Clarity User Training
Clarity User Training
Advanced
Code P008 Date: 20.2.2007
1
For advanced users
User settings and display options
Sequences
Multidetector systems
Export data
Report setup
Batch data reprocessing
Code P008 Date: 20.2.2007
2
For advanced users
Data archivation
Column Performance
Noise and Drift
Special calculations
Control modules
GLP a CFR21part11
Code P008 Date: 20.2.2007
3
User settings
User settings are stored in
individual users desktop *.dsk
files.
For this reason they are set in
the instrument window and not
in the main window
Code P008 Date: 20.2.2007
4
Labels and display options
Overlay
activation Math
Moving and operation
Scaling of s
Chromatograms
Text Label and Lines
for legends and arrows
Chromatogram/Create Label
Code P008 Date: 20.2.2007
5
Colored peak area in chromatogram
Peaks in calibration can
have a color assigned.
When these peaks are identified
in chromatogram they will be
automatically colored
Code P008 Date: 20.2.2007
6
Colored peak area in chromatogram
Highlight currently selected peaks
Simply select a peak in the
chromatogram result table.
The result table selection
overrides the color defined
in calibration
Code P008 Date: 20.2.2007
7
Colored peak area in chromatogram
Use Peak Color set in
Calibration file
Highlight peaks selected in
Results table
(signal color used)
Code P008 Date: 20.2.2007
8
Table customizing options
Context menu
invoked by
right mouse
click in table
Code P008 Date: 20.2.2007
9
Common columns in Summary table
Independent setting of columns
Common - for all compounds
Summary – compound specific
Code P008 Date: 20.2.2007
10
Summary table options
Code P008 Date: 20.2.2007
11
Graph display options
Context menu invoked
by right mouse click in
Chromatogram window
Axis properties
setting
Graph
properties
setting
Code P008 Date: 20.2.2007
12
Chromatogram display options
Setting for
individual
signals
Gradient display
options
Code P008 Date: 20.2.2007
13
Customizing Toolbars
Reset All
restores original
settings
Code P008 Date: 20.2.2007
14
Sequence table
Define a serie of samples to be measured
Essential for working with autosamplers
For controled autosamplers the Vial number and Injection
volume are used from this table
Allows modifications during a run for not yet processed
samples
Simply add samples by filling any field in the last empty row.
Other fields will be copied from previous. Several samples can
be added by copying from a table
Allows for import of sample data from text files
Allows reprocessing of last measured sample set
Code P008 Date: 20.2.2007
15
Sequence table window
File name, Automatic Individual
Start, End vial and Injections automatic recalibration Post-Run
number. Several vials and/or %variables for from settings for
injections defined on single row SV, EV, IV sequence each row
Tooltip with actual file
Use row, add new name
Row status rows during
indicator sequence run
Code P008 Date: 20.2.2007
16
Sequence options
Pasive sequence
Clarity only expects Start signal,
analysis time is set on autosampler
Active sequence
Clarity sends a Ready signal to
autosampler and waits for Start
signal
AS control (Active sequence)
Autosampler injects according to SV,
EV, IV and
Inj. Volume in sequence table
Initial value
for %n
Editation of template
method used on
selected row
Code P008 Date: 20.2.2007
17
Sequence control
Run sequence Resume sequence
starts complete sequence Pause continues from last
(resets the status) sequence injection
Skip
Vial
Stop sequence will stop the
sequence run. Repeated Repeat
Stop will stop the current Injection
analysis
Status of the sequence.
Stored at Pause, Abort, Stop
Can be changed by Edit/Reset Status for selected rows
Code P008 Date: 20.2.2007
18
Sequence import
Select the file to import
Assign the respective fields to be imported
Only Start vial (SV) and file name must be
set (marked in bold)
Other fields will be filled by default values
Code P008 Date: 20.2.2007
19
Export data
Export data to other programs to prevent typing
errors
Simple Copy and Paste (Ctrl C – Ctrl V)
Export Data settings common for
manual export from chromatogram window
automatic export from postrun at the end of each run
automatic export by batch reprocess of
chromatograms or sequence
Export of chromatograms in common formats
Code P008 Date: 20.2.2007
20
Export data - simplest
Copy a picture to MS
Office (including labels)
via Clipboard or store it
in *.emf format
Copy a selected part of table using
Copy (Ctrl C) and Paste (Ctrl V)
Code P008 Date: 20.2.2007
21
Export data - options
Export data:
manual - from Chromatogram window
automatic - according to Setting/Postrun
batch - from Analysis/Batch
Export time/signal
data pairs
(import to other
programs – for
example graphics)
Headers:
Column headers and
chromatogram info
Full format:
Filename, date and time
before each row
blank field = actual chromatogram name All exported
+ .txt extension data are in
Path can be used single file
Code P008 Date: 20.2.2007
22
Export data - example
Code P008 Date: 20.2.2007
23
Batch processing
print
Report style from Report Setup dialog will be used
do not forget to switch off the Overlay mode
export
reprocess of chromatograms according to instrument method
reprocess whole sequence Export according
to Settings -
electronic signature Export Data
File type
Chromatograms
Calibration standards
Sequence Files
Electronic signature for
series of chromatograms
Code P008 Date: 20.2.2007
24
Multidetector systems
Assignment of detectors to Clarity Instruments
in the System Configuration dialog
All signals stored in a single chromatogram
Result table, Integration table and Acquisition
parameters are individual for each chromatogram
Common calibration file
calibration curves are individual for each signal
Summary table for display of results from all
signals
Code P008 Date: 20.2.2007
25
Multi signal chromatograms
Connecting link marks signals from one
chromatogram. Click icon to set signal
active
Detector Delay
compensates for
Results are displayed individually, interdetector volume,
some properties like calibration are inactive for first signal
common for the whole chromatogram
Code P008 Date: 20.2.2007
26
Overlay
Ctrl + click will
hide/show the signal
Ctrl + click will lock In overlay, several
the tool for repe- chromatograms can be
titive use. compared. Maximum number Scale, move
can be set in user options and math
operations are
active only in
Doubleclick on overlay
legend will set
signal active
Summary table displays
results for several
chromatograms or
signals simultaneously
Code P008 Date: 20.2.2007
27
Report Setup - logo
Code P008 Date: 20.2.2007
28
Report Setup – tiled signals
Code P008 Date: 20.2.2007
29
Archiving and backup
To reduce the data size compressed archive
*.dgz files should be used
*.dgz archives always replace all data
(files deleted from project will be during
archiving deleted also from the archive).
It is advantageous to archive whole projects
– should be considered when planning
project structure
Code P008 Date: 20.2.2007
30
Archive data
Archive and Restore for copying or moving
projects
Selected file types
Without compressing = original format and structure
compressed *.dgz archives = single file
Code P008 Date: 20.2.2007
31
Performance parameters
Separation
parameters
Advanced tab in the
Method Setup (displays
template method
Values used for Capacity ratio and
Performance tab Efficiency calculation
(hidden in
default layout)
Calculation type
Code P008 Date: 20.2.2007
32
Noise monitoring
Actual noise monitoring in the
Data Acquisition window
Code P008 Date: 20.2.2007
33
Noise and Drift
Drift and Noise can
be used for in User
column calculations
Drift and Noise displayed
for selected intervals in
Result table header
Code P008 Date: 20.2.2007
34
Calculations
ISTD calculations
Normalized % calculations
Groups
calibration of peak group like single compound
example - isomers (xylen)
Scale factor
recalculation of result to other units
Uncalibrated response
quantification of unidentified peaks using Uncalibrated
response factor
Injection volume
Injected volume correction
for manual injection in GC
Code P008 Date: 20.2.2007
35
ISTD calculations
ISTD calculations are performed:
a) Amount for ISTD is set in both
calibration and sample
b) both are zero
ISTD compound is
calibrated
Only one ISTD
compound is allowed ISTD peak can
in calibration. If the be hidden in the
conditions are not Result table
met, no results will be ISTD amount in
calculated sample is given
Code P008 Date: 20.2.2007
36
Normalized % calculations
Amount % gives the
normalized % if all If Sample amount is non-
Total line is zero, it will replace the sum
calculated as a integrated peaks are
calibrated and Sample in the Amount column and
sum of all groups serve as a base for Amount
+ sum of peaks amount is zero
% calculation
not belonging to
any group.
Groups are
calibrated, thus
peaks should not
be both calibrated
and in a group to
get correct
Amount% values
Code P008 Date: 20.2.2007
37
Groups
1st step
Adding peaks to a group
in the integration table
2nd step
Setting a Group in
Calibration table.
The compound name
and amount is set as for
ordinary peaks
Code P008 Date: 20.2.2007
38
Scale Factor and Uncal.Response
Uncal. Response:
Used for quantification of
unidentified compounds. For
universal response detectors
Scale factor and Units after scaling: (ELSD, FID, RI)
Multiplies values in Amount column and
changes the units in column header.
Method specific parameter, multiplies also
the entered Sample Amount value
Dilution:
Multiplies values in Amount column.
Sample specific parameter (entered from
Sequence or Single Analysis window).
Code P008 Date: 20.2.2007
39
Injection Volume
Default Injection Volume must be set first in calibration
and Injection Volume set in standards before calibrating
Supposes linear detector response
Setting Injection
Can lead to incorrect results, if used improperly Volume for
sample and
standard
Setting
Default
Injection
Volume in
calibration
Code P008 Date: 20.2.2007
40
Calibration
Command
Calibration/Show
History displays
data used for
selected level
Deviation of
point from
calibration
curve in %
Residual sum of
deviation squares
Code P008 Date: 20.2.2007
41
Retention indexes, LOD, LOQ
Calibration
Retention (Kovats) indexes
Columns for Retention LOD and LOQ limits
(Kovats) indexes are
hidden in default
settings.
Use setup columns to
disdplay them
Amounts lower then
specified LOD or LOQ
limits are marked in
Peak type column
Code P008 Date: 20.2.2007
42
Control Modules
For selected instruments the direct control is
available. The advantages are
The instrument parameters are stored in a
method
It is not necessary to enter the parameters
twice, especially useful for autosamplers
The parameters are recorded in the
chromatogram
Code P008 Date: 20.2.2007
43
Control modules (GC)
Control modules enable full
control of connected devices
like gradient programe and
other settings. The used
settings are stored within a
chromatogram.
Code P008 Date: 20.2.2007
44
Control modules (LC, AS)
Code P008 Date: 20.2.2007
45
Control Modules
Option to Send/Confirm/Do not send method to
instruments after method change
Confirmation dialog
Send Method options
Code P008 Date: 20.2.2007
46
Control modules overview
Code P008 Date: 20.2.2007
47
Additional control
Other devices can be controlled by A/D board digital outputs:
gradient or detector program start
detector autozero
valve switching
fraction collector control
Synchronization in Active Sequence
Dig. output set in System Configuration, not in the Event table
Code P008 Date: 20.2.2007
48
Event table
Set events in the Event Table tab in the Method Setup
Named events
Digital event input type
Output event can trigger run/sequence commands:
Stop
Abort
Skip vial
Shutdown
Repeat injection
Code P008 Date: 20.2.2007
49
GLP a CFR21 part 11
Chromatogram history
All saved versions are stored in the chromatogram
Versions can be opened (including calibration)
Audit trails
Station, Chromatogram, Sequence, Calibration
User access
Can be restricted by User Accounts settings
Code P008 Date: 20.2.2007
50
GLP a CFR21 part 11
Electronic signatures
based on User Accounts
based on Certificates
Validation
Installation qualification - standard feature
Performance qualification - by optional Validation kit
Operation qualification - SST module for statistical
evaluation and pre-set limits control
Code P008 Date: 20.2.2007