SAFETY AND OPERATION INSTRUCTIONS
Document Sample


INSTRUCTION MANUAL
VACUUM/HEATED SLAB GEL DRYER
FB57737 AND FB57747
WARNING
THESE UNITS ARE CAPABLE OF DELIVERING POTENTIALLY LETHAL VOLTAGE AND
ARE TO BE OPERATED ONLY BY QUALIFIED TECHNICALLY TRAINED PERSONNEL.
PLEASE READ THE ENTIRE OPERATOR’S MANUAL THOROUGHLY BEFORE
OPERATING THIS UNIT.
THESE UNITS COMPLY WITH THE STATUTORY CE SAFETY DIRECTIVES:
73/23/EEC: LOW VOLTAGE DIRECTIVE: IEC1010-1:1990 plus AMENDMENT 1:1992
EN61010-1:1993 & BSEN61010-1:1993
89/336/EEC: EMC DIRECTIVE: EN55011 Emissions, EN50082-1 Immunity
THE FISHERBRAND GEL DRYERS ARE DESIGNED TO GIVE LONG SERVICE AND
REPRODUCIBLE RESULTS IN YOUR LABORATORY. A FEW MOMENTS SPENT
READING THESE INSTRUCTIONS WILL ENSURE THAT YOUR EXPECTATIONS ARE
REFLECTED IN THE SUCCESSFUL USE OF THE APPARATUS.
FIRST CHECK THAT THE APPARATUS HAS BEEN RECEIVED COMPLETE AND
UNDAMAGED FOLLOWING SHIPMENT. ANY FAULTS OR LOSSES MUST BE NOTIFIED
TO FISHER SCIENTIFIC UK IMMEDIATELY, FISHER SCIENTIFIC UK CANNOT ACCEPT
RESPONSIBILITY FOR GOODS RETURNED WITHOUT PRIOR NOTIFICATION.
REFER TO THE PACKING LIST AND CHECK THAT ALL COMPONENTS AND
ACCESSORIES ARE PRESENT.
PLEASE RETAIN ALL PACKAGING
MATERIALS UNTIL THE WARRANTY
PERIOD HAS EXPIRED.
1
SPECIFICATIONS:
Construction:
Size: 110 x 570 x 400 (H x W x D)
Mains Supply. 220 - 240V a/c., 50 - 60Hz.
Mains Fuse. 13 Amp.
Input Fuse. 6.3 Amp, 240V Antisurge.
Heating Element . 800 Watts.
Vacuum Timer. 0 - 5 Hours.
Temperature Timer. 0 - 5 Hours.
Digital Display. 0 - 950C.
Controller Accuracy. +/- 2%.
Weight: 9 Kg
Environmental Conditions:
This apparatus is intended for indoor use only.
This apparatus can be operated safely at an altitude of 2,000m.
The normal operating temperature range is between 4 oC and 65oC.
Maximum relative humidity 80% for temperatures up to 31oC decreasing linearly to 50%
relative humidity at 40oC.
The apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664. POLLUTION
DEGREE 2, states that: “Normally only non-conductive pollution occurs. Occasionally,
however, a temporary conductivity caused by condensation must be expected”.
2
PACKING LISTS:
No. Items Description Part Number Check
FB57737
1 Gel Dryer Unit -
1 Stainless Steel Screen FB57739
1 Mylar Sheet FB57741
1 Porous Polyethylene Sheet FB57743
1 Clear Silicone Rubber Overlay FB57745
1 Mains Lead - European -
1 Mains Lead - UK, 13 Amp Fuse -
2 Replacement Fuses, 6.3A Antisurge. -
1 Operator Manual -
FB57747
1 Gel Dryer Unit -
1 Stainless Steel Screen FB57749
1 Mylar Sheet FB57751
1 Porous Polyethylene Sheet FB57753
1 Clear Silicone Rubber Overlay FB57755
1 Mains Lead - European -
1 Mains Lead - UK, 13 Amp Fuse -
2 Replacement Fuses, 6.3A Antisurge. -
1 Operator Manual -
3
Operating the Slab Gel Dryer
A. Safety Considerations
A 6.3 Amp input fuse is located on the side panel.
A 13 Amp mains fuse is fitted to the power lead.
To replace a fuse isolate the control unit from the mains supply and open the fuse holder with a
screwdriver blade. The holder contains two fuses. The loose fuse is a spare. Always use the
recommended fuse and NEVER replace it with one of a different rating.
B. General Care and Maintenance
The unit can be cleaned as required when the mains supply is isolated.
Cleaning should be carried out with a cloth moistened with water or tissues impregnated
with 70% iso-propyl alcohol. NO OTHER CLEANING SOLUTIONS SHOULD BE USED.
There are no internal operator serviceable parts in this unit. If the power should fail, the unit
must be returned to the authorised service center.
C: Operation
To dry slab gels the procedure detailed below should be followed.
1. Vacuum System
The vacuum pump employed must be capable of pulling an air volume of between 1.5
and 6m3/hour. Check the manufacturers details of the vacuum pump to see if it is
resistant to chemicals and contains separators. If so, no chemical vapor trap or cold trap
are required to remove any fumes from entering the pump. If the pump is not resistant to
chemicals then a vapor trap and cold trap should be fitted. It is always advisable to
incorporate a cold trap, what ever type of vacuum pump is used.
Connect the vacuum pump to the vacuum connector, using strengthen vacuum tubing.
Connect the vacuum pump to the mains inlet, if the vacuum timer is required.
Connect the mains lead to the mains input socket.
2. Preparation of the Drying Plate
Rinse the aluminium platen and stainless steel screen with distilled water only. Dry with
clean paper towels. DO NOT use abrasives or solvents.
NOTE: Radioactive material can accumulate on the platen and stainless steel screen.
Clean these periodically with strong detergents to remove this material, rinse well with
distilled water and dry with clean paper towels.
It is important to follow all the required radiation safety procedures.
3. Preparation for Drying Gels.
Place the stainless steel screen into the top of the aluminium platen.
Cut a piece of 3mm filter paper to cover the stainless steel screen.
Cut a piece of 3mm filter paper, just larger than the gel to be dried, and place it onto the
filter paper on the stainless steel screen.
Place the gel onto the filter paper, preventing air bubbles.
Cover the gel with Saran wrap, preventing air bubbles.
Place the Mylar sheet on top of the sandwich when drying gels of less than 1.5mm thick.
If the gel is more than 1.5mm thick cover the sandwich with the porous polyethylene
sheet.
Cover the sandwich with the clear silicone rubber overlay sheet.
4
4. Drying the Gel.
Turn on the mains power.
Turn on the vacuum pump and the vacuum pump timer to the required time.
A seal should be created between the rubber overlay and the plate.
Set the required temperature using the digital display panel.
Press the ‘SET’ button, the previous temperature stored in the memory will be displayed.
NOTE: The temperature is initially set up at 80oC.
Adjust the required temperature by pressing the Temperature ‘UP’ and ‘DOWN’ buttons.
Press the ‘SET’ button and the new temperature will be stored in the memory.
After a few seconds the temperature display will return to the actual temperature.
The ‘Heating Element On’ light will illuminate when the heating element is on.
Set the ‘Heating Element Timer’ to the required time.
A standard sequencing gel will dry in approximately 40 minutes at 80oC.
NOTE: The Hysteresis control is preset. It is NOT necessary for the user to use this function.
5. After Drying.
To remove the gel after drying is completed.
Turn off Vacuum pump.
Turn off Vacuum pump timer.
Turn off Heating element timer.
Remove the silicone rubber overlay, this will break the vacuum.
Remove the stacking layers to reach the dried gel.
D. Troubleshooting:
Condition Probable Cause Remedy
No power The mains fuse has blown. See warning Below
No heat The heating element fuse See warning Below
has blown.
No vacuum Check the vacuum input fuse See warning Below
and vacuum pump fuses.
Gels crack Gels too thick or concentrated. Try to use thinner gels.
Decrease the drying
temperature and the
drying time.
WARNING
DO NOT attempt to remove the outer casing or make repairs to our electrical range of products,
should any unit fail.
Contact Fisher Scientific UK immediately if the need for repair or servicing should arise.
5
QUALITY CHECK LIST
Model............................. Serial Number..................................
1. Electrical Test Check............................
2. Labels Positioned Check............................
3. Labels Test/Serial No. Check............................
4. Unit Scratch/Blemish Free Check............................
5. Clean Silicone Overlay - Attached Check............................
6. Accessories - See Packing List Check............................
7. Instructions Check............................
ALL FISHERBRAND PRODUCTS ARE SUPPLIED HAVING PASSED RIGOROUS QUALITY
CONTROL PROCEDURES.
IF HOWEVER, YOU HAVE A QUERY OF ANY NATURE, PLEASE CALL FISHER SCIENTIFIC
UK 01509 231166 FOR COMPLETE SALES OR PRODUCT SUPPORT
SIGNED...........................................
QUALITY CONTROL ASSESSOR
6
WARRANTY
Fisher Scientific UK guarantees that the unit you have received has been thoroughly tested and
meets its published specification.
This unit (excluding all accessories) is warranted against defective material and workmanship for
a period of twelve (12) months from the date of shipment ex factory.
Fisher Scientific UK will repair all defective equipment returned during the warranty period
without charge, provided the equipment has been used under normal laboratory conditions and
in accordance with the operating limitations and maintenance procedures outlined in this
instruction manual and when not having been subject to accident, alteration, misuse or abuse.
No liability is accepted for loss or damage arising from the incorrect use of this unit. Liability to
Fisher Scientific UK is limited to the repair or replacement of the unit or refund of the purchase
price, option is with Fisher Scientific UK. Fisher Scientific UK is not liable for any consequential
damages.
Fisher Scientific UK reserves the right to alter the specification of its products without prior notice.
This will enable us to implement developments as soon as they arise.
Fisher Scientific UK products are for research use only.
A return authorisation must be obtained from Fisher Scientific UK before returning any product
for warranty repair on a freight prepaid basis.
WARNING
DO NOT attempt to remove the outer casing or make repairs to our electrical range of products,
should any unit fail.
Contact Fisher Scientific UK immediately if the need for repair or servicing should arise.