Zeiss Axio Imager Firmware Update Draft procedure

Document Sample
Zeiss Axio Imager Firmware Update Draft procedure Powered By Docstoc
					                                                                                                             No:924-10000-00
                                                                                                             Page: 1 of 22
                                                                                                             Revision: A
 Title: Zeiss Axio Imager Microscope Firmware Update Procedure

Owner: E.Pettigrew / R&D
 Rev     Co No.                                                   Revision History                                                     Effective Date



Owner Signature / Date                                                      Doc Control / Quality Assurance Approval / Date




          Zeiss Axio Imager Microscope Firmware Update Procedure

          Note:
          In order to update the Main Control firmware to version 01.002, the microscope
          must have at least firmware version 00.705 installed and the VIAS application
          must be version 2.0.0.1 or higher. Updating the firmware to 01.002 will enable the
          ability to save light and condenser settings using the Light Manager Classic
          mode function of the microscope.

          Make sure that all motorized components are attached and connected to the
          microscope. Select the CAN Node for the component you want to update and
          also select the correct firmware and descriptor file.

          It may take up to 30 seconds to populate the CAN Nodes list. Please wait until it
          finishes. Likewise the firmware and descriptor list can take a few seconds to
          populate.

          For each additional update, please start from the beginning by selecting:
           1. Connect Tab and Click Connect
           2. Firmware Tab and select the CAN Node that needs to be updated.

          You must turn the microscope power Off then ON after you update the firmware
          for All the components in the microscope.

          The MainControl, Nosepiece, Condenser, Power Supply, and LightControl
          firmware will be updated first using the procedure in section 1 Updating
          Firmware. The existing USB cable connection between the PC and microscope
          is used during this Procedure.

          The Focus Firmware will be updated next using a serial connection between the
          PC and the microscope. See the section 2 Focus Update for the proper
          procedure for updating.

          Please follow all instructions carefully.



           This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
                   express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 2 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      1. Updating Firmware
      Insert the Firmware CD into the CD-ROM Drive of PC.
      From the CD ROM, left click on the PhxTest folder. Double click on PhxTest.exe. The
      utility program will start.
      In the PhxTest Software window, select “Connect” tab. Select USB.
      Click “Connect” button (1) then select the “Firmware” tab (2).




      Next, specify where to get the new firmware data by clicking on the browse button to the
      right of the “Firmware Source Folder” textbox.

       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 3 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      Select the correct folder where the new firmware is located.
      Your Folder should be: CD ROM Drive (usually D):\Z1M1_Jun28_Firmware.
      Then click OK.




      1. Click Scan Stand. Wait for the CAN Nodes list to populate.
      2. Select MainControl in the CAN Nodes Field.
      3. Select 01_002 in the Firmware Field and click the download button.

       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 4 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      Please, wait until the process has been completed (see next image).




      The Main Control firmware update does not contain a descriptor file to be downloaded.
      The other components (Condenser, Nosepiece, Power Supply, and LightControl) will
      have descriptor files to be downloaded together with the firmware update. In that case,
      you must select the descriptor in addition to the firmware and then click on the button
      “Download”




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 5 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      The Condenser component will be updated next.




      Highlight Condenser. Click on the firmware number. Select condenser_b as the
      Descriptor. Click Download. After the download completes successfully, update the
      Nosepiece.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 6 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      After every download you will see a similar window to the one shown above.
      It indicates that the Descriptor for the Firmware of the Component updated was
      successful.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 7 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      The Nosepiece component will be updated next.




      Highlight Nosepiece. Click on the firmware number. Select nosepiece07 as the
      Descriptor. Click Download. After the download completes successfully, update the
      Power Supply.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 8 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      The Power Supply component will be updated next.




      Highlight Power Supply. Click on the firmware number. Select powersupply as the
      Descriptor. Click Download. After the download completes successfully, update the
      LightControl.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 9 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      The LightControl component will be updated next.




      Highlight LightControl. Click on the firmware number. Select
      lightcontrol_mot_ref_6_obj_7 as the Descriptor. Click Download. After the download
      completes successfully, update the Focus Z1. See section 2 to update the Focus
      firmware.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 10 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      2. Focus update
      Connect the RS232 cable (null-modem cable) between COM3 of the PC and the
      microscope focus PC board.




      Connect the RS 232 cable from the microscope to the RS 232 connector on PC.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 11 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      Connect the RS 232 cable from the PC to the Focus Board RS 232 connector.

      Start the PhxTest utility program. (As described in step 1) if not already open.




      Highlight Focus Z1. Click on the firmware number. Select focus_3ph_japanservo as
      the Descriptor. Click Download.

      Now follow the onscreen directions.




      Select COM 3.



       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 12 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      Wait until you see the “Download was successful” message. Then click “close”.




      Turn microscope off and click OK.




      Disconnect RS232 cable (null modem cable) from focus board and from the PC.
      Turn microscope on.

      The Microscope firmware is now updated.

      After the Firmware is updated you will have to reassign the objective types and turret
      position for each objective (see section 3 Resetting Objective Types and Turret
      Positions). Also the light manager setting has to be set to Classic and the Dazzle
      Protect setting set to off (see section 4 Dazzle Protect and Light Manager Settings).




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 13 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure


      3. Resetting Objective Types and Turret Positions




      From the Home screen, press Settings.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 14 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      The Components screen will open. Make sure the Objectives tab is selected. Press
      Objective 1




      Press via catalog no.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 15 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      Using the keypad, enter 1113-114. Press OK. You should see A-Plan 2.5x/0.06 under
      the resulting configuration and 000000-1113-114 under the Catalog number. Press
      Save. When the Replace objective 1? dialog box opens, press YES.




      Press Objective 2




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 16 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      Press via catalog no.




      Using the keypad, enter 441020. Press OK. You should see A-Plan 5x/0.12 under the
      resulting configuration and 441020-0000-000 under the Catalog number. Press Save.
      When the Replace objective 2? dialog box opens, press YES.




      Press Objective 3

       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 17 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      Press via catalog no.




      Using the keypad, enter 440030. Press OK. You should see Achroplan 10x/0.25 under
      the resulting configuration and 440030-0000-000 under the Catalog number. Press
      Save. When the Replace objective 3? dialog box opens, press YES.


       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 18 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      Press Objective 4




      Press via catalog no.




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 19 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      Using the keypad, enter 440040. Press OK. You should see Achroplan 20x/0.45 under
      the resulting configuration and 440040-0000-000 under the Catalog number. Press
      Save. When the Replace objective 4? dialog box opens, press YES.




      Press Objective 5


       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 20 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      Press via catalog no.




      Using the keypad, enter 440050. Press OK. You should see Achroplan 40x/0.65 under
      the resulting configuration and 440050-0000-000 under the Catalog number. Press
      Save. When the Replace objective 5? dialog box opens, press YES.


       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 21 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure




      The Objectives list should now look like this. The objectives are now reset. Press Home




       This document contains confidential and proprietary information and may not be copied or reproduced in whole or in part without the
               express written permission of TriPath Imaging, Inc, Burlington, NC. 27215. Guideline/Form No. 990-00947 Rev J.
                                                                                                         No:924-10000-00
                                                                                                         Page: 22 of 22
                                                                                                         Revision: A
Title: Zeiss Axio Imager Microscope Firmware Update Procedure

      4. Dazzle Protect and Light Manager Settings




      Press Extras. Press Dazzle Protect. Press Off. White is off.




      Press Light manager. Press Classic. When background is White, item is selected.
      Press Home.