, FL 33152). The
                                                                       urinary Ca, Mg, Zn, and Fe. Urinary copperwas determined
urine control was Urichem Level U (lot no. 523-022; Fisher
                                                                       with a Perkin-Elmer   Model 5000 graphite furnace equipped
Diagnostics, Div. of Fisher Scientific Co., Orangeburg, NY
                                                                       with a Model HGA 500 temperature controller and a Model
10962).
   Patients’ specimens. Serum and urine samples were select-
                                                                       AS 40 autosampler. These two instruments were used daily
                                                                       for the serum and urine analyses described here. Procedures
ed without conscious bias from among those sent to our
laboratory for routine analysis for metals. All serum sam-             for their use have been described elsewhere (13, 14).
pleswere drawn into acid-leached “Monovette” lO-mLplas-
tic syringes (no. 02.263.100; W. Sarstedt, Inc., Princeton, NJ         Results and Discussion
08540). These syringes had been disassembled, leached in
10 milL nitric acid for seven days, rinsed in water, air-                 Our goal was to accomplishsimultaneous multi-element
dried, reassembled, and sterilized with ethylene oxidebefore           analysis of serum or urine in a routine clinical laboratory
use for specimen collection. No beads were usedwith these              situation. To assess the viability of the ICP approach, we
syringes. After inserting the needle (no. 5175; Becton Dick-           evaluated critical factors such as detection limit, range of
inson Co., Rutherford,      NJ 07070) into the vein, we drew           calibration, accuracy, and precision after making the critical
several milliliters     of blood into a disposable syringe to          hardware or software choices(e.g., wavelengths, nebulizer,
cleanse the needle, then attached the Sarstedt syringe and             and calibration technique).
drew approximately 8 mL of blood. We let the blood clot for               Atomic emission spectra from an ICP are rich in atom and
10 mm, removed the plunger stem, and centriftiged the                  ion lines, some very intense and some very weak. An
syringe at 3400 rpm for 5 mm. The serum was then                       emission wavelength can usually be found that provides
decanted into an acid-leached 6-mL plastic screw-top vial              adequate sensitivity while avoiding matrix interferences.
(Sarstedt), and the tubes were capped tightly until assay              The high concentrations of Ca and Mg in urine and serum
that same day.                                                         warrant   the choice of weak ion wavelengths for these
   The 24-h urine specimens were collected into acid-washed            elements in order to avoid multiple sample dilution. The Ca
plastic bottles, then sufficient hydrochloric acid was added to        and Mg wavelengths listed in Methods are 260 and 400
adjust the urine pH to less than 2. The specimens were                 times weaker than the most intense lines, although re-
thoroughly mixed for 2 h, the total volume was measured,               searchers commonly choose the most-sensitive Ca and Mg
and aliquots were stored in smaller acid-leached bottles at            ICP wavelengths for biomedical analyses (12). ICP charac-
          until
-20 #{176}C assay.                                                     teristically has a wide dynamic concentration range, but for
                                                                       analyses involving the more sensitive wavelengths, addi-
Other Procedures                                                       tional dilutions must be made for very high analyte concen-
  Calibration.      Reagents    prepared    for urine or serum cali-   trations. Thus the choiceof the most sensitive wavelength,
bration are listed in Table 3. For the analysis of urine               when unnecessary, compromises and negatesthe real power
samples, we calibrated   the ICP with an aqueous acidic                of ICP-simultaneous     multi-element analysis.
standard prepared from stock solutions. All samples were                  When sera are diluted 10-foldand urine specimens dilut-
diluted with an equal volumeof 20 mL/L nitric acid contain-            ed twofold,detection limits are more than adequate,even in
ing yttrium      internal   standard.                                  the case of urine copper. The weak wavelengths we chosefor
  For serum specimens, triply              distilled   water was the   Ca and Mg are adequate for detecting these analytes in
preferred diluent, to avoid precipitation of protein. Calibrat-        biomedical    samples and provide linear dynamic ranges
ing standards,   controls, and samples were diluted 10-fold            sufficient to determine abnormally high analyte concentra-
with water before assay. The ICP was calibrated            with        tions with no further dilution. On the other hand, we chose
Cation-CalTM Calibration    Reference (American Dade). The             the most sensitive wavelengths for Cu, Zn, and Fe because
concentrations listed in Table 3 were confirmed by FAAS.               normal concentrationsof these analytes are in the .ug/L
  Atomic absorption spectrometers. We used a Perkin-Elmer              range. With the most sensitivewavelength selected for Fe,
Model 460 FAAS with an air-acetylene single-slot burner                the range is adequate to assay without dilution Fe concen-
for all serum results comparisons and for the analysis of              trations in urine from patients with hemochromatosis.

            CHEMISTRY,
1662 CLINICAL        Vol.32, No.9, 1986
                                Table 4. DetectIon Limits and Sensitivities Produced by Nebulizers
                                                  Detection limits, mg/L                                                         Sensltlvitya
       Nebulizer                 Ca           Mg            Cu             Zn        Fe              Ca                 Mg             Cu                Zn        Fe
Jarrell-Ash                    0.024        0.090         0.006         0.003      0.004           0.056            0.034             1.96           1.70         0.66
Perkin-Elmer                   0.114        0.172         0.002         0.006      0.010           0.010            0.019             1.05           0.40         0.16
Modified Perkin-Elmer         0.050         0.147       0.002           0.002      0.006           0.064            0.033             2.13           1.81         0.69
 a Calibration
             curve slope (or net intensity/elementconcentration in 1zg/L).

Nebulizers                                                                      elements-variations    in stray light may impinge on one or
   The nebulizer creates a fine mist of sample aerosol                          more of the analyte channels (2). These variables can also
suspended in a carrier gas, which is swept into the excita-                     affect the sample flow into the nebulizer.       For analyte
tion source. Unlike atomic absorption nebulizers with wide-                     concentrations >10-fold the detection limit, this variance in
bore orifices for gas and sample, nebulizers designed for ICP                   stray light signal or nebulization efficiency has little effect
have gas orifices small enough to produce a high-velocity jet                   on the final result. However, when analyte                         content is near
stream (>500 mIs) past the sample orifice and a sample                          the detection limit (as is the case for urinary copper), these
orifice large enough to allow the free aspiration of material                   contributions must be taken into consideration.
such as urine or diluted serum without clogging (15-18).                             Investigators have taken several approachesto this prob-
The efficiencyand uniformity of this nebulization process                       lem. Some simply do not report values for urinary copper
directly influences detection limits and sensitivity.                           (19,20),     whereas others make a complete extraction of the
   We evaluated pneumatically operated crossflow nebuliz-                       analytes from the urinary matrix, using a resin (21, 22).
era from Jarrell-Ash and Perkin-Elmer. Although the Jar-                        More typical solutions have either centered on making
rell-Ash nebulizer produced excellent detection limits and                      corrections by using expensive computer-controlled, stepper-
sensitivities (Table 4), the presence of any particulate mat-                   motor-driven, direct-reading instruments     to measure back-
ter in the sample, e.g., aggregated protein, clogged the                        ground excursions at the wavelength adjacent to the analyte
sample-uptake tube, necessitating stopping the ICP and                          wavelength or using an internal standard element to gauge
cleaning the aspiration tip. This considerably interrupted                      the signal fluctuations from the urine alone (23-25).
operation, with blockages occurring at least once for each 6                         Our approach is to add yttrium at a low concentrationto
h of operation.                                                                 each urine specimen as an internal standard (25). Added at
   The nebulizer from Perkin-Elmer offers a different ap-                       a concentration near the upper normal copper concentration
proach. It is completely demountable, and the inexpensive                       (50 zg/L), this providesa sensitive corrector of significant
gas and sample needles, made of molded plastic, are easily                      background shifts. Other investigators have used the inter-
replaced, because new needles fit prebored holes in the                         nal standard approach for background corrections (23,24),
nebulizer body. Sample-needle penetration and alignment                         but they addedthe internal standard at concentrations500-
are predetermined by a stop ring molded on the needle.                          fold greater than the expected analyte concentration. An
However, this system is not self-aspirating. The stock Per-                     internal standard addedat low concentrationcan correct for
kin-Elmer nebulizer produced sensitivities (calibration re-                     background shifts, as demonstrated by the close correlation
sponses) ranging from 18% to 54% of those attained with the                     of urinary copper measured by ICP and GFAAS shown in
Jarrell-Ash design (Table 4). When serum or urine samples                       Table 5. When we added yttrium at 5000 ugfL, however,
were pumped into this device, a visible stream of iridescent                    background changeswere not detectedand the correlation
particles traversed the aerosol channel of the ICP, which                       was not acceptable.
suggested that the stock Perkin-Elmer nebulizer produced                           ICP detection limits are 20- to 50-foldmore sensitivethan
large aerosol droplets. Under our conditions of operation,                      the normal serum concentrations of Cu, Zn, and Fe. Normal
these droplets were transported through the ICP too rapidly,                    concentrationsof stray-light-producing elements such as
sothat atomization was incomplete. The net result was poor                      calcium are significantly lower and much less variable in
analytical performance.                                                         diluted serum than in urine. We have alsoobserved that the
   To retain the detection limit and sensitivity performance                    nebulization   characteristics of diluted serum samples are
of the Jarrell-Ash design and the serviceability of the                         nearly identical. These observationssuggest that, unlike
Perkin-Elmer unit, we replaced the original gas orifice of                      urine analysis, internal standard correctionsare of minor
the Perkin-Elmer nebulizer with a glass capillary having a                      value in serum analysis.
diameter 75% of the original. We also extended the capillary                       Some investigators have suggested that sera must be
to bring the high-velocity gas stream closer to the rear of the                 digestedbeforeanalysis (20,26). Others have usedaqueous
sample orifice (Figure 2). The net result of this modification                  calibrating  solutions for serum analyses (24, 27). Tech-
is presented in Table 4. With this simple substitution and no
modification to the nebulizer body, sensitivities were im-                           Table 5. Regression Relationships for Patients’
proved fourfold and detection limits were improved by a                          Samples  Analyzed by Atomic Absorption  (a) and ICP (
factor of two. Overall performance of the device was equiva-                                       Urine (n    = 160)                           Serum (n = 169)
lent to or better than the Jarrell-Ash fixed-croasfiow nebu-
lizer, and the ease of maintenance of the stock Perkin-Elmer                                    Regression                                   Regression
nebulizer is retained. These modifications are critical to                                      equation                     r                equation           r
making our system sensitive and trouble-free.                                   Ca         y=   0.97x +       3.2       0.997         y = 0.98x + 0.17        0.996
                                                                                Mg         y=   0.91x- 1.8              0.95          y= 1.Olx- 0.05          0.98
 Calibration                                                                    Cu         y=   0.93x + 4.1             0.97          y = 1 .05x - 0.05       0.98
                                                                                Zn         y=   0.90x + 78.6            0.996         y = 0.98x + 0.03        0.98
   Because of the variations inherent with urine speci-                         Fe         y=   0.87x - 12.1            0.997         y = 0.99x + 0.00        0.997
 mens-salt content, total volume, organic matter, and other

                                                                                                    CLINICAL CHEMISTRY, Vol. 32,                  No.9, 1986 1663
       Table 6. Results for the Analysis of Serum and Urine Controls with the Modified Perkin-Elmer Nebullzer
                                                                                       Concn,   mg/L (!     ± 2   SD)
                                Ca                                        Mg                         Fe                        Zn                   Cu
Fisher Urichem Urine Control
1CP                        94   ±    4                            159     ±    10             0.08   ± 0.02             0.42   ±    0.04     0.20   ±   0.02
Certifiedvalue              104 ± 11                              163     ±    10             0.09   ± 0.028            0.48   ±    0.08     0.19   ±   0.05
Dade Lab-trolPlasma ProteinControl
ICR                          98.4 ± 2.1                            25.2    ±     2.2          0.64   ±    0.09          0.35   ±    0.05     0.92   ±   0.08
Certified value              99.0 ± 3.0                            25.0    ±     1.6          0.63   ±    0.11          0.31   ±    0.048    0.91   ±   0.08’
                                 by
  ‘values notcertified;determined FAAS. n =          10.


             400                                                                       accomplished simultaneously     by ICP from the detection
                                                                                       limit (which is 10-fold less than normal range) to 104-fold
                                                                                       above the detection limit (approximately 102-fold above the
                                                                                       normal range) with no sample dilution or pretreatment.
             300
       ‘5.                                                                             References
       c1                                                                              1. Scott RH, Fassel VA, Kniseley RN, Nixon DE. Inductively
                                                                                       coupled plasma-optical emission analytical spectrometry. A com-
       E                                                                               pact facility for trace analysis of solutions. Anal Chem 1974;46:75-.
             200
       E                                                                               80.
       C)                                                                              2. Larson GF, Fassel VA, Scott RH, Kniseley RN. Inductively
       C)                                                                              coupled plasma-optical emission analytical spectrometry. A study of
             100                                                                       some unterelement effects. Anal Chem1975;47:238-43.
                                     Y       0.9653x        +    2.9546                3. Fassel VA, Kniseley RN. Inductively coupled plasmas. Anal
                                     n   =   99                                        Chem 1974;46:1155A.
                                     R   =   0.996
                                                                                       4. Greenlleld S, Jones IL, McGeachin HMcD, Smith PB. Automatic

               0                                                                       multisample simultaneous multielement analysis with a HF plas-
                   0     100         200                   300             400         ma torch and direct reading spectrometer. Anal Chim Acta
                           Calcium, mg/ 24 h                                           1975;74:225-45.
                                                                                       5. Fassel VA, Kniseley RN. Inductively coupled plasma. Optical
FIg. 3. Correlation of caldum inacidified
                  plot                    human urine analyzed by                      emission spectroacopy. Anal Chem 1974;46:111OA.
atomic absorption spectroscopy (abscissa) and inductively coupled                      6. Fassel VA. Quantitative elemental analyses by plasma emission
plasma atomic emission spectroscopy (ordinate)                                         spectroscopy [Review]. Science 1978;202:183-91.
niques involving aqueous standardization       tend to yield                           7. Barnes RM. Recent advances in emission spectroecopy: induc-
                                                                                       tively coupled plasma discharges for spectrochemical analysis [Re-
results biased toward lower values. When we constructed                                view]. Crit Rev Anal Chem 1978;7:203-96.
calibration  curves from aqueous standards,     we observed                            8. Delves HT. The analysis of biological and clinical materials
higher slopes than those resulting from pooled sera or                                 [Review]. Prog Anal Atom Spectrosc 1981;4:1-48.
pinann protein, whereas plasma protein and pooled sera                                 9. Morrison GH. Elemental trace analysis of biological materials
calibration curves were virtually identical. Using protein-                            [Review]. Crit Rev Anal Chem 1979;8:287-320.
based standards, we have accomplished accurate analysis                                10. Mermet JM, Hubert J. Analysis of biological materials using
without digestion. For the sake of consistency in calculation                          plasma atomic emission spectroscopy [Review]. Prog Anal Atom
between serum and urine procedures, we use the internal                                Spectrosc  1982;5:1-33.
standard mode routine for both.                                                        11. Barnes RM. Determination of trace elements in biological
                                                                                       materials by ICP [Review]. ICP Inf Newsl 1984;1O:299-301.
Assays of Samples and Controls                                                         12. Schramel P, Liii G, Hasse S. Mineral and trace elements in
   To test the validity of our wavelength selections, calibra-                         urine. J Clin Chem Clin Biochem1985;23:293-301.
tion techniques, and nebulizer modifications,   we compared                            13. Pickup JF, Jackson MJ, Price EM, Brown SS. Assessment of
                                                                                       the reference method for determination of total calcium in serum.
results by ICP with those by FAAS and GFAAS for actual                                 Clin   Chem1974;20:1324-30.
clinical specimens; we alsoanalyzed certified, commercially                            14. Halls DJ, Fell OS, Dunbar PM. Determination of copper in
available urine and serum controls (Table 6).                                          urine by graphite furnace atomic absorption spectrometry.Cliii
   Regression relationships, with FAAS or GFAAS as the                                 Chim Acts 1981;114:21-7.
independent variable, are shown in Table 5 for urine and                               15. Wohlers CC, Hoffman CJ. Improvementsin nebulizer design
serum samples. Agreement between the ICP and FAAS                                      for the inductively coupledplasma.ICP Inf Newsl 1981;6:500-7.
results was excellent for all elements in serum. Close                                 16. Novak JW Jr, Lillie DE, Boorne AW, Browner RF. Fixed
correlation of urine results was obtained, although the slope                          croesfiow nebulizer for use with inductively coupled plasmas and
values suggest a bias by one method-probably          because                          flames. Anal Chem 1980;52:576-9.
additional dilution is not required for ICP. In Table 5 the                            17.  Gustavsson A. The determination of some nebulizer character-
data for urine constituents range from abnormally low to                               istics. Spectrochim Acta 1984;39B:743-6.
abnormally high. For urine copper, the comparison of ICP                               18. Fjjishiro Y, Kubota M, Ishida K A study of designs of a
results was made with those obtained from GFAAS. Figure                                                                                         S
                                                                                       croesfiownebulizer for ICP atomicemissionspectrometry. pectro-
                                                                                       chim Acts 1984;39B:617-20.
3 illustrates the near perfect correlation of these two tech-
                                                                                       19. DelvesHT, Bunker V, Husbands AP. A comparison of induc-
niques for urinary calcium, which is typical for the elements                          tively-coupledplasma optical emission spectroecopy and atomic
and sample types studied. With our choice of wavelengths,                              absorption spectroscopy for multielementalanalysesofdiets, faeces,
calibration technique, and modified nebulizer, accurate                                and urine. Proc 2nd hit Workshop,Trace Elem Anal Chem Med
analysis of urine and serum Ca, Mg, Cu, Zn, and Fe can be                              Biol. Berlin: Walter de Gruyter & Co. 1983:119-22.

1664   CLINICALCHEMISTRY, Vol. 32, No.9, 1986
20. SuddendorfRF, Smith GL. Inductively coupledplasma (ICP)            24. Uchida  H, Nqjiri Y, Haraguchi H, Fuwa K. Simultaneous
emission 8pectroecopy: concepts and clinical applications. Cliii Lab
                                                         J             multi-element analysis by inductively-coupled plasma emission
Autom 1983;3:256-62.                                                   spectrometry utili.ing micro-sampling techniques with internal
                                                                       standard. Anal Chim Acts 1981;123:57-63.
21. Barnes RM, Genna JS. Concentration       and spectrochemical       25. Nixon DE, McCall JT. A comparison of inductively coupled
determination oftracemetals in urine with a poly(dithiocarbamate)      plasma and atomicabsorption spectrometry for the direct determi-
resin and inductively coupled plasma-atomic emission spectrome-        nationofCa, Mg, Fe, Zn, and Cu in urine without preconcentration.
try. Anal Chem 1979;51:1065-70.                                         17thGreat Lakes Regional ACSMeeting, June 1-3, 1983;paper no.
                                                                       85. Washington, DC: Am. Chem.Soc.
22. Barnes RM, Fodor P, Inagaki K, Fodor M. Determination of           26. Mianzhi Z, Barnes RM. Determination of major, minor, and
trace elements in urine using inductively coupled plasma spectroe-     trace elements in human serum by using inductively coupled
copy with poly(dithiocarbamate) chelating resin. Spectrochim Acts      plasma-atomic emission spectroecopy.ApplSpectroec 1985;39:793-
1983;38B:245-57.                                                       6.
                                                                       27. Herber RPM, Pieters JH, Elgerema JW. A comparison of
23. Kimberly MM, Paschal DC. Screening for selected toxic ele-         inductively coupled plasma atomic emission spectrometry and
ments in urine by a sequential-scanning inductively-coupled plas-      electrothermal atomization atomic absorption spectrometry in the
ma atomic emission spectrometer. Anal Chim Acta 1985;174:203-.         determination ofcopperand zinc in serum. Freeenius Z Anal Chem
10.                                                                    1982;313:103-7.




                                                                                       CUNICAL CHEMISTRY, Vol. 32, No.9, 1986 1665