					ASADAL-M1 ESWL                                    Software Validation




                 SOFTWARE VALIDATION

                 Doc. No. : KMM-I-SV-28                  Ver1.0




             COMED Medical Systems Co.,Ltd.




                             1/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                                  Software Validation




                                        CONTENT

  1. General ---------------------------------------------------------------------------------        P.4
  2. Level of concern ----------------------------------------------------------------------          P.4
  3. Functional Requirement & System specifications -------------------------                         P.4
  4. Groups or unit Hardware -----------------------------------------------------------              P.5
      4.1. Glossary ---------------------------------------------------------------------------       P.5
      4.2. General System Block Diagram ---------------------------------------------                 P.5
      4.3. Operation Panel -----------------------------------------------------------------          P.6
      4.4. Reflector------------------------------------------------------------------------------    P.6
      4.5. I.I ------------------------------------------------------------------------------------   P.6

  5. Groups or units functions
      5.1. PLC-----------------------------------------------------------------------------------     P.7

       5.2. Operation Panel
            5.2.1. Block Diagram ---------------------------------------------------------- P.22
            5.2.2. Flow Chat ---------------------------------------------------------------- P.23
            5.2.3. Function Description --------------------------------------------------- P.23

       5.3. I.I
            5.3.1. Block Diagram ---------------------------------------------------------- P.24
            5.3.2. Flow Chat ---------------------------------------------------------------- P.25

       5.4. X-RAY
            5.4.1. Block Diagram ---------------------------------------------------------- P.26
            5.4.2. Flow Chat ---------------------------------------------------------------- P.27
            5.4.3. Function Description -------------------------------------------------- P.28

   6. Error Code ----------------------------------------------------------------------------- P.28




                                                 2/49         COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                                             Software Validation




APPENDIX

1. Program Structure
 1-1. Main -------------------------------------------------------------------------------------------------- P.31


2. Work List Window
 2-1. Main --------------------------------------------------------------------------------------------------- P.32
 2-2. Patient Information Control ---------------------------------------------------------------------- P.32
 2-3. Study Information Control ------------------------------------------------------------------------ P.33
 2-4. Component Function ------------------------------------------------------------------------------ P.34


3. Live Viewer
 3-1. Main --------------------------------------------------------------------------------------------------- P.35
 3-2. Image Control --------------------------------------------------------------------------------------- P.35


4. Still Viewer
 4-1. Main ---------------------------------------------------------------------------------------------------- P.37
 4-2. Image Control ---------------------------------------------------------------------------------------- P.33


5. X-ray Control
 5-1. Main ---------------------------------------------------------------------------------------------------- P.39
 5-2. Component Function ------------------------------------------------------------------------------- P.39


6. Shockwave Control
 6-1. Main ---------------------------------------------------------------------------------------------------- P.43
 6-2. Component Function ------------------------------------------------------------------------------- P.43


7. System Configuration
 7-1. INI Configuration ------------------------------------------------------------------------------------ P.45




                                                        3/49           COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                     Software Validation




Software and Hardware Description

    1. General


This appendix describes the functions, methods for testing and validation of the
software. All software is written in the ASSEMBLY language. This allows for a
modular program that is easily tested and verified for each mode of operation. This
allows for the self-diagnosis of all systems.


    2. Level of Concern

The level of concern for the software should be rated MINOR. Failure or latent design
flaws would not be expected to result in death or serious physical injury to patients. A
failure in the software or design flaw prevent proper exposure or produce an unusable
image on an X-ray film and display an error message on the control panel.


    3. Functional Requirements and System Specification


The PLC system has many safety features that prevent an X-ray exposure or will
terminate an exposure in event of PLC failure. One external Watchdog circuit which
separate time base is always active on the Main Board. The software in the ASADAL
consists of just one PLC system.
ASADAL has also hardware units that control directly warnings and alarms without
microcontroller. This hardware unit communicates to Main board the alarm status and
stops itself the function, the warnings status instead is only communicated to Main
board.




                                          4/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                   Software Validation




    4. Groups or Unit Hardware


4.1.   Glossary


WARNING : We define “warning” as any situation that is not dangerous. It is
           presented in the system as an anomaly and an audible “beep” will alert the
           user. The system will continue to work after a warning is sounded.
ALARM : We define “alarm” as any situation dangerous to the system. All alarm
           conditions will be shown on the LCD window. A continues audible signal
           will be generated and the system will discontinue to operate.
SPIE : We define “spie” as all symbols designed on the LCD Graphic Display
DATA : We define “data” as all numeric information and factors (i.e. - 100kV, 25mA ,
and Exposure Time).


4.2. General System Block Diagram



                                                       Monitor          CCD

                                                                        Camera
   Reflector           I.I TUBE
                                                                        PC System




                                                         Operation




                                                     HV Transformer
                                                            Panel


                                                        X-Ray

                                                        Tube



                                        5/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                      Software Validation




4.3. Operation Panel
This unit gives and gets all the relative information to the ASADAL from the operator.
Relative data to the Shockwave level, count, X-Ray kV, mA, Exposure Time, way of
work like as: bed status, C-Arm status and others.
The commands and data arrive from the PLC each 50msec by RS232 communication
cable.
It is compose from a Membrane and a graphic LCD display, graphic interface board
and input data.




4.4. Reflector
All commands arrive from Main Board by RS232 communication cable.




4.5. I.I (IMAGE Intensifier)
This board drives the Iris diaphragm on the camera, it drives Image rotation on the
camera, it select the Zoom of the Image Intensifier besides interface all signals from
the CCD Camera from Main Board.
All commands arrive from Main Board by RS232 communication cable.



4.6. CCU & CCD Camera
This board sends to Main board the kV balance signals. The kV balance signals are
used from the Main board to regulate the input light level present on CCD sensor.




                                          6/49       COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                          Software Validation




5. Groups or units function

5.1. PLC

5.1.1 Block Diagram




                           PLC POWER




       PC                                                    EEPROM
                              RAM
     SYSTEM
                      DRIVING PROGRAM



                              Analog Outputs


                      Fig 5-1 PLC Block Diagram




                                7/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                         Software Validation



5.1.1.1. Flow Chat(error check)




5.1.1.2. Functions Description
장비 사용시 이상있는 부분에 대해 메시지 및 부저음이 발생 됩니다.




                                  8/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.2.1. Flow Chat(BED on/off)




5.1.2.2. Functions Description
Bed System 의 사용의 구동 및 잠금 제어 합니다




                                 9/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.2.3. Flow Chat(BED DWON)




5.1.2.4. Functions Description
Bed System 의 Up mode Inter lock 제어 하며 Dwon mode 구동 합니다




                                 10/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.2.5. Flow Chat(BED UP)




5.1.2.6. Functions Description
Bed System 의 Dwon mode Inter lock 제어 하며 Up mode 구동 합니다




                                 11/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL
Software Validation


5.1.2.7. Flow Chat(BED MOVE)




5.1.2.8. Functions Description
Bed System 의 front 와 rear / right 와 left 동작을 제어합니다



                                 12/49    COMED Medical Systems Co.,
Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.3.1 Flow Chat(WATER)




5.1.3.2. Functions Description
16 진수 시그널을 이용 water,c-arm,trigger(manual mode) 를 제어 합니다.




                                 13/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.3.3. Flow Chat(WATER OUT)




5.1.3.4. Functions Description
Operation Panel 에서 Water Out 제어 합니다.




                                 14/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.3.5. Flow Chat(WATER IN)




5.1.3.6. Functions Description
Operation Panel 에서 Water In 제어 합니다.




                                 15/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                           Software Validation



5.1.4.1. Flow Chat (C-ARM ON/OFF)




5.1.4.2. Functions Description
c-arm 구동의 잠금과 해제 를 제어 합니다




                                    16/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                         Software Validation



5.1.4.3. Flow Chat (C-ARM MOVE)




5.1.4.4. Functions Description
c-arm 의 현재 위치를 감지 하고 방향 전환을 제어 합니다.




                                  17/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                        Software Validation



5.1.5.1. Flow Chat (X-RAY))




5.1.5.2. Functions Description
x-ray 사용에 필요한 On Off 기능 및 부가 기능들을 제어 합니다




                                 18/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                             Software Validation



 5.1.5.3. Flow Chat (KVP/MA DATA)




5.1.5.4. Functions Description
x-ray kVP,MA Data Level      Calibration




                                      19/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                            Software Validation



5.1.6.1. Flow Chat (SHOCK WAVE ON)




5.1.6.2. Functions Description
SHOCK WAVE On Off 및 Level 을 제어합니다.




                                     20/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                           Software Validation



5.1.6.3. Flow Chat (SHOCK WAVE   AUTO TRIGGER) )




5.1.6.4. Functions Description
SHOCK WAVE Auto Trigger mode




                                   21/49    COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                    Software Validation



5.2. Operation Panel


5.2.1. Block Diagram


                1920x1280
               LCD Graphic
                 Display

              Operation Pad


                                   Main Board


                       Figure 5-3 Operation Panel Block Diagram




                                         22/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                   Software Validation



5.2.2 Flow Chat




5.2.3. Functions Description
Control section for the whole equipment. It controls generating shockwave, checking X-
Ray, driving patient table.




                                         23/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                          Software Validation



5.3. I.I


5.3.1. Block Diagram




     Sensor DATA




                          PC SYSTEM
     CCD Camera
                             1394



                        X-Ray Control


                       Figure 5-7. : I.I




                             24/49         COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                            Software Validation



5.3.2 Flow Chat

                             Reset

                           Enable
                         Interrupts
                        RS232 Serial
                         Main Loop
                            port

                     Decode and verify
                    New RS232 message

                      New Command
                       from RS232

                      Answer to Main
                              Board
                        OK or FAULT
                         I.I. and CCD
                             Camera
                      New Command
                         I.I. and CCD
     Iris Command
                             Camera
                           Command
                     I.I. Zoom or CCD
        Iris New
                      New Command
       Command
                     Figure 5-8: I.I   Flow Chat




                                 25/49       COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                   Software Validation



5.4. X-RAY
5.4.1. Block Diagram

                                                    Filament status
               Rotor             OP PANNEL
               status


                                          Current
                         Rotor
                                          Generator
                         Speed
                                          Focus Size
                   Inverter


                                   TUBE



                        Figure 5.6 : Inverter Block Diagram




                                       26/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                          Software Validation



5.4.2. Flow Chat

                                                                              Radio
Fluoroscopy
                              RESET                                              gr
                                                                                 ap
                           kVp/mA Data                                           hy
                              setup
                       Filament Focus Select

                           Main Loop


                              Sensor
                              Signal?

                                                  PLC
                           Rotor Rotate
                               start
                              Rotor
                              Signal?

                         X-ray Exposure
                             Signal

                         kVp Feedback
                         Value
                         mA Feedback
                         Value
                              X-Ray                       Timer
          Timer 5min
                             Exposure                     Setting Time

              Alarm           X-Ray                         Alarm
                             Exposure




                              27/49       COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                       Software Validation



5.4.3. Function Description


The command and type of rotation speed directly receive from Main Board. If
Command is given of slowing or Fasting rotate the Rotor and to the heads of the rotate
anode motor there is not the right voltage it will be generate a Rotate Fault alarm for
the Main Board. The Main Board will stop immediately the RX Power.


6. Error Code
6.1 CP1
Cause : Over current(Over 10A) flowed into X-Ray HV Trans T1,T2 input lines
Measure : Returns to CP1 with Operation Desk Trip


6.2 CP2
Cause : Over current(Over 2A) flowed into STB OUT Line (mA slidacs)
Measure : Returns to CP2 with Operation Desk Trip


6.3 CP3
Cause : Over current(Over 10A) flowed into X-Ray Tube Rot Start 220V line
Measure : Returns to CP3 with Operation Desk Trip


6.4 CP4
Cause : Over current(Over 5A) flowed into X-Ray Tube Rot Run 45V line
Measure : Returns to CP4 with Operation Desk Trip


6.5 CP5
Cause : Over current(Over 5A) flowed into System load input 220V line
Measure : Returns to CP5 with Operation Desk Trip


6.6 CP6
Cause : Over current(Over 5A) flowed into Trigger PCB input 220V line
Measure : Returns to CP6 with Operation Desk Trip




                                           28/49       COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                   Software Validation



6.7 CP7
Cause : Over current(Over 5A) flowed into Shockwave Generator discharge M/C
Measure : Returns to CP7 with Operation Desk Trip


6.8 CP8
Cause : Over current(Over 5A) flowed into SMPC Power input power 0V line
Measure : Returns to CP8 with Operation Desk Trip


6.9 CP9
Cause : Over current(Over 5A) flowed into OP Head Outlet 220V line
Measure : Returns to CP9 with Operation Desk Trip




                                       29/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                           Software Validation




                 APPENDIX




                   30/49    COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                        Software Validation




1. Program Structure


                            1-1. Main




                Live View
                                                                 Still View
                Information
                                                                 Informati
                                        Work List
                                                                 on
                                        Information




                X-ray Control                           Shockwave Control
                Information                             Information

                                  [Image 1. Program ]




                                           31/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                Software Validation




   1.   Work List Window
        2-1. Main




                              [Image 2. Work List]
2-2. Component Function
                    상위 목록 - 환자이름(환자 ID)
   Tree View(①)
                    하위 목록 - Study Date(Database Key)
   Text View(②)     Tree 에서 선택되어진 목록의 내용을 보여준다.

                    새로운 환자 정보를 생성


                    새로운 Study 정보를 생성


                    생성된 환자 및 Study 정보를 수정


                    환자 및 Study 목록을 삭제

                    선택된 환자, Study 정보를 프로그램 등록 및 사용(등록 후 Work
                    List 창을 닫는다).

                    Work List 창을 닫는다.


                    검색어 종류를 선택할 수 있다(선택 : ID&Name, Date).


                    검색어를 입력한다.


                    입력된 검색어로 환자 및 Study 정보를 검색한다.




                                      32/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                    Software Validation



 2-3. Patient Information Control
   1) Main




                    [Image 3. New and Edit Patient Information]
   2) Component Function
                         ID 를 입력(수정이 필요한 경우 Patient 정보를 삭제 후 다시
                         생성하여야 한다.)

                         이름을 입력


                         생년월일을 입력(순서 : 월, 일, 년)


                         나이를 입력


                         주석을 입력


                         성별을 선택(Default : Male)


                         환자 정보 생성 및 수정을 완료한다.


                         환자 정보 생성 및 수정을 취소한다.




 2-4. Study Information Control

                                       33/49       COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                 Software Validation



  1) Main




                   [Image 4. New and Edit Study Information]




2) Component Function
                        Access Number 를 입력
                        Modality 를 입력 및 선택(입력 후 등록하면 자동으로 목록에
                        저장) - 참조 : 7-1
                        Body Part 를 입력 및 선택(입력 후 등록하면 자동으로 목록에
                        저장) - 참조 : 7-1
                        Refer Doctor 를 입력 및 선택(입력 후 등록하면 자동으로
                        목록에 저장) - 참조 : 7-1
                        Operator 를 입력 및 선택(입력 후 등록하면 자동으로 목록에
                        저장) - 참조 : 7-1
                        기재 사항 입력

                        Study 정보 생성 및 수정을 완료한다.


                        Study 정보 생성 및 수정을 취소한다.




                                     34/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                          Software Validation




3. Live Viewer
  3-1. Main



                                                   Last Image Hold
         Patient Information
                                                   State
         Information




                                                      View State
                                                      Information


                                    [Image 5. Default]
 3-2. Image Control
      1)X-Ray Shot



                                            Photographing State
                                            Information




         [Image 6. Photographing]
      - X-Ray 조사를 하면 Live Viewer 에서 영상을 확인 할 수 있다.


                                           35/49           COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                     Software Validation



2) Brightness and Contrast Control




                                     Brightness, Contrast
                                     Control


                                     Information
      [Image 7. Brightness and Contrast]
 - 마우스 왼쪽버튼을 사용하여 영상의 밝기와 대비를 조절한다.


3) Zoom Control




                                       Zoom Percent
                                       Information


                  [Image 8. Zoom]
- 마우스 우측버튼을 사용하여 Viewer 의 Zoom Percent(100, 166, 200)변경한다.

    2.   Still Viewer

                                           36/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                       Software Validation



        4-1. Main


                                       Image Number
  Patient Information                  Information
  Information




                                     Window     Level
                                     Value

                    [Image 9. Default]Information
4-2. Image Control
1) Window Level Control




             [Image 10. Window Level]
     - 마우스 좌측 버튼을 사용하여 값을 변경한다(좌우: Level, 상하: Width).
2) Magnify



                                          37/49         COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                           Software Validation




           [Image 11. Magnify View]
 - 마우스 중간(Wheel)버튼을 눌러 확대경을 나타낸다.


3) Zoom




                        [Image 12. Zoom In, Out]
 - 마우스 중간(Wheel)버튼을 Scroll 하여 확대형태를 조절한다.




                                 38/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                      Software Validation



4) Image Delete




[Image 13. Delete Message]
- 마우스 우측버튼을 눌러 Image Delete Message 를 나타낸다.




5. X-Ray Control
 5-1. Main




                              [Image 14. X-Ray Main]


 5-2. Component Function

                         X-Ray kvp Data display and level adjustment


                         X-Ray mA Data display and level adjustment


                         Graph representing X-Ray cumulative examination time


                         Mode function button to auto control for images’ brightness


                         Mode function button to fix tube current to 20mA


                         Flickering mark of indicating X-Ray examination


                         Memory function button to capture Live images


                                         39/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                Software Validation



                 Mode Function button to manage patients’ history



                 X-Ray Power


                 Function button to reverse images upside down


                 Function button to reverse images from side to side


                 Function button to invert black and white on images


                 Function button to control brightness of images

                 Function button to indicate position of stones



                 Function button to indicate position of stones with arrows


                 Function button to add comments.




                       [Image 15. Flip]




                                 40/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                     Software Validation




                 [Image 16. Reverse]




                 [Image 17. Invert]




                        41/49         COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                    Software Validation




                   [Image . Mark]




                 [Image 18. Arrow]




                         42/49       COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                      Software Validation




   [Image 19. Comment]
   -   Comment 버튼을 눌러 위와 같은 창이 나타나면 Comment 를 입력 또는
       수정한다.     (Still View 에서 선택된 Image 만 적용된다).


6. Shockwave Control
 6-1. Main




                           [Image 20. Shockwave Main]


 6-2. Component Function

                         Function button to indicate and control energy

                         Function button to choose 4 energy level of shockwaves

                         Shockwave System power On/Off button


                         Mode function button of auto output of shockwaves


                         Function button to start generating shockwaves


                         Function button to stop generating shockwaves.




                                         43/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                              Software Validation




                 Mode function button of shockwave’s manual output


                 Function button of shockwave’s manual generation.


                 Function button of Operating Count Reset


                 Function button to supply water inside of membrane


                 Function button to remove water inside of membrane


                 Function button to rotate Water System


                 Function button to air System

                 Setting and indicating of number of shockwaves which will be
                 treated
                 Indication of number of shockwaves which have treated
                 Setting and indicating of generating cycle of shockwaves
                 Indication of left treatment time

                 Function button to import images


                 Function button to export images


                 Button to choose prior images of Thumbnail.


                 Button to choose next images of Thumbnail.


                 Button to choose Still View size (1X1, 1X2, 2X2, 3X3).


                 To exit the program.




                                  44/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                       Software Validation



7. System Configuration
  7-1. INI File Configuration
    1) Modality




[Image 21. Study Information]
DEFAULT : Study 를 생성 할 경우 각 속성들의 목록에 기본 선택되어지는 Data 이다.
COUNT : 목록에 나타날 정보의 개수이다.
※ Data 의 목록은 Study 창에서 추가하면 1 에서부터 순차적으로 증가하며, Default 값을
변경은 직접 INI 파일에서 변경하여야 한다.


6) Institute Information




    [Image 22. Institute]
설치 장비에 대한 정보이며, 장비 설치하게 되면 정보를 입력하여 주어야 한다.




                                45/49   COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                             Software Validation




7) Shockwave




 [Image 23. Shockwave]
DEFAULT_COUNT : 프로그램 실행 할 경우                정해진 값을 PLC 로 전송하여 사용할
Shockw ave Count 를 설정한다(단위 : 1 회).
COUNT_MIN, MAX : 사용 가능한 Shockwav e Count 의 범위를 설정한다.
DEFAULT_INTERVAL : 프로그램 실행 할 경우 정해진 값을 PLC 로 전송하여 Shockwave
Interval Time 를 설정한다(단위 0.1 초).
INTERVAL_MIN, MAX : 사용 가능한 Shock wave Interval Time 의 범위를 설정한다.


8) Program




MASK_X : Live View 영상을 Masking 할 X 축 좌표 값이다.
MASK_Y : Live View 영상을 Masking 할 Y 축 좌표 값이다.
MASK_SIZE : Live View 영상을 Masking 할 X 축, Y 축 기준으로 그려질 반지름 값이다.
MASK_USED : Live View 영상을 Masking    유무를 나타낸다.
DEFAULT_FILTER : Recursive Filter 의 횟수를 지정한다.
DYNAMIC_FILTER : 사용안함
DYNAMIC_USED : 사용안함



                                    46/49     COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                     Software Validation



ABS_AREA : Live View 영상에서 ABS 의 기준이 되는 영역을 나타낸다(순서 좌에서 우 :
Left, Top, Right, Bottom).
LEVEL : Still View 에 Default 로 적용되어질 Level 값이다.
WIDTH : Still View 에 Default 로 적용되어질 Width 값이다.
BRIGHTNESS : Live View 에 Default 로 적용되어질 밝기 값이다.
CONTRAST : Live View 에 Default 로 적용되어질 대비 값이다.
KVP_MAX : Kvp 값이 정해진 값을 넘을 경우 더 이상 올라가지 않도록 구현 되어있다.
GUEST : 프로그램 실행하면 Work List 창을 실행 여부를 나타낸다(0 : Open).
MODE : 사용안함
TODAY : 오늘 날짜를 나타낸다.
DAY_COUNT : 오늘 시술 회수를 나타낸다.
MSG_COUNT, LIMIT_COUNT, TOTAL_COU NT : 사용회수가 TOTAL_COUNT + LIMIT_
COUNT 가 넘는다면 SW 가 동작하지 않고, MSG_COUNT 가 넘을 경우 경고 메시지를
나타낸다("SDS5000"의 값이 0 인 경우만 동작)
SDS5000 : 프로그램이 EM 으로 동작할지 여부를 나타낸다(0 : EM).
                         [Image 24. Program]




                   [Image 25. System Configuration]
- Program 에서 사용할 매개 변수를 설정할 수 있으며, INI 의 설정 값과 동일하다
    (단축키 : Alt + L).



                                         47/49        COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                                Software Validation




[Image 26. Shockwave Count Reset]
- INI 에서 SDS5000 의 매개 변수가 0 인 경우 나타나는 버튼이며, 누르게 되면
위와 같은 창이 나타난다. 창에서 “7321”를 입력 후 OK 버튼을 누르게 되면
현재 Shockwave Count 가 초기값으로 지정된다.
9) ABS Configuration




                             [Image 28. ABS Config]
AVERAGE_MIN, MAX : ABS 영역의 평균값의 범위를 지정한다.
DEFAULT_KVP, MA : 사용안함
MAX_KVP, MA : 사용안함
RECORD_COUNT : ABS 동작 Kvp, mA Dat a 의 개수를 나타내며, Data 는 1 부터
1 식증가하여 범위를 정한다.
※ 참조 : Image .




                                      48/49      COMED Medical Systems Co., Ltd
ASADAL-M1 ESWL                                   Software Validation




        [Image 29. ABS]
- ABS 에 적용될 매개 변수를 지정한다.
- Average Byte 의 Min, Max 값을 높이게 되면 ABS 동작할 경우 영상의 밝기가 높아져야
 하므로 X-Ray 조건이 높아지게 되며, 값을 낮추게 되면 ABS 동작할 경우 영상의 밝기가
 낮아지게 되므로 X-Ray 조건이 낮게 조사된다.




                            49/49   COMED Medical Systems Co., Ltd

				