Ultrasonic Examination of Pipeline Welds

W
Shared by: PItEY5iY
Categories
Tags
-
Stats
views:
28
posted:
6/11/2012
language:
English
pages:
17
Document Sample
scope of work template
							                                                                         Section: C1071
                                                                         Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                                                   Created: 2009-11-01

Table of Contents
1.  Scope ................................................................................................................................................... 1
2.  Referenced Codes and Standards ....................................................................................................... 1
3.  Definitions ............................................................................................................................................. 1
4.  General ................................................................................................................................................. 2
5.  Visual Observation ............................................................................................................................... 3
6.  Ultrasonic Inspection ............................................................................................................................ 3
7.  System Set-Up ..................................................................................................................................... 6
8.  Calibration Qualification ....................................................................................................................... 8
9.  Field Inspection .................................................................................................................................... 9
10. System Verification ............................................................................................................................... 9
11. Scanning Sensitivity ........................................................................................................................... 10
12. Re-Inspection ..................................................................................................................................... 10
13. Transducer Performance .................................................................................................................... 11
14. Evaluation of Indications .................................................................................................................... 11
15. Weld Inspection Record and Reports................................................................................................. 12
16. Additional Requirements .................................................................................................................... 13
17. Records .............................................................................................................................................. 13
Appendix A - Specification for Manual Ultrasonic Inspection of Pipeline Welds ........................................ 15


1. Scope
      This specification establishes minimum requirements for ultrasonic examination of GMAW and
      Manual Weld procedure qualification welds, performance qualification welds, production welds, and
      repair welds completed for the Company. Notwithstanding the requirements set forth herein,
      ultrasonic examination shall conform to the requirements of the most recent approved edition of 49
      CFR 192 and API 1104.

2. Referenced Codes and Standards
      In addition to this specification, the most recent edition of the following documents shall apply to the
      work.
           49 CFR 192, “Transmission of Natural And Other Gas By Pipeline”
           API 1104-19th Edition, “Welding of Pipelines and Related Facilities”
           API 1104-20th Edition, Addendum to API Standard “Welding of Pipelines and Related Facilities” –
            Revision of Appendix A – Alternative Acceptance Standards for Girth Welds
           SNT-TC-1A, “Personnel Qualification and Certification in Nondestructive Testing”
           ASTM E164, “Practice for Ultrasonic Contact Examination of Welds”
           ASTM E317, “Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo
            Testing Systems Without the Use of Electronic Measurement Instruments”
           ASTM E494, “Practice for Measuring Ultrasonic Velocity in Materials”
           ASTM E1316 “Standard Terminology for Non Destructive Examinations”
           ASME Section V, “Nondestructive Examination”
           BS 7706 1993 Calibration and setting up of the ultrasonic time of flight diffraction technique for
            the detection, location and sizing of flaws.
3. Definitions
      Definitions contained in ASME Section V, Article 5, Mandatory Appendix III apply to this specification.
      Company: The term “Company” as used in this specification shall mean Kinder Morgan Incorporated
      (KMI) including its engineering agencies, inspectors, and other authorized representatives.
      Contractor: The term “Contractor” as used in this specification shall mean the ultrasonic inspection
      contractor engaged by the Construction Contractor in work covered by this specification.
      Operator(s): The term “Operator(s)” as used in this specification or in any attached specification
      sheet shall mean operators of ultrasonic equipment. Ultrasonic operators shall be qualified and
Highlighting indicates revisions made as of the date on this standard.                                                                      Page 1 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

    certified in conformance with the requirements of Level II ASNT Recommended Practice SNT-TC-1A
    (at a minimum) and shall submit their records of certification to the Company Representative.
    Operator performance will be observed during welder tests. The operator must demonstrate an
    acceptable level of proficiency when operating the ultrasonic inspection equipment and interpret the
    inspection data in accordance with API 1104 requirements. The designated spread supervisor shall
    be responsible for the entire crew. The Company shall approve operators before the start of any
    work.
    Pipeline: The term “pipeline” as used in this specification means a pipeline (including laterals,
    branch connections, extensions, compressors, pumps, and related facilities) for the transmission of
    natural gas.

4. General
    4.1. General Conditions
          4.1.1.   The Company shall have the right to nondestructively examine any weld while the weld
                   is being made and after the weld has been completed. The examination shall be used
                   to assess conformance with 49 CFR 192, API 1104, and Section 14 - Evaluation of
                   Indications, of this specification. The Company shall also have the right to order the
                   removal of any weld for the purpose of destructive tests. The Company is the final
                   authority for the acceptance of welds.
          4.1.2.   The Contractor shall provide ultrasonic inspection service for all automatic girth welds
                   and any other welds designated by the Company during the construction of a pipeline
                   or facility.
          4.1.3.   Pulse-echo and/or pitch-catch transmission techniques with focused probes shall be
                   used with a minimum of A-scan and B-scan presentations and a fully automatic
                   recording system to indicate accurately circumferential location, circumferential length,
                   through-thickness location, and through-thickness height of defects as well as the
                   continuity of acoustic coupling. The instrument shall be capable of both pulse-echo
                   and pitch-catch operating modes.
          4.1.4.   The Contractor shall provide the Company with:
                      A hard-copy recording of the inspection
                      The judgment of acceptability of each weld examined
                      The recordings of the calibration standard scans
                      The electronic version of welds with applicable viewing software
          4.1.5.   The inspection shall use an automatic ultrasonic system capable of maintaining
                   production rates established by the Pipeline Construction Contractor. The system shall
                   be self-contained, have an independent power supply, and be mounted on an all-
                   terrain vehicle. Manual ultrasonic inspections may also be required for weld repairs or
                   other welds as directed by the Company.
          4.1.6.   The principle mainline welding technique shall be Automatic Gas Metal Arc Welding.
                   Manual welds and repair welds will be completed by Shielded Metal Arc Welding, MIG
                   and/or Flux-Core Welding combinations.
          4.1.7.   Automatic ultrasonic testing (AUT) tie-in rigs may be used to inspect Manual field welds
                   and repairs if the AUT Manual Weld procedure has been qualified prior to field
                   inspection commencement and as a minimum has met the requirements of this
                   specification and API 1104.
          4.1.8.   The Contractor shall designate a Job Supervisor responsible for the conduct and
                   performance of all ultrasonic personnel and for maintaining the reliability of all
                   equipment and supplies. The Job Supervisor shall work closely with the Company’s
                   Senior Welding Inspector and Field Auditor in the documentation of completed
                   inspections, and shall be responsible for all reports, interpretations, and evaluation.
                   The Job Supervisor shall ensure that interpretations are consistent, informative, and in

Highlighting indicates revisions made as of the date on this standard.                         Page 2 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

                   accordance with API 1104. The Job Supervisor shall be capable of technical support
                   for automatic UT operations.
          4.1.9.   Unless otherwise directed by the Company, Contractor shall have all necessary
                   equipment and personnel on site to maintain a production rate that ensures all welds
                   completed before 4:30 PM on any working day are inspected, interpreted, and included
                   in the Ultrasonic Inspection Report for that working day.
          4.1.10. Contractor shall have all necessary spare equipment to ensure continuance of the work
                  without interruption or delays. This will be routinely inspected by the Company Field
                  Auditor. Nonconformance reports shall be reported to the Pipeline Contractor and NDE
                  Contractor immediately.

5. Visual Observation
    GMAW Welds shall be scribed when ever possible to ensure that the search unit array is adjusted to
    the same distance from the weld centerline to duplicate the examination of the reference standard.
    Longitudinal or helical weld seams shall be ground flush and weld spatter shall be removed prior to
    AUT Mainline arrival for all weld processes to a minimum distance as defined in the contractor’s AUT
    procedure.
    Note: Tie-Ins may be excepted, but require centered root geometry readings in the AUT strip chart.
    Any weld impediments such as Grease Markings within 6” of weld, mud, longseams etc observed
    during setup and scanning that effect the AUT scan integrity shall be reported to the Company
    Representative and rectified.

6. Ultrasonic Inspection
    6.1. Introduction
          6.1.1.   Mainline girth welds made by automatic welding methods and Manual Tie-in welds (if
                   applicable) shall be fully examined by an automatic ultrasonic inspection system for
                   100% of the circumference, in conformance with API 1104. The inspection system
                   shall be capable of locating and accurately accessing the through-wall height and
                   circumferential length of defects, in a manner that permits the use of the acceptance
                                                                                             th
                   criteria defined by the Company in conformance with API 1104, 20 Edition –
                   Appended Appendix A. SMAW welds shall be inspected in conformance with
                   Company Construction Standard C1066 Weld Inspections and Testing, C1070 –
                                                                                      th
                   Non- Destructive Examination Requirements and API 1104, 19 Edition - Section
                   9.6 (workmanship) criteria.
          6.1.2.   To ensure the consistent production of high-quality ultrasonic inspections, ultrasonic
                   personnel shall be qualified in conformance with Subsection 6.2 - Operator
                   Qualification (follows), and ultrasonic procedures shall be qualified in conformance with
                   Subsection 6.3 - Procedure Qualification (below).
    6.2. Operator Qualification
          6.2.1.   A nationally certified ASNT Level III in the ultrasonic method shall develop the
                   application technique and prepare and approve the testing procedure. The procedure
                   shall be submitted to the Company for review and approval before production welding.
          6.2.2.   Ultrasonic operators shall be qualified and certified by examination in conformance with
                   Level II requirements of the most recent edition of ASNT Recommended Practice
                   SNT-TC-1A, and shall submit their records of certification to the Company
                   Representative. Additionally, operators shall have received a minimum of 40 hours of
                   training on automatic testing, including practical and theoretical aspects pertinent to the
                   equipment and configurations to be examined. This training shall be documented.
          6.2.3.   Before the initiation of any work, the Company shall approve ultrasonic operators and
                   the automatic ultrasonic inspection system. AUT systems shall qualify for this pipeline
                   inspection by demonstrating its ability to detect and characterize typical weld
Highlighting indicates revisions made as of the date on this standard.                           Page 3 of 17
                                                 Section: C1071
                                                 Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                           Created: 2009-11-01

                   indications and determine the acceptance in conformance with the criteria of API 1104,
                   Appendix A and/or Section 9. The characterization shall include the ability to locate the
                   indication within the weld and determine through-wall height. The company shall
                   provide qualification test pieces consisting of one or more pipe samples with defects.
                   The AUT qualification test shall be administered and witnessed by the Company during
                   weld production welding.
          6.2.4.   Ultrasonic operators shall qualify for manual inspection of repaired welds using
                   qualification test pieces containing defects in repaired welds. Operators shall
                   demonstrate ability to detect and characterize typical indications - and determine their
                   acceptance in conformance with API 1104, Section 9, and shall be approved by the
                   Company.
          6.2.5.   Only qualified ASNT level II (or equivalent) automated ultrasonic and manual ultrasonic
                   operators in conformance with this specification shall be permitted to inspect welds and
                   mark out defect areas using the ultrasonic technique.
    6.3. Procedure Qualification
          6.3.1.   Ultrasonic inspection shall be performed in conformance with a documented, detailed
                   procedure approved by the Company. Approved procedures shall include the
                   applicable information stated in this specification, information required by API 1104
                   Section 11, information stated in ASME Section V, Article 5, Paragraph T-522, and a
                   description of the methodology used to investigate indications.
          6.3.2.   The Contractor shall demonstrate the effectiveness of proposed procedures before
                   being approved for use in the performance of the Work. The Contractor shall not use a
                   procedure in the performance of the work that has not been approved by the Company.
                   Qualifying inspections shall be produced under site conditions and in the presence of
                   the Company Representative for each proposed procedure and/or technique. Such
                   procedure acceptance testing shall be performed using Contractor weld procedure or
                   welder qualification test welds. The qualification process shall occur before the
                   commencement of production welding.
          6.3.3.   The Contractor shall provide an automatic ultrasonic inspection procedure that allows
                   for full volumetric examination of a weld in a single pass, from both sides of the weld.
                   The procedure shall allow for edge misalignment and weld shrinkage, and shall provide
                   for characterization of defect location in the weld thickness direction, through-wall
                   defect height, and circumferential defect length and depth. Exception: AUT procedures
                   for tie-ins welds of unequal wall thickness may be inspected with multiple AUT scans or
                   passes when qualified. Defect location and sizing shall permit use of Company-defined
                   acceptance criteria based on the Engineering Critical Assessment provisions of API
                   1104, Appendix A (GMAW only). It shall include, but is not limited to, the following
                   requirements:
                      An automatic variable speed scanner mounted on automatic welding bands or
                       other tracking devices
                      An encoder capable of accurately indicating any defect location within the girth
                       weld
                      Independently loaded ultrasonic transducers mounted in an array that provides
                       independent inspection of the weld from both sides
                      Provisions for adjustments and maintaining the alignment of the transducers
                      Provisions for recording the continuity of the coupling
                      Provisions for ensuring the mechanical reliability of the equipment
                      A technique summary sheet stating beam angles, wave types, transducer
                       frequencies, beam size profiles, including sketches for each geometry to be
                       examined and other relevant information
                      A record of analog and/or digital signals from the multiple channels to a common
                       distance of rotation


Highlighting indicates revisions made as of the date on this standard.                         Page 4 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

                      Provisions for a permanent copy of scans in an easily-interpreted format to meet
                       archival and audit needs
                      Construction and accuracy details of the calibration piece
                      Calibration procedures to be used in the field
                      Calibration checks that shall be established and verified on a time or weld cycle
                       defined by the company, with system performance between calibration checks
                       continually monitored for degradation
          6.3.4.   A minimum of two welds per welding procedure prepared from production pipe
                   containing defects and acceptable imperfections shall be examined. The welds shall
                   be radiographed in accordance with Company Construction Standards C1066 - Weld
                   Inspections and Testing and Construction Standard C1070 - Non-Destructive
                   Examination Requirements. The UT procedure shall be applied within the detailed
                   temperature ranges and the results documented and compared with the radiographs.
                   Differences in detection results shall be noted and documented. At the discretion of
                   the Company, destructive testing of the weld sample may be performed to confirm
                   results.
          6.3.5.   The test details, including a notation of all the defects (along with a hard-copy recording
                   of the inspection) shall be submitted to the Company for approval. The Company shall
                   notify the Contractor of its approval of each procedure and/or technique validated.
                   Upon approval from the Company, the procedure becomes a mandatory requirement
                   and changes are only permitted with further approval from the Company.
          6.3.6.   The Contractor shall prepare a procedure for manual inspection of manually repaired
                   welds. ASTM E164 and API 1104, Section 11.4 shall be used as guidance in
                   developing the procedure. Appendix A of this document provides some details of the
                   required contents of this procedure.
    6.4. System Performance
          The Contractor shall successfully demonstrate system performance using project pipe. The
          demonstration shall include integration of hardware, software, and personnel in a manner that
          permits reliable detection and accurate sizing weld imperfections in conformance with API
          1104, Appendix A. The Contractor shall demonstrate the accuracy of the system for accessing
          through-wall height imperfections. The accuracy shall be verified by destructive testing results.
          Accuracy shall be demonstrated on the thinnest and the thickest wall thickness of project pipe.
          The Company shall be sole judge of system performance for the intended use.
    6.5. Calibration
          6.5.1.   The calibration block and included reflectors shall provide the required sizing accuracy,
                   the ability to provide defect location in the weld, and the determination of surface
                   interaction per API 1104, Appendix A.
          6.5.2.   Calibration standards shall be used to qualify the inspection system and to monitor
                   system performance on an ongoing basis.                 Calibration standards shall be
                   manufactured for a section of project-specific pipe supplied by the Company, and the
                   Company shall provide the Contractor with geometric details of weld bevels intended
                   for the project. The Contractor shall submit a calibration standard design to the
                   Company for approval before manufacture. No design changes shall be made without
                   prior approval from the Company. Calibration standard reflector dimensions and
                   locations within the standard shall be certified with instruments traceable to a national
                   standard(e.g. NIST or equivalent). The block shall be uniquely identified and traceable.
                   The unique number, wall thickness, diameter, steel manufacturer, shear wave acoustic
                   velocity in the direction of testing, temperature, and velocity measurement shall be
                   documented. The acoustic velocity shall be measured in conformance with ASTM
                   E494 or equivalent approved by the Company. The velocity results shall be
                   documented.


Highlighting indicates revisions made as of the date on this standard.                           Page 5 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

          6.5.3.   Calibration reflectors shall be flat bottom holes (FBH) for fusion (planar) defects.
                   Surface notches shall be used in line with the anticipated fusion line. These notches
                   shall be N10 notches as defined in API 1104 Section 11. Root notches shall be of the
                   same angle as anticipated fusion line of the root of the weld examined. Porosity
                   reflectors shall be flat-bottomed holes; other informational reflectors may be used but
                   must not interfere with calibration reflectors. A through-wall hole or notch should be
                   machined in the center of the block or ends. Calibrations for primary sensitivity shall
                   use flat bottom holes. Time of flight diffraction (TOFD) EDM notches for ID and OD
                   shall be used. For root calibration, a root notch may be used. Reflectors for planar
                   indications shall be located on the anticipated fusion line.
          For manual ultrasonic inspection of shielded metal arc welds, N10 notches shall be milled on
          the outside surface and inside surface of the block.
          6.5.4.   Machining tolerances for all calibration parameters shall be less than or equal to the
                   following:
                   6.5.4.1. Hole diameter                         ± 0.2 mm
                   6.5.4.2. Flatness of FBH                       ± 0.1 mm
                   6.5.4.3. Angles                                ± 1 degree
                   6.5.4.4. Notch Depth                           ± 0.1 mm
                   6.5.4.5. Notch Length                          ± 10%
                   6.5.4.6. Center location                       ± 0.1 mm
                   6.5.4.7. Hole Depth                            ± 0.2 mm
          6.5.5.   Project pipe will be available for procedure development and calibration block design
                   before weld procedure qualification or AUT qualification commences.
    6.6. Ultrasonic Equipment
          The scanning system shall provide enough inspection channels to ensure the complete
          volumetric examination of the weld through thickness in one circumferential scan, including
          transverse scans. The instrument shall provide a linear “A” scan presentation for each channel
          selected. A selectable “A” scan presentation shall be available during scanning. The
          inspection channels shall allow assessment of the full volume of the weld scanned, and all
          defects present detected, located, and sized in the through-wall direction. Instrument linearity
          shall be determined in conformance with ASTM E317, within 3 months of the intended
          end use date. A Company-approved form shall be used to document instrument linearity
          verification. The Contractor shall retain a current copy of the calibration certificate at the
          worksite. Each transducer’s performance shall be documented before use on a Contractor-
          developed form approved by the Company.
          6.6.1.   The transducer array design shall be specific to the project where the inspection is to
                   be performed.
          6.6.2.   All search units shall be contoured to match the curvature of the pipe surface.

7. System Set-Up
    7.1. Transducer Positioning and Primary Reference Sensitivity
          7.1.1.   Primary Reference Level
                   The system shall be optimized for field inspection using the calibration standard
                   established in Subsection 6.5 – Calibration, in this specification. Each transducer shall
                   be positioned at its operating distance away for the simulated weld centerline on the
                   calibrated standard and adjusted to provide a peak signal from its target reference
                   reflector in the transducer’s inspection zone. The peak signal response shall be
                   adjusted to ideally 80% of the full screen height for each channel. This gain level shall

Highlighting indicates revisions made as of the date on this standard.                         Page 6 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

                   be the primary reference for that transducer. The primary reference level shall be
                   recorded in the procedure (see Subsection 6.3 - Procedure Qualification (above)).
                   Variation and corrective actions to maintain calibration shall be recorded on a
                   Company pre-approved form. Whereas for intitial standardization, the channels shall
                   be adusted to ideally 80% full screen height, a satisfactory standardization during
                   inspection may indicate 70 to 99% full screen height. The noise shall be at least 20dB
                   weaker than the signal at the target path, and electronic noise shall be 40 dB weaker.
                   The transducer operating distance relative to a marked index point shall also be
                   recorded on the form.
          7.1.2.   Gate Settings
                   Using the calibrated standard, each detection gate shall be set to cover a sound path
                   which starts at least 1 mm before the weld heat-affected zone or 3 mm before the weld
                   preparation, whichever is greater, and ends at least 1 mm past the weld centerline.
                   The gate start position and gate length (with respect to the weld preparation for each
                   detection channel) shall be recorded in the procedure.
          7.1.3.   Gate Recording Threshold
                   The recording threshold of each detection channel shall be equivalent to the
                   requirements of API 1104 Specific targets for porosity detection should be determined
                   by a NDT level III in the ultrasonic method and shall be validated at AUT qualification.
                   7.1.3.1. AUT system qualification, defect assessment, and verification trials shall be
                            performed during and after Weld Procedure Qualifications.
                   7.1.3.2. AUT operators/scanner technicians shall notify Visual Company Inspectors of
                            any suspect visual calls noted on the outer surface or inner surface. The area
                            will be noted on the NDE Daily Sheet as a visual inspection request (VIR) with
                            applicable area and the visual inspector will measure and confirm the area
                            with mechanical measurement. The visual inspector shall accept or reject the
                            visual indication in the applicable weld area and sign the report.
          7.1.4.   Chart Recorder Set-Up
                   Channel output signals shall be arranged on the AUT strip chart record in an approved
                   order. In addition, time delays shall be applied to the signals to compensate for
                   different transducers circumferential positions relative to the circumferential zero point
                   and subsequent distance markers. Details of the delays applied and the chart
                   arrangement shall be recorded in the procedure. Variations and corrective actions to
                   maintain calibration shall be recorded on a Company pre-approved form.
          7.1.5.   Circumferential Scanning Velocity
                   Where W c is the narrowest -6dB beam width at the applicable operating distance(s) of
                   all transducers in conformance with the design requirements, and PRF is the pulse
                   repetition frequency per transducer, the circumferential scanning velocity, Vc shall be
                   limited by the equation:



                       Vc < W c PRF
                                  3

          7.1.6.   Circumferential Scanning Direction
                   The circumferential scanning direction shall be clockwise when viewed from the
                   upstream side of the pipe in the direction of construction. Drag sections and other
                   instances when the pipe is scanned in reverse shall be documented “REVERSE
                   SCAN” on the weld chart and NDE Weld Form.


Highlighting indicates revisions made as of the date on this standard.                          Page 7 of 17
                                                   Section: C1071
                                                   Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                             Created: 2009-11-01

    7.2. Transfer Value
          A comparison of transfer value shall be made between the calibration block and the pipe weld
          to be tested. The value shall be taken through the weld and the result shall be added or
          subtracted, as applicable to or from the reference level. This task shall be performed at the
          weld procedure qualification and or prior to any field work. The result of the transfer value shall
          be documented with the weld identification (Weld ID) and probes used.
    7.3. Operating Log
          A log shall be kept of adjustments, changes to measurements, dB changes, or other changes
          from those stated in the weld specific technique (instruction) setup. Major changes shall be
          documented (e.g. wedge change outs, transducer changes, dB calibration settings). Essential
          variables include: temperature, probe angles, probe frequency, probe offsets and channels per
          weld.

8. Calibration Qualification
    8.1. Detection Channels
          8.1.1.   With the system optimized, the calibration standard shall be scanned in the multiplexed
                   mode. The recording medium should indicate 70-99% (within specified tolerance) of
                   full screen height signals from each calibration reflector recorded in their correct
                   position assigned on the recorder chart. The actual calibration block reflector spacing
                   shall be compared to the AUT recorded reflectors and should have a circumferential
                   positional accuracy within +/-2mm (relative to each other), when approved by the
                   Company, shall be used as the inspection quality standard to which subsequently-
                   produced calibration charts shall be judged for acceptability. Due care shall be taken
                   with respect to the start / finish distance, the circumferential reading shall be within ± 1
                   cm of actual measured pipe circumference. Also the start position shall be clearly
                   marked on the pipe at approximately 12:00 with a permanent marker and the probe
                   array lead transducer shall be started at that location.
          8.1.2.   Calibrations shall be performed at the same scanning speed as the weld examination.
                   As this is a dynamic scan, no stop/starting or rocking is allowed.
          8.1.3.   Calibrations will provide sufficient but not saturated overtrace weld coverage to
                   adjacent zonal areas. The adjacent inspection zone coverage from primary target
                   location to adjacent reflector target separation should provide a response no less than
                   14dB but greater than 6dB.
    8.2. Coupling Monitor Channels
          8.2.1.   When using separate coupling monitors, amplifier gain shall be adjusted (when on the
                   calibration standard) to produce a maximum echo height of 80% from the full “vee”
                   path. When in through transmission mode, amplifier gain shall be adjusted to produce
                   a maximum echo height of 80% FSH and up to a maximum of +12dB transmitted
                   signal. The recording gate threshold level shall be a minimum of 20% of full screen
                   height to a maximum 80% of full screen height.
          8.2.2.   The re-inspection of the calibration standard (with its surface wiped dry) shall produce a
                   record showing a lack of coupling (i.e., absence of recording signal).
    8.3. Measurement to Assess Transducer Wear
          8.3.1.   Transducers shall be certified as meeting the performance requirements of British
                   Electrical Supply Industry Standard ESI 98-2.
          8.3.2.   The Contractor shall provide baseline height measurement for the transducer case
                   before the start of field weld examinations. An accurate (± 0.1 mm) measurement shall
                   be made of the case height at each of the corners of each transducer. These
                   measurements shall be recorded on a Company-approved form for a comparison of


Highlighting indicates revisions made as of the date on this standard.                            Page 8 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

                   periodic measurements taken throughout the project (to assess transducer wear). The
                   frequency of checks shall be defined in the AUT procedure.

9. Field Inspection
    9.1. Weld Marking
          9.1.1.   The Contractor shall permanently identify each weld with a unique weld identification
                   (Weld ID) following the Weld ID system provided by the Company.
          9.1.2.   The Weld ID shall be positioned on the pipe surface on the work side of the pipe and
                   on the downstream side of the weld in a manner approved by the Company
                   Representative.
          9.1.3.   Applied using a white or black permanent marker, Weld IDs shall be large enough to
                   be clearly identified from work-side of the Right-of-Way (ROW) (e.g. a distance of 3
                   meters). The Weld ID shall be located at a distance from the weld that ensures that the
                   ID shall not be covered or obscured by other operations and on the pipe such that the
                   Weld ID is visible while viewing the pipe when laid in the ditch.
    9.2. Coating Cutback
          Transducers shall not be allowed to run on coating (partially or fully, in any way) during
          production weld inspection. Coating cutback distances must be defined in the AUT procedure
          for Manual and GMAW inspections.
    9.3. Surface Condition
          The scanning surface shall be free of weld spatter and other irregularities that could interfere
          with the movement of the transducers, the coupling, or the transmission of acoustic energy into
          the material. Refer to Section 5.0 – Visual Observation.
    9.4. Reference Line
          9.4.1.   Before welding, the Contractor wherever possible shall scribe a reference line on the
                   pipe surface on the intended band side of pipe - at a known distance from the weld
                   centerline - with an accuracy of ± 0.5 mm from the centerline of the weld preparation.
          9.4.2.   The reference line shall be used to ensure that the band is adjusted to the same
                   distance from the centerline as to replicate the calibration standard. The tolerance to
                   ideal positioning shall not exceed ± 0.5 mm. The 12 o’clock position (not 1 o’clock or
                   2 o’clock) shall also be clearly marked on the pipe to indicate the start position (0 cm)
                   position of the scan.

10. System Verification
    10.1. Sensitivity
          10.1.1. The calibration standard shall be used daily:
                       Before beginning pipe weld inspections
                       After repairs
                       Thereafter at intervals not exceeding 1 hour or 10 welds, whichever comes first
                       At the conclusion of the shift
          10.1.2. The percent of screen reflections from the calibration standard shall be recorded on a
                  Company-approved Sensitivity Record Form. Any gain changes required to maintain
                  required sensitivity during the shift as per Section 7.0 - System Set-Up (in this
                  specification) shall be recorded on the same Company-approved form. The hard-copy
                  record of the calibration standard scan shall be included sequentially with the weld
                  inspection charts. The last weld inspected - and the time when it was performed, shall
                  appear on each calibration chart. These charts shall be made available to the
                  Company when requested.


Highlighting indicates revisions made as of the date on this standard.                         Page 9 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

    10.2. Circumferential Position Accuracy
          The position accuracy of chart distance markers shall be validated twice during each working
          shift. The scanner shall travel from the zero position with the scanning frame coincident with
          the pipe. At all o’clock positions, the index marks on the scanning frame and the pipe must be
          aligned. The chart shall then be compared to circumferential distance measured with a
          diameter tape. Chart accuracy shall be within ±1 cm or better. Results shall be recorded on a
          Company-approved form.
    10.3. Transducer Surface Position Accuracy
          Shifts in the refracted angle - resulting from the change in velocity within the wedge line
          materials or pipe due to temperature, strain, or compositional variances - shall be corrected for
          by adjustment of the transducer surface position. Care shall be taken to ensure that the gate
          start, length, and height positions are still set following any such adjustment with each
          calibration performed.
    10.4. Temperature Control
          10.4.1. It shall be the NDE Contractor’s responsibility to maintain the AUT system temperature
                  differential. The temperature differences between the probes on the welds (to be
                  inspected) and on the calibration block shall be monitored using a thermocouple
                  mounted in a probe housing of the probe array. The temperature shall be monitored
                  using a thermocouple and recording device. The temperature shall be recorded on all
                  the hard-copy printouts. These steps are mandatory unless an alternative procedure
                  can be validated before field inspection to ensure inspection overlap coverage is not
                  compromised.
          10.4.2. For testing at temperature extremes in hot or cold conditions, the calibration block may
                  require heating or cooling to keep the array within a 18°F (10° Celsius) variance when
                  completing ‘calibration in’, weld scan, and ‘calibration out’ cycles. This shall be
                  documented on all hard-copy calibration and weld printouts.

11. Scanning Sensitivity
    The calibration standard shall be used to verify scanning sensitivity:
       At the start of each shift
       Thereafter at intervals not exceeding 1 hour or 10 welds, whichever comes first
       Weld Repair calls
       At the conclusion of each shift
       Change out of AUT essential variables – Section 7.3 – Operating Log
       At any other intervals defined by the Company
    At a minimum, the first 50 production welds shall have calibrations both before and after welding to
    ‘wear in’ the system (unless Contractor procedure is more stringent). Records for each calibration
    standard scan shall be included, sequentially, with the weld inspection data. During production weld
    inspection, at the Company’s discretion, the system may be operated at a higher gain to ensure
    detection of defects and to compensate for differences in coupling efficiencies between the calibration
    standard and the production pipe (this will be established at the welder qualifications and then
    implemented into the AUT procedure). An acceptable calibration during production indicates values
    between 70-99% FSH(unless otherwise approved). Calibration outside of this range shall require re-
    calibration of the system unless qualified or compensated for during AUT qualifications.

12. Re-Inspection
    12.1. Sensitivity
          Welds examined at a sensitivity varying more than the scanning sensitivity as defined in Section
          11 - Scanning Sensitivity , shall be re-inspected or reevaluated at acceptable procedure defined
          settings from last acceptable calibration point.

Highlighting indicates revisions made as of the date on this standard.                       Page 10 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

    12.2. Coupling Losses
          If not covered by TOFD, welds examined with coupling losses in more than 1 adjacent channel
          shall be manually scanned with MUT to verify that no defects in the couplant loss area exist,
          and/or to identify why the couplant loss condition exists (e.g. mud on the pipe, excessive hi-lo,
          or an actual defect). Individual losses of couplant shall not exceed 12mm.
    12.3. Scanner Slippage
          Scans shall not be acceptable where any slippage occurs. Where slippage is noted, the
          scanner shall be adjusted to correct the problem and the scan shall be repeated, with no
          additional compensation to the Contractor.
    12.4. Defect
          A weld originally called for repair by automatic ultrasonic inspection shall be re-tested by
          manual ultrasonic testing per Appendix A (of C1071) after the repair has been completed.
    12.5. Temperature Variance
          Welds examined with wedge temperature difference greater than ±18°F (+10° Celsius) from the
          temperature used for calibration shall be re-tested, unless qualified before field work.
    12.6. Other Conditions
          When conditions exist where data collection is impaired or data is lost, the weld shall be re-
          tested. In addition, 100% of all welds shall be inspected.

13. Transducer Performance
    Transducers shall be examined for wear as per Subsection 8.3 - Measurement to Assess Transducer
    Wear (in this specification), whenever wear is apparent, but not greater than at an interval of 500
    welds. The signal-to-noise value for each transducer shall also be compared to the value obtained in
    conformance with Section 7 - System Set-Up - at the start of the project. Case heights and the
    signal-to-noise value for each transducer shall be recorded on Company-approved forms.
    13.1. Transducer Re-surfacing/Replacement
          The transducer contact face shall be re-surfaced, or the transducers replaced to correct any of
          the following:
          13.1.1. Beam angle changes of ±1.5 degrees for angles less than 45 degrees or ± 2 degrees
                  for angles greater than 45 degrees.
          13.1.2. Squint angles exceeding 1.5 degrees for single crystal transducers and 2 degrees for
                  twin crystal transducers.
          13.1.3. A signal-to-noise reading 6 dB less than the value obtained at the start of the project.
          13.1.4. Scores in the transducer wear face, which exceed 0.5 mm in depth.
          13.1.5. Balance from corresponding upstream and downstream channels shall not differ by
                  more than 6dB.

14. Evaluation of Indications
    14.1. GMAW Welds
          All identical wall GMAW welds will be inspected to API 1104 Appendix A “Alternative
          Acceptance Criteria”. Indications from GMAW weld imperfections shall be evaluated according
          to defect acceptance criteria defined by the Company using an Engineering Critical Assessment
          carried out according to API 1104, Appendix A. Evaluations shall be completed immediately
          after inspection of the weld. Failure to meet the API 1104 – Appendix A “Alternative
          Acceptance Criteria” requires Manual UT.



Highlighting indicates revisions made as of the date on this standard.                        Page 11 of 17
                                                 Section: C1071
                                                 Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                           Created: 2009-11-01

    14.2. Manual UT Inspection on Repairs
          14.2.1. To verify that the initial GMAW defect has been completely removed.
                                                                               th
          14.2.2. To verify that the Manual repair conforms to API 1104 19 Edition Section 9.6. See
                  Appendix A - Specification for Manual Ultrasonic Inspection of Pipeline Welds of this
                  document for more details.
                                                      th
          14.2.3. All welds evaluated to API 1104 19 Edition Section 9.6 with individual indications with
                  a vertical height (through-wall) dimension determined to be greater than one quarter of
                  the wall thickness and multiple indications at the same circumferential location with a
                  summed vertical height (through-wall) dimension exceeding one half the wall thickness
                  shall be considered defects.

15. Weld Inspection Record and Reports
    15.1. Weld Inspection Record
          The Weld Inspection Record shall consist of a complete strip chart type record showing the
          reference point, the direction of scanning, the Weld ID, date and time of inspection, and the
          Operator signature. The search unit number in each channel shall also be recorded. The
          Company may approve alternate record formats.
    15.2. Interpretation of Weld Inspection Record
          Areas on the AUT chart with a corresponding pulse echo response (exceeding evaluation
          threshold) will be evaluated and given a disposition by the operator as to relevant or non-
          relevant indication. For example: This area can be highlighted on the strip chart and have an
          associated comment in the ‘comments’ section of the AUT chart or have any other code
          compliant technique for dispositioning these pulse echo responses such as go/no go bars.
    15.3. Weld Acceptance from the Inspection Record
          15.3.1. A weld shall be considered acceptable when the weld inspection shows discontinuities
                  that do not exceed the tolerable values as given by the Company.
          15.3.2. The Contractor shall submit ultrasonic reports to the Company daily, as directed by the
                  Company, using forms approved by the Company. Contractor shall ensure that
                  chainage is shown on the documents at maximum of 10 joint spacing (~800 foot 250
                  meter) intervals. Contractor shall ensure that all electronic data files and hard-copy
                  graphs are further identified with the following:
                     Spread Identification
                     Project Name
                     Work Order Number
                     Date
                     Inspector’s Name
                     Inspection Unit Number
                     Weld Number
          15.3.3. All completed hard-copy graphs shall be catalogued and packaged in corrugated
                  cardboard boxes supplied by the Contractor. Boxes of mainline hard-copy graphs shall
                  be number sequentially in folders of 100 with corresponding Calibrations.
          15.3.4. Clear and legible interpretation NDE sheets shall be included inside the boxes for those
                  100 graphs catalogued in the corresponding folder. Each corrugated box shall hold
                  approximately 2500 welds, including associated NDE sheet judging acceptance.
          15.3.5. Contractor shall supply labels for the recorder chart boxes. The following information is
                  required on the labels:
                     Project Name
                     Work Order Number
                     Weld IDs of contained Inspection Reports
Highlighting indicates revisions made as of the date on this standard.                       Page 12 of 17
                                                 Section: C1071
                                                 Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                           Created: 2009-11-01

                     Box Number
                     Name of Ultrasonic Technician responsible for Weld Interpretations
    15.4. Data Management
          Backups of all digital data shall be made on a daily basis with a minimum of two backups of
          each file on separate mass storage media. The working computer system disk shall not be
          considered as a backup.
    15.5. Archives
          At a recurring frequency approved by the Company (bi-weekly at a maximum), all electronic
          data files shall be archived on CD-ROM and presented to the Company. For these files, the
          Contractor shall supply MS-DOS or MS WINDOWS® viewing software and instructions
          for its use.

16. Additional Requirements
    16.1. Weld Repair List
          16.1.1. Weld repair lists shall be provided to the Company in electronic and hard-copy format
                  each working day by 6:00 A.M.
          16.1.2. Repair lists should report the weld quality, including outstanding weld repairs and/or
                  weld cutouts, of all welds completed by the prime Contractor’s mainline welding crew,
                  by 4:30 P.M. each day or an agreed upon time.
          16.1.3. The Contractor shall have all necessary equipment and personnel on site to maintain a
                  production rate that ensures any weld completed before 5:30 P.M. on any working day
                  shall be inspected, interpreted, and included in the Ultrasonic Report for that working
                  day.
    16.2. Destructive Testing Equipment
          ‘Cut out’ Pieces shall be tracked and verified by a Company Representative. Samples shall be
          sent to an approved Materials Testing Vendor for sample verification when deemed necessary.
    16.3. Training
          Contractor shall provide training (one-day minimum) for Company Inspectors on the techniques
          of ultrasonic inspection. This training shall be at a date and location determined by the
          Company and shall occur before production welding starting on the project.

17. Records
    17.1. Record Retention
          The following records shall be kept:
          17.1.1. Scanning Calibration, including:
                     Material and manufacture
          17.1.2. Standard Records, including:
                     Date of manufacture of the standard
                     Actual dimensions and angles of reference reflectors
                     Actual surface roughness and outer profile
          17.1.3. Search Unit Records, including:
                     Make, date of manufacturing, and serial number
                     Nominal dimensions
                     Nominal frequency
                     Height and wear


Highlighting indicates revisions made as of the date on this standard.                     Page 13 of 17
                                                 Section: C1071
                                                 Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                           Created: 2009-11-01

          17.1.4. Ultrasonic System Records, including:
                     System description
                     Make and model number of units
                     operating procedure
                     technique detail
                     operator log
    17.2. Report Format
          Equipment performance logbooks, daily reports, and repair lists shall be presented to the
          Company in both hard-copy and electronic format. The format shall be comma-delimited ASCII,
          Excel, Word etc and be capable of displaying and printing the Electronic Report, Daily Repair
          Summary, and Repair List.
          Evaluation of the acceptability of every weld inspection shall be reported on the "NDE Daily
          Report". The NDE Daily Report form shall be the only form used for NDE (UT, VT, RT, MT, and
          PT) for weld accepted/reject status. The Contractor shall ensure that station number or
          GPS coordinates are recorded with all applicable weld information on NDE Daily Report
          for all Tie-ins, all repair welds, start of new mainline and minigang pipe sections and
          every 800 foot (250m) interval thereafter (where applicable).
          All completed ultrasonographs shall be catalogued and packaged on a daily basis. The Repair
          List shall consist of the following fields:
             Date
             Weld ID
             Imperfection type
             Length in mm
             Position in mm from zero
             Upstream or down stream
             Depth in mm to the bottom of the imperfection
             Station number or GPS.
          The Repair List shall retain all non-status welds and/or weld repairs until repairs are cleared or
          status is determined.
          17.2.1. Reporting Production Statistics
                  Upon daily work completion, field personnel responsible for GENERAL NDE Reports
                  shall turn in completed forms to the Coordinator. The Coordinator will subsequently
                  distribute repair lists, production rates, and outstanding (non-status and un-repaired)
                  welds.
                  Database statistics from previous day’s performance will also be provided in a hard-
                  copy form. At a minimum, the report shall include (but is not be limited to) the daily
                  repair % and ‘to-date’ repair %. The Contractor shall e-mail repair information to a list
                  of Company personnel each day (to update the production history), in a graphical form
                  accepted by the Company.
                  These reports shall be delivered to the Senior Weld Inspector or a designated
                  representative by 6:00 AM unless another specific time can be mutually agreed upon.
          17.2.2. Other Reporting Requirements
                  The Contractor shall report arc burns and other pipe imperfections observed visually, or
                  where the weld does not meet the cap width or reinforcement criteria of the welding
                  procedure. Indications not specified by ultrasonic acceptance criteria shall be noted as
                  V.I.R. (Visual Inspection Required). The Contractor is not responsible for specifying
                  the accept/reject disposition of visual indications. The Company Representative/
                  Inspector shall be responsible for the disposition and sign-off of any visually-noted
                  imperfections determined by the Contractor.

Highlighting indicates revisions made as of the date on this standard.                        Page 14 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

Appendix A - Specification for Manual Ultrasonic Inspection of Pipeline Welds
1. Scope
    This specification establishes the requirements for manual ultrasonic examination of pipeline GMAW
    weld verifications and manual repair girth welds. When directed by the Company and/or qualified,
    this inspection technique can be used to verify other radiographed welds, longseam welds, loss of
    couplant indications on tieins and assist AUT tie-in inspection when repair areas are outside of
    covered scope. Referenced paragraphs in the main Ultrasonic Specification are indicated.

2. Requirements
    2.1. Personnel
          2.1.1.   Personnel performing manual ultrasonic inspection shall be certified to a minimum of
                   SNT-TC-1A Level II, in ultrasonic testing of butt welds in steel. Personnel shall have
                   an additional 16 hours of contractor in-house training on manual inspection of girth
                   welds. Additionally, a further 16 hours of training is required on the manual
                   examination of pipeline girth welds using procedure-approved probes in a manual
                   zonal inspection approach. Personnel shall demonstrate proficiency in calibrating
                   portable A-scan instruments on the sensitivity calibration blocks with the selection of
                   probes to be used to inspect different zones. Personnel must demonstrate proficiency
                   in setting and interpreting both the amplitude scale and the time gates to determine
                   location of received indications within the weld. Personnel must demonstrate
                   proficiency in:
                      Defining probe standoffs
                      Defining the location(s) of magnetic tape(s) to be used to keep the probe a set
                       distance from the weld centerline
                      Ability to recognize linear versus volumetric echodynamic patterns
                      Differentiate between geometry and actual weld anomalies
                   This operator training shall be documented and accompanied by a contractor in-house
                   certifying exam.
          2.1.2.   The Company shall approve operators as described in Subsection 6.2.3. Manual
                   ultrasonic operators shall be required to take the manual qualification test only.
    2.2. Procedures
          A procedure, specific techniques, and instructions for each thickness and variation shall be
          prepared (see paragraphs 6.3.2, 6.3.3, 6.3.4, and 6.3.5 of this specification). The procedure
          shall meet the requirements of API 1104, Section 11.4 .
    2.3. Equipment
          2.3.1.   Ultrasonic equipment shall conform to the requirements of ASME Section V, Article 5.
          2.3.2.   Instrument “A” scan presentation shall conform to the requirements of ASME Section V,
                   Article 5.
          2.3.3.   Instrument linearity shall meet requirements of Subsection 6.6 - Ultrasonic Equipment,
                   of this specification.
          2.3.4.   Ultrasonic probes shall conform to the requirements of Subsection 6.6 - Ultrasonic
                   Equipment, of this specification.
          2.3.5.   Sensitivity calibration blocks (sensitivity) shall meet the requirements of Paragraph 6.5
                   – Calibration, of this specification. The calibration blocks shall also conform to the
                   requirements of the proposed API Sensitivity Reference Standard.
          2.3.6.   Reference blocks for angle beam distance calibration shall conform to the requirements
                   of ASTM E164 Appendix 1.

Highlighting indicates revisions made as of the date on this standard.                        Page 15 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

          2.3.7.   Coupling shall be a water-based product. Water or gel (for which an MSDS sheet must
                   be provided) may be used - but shall be completely removed upon completion of the
                   test.
    2.4. Calibration
          2.4.1.   MUT – AUT defect verification - zonal calibration shall conform to the requirements of
                   Section 7 - System Set-Up, of this specification, except that individual probes shall be
                   set up according to the parameters used by the automatic ultrasonic unit. The
                   “standoff distances” shall be noted if different from the technique detail. Transfer value
                   will be checked and documented for each weld procedure.
                                                                                        th
          2.4.2.   MUT – Weld Repair assessment shall conform to API 1104, 19 Edition – Section
                   11.4.5 – API Sensitivity Reference Standard. Manual ultrasonic testing shall be based
                   on a two or three point reference level Distance Amplitude Correction(DAC). Transfer
                   value will be checked and documented for each weld procedure.
    2.5. Scanning
          2.5.1.   The repaired area shall be scanned ± 50mm either side of the repaired area. Tested
                   acceptable areas shall be marked to indicate acceptance or rejection status and noted
                   as such on the AUT report. If defects are present then specific MUT reports shall be
                   required to describe the unrepaired defect. This form shall be approved by the
                   company.
          2.5.2.   GMAW AUT transducer setups used shall mimic sensitivities from AUT calibration
                   block when verifying GMAW indications and ensuring all GMAW defects have been
                   removed in designated repair areas.
          2.5.3.   MUT fixed scans (i.e., the standoff distance for the probe and zone shall be determined
                   and the probe shall then be set at this distance) with AUT transducer arrays can be
                   used to assist the operator in GMAW defect verifications and removal confirmation. A
                   magnetic strip can be used to keep the probe at the set distance for the length of
                   repair. The adjacent zones may be examined in the same way for GMAW confirmation
                   or removal.
          2.5.4.   The probes used shall be the same probe specification as the original mechanized test
                   (i.e., probe beam angle shall be ±1.5 degrees for angles less than 45 degrees, or ± 2.0
                   degrees for angles greater than 45 degrees, and probe frequency, focus, crystal size
                   and configuration shall match.).
          2.5.5.   The manual repair weld manual UT setup sensitivities shall be set off of N10 ID/OD
                   notches with DAC curves. All interpreted indications shall be recorded on the report
                   sheet.
          2.5.6.   The scans for manual repair welds shall be in accordance with ASME V article V and
                   acceptable to the Company. A minimum of three angles shall be used to confirm the
                   weld volume of the repair area. Ie. 45, 60 and 70 Degree.
          2.5.7.   When manual ultrasonic inspection of a repair weld is required, naturally focus probes
                   are to be used. Root inspection focused transducers may be permitted for bead
                   inspection. The test shall be conducted with a minimum of three different angles to a
                   procedure based on ASME V article 5 and acceptable to the Company.
    2.6. Recording
          2.6.1.   The “AUT” zonal defect verification setup – established with AUT sensitivies from the
                   AUT procedure will be used for defect verification removal. Threshold gates when
                   exceeded shall be recorded. Ensure all GMAW defects have been removed in their
                   entirety.
          2.6.2.   All indications shall be noted above 50% of DAC with 6dB added (scanning and
                   evaluation sensitivity) inspected with the 45, 60, and 70 degree volumetric inspection.

Highlighting indicates revisions made as of the date on this standard.                         Page 16 of 17
                                                  Section: C1071
                                                  Title:   Ultrasonic Examination of Pipeline Welds
CONSTRUCTION STANDARDS                            Created: 2009-11-01

          2.6.3.   The acceptance criteria for GMAW/SMAW/FCAW repair areas performed by MUT shall
                                               th
                   be defined by API 1104, 19 Edition - Section 9.6.
                                                                         th
          2.6.4.   In addition, all welds evaluated to API 1104 19 Edition Section 9.6 - Individual
                   indications with a vertical height (through-wall) dimension determined to be greater
                   than one quarter of the wall thickness and multiple indications at the same
                   circumferential location with a summed vertical height (through-wall) dimension
                   exceeding one half the wall thickness shall be considered defective.
          2.6.5.   Discontinuities and defects shall be noted numerically and diagrammatically on the
                   report. Reports shall be pre-approved before fieldwork.
    2.7. Reporting
          2.7.1.   At a minimum, reports shall include:
                      Operator Name and Date of Test
                      Technique number or unique ID
                      Zone(s) examined
                      Thickness
                      Gate start and length
                      Repair or verification or other detail
                      Position(s) and length(s)
                      Coupling used (and removed)
                      Probe(s) used (type, size, frequency) and standoff and gain setting
                      Unique number of calibration (sensitivity) block used
                      Defect/Discontinuity description, depth, location, size length, zone, and amplitude
                      UT instrument serial number
          2.7.2.   All welds tested shall be reported on a separate report sheet(s) approved by the
                   Company. The reports shall be handed to the Senior Weld Inspector on a daily basis
                   to meet the requirements of Section 15 - Weld Inspection Record and Reports. The
                   designated Job Supervisor shall verify the areas scanned against the interpretations
                   made by both automatic ultrasonic inspection and manual ultrasonic inspection. The
                   Job Supervisor shall record and store said data for review and transmittal to the
                   Company. Reports shall be kept current and up-to-date with weld repair production.




Highlighting indicates revisions made as of the date on this standard.                        Page 17 of 17

						
Other docs by PItEY5iY
TRAUMATISMOS DE TORAX
Views: 61  |  Downloads: 0
Diapositiva 1
Views: 0  |  Downloads: 0
PlanPR04 05
Views: 2  |  Downloads: 0
Rutherford k�s�rletei
Views: 3  |  Downloads: 0
Vis Render2006 Review
Views: 0  |  Downloads: 0
�VOD DO POC�TACOV�CH SIET�
Views: 6  |  Downloads: 0
5 Valvulopatia aortica y tricuspidea
Views: 111  |  Downloads: 0