Embed
Email

Blind Duplicate Patient Testing on the HmX

Document Sample

Shared by: linzhengnd
Categories
Tags
Stats
views:
0
posted:
12/12/2011
language:
pages:
4
Lab Name Department of Pathology

Address Automated Hematology







Blind Duplicate Patient Testing on the HmX



Author: Document Number: Equ30-42

Heidi Hanes Effective (or Post) 15-June-2010

Date:

Review History Date of last review: NA

Reviewed by: Heidi

SMILE Comments: This document is provided as an example only. It must be

revised to accurately reflect your lab’s specific processes and/or specific protocol

requirements. Users are directed to countercheck facts when considering their

use in other applications. If you have any questions contact SMILE.



Document: HEA.2620 Effective: _____________

Version 1.0



PURPOSE

This procedure provides instruction for using patient specimens as an additional

quality control process for both the primary and secondary modes on the GenS

and HmX Hematology Analyzers.



POLICY

They are:

 The new high and low patient specimen is run through both the primary and

secondary mode of the HmX. This acts both as a QC for the secondary

mode and a correlation of the two modes.

 Both specimens are run through the primary mode only at designated times

throughout the shifts.

 Samples are run the following morning until new patient controls are chosen

from the morning run.

 Refrigerate samples when not in use.



NOTE: The same samples will be used on both the GenS and the HmX

instruments each day.





SAMPLES

Use the following table to select patient samples for use as blind duplicates:

Sample Type WBC HGB

3 3

High 5.0-10.0 x 10 /mm 12.0-15.0 mg/dL

Low 10.0-15.0 x103 /mm3 7.0-10.0 mg/dL



PROCEDURE

Step Action

1 All monitored parameters of blind duplicates should agree within + 2 SD



Blind Duplicate Patient Testing on the HmX Page 1 of 4

HEA.2620 Version 1.0

Lab Name Department of Pathology

Address Automated Hematology







with back up method.



If Then

 Any parameter  Rerun

exceeds limits

 Still out  Run on alternative method to determine if an

instrument problem or sample problem

If Then

 Same results on  Sample problem

both instruments  Mark “Sample Deterioration”

Attach to Hct result.

 Different results  Instrument problem

 Run on alternate instrument until

problem resolved

 Call Coulter Service.

 Consistent problem  Inform QC tech.

with Blind

Duplicates





RESULTS

A. The parameters below are monitored on the HmX to ensure proper instrument

function and correlation between instruments and modes. Each parameter

should agree within the following ranges:



High Low

WBC ± 0.6 ± 1.4

RBC ± 0.20 ± 0.20

HGB ± 0.4 ± 0.4

HCT ± 2.0 ± 2.0

MCV ± 3.0 ± 3.0

PLT ± 60.0 ± 60.0





B. Control decisions are as follows:



If Then

 All parameters on the high and low  Proceed with patient testing

samples read within ± 2 SD when

compared to the results of each

original run



OR



 One parameter of either the high or

low reads with ± 2-3 SD when

compared to the results of the

original run and all others are within



Blind Duplicate Patient Testing on the HmX Page 2 of 4

HEA.2620 Version 1.0

Lab Name Department of Pathology

Address Automated Hematology







+ 2 SD.

 Any parameter on either the high or  Do not proceed with patient testing

low sample reads greater than + 3  Repeat sample.

SD If Then

 Acceptable  Proceed with patient

testing

 Still out  Use back up

methods

 Results still the  Sample deterioration,

same on proceed with patient

backup testing

 Results  Indicates instrument

acceptable on problem, use back

backup up instrument

 Inform QC tech or

supervisor

 A single parameter on both high and  Repeat test on each sample:

low is greater than + 2 SD

If Then

 Acceptable  Proceed with patient

testing

 Still out  Use back up

methods

 Results still the  Sample deterioration,

same on proceed with patient

backup testing

 Results  Indicates instrument

acceptable on problem, use back

backup up instrument

 Inform QC tech or

supervisor.



C. Run 5C controls.



If Then

 Similar problems with 5C as with  Call Coulter Hotline

patient controls  Use back up instrument

 5C controls acceptable  Run 10-12 patients on back –up

instrument and compare results.

If Then

 Parameters  Proceed with patient

match testing

 Monitor moving

averages

 Moving  Call coulter Hotline

averages show  Use backup

same problem instrument

 Inform QC Tech and

Supervisor of any

unresolved or



Blind Duplicate Patient Testing on the HmX Page 3 of 4

HEA.2620 Version 1.0

Lab Name Department of Pathology

Address Automated Hematology







recurring problems.



D. Handling insufficient or deteriorated samples:



If Then

 Insufficient sample remaining  Go to function MEM

 Go to Test

 Use test modes to ensure correct

method (workload) assignment

 Enter CBC

 Enter blind dup acc. number, B-

Acc No.

 At prompt, enter 0.0-QNS for all

parameters and accept.

 Sample deteriorated  Attach onto Hematocrit, “Sample

Deteriorated”

 Both high and low samples become  Select 2 new samples from

deteriorated before 24 hours are previous run

completed  Follow protocol for selection of new

blind patient duplicates

 Follow Procedure A-C above.





REFERENCE MMC Laboratory QC Policy







Approval Signature:__________________________ Date: ______________









Blind Duplicate Patient Testing on the HmX Page 4 of 4

HEA.2620 Version 1.0



Related docs
Other docs by linzhengnd
Just Wright
Views: 0  |  Downloads: 0
Property and Insurance Services _PAIS_
Views: 0  |  Downloads: 0
Module 8 Unit3
Views: 0  |  Downloads: 0
Navy Response to EPA Comments on the
Views: 0  |  Downloads: 0
EALING LOCAL BUSINESS PARTNERSHIP
Views: 3  |  Downloads: 0
Hotel Sankara Nairobi
Views: 0  |  Downloads: 0
SERGIO NELSON MANNHEIMER
Views: 0  |  Downloads: 0
C-3 Rides 2010 v3a
Views: 0  |  Downloads: 0
Quality of life_ quality for life.
Views: 0  |  Downloads: 0
Intermodal freight transport-wikipedia
Views: 1  |  Downloads: 0
By registering with docstoc.com you agree to our
privacy policy

You are almost ready to download!

You are almost ready to download!