Embed
Email

CM200

Document Sample

Shared by: Kerala g
Categories
Tags
Stats
views:
0
posted:
12/14/2011
language:
pages:
12
CM 200 Instructions 1









Asturias Lab Instructions for CM200

Last updated October 2004







SOME GENERAL THINGS TO REMEMBER:



The day before imaging

1. Fill up two liquid nitrogen dewars (ice crystals will settle to bottom of the dewar if the

nitrogen sits overnight, thereby reducing the chances of sample contamination)

2. Place cryostage into the pumping station. Consecutively open all the valves (4 total, start

with the one closest to the turbo pump, then open the two butterfly valves, and last open

the valve on the cryostage).





While working on the microscope

1. Be VERY careful with the cryostage; never bounce into it. Be especially careful

when adding liquid nitrogen.

2. The objective aperture needs to be taken out temporarily during alignment. Make

sure it is back in before taking pictures.

3. At any time when the beam is not required (filling up the dewars, taking a break,

developing film, etc) close the gun valve and sample shield (small knob on the

cryostage handle should be pushed towards the column). Open and close gun

valve VERY slowly, or the vacuum system will sense a disturbance and shut

down.





Before taking pictures, check

1. Objective aperture (it should be in, i.e., turned to the left, pointing away from the

door)

2. CM/CCD mode switch should be in CM if you intent to collect imag es on film)

3. Nitrogen level in cryostage and cold trap dewar.

4. Eucentric height of the specimen (Z-height, use A-axis wobbler and minimize

sweeping movement of specimen)

5. Check exposure time and the size and position of the beam in exposure mode





While taking pictures check:

1. Eucentric height of the specimen (every time you change squares on the grid)

2. Exposure time and exposure area (often during first 15 minutes, then

approximately every 10-15 pictures)

3. Nitrogen level in cryostage and dewar (approximately every 30 minutes)

CM 200 Instructions 2

GENERAL INFORMATION





We operate the microscope in TEM

BRIGHT-FIELD -LOW DOSE, which

facilitates imaging previously

unexposed areas of a beam-sensitive

specimen. Within the Low-dose there

are three individual modes that we

use (settings are independent for

different modes):



SEARCH MODE

Function: checking a sample under

low magnification and low electron

dose conditions to identify areas

suitable for imaging.

Magnification: 2450x

Spotsize: usually 3-4; the largest

value (to lower beam intensity and

thus minimize damage) that allows

good overview of specimen areas of

interest.



EXPOSURE MODE

Function: data collection.

Magnification: 58kx, which equals

66kx when the screen is lifted.

Spotsize: usually 5 (in general

between 4-7), choose it so that the

required exposure time is between

0.5 – 1.0 s. Then set the manual

exposure time within 20% of the

required exposure time (manual exposure time can only be set to certain predetermined values.

Remember that it is always better to have slightly underexposed pictures, as they are easier to

scan).



FOCUS MODE

Function: focusing on areas of the specimen adjacent to the area selected for data collection.

Magnification: 135kx

Spotsize: 4-6

Positions for S1 and S2: 180 degrees apart, and 2.5-3.0 μm apart from image center.

Focus step size used for focusing: 4, 3 (rarely 5-7 when very far from focus).

Focusing:

Focus first at one (S1 or S2) position. Remember that when overfocused the carbon film

support grains appear dark. Underfocused grains appear light. To focus find value of

minimum contrast which is in between overfocused and underfocused values.

Press RESET DEFOCUS button, then focus again using S2.

The mean value between SI and S2 is used as the optimal focus value from which you defocus

(usually we underfocus -0.386 um up to -1.96 um).

See Troubleshooting if you have problems focusing.

CM 200 Instructions 3

TRANSMISSION ELECTRON MICROSCOPE (TEM)

Ray diagram for a complete electron microscope. Filament F, condenser 1 lens C1, condenser

2 lens C2, condenser aperture CA, specimen S, objective lens O, objective aperture OA (in the

back focal plane). 1st intermediate image and selector aperture SA. Intermediate lens PI,

second intermediate image h, projector lens P2 and final image on the fluorescent screen F.

CM 200 Instructions 4

OPERATION PROCEDURES

PREPARATION



1) PRESTART

Fill thermo-cold dewar for cold trap with liquid nitrogen up to the top, and cover the dewar

with the Styrofoam cover to avoid nitrogen evaporation and formation of ice. Record the

values of P3 (Vacuum page on scope, normal value is between 30 and 80), room

temperature (usually ~ 68F), and relative humidity (usually ~ 68%) into the EM logbook.



2) CHECK VACUUM STATUS

Check that UV and HVAC lights (located on right side of control panel), if HVAC light is not

on check that the camera chamber is closed or see if film has been changed recently (TEM LOW DOSE->PARAMETERS



Set "GUN LENS" to 3



Set "VOLTAGE" to 120 kV



Turn High Tension ON (hold down "HIGH TENSION" button on console for ~1 second)



Go to Configuration Page MODES->CONFIGURATION->DISPLAY check extraction

voltage limit, it should be 3.81. DO NOT SET HIGHER THAN THIS VALUE!

Turn FILAMENT KNOB slowly clockwise until it reaches the extraction limit of 3.8kV

(actually 3.81). "STANDBY" will change to "OPERATION" as soon as the filament knob is

turned.



Slowly open GUN VALVE. Check to see if you can find the beam at low mag (~2500X). If

no beam is present go to lower magnification and adjust the beam shift. After finding the

beam close GUN VALVE and proceed.



4) LOADING THE SPECIMEN

4.1) LOADING the specimen into the cryostage



Grease holder’s O-ring (ONLY USE WHITE FOMBLIN grease, that can be found in a

syringe on the desktop, or the drawer in the F20 room). Put a very small amount of

grease on your finger, and rub it on the o-ring on the cryostage rod. Use only enough

grease to leave make the o-ring look shiny, NOT MORE.



Remove clip ring and old sample from rod tip PRIOR TO COOLING!



Connect cryo-holder to the temperature meter, and make sure that heater is off (panel

with the current reading is not lit). Add liquid nitrogen to cryo-holder’s dewar and make

sure temperature of tip starts going down (may take a minute). As soon as temperature

starts to go down, add LN2 to sample tip (BEFORE it is cold enough to allow moisture to

condense and contaminate the tip). Wait until the tip is at -160 C. Load the specimen.



4.2) LOADING the specimen into the MICROSCOPE

CM 200 Instructions 5



Make sure that the gun valve is CLOSED.



For cryo samples set the goniometer tilt to -55. This is accomplished using the Tilt-

knob on the left panel (clockwise is ‘+’, counterclockwise is ‘-‘, it works as ‘gas pedal’, the

more you turn knob, the faster the value changes, turn the knob to original position when

the correct value is reached) or by loading one of the saved positions from the

compustage page.



Press VACUUM ON (right panel) in order to pre-pump airlock before inserting the

specimen.



Move the mounting station with the cryostage to the microscope. Pull the cryostage out

of the mounting station, and turn it 90 º CW, so that the small pin on the top of the tip is

at the 3 o’clock position. Insert the specimen holder into the airlock entry and slide it in

until a stop is reached. A RED airlock indicator LIGHT will turn on. Wait for pumping.

When the red light goes off, press RESET AB on the Compustage page, wait while the

stage rotates, then rotate the holder 90º CCW and allow it to slide slowly into the

column. Maintain a firm grip on the holder (hold on to the back end of the rod sticking out

of the dewar) during insertion so that the holder does not enter the column too quickly

(due to the suction by the vacuum) and that no damage to goniometer and/or the holder

can occur.



Allow 10-15 minutes for vacuum to recover, and for cryostage to reach thermal

equilibrium. Attempting to examine the specimen too soon will result in specimen

contamination, and it will not be possible to collect any images, as the specimen will be

drifting considerably.





MICROSCOPE ALIGNMENT



1) CHECK COLUMN OF CM200

ALL APERTURES should be set on 4 (Except Diffraction Aperture, which is not used and

must be placed in OUT position (turned to the right, towards the door).



SLOWLY open the GUN VALVE, open the sample shield, go to SEARCH MODE, and find

an empty area.



2) SET MODES

Search mode: set magnification to 2450X



Exposure mode: set magnification TEMPORARILY to ~25KX



Focus mode: set magnification to 135X

If you cannot find the beam in focus mode, for each focus position, lower the magnification

until the beam appears on the screen, center with the SHIFT X/Y knobs (beam deflectors),

and go back to 135KX

Check FOCAL POINTS: go to focus SI or S2

SI: LDrad (3.00 um) LD rot (325º)

S2: LDrad (3.00 um) LD rot (145º)

The focal points should be about 3um away from center of exposure area, and

separated by 180º



LDrad is controlled by MULTIFUNCTION knob Y (linear image shift).

CM 200 Instructions 6

LDrot is controlled by MULTIFUNCTION knob X (rotational image shift).



3) CENTER THE CONDENSER APERTURE

Go to EXPOSURE mode (~25KX). Go to crossover (smallest bright spot) and center the

beam with the BEAM DEFLECTORS (SHIFT X/Y). Expand beam by turning INTENSITY knob

clockwise. Remember that you always want to operate in the region where the beam expands

when you turn the INTENSITY knob CLOCKWISE.

If condenser aperture is not centered, the circular beam will not be centered as you increase its

radius. Go to crossover and center the beam with SHIFT X/Y. Increase the size of the beam

until it is about the size of the big circle on the screen and center it by moving the condenser

aperture with the controls on the aperture mount. Go back to crossover and re-center the beam

using the SHIFT X/Y knobs. Repeat until the beam is centered at crossover and expands to

cover the screen uniformly.



NOTE: If the beam is not circular, press the STIGMATOR alignment knob, and adjust the

CONDENSER stigmators until the beam is circular.



4) ADJUST EUCENTRIC HEIGHT

To adjust the height of the sample go to SEARCH MODE, expand the beam, find a

recognizable feature, and center it on the screen. Press COMPUSTAGE to switch over to

the Compustage Register Control page, then press A-WOBBLER (Angle or A axis-wobbler)

- the specimen is now being tilted continuously from +15 degrees to -15 degrees, back and

forth.



Minimize image movement using the Z-joystick (if z-movement does not work, activate Z-

joystick on Compustage Control page). Pay attention to the movement of the compustage

and the image. The image will jump a little when the direction of motion changes but should

be steady in between, and the recognizable feature should stay in the same place. When

this is accomplished, key A-WOBBLER again to stop sweeping motion.



Go to EXPOSURE MODE and repeat the process for fine adjustments. At this point the

specimen should be set to the right height, and you may want to press the AUTO FOCUS

knob to get close to the correct FOCUS value.





3) ALIGNMENT OF THE GUN

Before starting check to make sure that you are adequately close to FOCUS!



Gun Shift Alignment

Go to EXPOSURE MODE, where the magnification should be set between 25K and

55K. Dim the lights in the room.

Press the ALIGNMENT button and key GUN SHIFT

Select spot size 9, and use the INTENSITY knob to bring the beam to crossover. If you

can’t see the beam, go down in spot size until you can see the beam, and then center it

using SHIFT X/Y. If necessary, go back up to spot size 9 and center the beam again

with SHIFT X/Y.

Select spot size 1, bring the beam to crossover, and center it with MULTIFUNCTION

X/Y knobs (beam shift should be highlighted in the alignment page).

Repeat the two steps above until the beam remains centered when you go back and

forth between spot sizes 1 and 9.



Gun Tilt Alignment (55k MAG / Spot-Size=1)

The correct gun tilt alignment will give a small, very intense spot in the center of the

crossover beam.

CM 200 Instructions 7



PRESS GUN TILT on the Alignment page.



Bring the beam to crossover; and you will see the bright spot. Ideally it would be in the

center of the crossover beam, but usually you have to adjust its position. Adjust gun tilt

slowly with the MULTIFUNCTION X/Y knobs until the bright spot is centered. When you

reach the edge of the screen, center spot using GUN SHIFT and MULTIFUNCTION X/Y

knobs. Repeat until the bright spot is perfectly centered. Exit Alignment page.



NOTE: Gun Alignment will change and need to be re-adjusted if the high tension setting

or the extraction voltage are changed.





6) ADJUST ROTATION CENTER

Function: Aligns the objective-lens rotation center to minimize image displacement as the

objective lens’ current is modulated.

Magnification ~50K (EXPOSURE MODE). Choose the same spot size as you plan to use

for taking pictures (usually 5-6).

NOTE: Intermediate magnifications in SA or LM range have different rotation centers. Also,

rotation center changes when you change stop size.



In SEARCH mode, find a recognizable feature on the sample and center it on the screen.

Bring the beam to crossover and check to make sure that you are close to focus.



Select ROT-CENTER on ALIGNMENT Selection page



Select FOCUS STEP 5 or 6 (this step size controls the amplitude of motion). Using the

MULTIFUNCTION X/Y knobs, minimize image movement so that it will move in and out (like

a cork in the sea), but not up/down or left/right.



7) ADJUST PIVOT POINTS

The alignment of the pivot points determines the relationship between two deflection coils

used. You want to make sure that the beam pivots around the correct point. Adjust in

EXPOSURE mode with the spot size the same as you plan to use for taking pictures

(usually 5-6).



Take the OBJECTIVE APERTURE OUT.



Press ALIGNMENT BUTTON (on the right of TEM control panel).



Press PIVOT POINT X.



The two circular beams need to completely overlap, to achieve this use the X/Y

MULTIFUNCTION knobs.



Repeat for PIVOT POINT Y. If significant adjustment was needed, re-check the rotation

center adjustment.



VERY IMPORTANT: put the OBJECTIVE APERTURE BACK IN (positioned to the left). This

is essential to provide contrast when imaging unstained specimens.



8) CENTER OBJECTIVE APERTURE

In SEARCH mode find an area with thick ice or some other diffracting material. Go to

EXPOSURE mode

CM 200 Instructions 8



Press D BUTTON (diffraction mode button) on right panel.



You should see a dim halo around the central bright spot. If necessary, adjust camera

length (MAGNIFICATION) or INTENSITY to make the halo more clear. Then, center the

halo around the small bright circle using the OBJECTIVE APERTURE centering knobs on

the aperture mount.



Press D BUTTON again to exit diffraction mode and spread the beam.



9) CORRECTING CONDENSER STIGMATOR (BEAM ASTIGMATISM)

Function: to make the beam circular.



NOTE: This procedure is usually done throughout the alignment when the beam becomes

non-circular, but it needs to be repeated with precision when the microscope has been

aligned. Also, note that stigmator settings vary from one mode to another and between

various spot sizes, so you need to correct it separately in each mode you use.



Press STIG BUTTON on the right-hand side of the control panel. This opens the

STIGMATOR control page.



Select COND (condenser lens). Make sure not to inadvertently change the settings for the

objective stigmators, which will be adjusted next.



Use MULTIFUNCTION X/Y knobs to correct for astigmatism until the beam is circular and

remains circular when it is being spread or condensed with the INTENSITY knob. After the

beam is circular exit the STIG page.



10) CORRECTING OBJECTIVE STIGMATOR (IMAGE ASTIGMATISM) WITH CCD CAMERA



Turn switch on the left of the microscope into the CCD camera position.



Turn computer monitor on and start the CCD camera program.



Before taking pictures with the camera, create a FLATFIELD to correct the CCD. Find an

empty square (no carbon), go to EXPOSURE mode, lift the screen, and unblank the beam.

On the camera dialog box menu click on Flatfield –> Create Flatfield –> Store Flatfield.



To take a picture: find an appropriate spot, focus, defocus (to desired underfocus value),

go to EXPOSURE MODE, lift screen, unblank beam, and press EXPOSURE BUTTON on

the camera dialog box menu. Exposure time is usually set to 500ms, but shorter times can

also be used (e.g., 300ms).

To check the power spectrum of an image click Power Spectrum on the camera dialog box

menu.



To correct objective stigmator: Find an area where the carbon film/ice look good. Focus,

and then underfocus about 1um. Go to EXPOSURE mode. Press the STIG button, and

select OBJ. On the camera dialog box menu select Continues Picture mode, displaying the

Power Spectrum. Adjust OBJ STIGMATOR with MULTIFUNCTIONAL X/Y knobs so that

Power Spectrum looks as circular as possible. Go to lower defocus (~200nm) to adjust

OBJECTIVE STIGMATOR with higher precision. In the end Power Spectrum should be

perfectly symmetric and circular.

CM 200 Instructions 9

11) FINAL CHECK OF THE LOW-DOSE MODE



Check the settings in LOW DOSE mode one last time to make sure that they are correct.

The main thing to check is the magnification in EXPOSURE more (60KX), the manual

exposure time (should match the meter reading for an area that has ice of the right

thickness), and the size and centering of the beam in exposure and focus modes.



TAKING PICTURES ON FILM



Areas where the ice thickness and quality are adequate should first be identified using the CCD

camera. It is a good idea to scan the entire grid in SEARCH mode, collect CCD images to

check for particles, and record in the compustage memory the positions (x,y, and eucentric

height) of squares where the particles look good. NOTE: To get adequate contrast in SEARCH

mode, the focus value in this mode should be set to about -60μm!!



IMPORTANT: Before switching to film, make sure the switch is in TM position (not CCD!!).

If you forget to set the switch properly, your pictures will be BLANK!



1) Find a good spot in SEARCH mode. Check eucentric height.

2) In EXPOSURE mode check that the beam is centered on the four corners of the film outline,

and that the exposure meter time is adequate (between 0.5 – 1 s) and within 20% of the

manual exposure time setting.

NOTE: Pay attention to whether you are overexposing or underexposing it will be important

when you develop film (see note on p.1)

3) FOCUS at focus step 4, 3 in S1 and S2 positions. Set focus to zero (press RESET

DEFOCUS)

4) UNDERFOCUS (‘—‘) turn counterclockwise to get to the needed value (usually between –

0.78 um – 1.92 um).

5) LIFT the fluorescent screen.

6) EXPOSURE BUTTON light comes on. Wait until everything has calmed down before

carefully pressing it. The panel light goes off, sit quietly, listen to film fall into the receiving box

and look for exposure light to come on. When panel lights come on put the SCREEN down.

7) Repeat.



REMEMBER TO CHECK THE FOLLOWING

Eucentric height of the specimen

At least every time when changing squares on the specimen grid

Exposure time and exposure beam position

Often during first 15 minutes, then approximately every 10-15 pictures

Nitrogen level

In cryostage and dewar (approximately every 30 minutes)



SHUTING DOWN MICROSCOPE TO CHANGE FILM



1) CLOSE gun valve and specimen shield, and set FEG to STANDBY mode



2) TURN OFF high tension



3) EXCHANGE film cassette

Go to adjacent room and open nitrogen gas valve.

Go to the Vacuum page. Press CAM AIR BUTTON, when lid opens take box out and place

new one in from film desiccator. Press CAM AIR BUTTON again so that vacuum pumps will

turn on. Close nitrogen gas valve in the adjacent room.

DO NOT walk away from the microscope until VACUUM STATUS goes back to READY, which

CM 200 Instructions 10

should happen in about 5 minutes.



4) Reset camera film counter

Press TEM CAMERA, then CAM INTT, then RESET. Hit RESET, it will now show 55

pictures (the stock of a new cassette).





SHUTING DOWN MICROSCOPE AT THE END OF THE SESSION



Follow steps 1-4 above as required, then:



5) EMPTY cold trap dewar

By pouring liquid nitrogen on the floor and let it dry, place paper towel in place of the dewar.



5) REMOVE cryostage

Pull sample holder out as far as it will go, then rotate it clockwise to 5 o'clock position (keep

Styrofoam container underneath because liquid nitrogen will pour out), continue to carefully

extract the holder from the airlock.

Cover holder tip with protective tube and connect cryostage to the heater. Set heater

selector to DEWAR (yellow light should come on), and let the cryostage warm up unitl the

temperature reads ~ 60C. Take the cryostage to the pumping station, and open vacuum for

the tip only to dry it out.



7) WRITE final P3 value, etc. into logbook

CM 200 Instructions 11

FILM DEVELOPING (in the dark room)

1) Sign logbook, check that there have been no more than 250 images developed including

yours, otherwise change DEVELOPER (see instructions below).

Check FIXER with a few drops of Hypofix solution. If white cloudy trace does not disappear the

fixer is not good and you need to change it (see instructions below).



2) In the dark (only red lights are allowed) put film into the special containers.



3) Put racks with film into DEVELOPER for 9-12 minutes. Actual time depends on whether your

images were under or over exposed. For underexposed images use 12 minutes, for

overexposed use 9 minutes.



4) While waiting for developer put new film into the cassette, when putting the stack remember

that film side should be up, also remember to check the position of the cut corner (it should be

top right).



5) Wash in WATER for 30 sec to 1 minute.



6) Put containers into FIXER for 4-5 minutes.



7) Put in WATER for 20-30 minutes.



8) FOTOFLOW rinse, shake off excess.



9) Push film up a bit on the rack and put in drying oven for 60 minutes.



10) Put cassette (next cassette and boxes with negatives) into desiccator and pump for 30

minutes.





CHANGING DEVELOPER

We use Developer D19 (full strength). Use one pack per 3.8 liters (one bath). Mix and stir for 10

minutes.





CHANGING FIXER.

New fixer is located in the cabinets across the tab. Read instructions on the bottle. One bottle

should be good for one bath (3.8 liters).



Developer and Fixer WASTE are disposed of in the same large container.

CM 200 Instructions 12

TROUBLESHOOTING

Problem

The extraction limit is set too high/low. How do I get to the right value?

Solution

Click so that the extraction limit is not lit; rotate the filament knob clockwise until you reach

desired value of extraction limit. Click the extraction limit again it will become lit and now it will

be the same value as you dialed. (If you want to decrease the value you dialed rotate the

filament knob counter clockwise)



Problem

Image moves.

Solution

Tap lightly the sample holder and/or touch joystick X/Y slightly.



Problem

Message is flashing on the screen.

Solution

Erase by pressing RESET button.



Problem

Focus seems to be way off. Auto-focus does not work. How do I focus?

Solution

First check the eucentric height. Once it is set properly, AUTO FOCUS should work. If it

doesn’t, go to the lower MAG. Find some dirt on the specimen and look at the Fresnel fringes,

when they start to disappear it means that you are close to focus. Find the position of minimal

Fresnel fringes, this should be close to focus. Go back to high magnification and focus as

usual.



Problem

Increasing or decreasing the magnification produces an appreciable movement of the beam or

the image.

Solution

Column Alignment is required.



Problem

A shift of the beam is accompanied by a tilt, or vice versa.

Solution

Do Pivot Point Alignment.



Problem

Joystick doesn’t move right.

Solution

Check that JOYSTICK POWER is on 2.00 and SEPARATELY should be off for simultaneous

XY movement.



Related docs
Other docs by Kerala g
union-budget-2012-13-highlights
Views: 102  |  Downloads: 0
notification M.Tech_05-03-09
Views: 59  |  Downloads: 0
India_Customs Regulation 1
Views: 56  |  Downloads: 0
CE Notification 39-2011-12.9.2011
Views: 54  |  Downloads: 0
STATISTICS
Views: 72  |  Downloads: 0
A Hero (R.K. Narayan)
Views: 91  |  Downloads: 6
RRBPatna-Info-HN
Views: 116  |  Downloads: 0
RRB-Notice-Para
Views: 113  |  Downloads: 0
By registering with docstoc.com you agree to our
privacy policy

You are almost ready to download!

You are almost ready to download!