					Model 3051 Transmitter for Flow, Level, and Pressure Measurement

Ordering Information
TABLE 7. Model 3051C Differential, Gage, and Absolute Pressure Transmitters — = Not Applicable • = Applicable
Model
3051CD 3051CG 3051CA

Transmitter Type (Select One)
Differential Pressure Transmitter Gage Pressure Transmitter Absolute Pressure Transmitter

CD
• — —

CG
— • —

CA
— — •

PRESSURE RANGES (RANGE/MIN. SPAN) Code
0(1)

Model 3051CD

Model 3051CG

Model 3051CA
0 to 5 psia/0.167 psia (0 to 259 mmHga/8,6 mmHga) 0 to 30 psia/0.3 psia (0 to 207 kPa/2,1 kPa) 0 to 150 psia/1.5 psia (0 to 1 034 kPa/10,34 kPa) 0 to 800 psia/8 psia (0 to 5 516 kPa/55,16 kPa) 0 to 4000 psia/40 psia (0 to 27 580 kPa/276 kPa) Not Applicable

CD
• • • • • •

CG
— — • • • •

CA
• • • • • —

–3 to 3 inH2O/0.1 inH2O Not Applicable (–747 to 747 Pa/25 Pa) –25 to 25 inH2O/0.5 inH2O Not Applicable 1 (–6,22 to 6,22 kPa/0.12 kPa) –250 to 250 inH2O/2.5 inH2O –250 to 250 inH2O/2.5 inH2O 2 (–62,2 to 62,2 kPa/0,6 kPa) (–62,2 to 62,2 kPa/0,6 kPa) –1000 to 1000 inH2O/10 inH2O –407 to 1000inH2O/10in H2O 3 (–248 to 248 kPa/2,5 kPa) (–101 to 249 kPa/2,5 kPa) –300 to 300 psi/3 psi –14.7 to 300 psi/3 psi 4 (–2 070 to 2 070 kPa/20,7 kPa) (–101 to 2 070 kPa/20,7 kPa) –2000 to 2000 psi/20 psi –14.7 to 2000 psig/20 psi 5 (–13 800 to13 800 kPa/ (–101 to 13 800 kPa/138 kPa) 138 kPa) NOTE: 3051CG lower range limit varies with atmospheric pressure.

Code
A M

Output
4–20 mA with Digital Signal Based on HART Protocol Low-Power, 1–5 V dc with Digital Signal Based on HART Protocol (See Option C2 for 0.8–3.2 V dc) NOTE: Not available with hazardous locations certification Options Codes I1, N1, E4, and K6.

CD
• •

CG
• •

CA
• •

MATERIALS OF CONSTRUCTION Code Process Flange Type Coplanar Coplanar Coplanar Coplanar Coplanar Coplanar Flange Material Drain/Vent Flange Adapters CD
• • • • • • •

CG
• • • • • • •

CA
• • • • • • •

2 SST SST SST Hastelloy C Hastelloy C Hastelloy C 3 Monel Monel Monel 4 Plated CS SST Plated CS 5 SST Hastelloy C SST 7 Plated CS Hastelloy C Plated CS 8 0 Alternate Flange—See Options H2, H3, H4, H7, F1, G1, G2, FA, FB, FC, FD, or S5 NOTE: Materials of Construction Codes 3, 7, and 8 meet NACE material recommendations per MR 01-75.

Code
2 3 4 5 6 7

Isolating Diaphragm
316L SST Hastelloy C-276 (Meets NACE material recommendations per MR 01-75) Monel Tantalum (Available on Model 3051CD and CG, Ranges 2–5 only. Not available on Model 3051CA) Gold-plated Monel (Use in combination with O-ring Option Code B.) Gold-plated SST

CD
• • • • • •

CG
• • • • • •

CA
• • • — • •

Code
A B

O-ring
Glass-filled TFE Graphite-filled TFE

CD
• •

CG
• •

CA
• •

Code
1 2

Fill Fluid
Silicone Inert fill (Halocarbon)

CD
• •

CG
• •

CA
• —

Code
A B C D J K L

Housing Material
Polyurethane-covered Aluminum Polyurethane-covered Aluminum Polyurethane-covered Aluminum Polyurethane-covered Aluminum SST SST SST

Conduit Entry Size
½–14 NPT M20 × 1.5 (CM20) PG 13.5 G½ ½–14 NPT M20 × 1.5 (CM20) PG 13.5

CD
• • • • • • •

CG
• • • • • • •

CA
• • • • • • •

1 Http://www.ganhui.cn

E-mail: info@ganhui.cn

Tel: 021-50420548

Cpmpany name: shanghai ganhui CO.,LTD

Rosemount Inc. TABLE 7. (Continued) Model 3051C Differential, Gage, and Absolute Pressure Transmitters — = Not Applicable • = Applicable
Code Alternate Flange Options (Requires Materials of Construction Code 0) CD
• • • • • • • • • • • • •

CG
• • • • • • • • • • • • •

CA
• • • • • • • • • • • • •

H2 Traditional Flange, 316 SST, SST Drain/Vent, SST Flange Adapter H3 Traditional Flange, Hastelloy C, Hastelloy C Drain/Vent, Hastelloy C Flange Adapter H4 Traditional Flange, Monel, Monel Drain/Vent, Monel Flange Adapter H7 Traditional Flange, 316 SST, Hastelloy C Drain/Vent, 316 SST Flange Adapter HJ DIN Compliant Traditional Flange, SST, 7/16 in. Adapter/Manifold Bolting HK DIN Compliant Traditional Flange, SST, 10 mm Adapter/Manifold Bolting HL DIN Compliant Traditional Flange, SST, 12mm Adapter/Manifold Bolting FA Level Flange, SST, 2 in., ANSI Class 150, Vertical Mount FB Level Flange, SST, 2 in., ANSI Class 300, Vertical Mount FC Level Flange, SST, 3 in., ANSI Class 150, Vertical Mount FD Level Flange, SST, 3 in., ANSI Class 300, Vertical Mount FP DIN Level Flange, SST, DN 50, PN 40, Vertical Mount FQ DIN Level Flange, SST, DN 80, PN 40, Vertical Mount NOTE: Option Codes H3 and H7 meet NACE material recommendations per MR 01-75.

Code
S5

Integral Mount Manifold Options
Assemble to Model 305 Integral Manifold

CD
•

CG
•

CA
•

Code
S4

Integral Mount Primary Elements (Optional)
Factory Assembly to Rosemount Primary Element (Annubar or Model 1195 Integral Orifice) NOTE: With the primary element installed, the maximum operating pressure will equal the lesser of either the transmitter or the primary element. Option is available for factory assembly to range 1–4 transmitters only.

CD
•

CG
•

CA
•

Code
S1 S2

Diaphragm Seal Assemblies (Optional) NOTE: Standard flange and adapter bolts are austenitic 316 SST.
One Diaphragm Seal (Direct Mount or Capillary Connection Type) Two Diaphragm Seals (Direct Mount or Capillary Connection Type)

CD
• •

CG
• —

CA
• —

Code
S7 S8 S0 S9

Optional All Welded Diaphragm Seal Systems (for high vacuum applications) NOTE: Standard flange and adapter bolts are austenitic 316 SST.
One Diaphragm Seal, All-Welded System (Capillary Connection Type) Two Diaphragm Seals, All-Welded System (Capillary Connection Type) One Diaphragm Seal, All-Welded System (Direct Mount Connection Type) Two Diaphragm Seals, All-Welded System (One Direct Mount and One Capillary Connection Type)

CD
• • • •

CG
• — • —

CA
• — • —

Code
B4 B1 B2 B3 B7 B8 B9 BA BC

Mounting Bracket Options Coplanar Flange Bracket for 2-in. Pipe or Panel Mounting, all SST
Traditional Flange Bracket for 2-in. Pipe Mounting, CS Bolts Traditional Flange Bracket for Panel Mounting, CS Bolts Traditional Flange Flat Bracket for 2-in. Pipe Mounting, CS Bolts B1 Bracket with Series 300 SST Bolts B2 Bracket with Series 300 SST Bolts B3 Bracket with Series 300 SST Bolts SST B1 Bracket with Series 300 SST Bolts SST B3 Bracket with Series 300 SST Bolts

CD
• • • • • • • • •

CG
• • • • • • • • •

CA
• • • • • • • • •

Code
E5 I5 K5 I1 N1 E8 E4 C6 K6 KB I7 E7 N7

Hazardous Locations Certification Options
FM Explosionproof Approval FM Non-incendive and Intrinsic Safety Approval FM Explosionproof and Intrinsic Safety Approval BASEEFA/CENELEC Intrinsic Safety Certification NOTE: Not available with low-power Option Code M. BASEEFA Type N Certification NOTE: Not available with low-power Option Code M. CESI/CENELEC Flameproof Certification JIS Flameproof Certification NOTE: Not available with low-power Option Code M. Canadian Standards Association (CSA) Explosionproof and Intrinsic Safety Approval
(Requires 42.4 V dc maximum power supply)

CD
• • • • • • • • • • • • •

CG
• • • • • • • • • • • • •

CA
• • • • • • • • • • • • •

Combination of CSA and CENELEC Explosionproof and Intrinsic Safety Approval NOTE: Not available with low-power Option Code M. Combination of FM and CSA Explosionproof and Intrinsic Safety Approvals NOTE: Not available with low-power Option Code M. SAA Intrinsic Safety Certification SAA Flameproof Certification SAA Type N Certification

2 Http://www.ganhui.cn

E-mail: info@ganhui.cn

Tel: 021-50420548

Cpmpany name: shanghai ganhui CO.,LTD

Model 3051 Transmitter for Flow, Level, and Pressure Measurement TABLE 7. (Continued) Model 3051C Differential, Gage, and Absolute Pressure Transmitters — = Not Applicable • = Applicable
Code
L4 L5 L6

Bolting Options
Austenitic 316 SST Bolts ANSI/ASTM-A-193-B7M Bolts Monel Bolts

CD
• • •

CG
• • •

CA
• • •

Code
M5 M6

Meters (Optional)
LCD Meter for Aluminum Housing (Housing Codes A, B, C, and D only) LCD Meter for SST Housing (Housing Codes J, K, and L only)

CD
• •

CG
• •

CA
• •

Code
Q4 Q8

Other Options
Calibration Data Sheet Material Traceability Certification per EN 10204 3.1B NOTE: This option is available for the sensor module housing and Coplanar or traditional flanges and adapters (Model 3051C), and for the sensor module housing and low-volume Coplanar flange and adapter (Model 3051C with Option Code S1). Local Zero Adjustment Only NOTE: Local zero and span adjustments are standard unless Option Code J1 or J3 is specified. No Local Zero or Span Adjustment NOTE: Local zero and span adjustments are standard unless Option Code J1 or J3 is specified. Transient Protection Terminal Block NOTE: Not available with hazardous locations certification Option Code I1 or K6. Custom Software Configuration (Completed CDS 00806-0100-4001 required with order) 0.8–3.2 V dc Output with Digital Signal Based on HART Protocol (Output Code M only) Gage Calibration (Model 3051CA4 only) Analog Output Levels Compliant with NAMUR Recommendation NE43, 27-June-1996 NOTE: NAMUR-Compliant operation is pre-set at the factory and cannot be changed to standard operation in the field. Analog Output Levels Compliant with NAMUR Recommendation NE43, 27-June-1996: Alarm Configuration–Low NOTE: NAMUR-Compliant operation is pre-set at the factory and cannot be changed to standard operation in the field. Hydrostatic Testing Cleaning for Special Service Cleaning for <1 PPM Chlorine/Fluorine ¼–18 NPT Process Connections (No flange adapters) Coplanar Flange Without Drain/Vent Ports Ceramic Ball Drain/Vents JIS Process Connection—RC ¼ Flange with RC ½ Flange Adapter 4500 psig Static Pressure Limit (Model 3051CD Ranges 2–5 only) External Ground Screw Assembly

CD
• •

CG
• •

CA
• •

J1 J3 T1 C1 C2 C3 C4

• • • • • — •

• • • • • — •

• • • • • • •

CN

•

•

•

P1 P2 P3 D3 D7 D8 D9 P9 V5

• • • • • • • • •

• • • • • • • — •

• • • • • • • — •

Typical Model Number:

3051CD

2

A

2

2

A

1

A

B4

(1) NOTE: Model 3051CD0 is available only with Output Code A, Process Flange Code 0 (Alternate Flange H2), Isolating Diaphragm Code 2, O-ring Code A, and Bolting Option L4. For Multiple Options selections, contact your Rosemount representative or see Rosemount PDS 00813-0600-4001.

3

Http://www.ganhui.cn

E-mail: info@ganhui.cn

Tel: 021-50420548

Cpmpany name: shanghai ganhui CO.,LTD


				