Transducer Interface
Designed for use with DC strain gauge amplifiers with excitation of 3V DC or more.
A drive voltage of 0.9V which protects the sensor from electrolytic damage while it is implanted.
Product Overview
The S13 interface is designed for use with DC strain gauge amplifiers with excitation of 3V DC
or more. It supplies the Gaeltec Devices miniature pressure sensor with a drive voltage of 0.9V
which protects the sensor from electrolytic damage while it is implanted.
The unit has trimpot controls for adjustment of gain and zero and is supplied adjusted for the
particular drive voltage of your amplifier.
Up to three S13 circuits can be fitted into one case. The case dimensions are 100 x 50 x 25mm.
© 2012 Gaeltec Devices Ltd - Miniature Pressure Sensors and Pressure Monitoring Equipment
Telephone: +44 (0)1470 521 385 Fax: +44 (0)1470 521 369 Email us: web@gaeltec.com
Gaeltec Devices Ltd, Glendale Road, Dunvegan, Isle of Skye, Scotland IV55 8GU
The transducer connection is via a high quality Lemo insert in a lighweight nylon plugshell on
2 metres of PVC cable. The monitor connecting lead can be bare ended or fitted at cost with a
customer specified connector.
The S13 does not provide patient isolation. The monitor must be patient isolated.
Where to Buy
The Transducer Interface can be ordered direct through Gaeltec Devices Ltd or one of many
approved distributors Worldwide.
© 2012 Gaeltec Devices Ltd - Miniature Pressure Sensors and Pressure Monitoring Equipment
Telephone: +44 (0)1470 521 385 Fax: +44 (0)1470 521 369 Email us: web@gaeltec.com
Gaeltec Devices Ltd, Glendale Road, Dunvegan, Isle of Skye, Scotland IV55 8GU