GLUCOWATCH                 ®   BIOGRAPHER                    TM

                      G2
   (automatic noninvasive blood glucose monitor)



                            Is it really worth a
                              diabetic’s time
                              (and money)?


                                                      Alex Yeung
                                                    ChE 4T3/6T3
                                                             1
                                                   March 31, 2004
                 Outline

    What is diabetes?
    History of BG testing
    Technical aspects
    Clinical results
    Problems
    Conclusions             Figure 1. GlucoWatch® G2
                              BiographerTM (GW2B)1




2
                    Diabetes 101

    Diabetes Mellitus in Canada
    – Prevalence: >2.25 million
    – Cause: absent/insufficient
             insulin production;
             cells can’t utilize
             ingested glucose
    – Cost: 9 BILLION dollars/year2

    Diabetic Complications
    – macrovascular
    – microvascular
    – neurological
                                      Figure 2. Systemic effects of
                                           diabetes mellitus3

3
                      Diabetes 101
                 What can be done?!
                       Landmark studies by the
           Diabetes Control and Complications Trial Group and
    United Kingdom Prospective Diabetes Study Group concluded that:


                   reducing high        reduction in risk, and
                     BG levels
                  (hyperglycemia)   =    severity, of diabetes
                                            complications4
                    in diabetics


                 Hypoglycemia (low BG levels) can be dangerous
                   too as they can cause comatose and death
4
          Traditional BG Testing
             a)                                 b)
                                                                  apply
                                                                  blood
                                                                  here



         Figure 3. a) Prick finger5 b) Use biosensor to measure BG6

                    The ‘fingerstick’ method
                  – painful, cumbersome, inconvenient
                  – 4+ BG tests recommended per day

    Potential alternatives
    – boronic acid-containing fluorophores embedded in contact
      lenses7, skin tissue measurements utilizing optical
      coherence tomography8 or ultrasound9

    But, these methods are expensive, inaccurate, too large, so…
5
           Solution? The GW2B
    GlucoWatch® G2
    BiographerTM (GW2B)1
    – 2nd generation device by
      Cygnus Inc.
    – FDA approved in 2002 for
      diabetics 7+ years old
    – performs noninvasive,
      transdermal, near-continuous,
      BG analysis
    – tracks BG vales at 10-min           Figure 4. GlucoWatch® G2
                                       BiographerTM; available in the US
      intervals for 13 hr, stores                  and UK10
      BG data, identifies BG trends,
      alerts user of hypo- and
      hyperglycemia
6
     Technical Aspects of GW2B
    AutoSensor Component
    – two cross-linked polymeric
      hydrogel disks; must be
      replaced every 15 hr
    – disks contain electrolyte and
      glucose oxidase enzyme
      (GOx)


    Biographer Component
    – bulk of device
    – stores, analyzes, and displays   Figure 5. Components of GW2B
      data from AutoSensor using            (upside-down view)11

      mixture of experts algorithm
7                                                        (mech eng)
     Technical Aspects of GW2B

    Employs Reverse
    Iontophoresis
    1. apply 0.3 mA/cm2 current
       across skin; electromigratory
       and electroosmotic effects
       result
    2. interstitial neutral molecules
       (glucose) migrate with
       cations to cathodic               Figure 6. Molecular flow from
       hydrogel disk                    interstitial fluid of skin to GW2B
                                            (cross-sectional view)12



8                                                                (elec eng)
       Technical Aspects of GW2B

    Glucose selectivity           Biographer               Microprocessor, Data
                                                             memory, Display
    – Size exclusion: skin’s                                    IONTO(phoresis)
      MW cutoff is ~500 Da                                        electronics

         MWGlucose = 180 Da                                       Biosensor
         MWHb = 64.5 kDa                                          electronics
                                  AutoSensor
                                                        GOx          Gel       GOx
    – Charge exclusion:                               (anode)
                                                                    pads    (cathode)
      current is preferentially                Skin
      carried by in vivo Na+
                                                                Na+, glucose
          Uric and ascorbic
          acid migrate to anode
                                                      Figure 7. Electronics and
                                                          structure of GW2B
                                                       (cross-sectional view)13

9                                                                    (mech/elec eng)
        Technical Aspects of GW2B
     1. Oxidation reaction in cathodic hydrogel disk:
                         GOx
                      
       Glucose  O2   Gluconic Acid  H2O2

     2. Electrocatalytic oxidation in Pt/C composite
     electrode adjacent to hydrogel disk:
                                                 
                      2H  O 2  2e
               H 2O2 
      3. Electrons (current) are detected by Ag and Ag/Cl
      electrodes; integration performed to yield net
      charge, QA
10                                                      (chem eng)
        Technical Aspects of GW2B
     4. Process is repeated using the other hydrogel disk,
     and BG value is calculated from the data:


       BG(calibra tion value)
                               Blood Glucose Value
         Q A  Q B  2000

            -BG(calibration value) is obtained from fingerstick
            initially used to calibrate GW2B

            -QA: net charge from first hydrogel disk
            -QB: net charge from second hydrogel disk
            -2000 nCoulombs: empirically determined offset
            constant used to optimize biographer readings
11
     Clinical Results of GW2B




     Figure 7a-f. Sample BG measurements obtained from home
       environment study comparing biographer and fingerstick
                            readings14

            Biographer and fingerstick BG profiles
               correlate well with one another
12
             Clinical Results of GW2B
                                                Biographer and
                                             fingerstick BG results
                                            are correlated because:

                                              obtained correlation
                                              coefficient of 0.80
                                              Deming slope (linear
                                              regression) is near unity
                                              94.2% distribution of
                                              paired biographer and
                                              fingerstick results in
                                              clinically acceptable
      Figure 8. Overall results obtained
        from home environment study           regions of Clarke Error
     comparing biographer and fingerstick     Grid Analysis
                 readings14
13
     Clinical Results of GW2B

          Biographer readings (mmol/L)




                                                 Comparative readings (mmol/L)
       Figure 9. Clarke Error Grid Analysis comparing 2167 paired
     biographer versus serum BG values in clinical study. Regions A
             and B are the therapeutically relevant areas.15

                                         96.8% of the biographer’s data falls into
14
                                          clinically acceptable regions A and B
            Problems with GW2B
     Inaccurate hypoglycemic BG analyses:
     – 192 GW2B glucose tests performed on 89 hypoglycemic
       patients (BG ≤ 60 mg/dl)
     – Results:
          26 mg/dl difference between GW2B and serum BG results
          alarming of prehypoglycemia when alarm was set to 60 mg/dl
          happened only 23% of the time with false-alarm rate of 51%16


     Skin irritation:
     – 124 patients; 5 day biographer use
     – Results:
          87% of subjects: minimal/no skin irritation
          0.06% of subjects: intense erythema
                                                        Figure 10. Example of
          0.8% of subjects: strong edema13
                                                          forearm erythema
15
             Problems with GW2B
     Very few large-scale trials have been conducted
     – Demographic effects? Drug effects?
          1000 mg of acetaminophen consumed prior to device use does
          not affect BG results17, but what about other topical drugs…?
     Studies with biographer have not exceeded 6 wks
     – long-term alterations to skin structure and function?
     GW2B requires a two hour warm-up period and
     fingerstick-involving calibration
     – research into using Na+ in skin as an internal standard is
       underway

              Problems with novel device are expected.
                As well, GW2B can be easily removed
                       should problems arise.
16
       Interesting GW2B Analysis
     Cost of GW2B:
     – $698US, $85US for 16 AutoSensors (~10 d)
     – GW2B’s lifespan is ~4 years
     – not covered by insurance
     But, Monte-Carlo simulation study by Eastman
     et al.18 predicted that life-long biographer use
     would:
     – delay serious diabetes complications by 4.1 yr
     – prevent one case of blindness and 1.4 cases of renal
       failure in every 18 users
     – cause significant gains in terms of long-term cost
       savings

17
                   Conclusions
     GW2B’s chemical/electronic mechanisms have been
     comprehensively researched and optimized
     clinical trials have validated the GW2B’s accuracy and
     sensitivity with clinically acceptable results
     although slight problems exist (e.g. inaccurate
     hypoglycemic readings), GW2B provides real-time BG
     monitoring that will enable more aggressive insulin, diet,
     and exercise regimes to control diabetes
     more widespread use and studies will provide data for
     future biographer versions
     GW2B’s attributes (1/2 pancreas) could be used in the
     development of a true artificial pancreas!
     Better treatment offered right now, and possibly cure for
                     diabetes in the future?...
18
               References
     1. www.glucowatch.com
     2. Health Canada. Diabetes in Canada. 1999.
     3. www.abc.net.au
     4. Lancet 352:837-853 (1998).
     5. userweb.port.ac.uk
     6. www.lifescan.com
     7. Anal Chem 76(3):610-618 (2004).
     8. Phys Med Biol. 48(10):1371-90 (2003).
     9. Diabetes Technol Ther 4(4):489-97 (2002)
     10. www.diabetes.org.uk
     11. Clin Chem 45(9):1681-1683 (1999).
     12. www.glucowatch.com
     13. Annals of Medicine 32(9):632–641 (2000).
     14. Biosens Bioelectron 16(9-12):621-9 (2001).
     15. JAMA 282(19):1839-44 (1999).
     16. Diabetes Technol Ther 5(5):791-800 (2003).
     17. Diabetes Technol Ther 2(2):199-207 (2000).
     18. Pediatr Diabetes 4(2):82-6 (2003).
19
               Questions?




20 CONGRATS!

						
Related docs
Other docs by HC12080703564
SCH 3UO/E COURSE OUTLINE
Views: 4  |  Downloads: 0
Transportation Asset Management
Views: 3  |  Downloads: 0
The Fusion - DOC
Views: 15  |  Downloads: 0
Curriculum Vitae kk 2008
Views: 5  |  Downloads: 0
PowerPoint Presentation
Views: 1  |  Downloads: 0
ERFC Lpa JAMA 2009 Jul 22
Views: 2  |  Downloads: 0
PowerPoint Presentation
Views: 1  |  Downloads: 0
INSERT HEADLINE HERE
Views: 2  |  Downloads: 0