Aethalometer® Product Specifications Rack Mount Chassis Models AE 22 and AE 31
Document Sample


Aethalometer® Product Specifications
Rack Mount Chassis - Models AE-22 and
AE-31
Time Resolution (User specified)
AE-22: 1 hour to 15 seconds
AE-31: 1 hour to 2 minutes
Intercomparison
Aethalometer readings compare 1:1 with chemical analy-
sis for EC on parallel filter samples.
Sensitivity
Proportional to flow rate, inversely proportional to time
resolution - approximately 0.1 ug/m3 @ 1-minute resolu-
tion @ 3 LPM.
Noise Cancellation:
Short-timebase data may be integrated over longer peri-
ods to recover data exactly equivalent to higher-sensitiv-
Measurement Principle ity operation at longer timebases.
Incremental optical transmission absorption due to aero-
sol deposit collected continuously on quartz fiber filter. Flow Rate
2 to 6 LPM optimum, monitored by mass flow meter.
Analytical Method Internal pump provides 2 to 6 LPM, stabilized by closed-
AE-22 (‘Dual-Wave’): loop control.
Combination optical absorption measurement at 880 nm
(‘EC’), and 370 nm (designated as ‘UVPM’, interpreted as Aerosol Sizing
an indicator of aromatic organic Instrument collects all aerosol presented at inlet port.
compounds such as are found in tobacco smoke, PAH Size-segregating inlets (impactors, cyclones etc.) may be
mixtures, wood and biomass-burning smoke, etc.). attached.
Note: In almost all situations, the vast majority of EC aero-
AE-31 (‘Spectrum’): sol is in the PM2.5 submicron aerodynamic size range (i.e.
Simultaneous measurement at 7 wavelengths for studies less than 2.5 μm).
of aerosol optical absorption, atmospheric optics, radia-
tive transfer etc.: 370, 470, 520, 590, 660, Sampling Medium
880 and 950 nm. Special ‘alternative wavelength set’ opti- Reinforced quartz fiber tape. Tape advances automatically
cal sources are available upon request. to avoid optical saturation, advances 1 cm typically once
every few hours depending on concentration and flow
Interpretation rate. 15-meter roll of tape usually lasts from months to
Optical absorption is interpreted in real time as concen- years at most locations.
tration of Elemental Carbon (‘EC’, also called ‘Black Carbon
– BC’). Sample Retention
Aerosol is collected on quartz fiber tape. Subsequent
Specificity chemical analysis can be performed on the sample spots
No other aerosol species has optical absorption even following collection.
0.001 times that of Elemental Carbon in the visible range.
2020 Stuart Street ph 510.845.2801 www.mageescientific.com
Berkeley, CA 94703 fax 510.845.7137
Specifications - Rack Mount Aethalometer, Page 1 of 2, Jun 07
Product Specifications
Sample Collection Area Periodic Maintenance
Sample is collected on a defined spot on the filter fiber We recommend a flow rate verification every three
tape. months, and cleaning the inlet assembly once per year.
Options: ‘High Sensitivity’ option provides a small, con-
centrating spot for optimal performance at locations of Reliability
low concentration Aethalometers have been operating at hundreds of
‘Extended Range’ option provides a larger, dispersing urban, regional, rural and remote locations on all conti-
spot for optimal performance in areas of high concentra- nents, continuously since 1986.
tion.
Size
Internal Control Instrument constructed in fully-enclosed 19” rack mount
All functions controlled by embedded single-board chassis, 10.5” (6U) high, 12” deep. Additional benchtop
computer. Program and operating parameters stored in cabinet optional.
non-volatile RAM.
Weight
Data Output Approx. 40 lbs (18 kg).
Digital data is transmitted from rear COM port. Analog
output connector represents BC data as 0~5 volt DC sig- Power Requirement
nal, scalable in software. 100~240 VAC 50~400 Hz, autoswitching : 40 watts.
Data Recording Operation Environment
Data are written to recording medium once every time- Specified for typical indoor conditions. Not inherently
base period. weatherproof - additional enclosure recommended.
Option
Floppy Disk is standard. Temperature
Flash memory drive may be substituted 0 ~ 40 ºC operating. Electronics have active temperature
stabilization.
Display and Interface
4-line display screen (date, time, BC etc.); status indicated Vibration
by color-coded LED’s (OK, initializing, problem etc.); key- No effect for vibration up to levels typical of car at high-
pad for menu selection entries. way speed.
Attention Required Initialization
Instrument is completely automatic. No user attention is Approximately 15 minutes required from power-up to
required at all while operating. Starts and runs automati- valid data.
cally upon switching on. Recovers automatically from
power interruptions.
Consumables Required
The only consumable required is filter tape, approximate-
ly 1 roll per year.
2020 Stuart Street ph 510.845.2801 www.mageescientific.com
Berkeley, CA 94703 fax 510.845.7137
Specifications - Rack Mount Aethalometer, Page 2 of 2, Jun 07