					Application Bulletin 286 e

Installation Instructions of TitrIC 2
861 Dual Channel Advanced Compact IC with 855 Robotic
Titrosampler




                                                                                Metrohm




                                                                              800 Dosino




                                                     Metrohm       Metrohm

                                                                             20 mL
                                                                                VENT



                                                  800 Dosino    800 Dosino




                                                  50 mL         10 mL        HNO3
                                                     VENT          VENT




                       855              712                    3 x 800                     2.861.0010   2.861.0020

                The TitrIC 2 System with two 861 Compact ICs and an 855 Robotic Titrosampler
The TitrIC 2 System is used for the fully automatic analysis of water samples using direct measure-
ment, titration and ion chromatography. The following measured values are determined in a very short
time: temperature, conductivity, pH, acid capacity and, in parallel, the concentrations of the individual
anions and cations. Further Metrohm instruments can be incorporated in the existing system at any
time and used to measure additional parameters.

Analytical sequence
      1.      Conductivity is measured in the sample vessel on the sample changer.
      2.      First sample transfer to the two 861 Advanced Compact ICs.
      3.      The IC measurements are started automatically.
      4.      Second sample transfer to the external cell of the 855 Robotic Titrosampler.
      5.      Temperature, pH and acid capacity (p and m values) are determined.
The whole procedure is controlled by the TitrIC software: The user enters the sample position and
sample identification, IC system and tiamo method, after which TitrIC transfers all the relevant
information to tiamo and IC Net. When the measurement has been concluded, TitrIC automatically
generates a joint report containing all results. This report can be printed out or produced as an Excel
table.




Version 1.0
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 2/16


Table of contents
1     Delivery package............................................................................................................................... 3
2     Installation ......................................................................................................................................... 3
    2.1     855 Robotic Titrosampler ........................................................................................................... 4
    2.2     712 Conductometer (2.712.0010) .............................................................................................. 4
    2.3     Bottle holder (6.2061.010).......................................................................................................... 4
    2.4     Dosino holder (6.2055.100)........................................................................................................ 4
    2.5     861 Advanced Compact IC (2.861.00x0) ................................................................................... 4
3     Cable connections ............................................................................................................................ 5
4     Tubing connections........................................................................................................................... 7
    4.1     External cell ................................................................................................................................ 7
    4.2     Installation of the 10 mL and 50 mL Dosing Units...................................................................... 8
    4.3     Overview of all the tubing connections....................................................................................... 9
5     Miscellaneous ................................................................................................................................. 11
    5.1     Stirrer ........................................................................................................................................ 11
    5.2     Length of the aspiration tips in the external cell ....................................................................... 11
    5.3     Electrodes................................................................................................................................. 11
6     Software .......................................................................................................................................... 12
    6.1 Installation of the software........................................................................................................ 12
    6.2 Configuration of tiamo .............................................................................................................. 12
      6.2.1 Devices .............................................................................................................................. 12
      6.2.2 Swing Head configuration .................................................................................................. 12
      6.2.3 Special beaker definition.................................................................................................... 13
      6.2.4 Lift positions ....................................................................................................................... 13
      6.2.5 Solutions ............................................................................................................................ 13
      6.2.6 Electrodes .......................................................................................................................... 14
      6.2.7 Methods ............................................................................................................................. 14
      6.2.8 Configuration of the export template.................................................................................. 15
    6.3 Configuration of IC Net ............................................................................................................. 16
    6.4 Configuration of TitrIC .............................................................................................................. 16
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 3/16


1 Delivery package
- delivered with TitrIC 2 package:
Number     Article no.           Article designation
IC
1          2.861.0020            861 Advanced Compact IC «MSM II» (anions)
1          2.861.0010            861 Advanced Compact IC (cations)
1          2.145.0320            Edgeport/4 USB Converter
Sample processor
1          2.855.0020           855 Robotic Titrosampler
1          6.1462.050           Titration head
Conductometry
1          2.712.0010           712 Conductometer
1          6.0912.110           Conductivity measuring cell made of PP, with Pt 1000
1          6.2125.110           Cable to Edgeport USB Converter
p and m value titration (pH measurement)
1          6.0257.000            Aquatrode; combined pH glass electrode (with temperature measurement)
1          6.2103.130            Adapter red
1          6.2103.140            Adapter black
Liquid Handling
1          6.3032.250            807 Dosing Unit, glass, 50 mL
1          6.3032.220            807 Dosing Unit, glass, 20 mL
1          6.3032.210            807 Dosing Unit, glass, 10 mL
3          2.800.0010            800 Dosino
1          6.2055.100            Dosino holder
1          6.2061.010            Bottle holder Dosino
2          6.1608.030            Glass bottle GL45 1L (round)
2          6.5620.000            Connecting set for Dosing Unit: Dosino port 4
2          6.2744.080            Coupling M6-UNF 10/32 (Dosing Unit – IC screw)
1          6.2744.070            Pressure screw made of PEEK, short
2          6.1543.080            Aspiration tip 1.5 / 240 mm
1          6.1602.160            Eluent bottle attachment GL45
1          6.1609.000            Absorber tube
External cell
1          2.802.0020            802 Rod Stirrer
1          6.2001.070            Stand support for 804
1          6.1414.060            Titration vessel lid micro
1          6.1415.220            Titration vessel
7          6.2730.030            Stopper with nipple and O-ring
2          6.2730.080            Screw nipple
1          6.2730.020            Septum stopper
1          6.2730.070            Screw nipple M16/8mm
1          6.2730.060            Screw nipple for electrodes
Connection tubings
2          6.1805.110            Connection tubing FEP, L = 80 cm, 2 x M6 thread, i.d. = 2 mm
1          6.1805.060            Connection tubing FEP, L = 60 cm, 2 x M6 thread, i.d. = 2 mm
3          6.1805.030            Connection tubing FEP, L = 150 cm, 2 x M6 thread, i.d.= 2 mm
1          6.1803.020            PTFE capillary, L = 5 m, i.d.= 0.97 mm
Software
1          6.6056.112            tiamo 1.1 full

- optional accessories:
1          6.2041.xxx            Sample rack made of PVC
1          6.1432.xxx            Sample beaker
1          6.1006.xxx            Anion column Metrosep A Supp x
1          6.1006.xxx            Guard column Metrosep A Supp x
1          6.1010.xxx            Cation column Metrosep C x
1          6.1010.xxx            Guard column Metrosep C x
1          2.853.0010            853 MCS CO2 Suppressor
1          2.861.0500            Column Oven for 861
1          6.2323.000            Storage solution
1          6.2308.020            Electrolyte solution (3 mol/L KCl)
1          6.2301.060            Conductivity standard
1          6.2307.100/110/120    Buffer solutions 500 mL, pH 4 / 7 / 9
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 4/16


2 Installation
The following is a detailed description of the TitrIC 2 System's installation.


We strongly recommend that the individual steps are carried out in the order given below!

2.1 855 Robotic Titrosampler
The sample processor is placed at the side at which it is most easily accessible. Please note that the
following procedure refers exclusively to the version in which the 855 Robotic Titrosampler is located
at the left-hand side of the system.
         On the right-hand side of the 855 Robotic Titrosampler, attach the stand support
         (6.2001.070) without the horizontal brace to the tower of the 855 (hexagon screw beneath
         membrane pump 1). The brace will be installed later.
         Do not screw the titration head (6.1462.070) onto the Swing Head!
         The Sample Rack supplied is screwed onto the 855.
         In the vicinity of the 855 the two canisters (standard equipment of 855) are placed on the
         floor: One 10 L canister is used as a waste container, the second as a reservoir for deionized
         (DI) water used for rinsing the external cell.
         From the smaller opening of the canisters containing DI water, an M8 PTFE tubing connection
         of suitable length is led to the connection box made of white plastic of membrane pump 1 and
         connected.
         From the front opening of the waste canister an M8 PTFE tubing connection of suitable length
         is led directly to membrane pump 2 (black plastic) and connected.
A more detailed description is given in the Instructions for Use of the 855 Robotic Titrosampler.

2.2 712 Conductometer (2.712.0010)
The 712 Conductometer is placed directly to the right of the 855 Robotic Titrosampler and behind the
stand support. For detailed information about the usage check the Instructions of Use of the 712. It is
recommended to carry out the calibration of the cell constant manually and before tiamo is started up.
Consult thereto the Installation Instructions of the 712 Conductometer, chapter 4.4.3. But firstly the
712 has to be registered in the configuration of tiamo so that the calibration can be stored. Thereto the
calibration should be made later when the 712 is registered and tiamo shut down.

2.3 Bottle holder (6.2061.010)
Directly to the right of the 712 Conductometer, place the bottle holder for the bottles with the titration
solution (0.1 M HCl, at the rear, with 20 mL Dosing Unit and 800 Dosino) and the acidification solution
(1 M HNO3, at the front, with eluent bottle attachment).

2.4 Dosino holder (6.2055.100)
The Dosino holder (6.2055.100) is placed in front of the bottle holder and the 712 Conductometer; the
50 mL (6.3032.250) and 10 mL (6.3032.210) Dosing Units are screwed onto this holder at a later
stage.

2.5 861 Advanced Compact IC (2.861.00x0)
Place the 861 Advanced Compact IC for cations (2.861.0010) to the right of the bottle holder. To the
right of the 861 for cations, place the 861 Advanced Compact IC «MSM» for anions (2.861.0020). The
two instruments can already be installed. Please consult the Instructions for Use for the Advanced
Compact IC instruments.
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 5/16


3 Cable connections

The numbers given here (                ) refer to the drawing on the next page.


First all the power cables are connected to the instruments. Both the instruments and the PC remain
switched off.

        USB cable from Edgeport to PC – supplied with USB Converter Edgeport/4 (2.145.0320)

        RS cable (6.2125.110) from 712 Conductometer to Edgeport port 1

        RS cable (6.2134.100) from 2.861.0020 IC to Edgeport port 2 – supplied with 861

        RS cable (6.2134.100) from 2.861.0010 IC to Edgeport port 3 – supplied with 861

        USB cable (6.2151.000) from PC to 855 Controller – supplied with 855, do not connect!

        MSB cable from Dosino 800, 20 mL (0.1M HCl) to 855 MSB 1, do not connect!

        MSB cable from Dosino 800, 10 mL (cations) to 855 MSB 3, do not connect!

        MSB cable from Dosino 800, 50 mL (anions) to 855 MSB 2, do not connect!
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 6/16




                                                         Waste B       Waste A                                                   Waste B       Waste A




                                                                                                                                                                                                                          Stirrer




                            Analog Output                                                           Analog Output
                                             Remote                                  Fuse                            Remote                                  Fuse
                                                                           100-120V: 1,0A(T)                                                       100-120V: 1,0A(T)
                                                                           220-240V: 0,5A(T)                                                       220-240V: 0,5A(T)
                            Detector Block      RS 232                                              Detector Block      RS 232


                                                              Made by Metrohm Herisau Switzerland                                     Made by Metrohm Herisau Switzerland      RS 232 C
                                                                                                                                                                                                                      1 MSB

  PC                                                                                                                                                                            Remote
                                                                                                                                                                                                       USB   Contr.   2      3




                                                               6.2134.100
                                                                                                                              6.2125.110

                                                                                                                                                                                          6.2151.000
                                                                                                                                              Cable connections of the TitrIC 2 System
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 7/16


4 Tubing connections
4.1 External cell

  The numbers given here (              ,…)   and «opening 1,...» refer to the drawing below.
In the next step the external cell is prepared. The following components are inserted in the titration
vessel lid (6.1414.060):
The small 6.2730.030 stoppers (with the black rubber ring) are screwed loosely into openings 1, 2, 3, 5
and 7 and fixed in position when the buret tips, etc. are inserted.
The large 6.2730.080 stoppers are screwed into openings 4 and 8 and the screw nipple (6.2730.070)
into opening 9.
The titration vessel lid is attached to the stand support (6.2001.070) using opening 10, fixed at the
required height with the clamping ring and the horizontal brace of the stand support is mounted on the
855 with the socket head screw.

                                   6.2730.030
                                     stopper                 10



                                                                                    6.2730.080
                                                                                    screw nipple
                                                        1     2
                             6.2730.070                             3
                            screw nipple

                                                    8         9         4

                                                    7                   5
                                                              6
                        6.2740.020                                                  6.2730.020
                      rinsing nozzle                                               septum stopper


                       Numbers and connections on the titration vessel lid (6.1414.060)

          The PTFE aspiration tip (6.1543.170, M8, supplied with 855) is inserted in opening 1 and
          connected to membrane pump 2 of the 855 with the M8 PTFE tubing connection (6.1805.510,
          supplied with 855).
          The rinsing nozzle (6.2740.020, supplied with 855) is inserted into opening 2, which is
          connected to membrane pump 1 of the 855 with the FEP tubing connection (6.1805.060,
          supplied with 855).
          The buret tip (6.1543.200) is inserted through opening 3 and connected to port 1 of the 20 mL
          Dosing Unit (0.1 M HCl) with the FEP tubing connection (6.1805.100, supplied with 807).
          This is where the pH electrode will later be inserted.
          The rinsing nozzle (6.2740.020, supplied with 855) is inserted into opening 5, which is
          connected to membrane pump 1 of the 855 with the FEP tubing connection (6.1805.060).
          The septum stopper (6.2730.020) is pushed into opening 6.
          The rinsing nozzle (6.2740.020, supplied with 855) is inserted into opening 7, which is
          connected to membrane pump 1 of the 855 with the FEP tubing connection (6.1805.060).
          This is where the 802 Rod Stirrer will later be inserted.
          After the screw nipple (6.2730.070) has been screwed into opening 9, the aspiration tip
          (6.1543.100) is pushed through it. An FEP tubing connection (6.1805.060, supplied with 855)
          will be connected to the connection set of Dosing Unit port 4 of the 50 mL Dosing Unit.
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 8/16


4.2 Installation of the 10 mL and 50 mL Dosing Units




                                                             6.5620.000

                                  Luer: Port 4
                                                                  Port 3
                           6.1805.030                                      6.2744.070



                                 Port 1                       6.2744.080

                                 M6: Port 2                                6.1803.020

                                                            6.1805.110


                      Connections on the bottoms of the 10 mL and 50 mL Dosing Units


         Port 1 is used for emptying the Dosing Units into the waste canister.
         Port 2 is used by the 50 mL and 10 mL Dosing Unit for taking in the sample via the titration
         arm.
         Port 3 is used for sample transfer to the IC system. The transition from M6 to the IC screw
         size (1/16 in.) is achieved with the 6.2744.080 adapter.
         Port 4 is used for the 50 mL and 10 mL Dosing Unit. This requires a coupling between Luer
         and M6: The PEEK connection (6.5620.000) is attached to the Luer connection on the bottom
         of the Dosing Unit. With the 50 mL Dosing Unit this port is used to realize «liquid handling» to
         the external cell; in contrast with the 10 mL Dosing Unit it is used to aspirate the nitric acid for
         acidification.
Once the tubing connections have been made, the Dosing Units are attached to the Dosino holder
(6.2055.100) with the screw attachment and the distance ring of the Connecting Set for Dosing Unit
(6.5620.000).
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 9/16



4.3 Overview of all the tubing connections



                                        6.1805.060



                                          6.1805.510



                                                       6.1805.110


                                                       6.1543.080


                                                        6.1805.060                6.1805.030
                                                                                                         Metrohm




                                                                                   6.1805.100
                                                                                                       800 Dosino




                                                                        Metrohm          Metrohm

                                                                                                      20 mL
                                                                                                         VENT



                                                                     800 Dosino       800 Dosino




                                                                     50 mL           10 mL           HNO3
                                                                        VENT             VENT




                                                                         6.1618.020                6.2055.100       6.2061.010

                                              Overview of the tubing connections in the TitrIC 2 System
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 10/16


      The three 60 cm FEP tubing connections (6.1850.060, supplied with 855) are used to connect
      membrane pump 1 of the 855 and the three rinsing nozzles (6.2740.020, supplied with 855) to
      openings 2, 5 and 7 of the external cell (see page 7).

      The 60 cm M8 PTFE tubing connection (6.1805.510, supplied with 855) is used to connect
      membrane pump 2 of the 855 and the M8 PTFE aspiration tip (6.1543.170, supplied with 855) to
      opening 1 of the external cell (see page 7).

      The two buret tips (6.1543.080) are each attached by a stopper (6.2730.030) to the titration head
      (6.1462.070).

      The two 80 cm FEP tubing connections (6.1850.110) are used to connect the buret tips
      (6.1543.200) at the titration head (6.1462.070) and port 2 of the 10 mL and 50 mL Dosing Units.

      The 60 cm FEP tubing connection (6.1850.060) is used to connect port 4 of 50 mL Dosing Unit and
      opening 9 of the external cell (see page 7).

      The 40 cm FEP tubing connection (6.1850.100, supplied with 807) is used to connect port 4 of the
      10 mL Dosing Unit and the M6 opening of the eluent bottle attachment (6.1602.160); the FEP
      aspiration tubing (6.1829.010, supplied with 807) should previously have been inserted. The
      absorber tube (6.1609.000) is filled as usual with lime and cotton wool and inserted into the
      opening provided for it in the eluent bottle attachment in order to protect the nitric acid from CO2.


      The 40 cm FEP tubing connection (6.1850.100, supplied with 807) is used to connect port 1 of the
      20 mL Dosing Unit and the buret tip in opening 3 of the external cell (see page 7).


      The three 150 cm FEP tubing connections (6.1805.030) are connected to port 1 of the 10 mL and
      the 50 mL Dosing Units, as well as port 3 of the 20 mL Dosing Unit; all three tubing connections
      are then connected to the 10 L PE waste canister (6.1621.000).

      Suitable lengths of the 0.97 mm PTFE capillary (6.1803.020, supplied with 861) are used to
      connect port 3 of the 10 mL and the 50 mL Dosing Units (using the screws 6.2744.080 and
      6.2744.070) and the injection valves of the two 861 Advanced Compact IC (position 1). The 10 mL
      Dosing Unit must be connected to the 861 Advanced Compact IC for cations (2.861.0010) and the
      50 mL Dosing Unit to the 861 Advanced Compact IC for anions (2.861.0020).
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 11/16


5 Miscellaneous
5.1 Stirrer
The 802 Rod Stirrer is inserted through opening 8 of the external cell and connected to the tower of the
855. The propeller (6.1909.020) is pushed on from below.

5.2 Length of the aspiration tips in the external cell
         Clamp the titration vessel to the titration vessel lid.
         The M8 aspiration tip in opening 1 must reach right to the base of the titration vessel as this is
         the only way to ensure that the external cell is cleaned properly.
         The buret tips in openings 3 and 4 should reach the middle of the titration vessel.

5.3 Electrodes
         The pH electrode (Aquatrode Plus with Pt 1000; 6.0257.000) is inserted into opening 9 together
         with the SGJ sleeve (6.1236.050). The cable is connected to the 855 Robotic Titrosampler at
         the rear top right. The plug of the gray cable is connected to the opening marked Ind.. The red
         plugs with the adapters 6.2103.140 and 6.2103.130 are inserted in the openings marked Temp.
         If the electrode is not to be used for some time then it should be returned to its holder and kept
         in the storage solution (6.2323.000).
         Conductivity cell (6.0912.110): The measuring cell cable is led through the guide chain of the
         855 and the two black plugs are inserted into the openings marked Cond. Cell of 712
         Conductometer, the two red plugs into those marked Pt 100/1000. Later (as soon as the titration
         head has been mounted) the measuring cell with the SGJ sleeve (6.1236.050) will be inserted
         in the SGJ 14 opening of the titration head.
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 12/16


6 Software
6.1 Installation of the software
         Edgeport driver – installation (Inside Out NETWORKS-CD): Connect the Edgeport with the
         USB cable (included in 2.145.0320) with the PC and then insert the CD. It’s not necessary to
         start windows update, click on automatic installation of the software.
         tiamo 1.1 – installation (tiamo-CD): All the standard directories proposed by the program should
         be accepted.
         IC Net 2.3 – installation (Metrodata-CD): All the standard directories proposed by the program
         should be accepted. Restart windows.
         TitrIC 1.0 – installation (Metrodata-CD): TitrIC 1.0 must only be installed after tiamo 1.1 and
         IC Net 2.3 have already been installed. After installation the standard TitrIC directory can be
         found under C:\Programme\ Metrohm\TitrIC, the TitrIC.exe file under C:\Programme\Metrohm\tiamo\bin.
   All programs must be shut down!

6.2 Configuration of tiamo
  The Controller cable (6.2151.000), which is already connected to the 855, is now connected to an
USB interface on the PC. Wait a few seconds. The driver installation for the 855 follows. It’s not
necessary to start windows update, click on automatic installation of the software for both.

   Switch on the 712 Conductometer

   tiamo 1.1 is started

6.2.1 Devices
Connected devices are automatically recognized when tiamo is started. After confirmation of the auto-
matically generated request the devices are stored in the configuration, keep the proposed device
names. The 712 Conductometer is not recognized automatically and must therefore be added with the
name 712_1 via Configuration / Devices / Edit / New / Meter / 712 Conductometer, as well as selected COM 5 under
the entry RS 232 and confirmed with <OK>. Depending on the PC, the Edgeport port 1 could be assigned
to a different COM port. This is checked by clicking on Start / System control / System / Hardware / Device
manager / Connections (COM and LPT) and checking which COM port has been assigned to port 1 of the
Edgeport.
6.2.2 Swing Head configuration
Under                      top left, double-click on 855_1 under Devices, click on the entry Tower 1, set the
        Configuration at the
Axial distance to 196.0 mm and in the middle of the window under Swing Head click on Configuration and Yes
and then enter the following values:




                                        Configuration of the Swing Head
Confirm the entries twice with <OK> and then mark 855_1 under Devices and click on Edit /      Initialize.   Wait a
few seconds. The titration head (6.1462.070) can now be mounted on the Swing Head.
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 13/16


6.2.3 Special beaker definition
Under Configuration mark 855_1, select Edit / Properties / Rack / Rack data / Special beaker, Special beaker 1, click
on Edit and define the position to be used for rinsing as Rack position. The last position is normally used
as the rinsing position. A sample vessel containing DI water should always stand in this position.
Confirm by clicking on <OK> twice.


6.2.4 Lift positions
Under Configuration / Tools / Manual operation click on 855 Sample Changer / Tower 1 in the left-hand column
and use the entries Move and Assign position (See 855 Instructions for Use, page 67) to define suitable
values (click on <Assign> each time) for:
         Shift position                 for changing the rack position (approx. 53 mm)
         Work position                  for aspirating of the sample (approx. 173 mm)
         Special position               for dipping for the conductivity measurement (approx. 130 mm)
For the special position you must define the work position again, then at the right beside               Work position
click on Tower and select Special position 1.


6.2.5 Solutions
The solutions to be used must, as usual, be defined in tiamo. Connect the Dosino of the 20 mL Dosing
unit to MSB 1 of the 855, the Dosino of the 50 mL Dosing unit to MSB 2 and the Dosino of the 10 mL
Dosing unit to MSB 3. The user is automatically requested to save the solutions – click on <Yes>. In
order to ensure compatibility with the methods provided, they must be named as described below.
         At the left click on Configuration, then at the top right under Titrants/Solutions double-click on the
         appropriate (empty) entries and enter the following under Solution name and Concentration:




                                         Configuration table for titrants/solutions


Now the tubing lengths and ports are defined for each Dosing unit under                Edit/Properties…/ Dosing unit   as
follows:


                                    anion                           cation                        HCl

                            Port            Length          Port             Length       Port          Length
        Dosing Port
        Prep/Empty      Special Port           -        Dosing Port 1          -      Dosing Port 1       -
       Dosing Port 1       Port 4            60 cm         Port 1            60 cm       Port 1         40 cm
       Dosing Port 2       Port 3            0 cm          Port 3            0 cm        Port 3         0 cm
          Fill Port        Port 2            80 cm         Port 2            80 cm       Port 2         25 cm
       Special Port        Port 1            0 cm          Port 4            40 cm       Port 4         0 cm


                                       Configuration table of the three dosing units
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 14/16


6.2.6 Electrodes
         pH electrode: Under         Configuration/Sensors click on Edit/New/pH electrode and enter under Sensor
         name Aquatrode Plus,       type in the desired sensor information (e.g. serial number) and click on
         <Ok>.

         Conductivity sensor: Under Configuration/Sensors click on Edit/New/other sensor and enter            under
         Sensor name Conductivity sensor type in the desired sensor information and click on <Ok>.



6.2.7 Methods
The following seven methods are provided:
                            With this method all the parameters possible with the standard instruments
         TitrIC 2 - complete run:
         are measured and the IC measurement is initiated.
         TitrIC 2 - titration and measurement: Only titration and direct measurements are carried out; the IC
         measurements are not initiated. In TitrIC under Settings/IC Net settings the IC Anions and IC Cations
         have to be deactivated, or the method is started up only from tiamo.
                                                  This method is used for preparing all titration instruments,
         TitrIC 2 - preparation of titration equipment:
         i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
         rinsing the external cell. Start this method only from tiamo – not from TitrIC!
         TitrIC 2 - IC calibration - anion & cation: This method is used for calibrating the IC measurements
         with standard solutions without having to carry out the titration part of a typical TitrIC
         measurement; this means that a lot of solution can be saved.
         TitrIC 2 – semi automated pH calibration: With this method the pH electrode is calibrated automated;
         the buffer solutions are poored into the titration vessel. Start this method only from tiamo – not
         from TitrIC!
         TitrIC 2 - pH calibration: Here the Aquatrode is taken out of the titration vessel and placed into the
         buffer solutions. Start this method only from tiamo!
         TitrIC 2 - titer determination:   This method is used for the titer determination. Start this method only
         from tiamo!
         An automated conductivity cell calibration method for tiamo can be delivered if desired
         (titric@metrohm.com); it is recommend to do it manually.



    Import: Click on Method at the left, then select under File / Method manager… / Edit          / Import…   in the
    directory C:\Program Files\Metrohm\tiamo\methods_titric all methods and click on <Open>.

         Note!
         Alterations of these methods should only be carried out by a person at the Administrator level
         and who is thoroughly familiar with tiamo. Never forget that in TitrIC this new method must be
         selected in the sample table under Methods tiamo.
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 15/16


6.2.8 Configuration of the export template
Export template: tiamo must transfer its results to TitrIC in a particular way, which means that the
export template must be generated exactly in the following form:
Under Database / Tools / Templates / Export templates / New enter the following:




                                    Configuration of the export template in TitrIC

then click on Select fields and enter the following:




                                        Input window for the export template
The entry Start counter is obtained at the left by clicking on Determination/Identification, ID1.Name and ID1.value
are obtained at the left by clicking on Sample / Identifications, RS0x by clicking on Results / RS0xxx. By
clicking on the upper double arrow in the center the entries can be moved to the right. Of course
another number of results than five is possible.

    Shut down tiamo 1.1 !
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 16/16


6.3 Configuration of IC Net
   Start IC Net 2.3

         Under     Options / Global Preferences the following must be marked: under If method changed - Don’t ask
         to save   and System file overwrite - Always.
         With the TitrIC software predefined IC systems are also installed. These are found under
         C:\Program Files\Metrohm\IC Net 2.3\IC Net\Systems\Anion\anion.smt and C:\Program Files\Metrohm\IC Net
         2.3\IC Net\Systems\Cation\cation.smt. Open the two systems, click on Control / Connect to workplace
         and choose the right COM-Port. If you don’t know to which COM-Ports the systems are
         connected to you have to try out until you have communication between PC and the 861. Work
         should always be carried out with these systems, as otherwise the result transfer to TitrIC will
         be disturbed. This is why, if the systems have to be altered, the existing system is always
         altered and saved under the same name.
         The suppressor step is not made at the start of the measurement in the usual way, but with Fill
         directly after the end of the chromatogram. These means that after long pauses between
         measurements the suppression may no longer function. This is why after longer pauses you
         should carry out a suppressor step manually and ensure that the injection valve is in the Fill
         position.
         Depending on the application and column it may be necessary to adapt the time program in the
         IC system: Inject should always be carried out right at the start (0.10 min) of the time program,
         Fill right at the end. In this way you can start to measure with a stable baseline immediately
         after the sample loop has been filled by the Dosinos.
         If a calibration is carried out later and the new calibration data are to be applied to a sample that
         has already been measured then this is possible by using TitrIC – see the online help of the
         TitrIC software under Database / Recalculate / Recalculate for the results of several samples.

   Shut down IC Net 2.3



6.4 Configuration of TitrIC
   TitrIC is started (tiamo and IC Net are started automatically at the same time)

         All paths have already been entered as defaults and the system is ready for work. If other paths
         have been used instead of the default paths then this must be altered in the corresponding
         dialog under Workplace / Settings / ….
         TitrIC starts tiamo and IC Net at the same time. However, the tiamo window is not visible, but
         can be brought to the foreground in TitrIC via View / Show tiamo.
         If TitrIC is shut down then tiamo will also be shut down automatically – but IC Net must always
         be shut down individually. Always shut down TitrIC as the first program! If tiamo is accidentally
         shut down first then TitrIC is also no longer visible; the program will continue to run in the
         background. It can then be shut down with a right-hand click on an empty position in the Taskbar
         and a left-hand click on Task manager / Processes / TitrIC.exe / End process.
         For further information please consult the TitrIC online help (? / Show Help) or contact the
         responsible person at your local Metrohm agency. You can also report your problems to the
         following support email: titric@metrohm.com.



                           Always start and shut down TitrIC first!

				