Total Hydrocarbons or Volatile Organic Compounds Flame Ionisation by usn16817

VIEWS: 9 PAGES: 2

									HCT Analyseur hydrocarbures    15/12/00 15:09     Page 1




                 Total Hydrocarbons or
               Volatile Organic Compounds
               Flame Ionisation Detection
            MODEL HCT 2000 G
                                      FRENCH STANDARD AFNOR X 43.301




        AMBIENT AIR                                               • A 4 x 40 character digital display and 16-keyboard
        Series 2000 G is a line of ambient air analysers using      permit a dialogue between operator and analyser.
        the same printed circuit boards. “Physics” measure-         Several menus are available for maintenance aid,
        ment principles associated to digital electronics           calibration, zeroing...
        designed around a 16-bits microprocessor gives the        • Continuous microprocessor self-diagnostic with
        operator the necessary security for the monitoring          display and remote transmission
        and control of our environment.                           • Digital signal processing. This new and advanced
                                                                    design improves accuracy while maintaining a very
        ANALYSER ENSURING                                           short response time.
        ENVIRONMENTAL PROTECTION                                  • Internal measurement storage memory with data trans-
        The HCT 2000 G is an advanced design analyser               fer to computer or printer via RS 232 C output.
        that incorporates the latest achievements in conti-       • OPTION : NON-METHANE HYDROCARBONS MEASUREMENT
        nuous hydrocarbons measurement.




          A I R      Q U A L I T Y          U N D E R            S T R I C T      S U R V E I L L A N C E
 HCT Analyseur hydrocarbures                 15/12/00 15:09           Page 2




                                         MODEL HCT 2000 G
                                                         FRENCH STANDARD AFNOR X 43.301


                                                                          PRINCIPLE
                  The main charasteristic of this F.l.D. detector is that it               the flow remains strictly constant even if the inlet fil-
                  uses a sampling pump which is located downstream of                      ter becomes partially blocked or the pump perfor-
                  the burner ; absorption or condensation risks ups-                       mance changes due to deposits on the valves. This
                  tream of the analyser are thus avoided. Moreover the                     novel arrangement contributes significantly to the
                  burner design is such that ambient air will be used as                   avoidance of drifts. On option we add in the same
                  an auxiliary combustive thus avoiding the use of a                       cabinet a little chromatographic column wich separates
                  bottle or of a pure air generator. To control the sample                 methane from the other hydrocarbons. In such a case
                  flow a specially designed depression regulator sys-                      automatic zero is carried out at every cycle using
                  tem maintains a constant pressure drop across the                        internal zero air generator.
                  burner. As the burner is held at constant temperature




                                                  TECHNICAL SPECIFICATIONS
                  • Ranges : 10 ppm to 10 000 of methane (can be                           • Hydrogen flowrate = 4,5 l/h.
                    calibrated with any organic compound).                                 • Auxiliary air flowrate = 18 to 60 l/h.
                  • Lower detectable limit : < 0.020 ppm                                   • Solenoid valves for zero/span (remote or manual
                  • Response time : from some seconds to 1 minute.                           control).
                  • Zero drift : < 0,2 ppm per week.                                       • Zero air generator on option.
                                                                                                                                                       Specifications subject to change without prior notice




                  • Span drift : < 1 % per week.                                           • Analog output : 4 - 20 mA with galvanic insulation
                  • Linearity : ± 1 %.                                                       (other outputs on request).
                  • Temperature range : 0 - 40° C.                                         • Digital RS 232 C output. Printer interface.
                  • Humidity : 0 - 96 % non condensing.
                  • Power consumption : 300 V.A.
                  • Power supply : 220 V - 240 V / 50 Hz.                                  SPECIFICATIONS ARE PER US E.P.A. AND GERMAN U.B.A.
                  • Mounting : standard 19” - 4 units rack,                                TEST CRITERIA . A LL US E.P.A. SPECIFICATIONS EXCEEDED.
                    H = 180 mm, W = 480mm,
                    D = 520 mm - 600 mm (non-méthane).
                  • Weight : 24 kg.
                  • Internal pump.
                  • Sample flowrate = 4 l/h.
 04 42 97 63 90