Docstoc

ENA-B service stuff

Document Sample
ENA-B service stuff Powered By Docstoc
					(copied from ENA service web site Jan. 29, 2008)

Filename: ENA-B service stuff.doc


                                                                                Last Update : 07/08/2007 21:10:04



Agilent E5070B/E5071B
RF Network Analyzer
System Recovery
The system recovery disk (3.5 inch floppy disk) is furnished with the instrument. If you do not have it, you
can make it using the following procedure.

    1.   Copy the following files to the specific directory (e.g. C:\temp) on your PC.
            o File Download : enarcvry.bin (1.44MB, Nov.'05) copydisk.exe (21KB, Nov.'05)

    2.   Insert a floppy disk into the FDD of your PC.
    3.   Run the DOS prompt, and change the current directory to the specific directory which includes the
         above two files.
    4.   Perform the following command. Then, the system recovery disk will be made automatically.

copydisk enarcvry.bin a:
NOTE : Try to perform the System Recovery to recover the operating system before replacing the A27
HDD. For details, see Chapter 12 "System Recovery" on the E5070B/71B User/s Guide or Appendix B
"System Recovery" on the E5070B/71B Service Manual.

System Update using Symantec Ghost <Agilent Service Center Use Only>
        Procedure : See service note of E5070B-10B or E5071B-10B
        OS Image : Kobe sends latest version of DVD disk to responsible service centers, when the major
         revision is changed. (for example, 5.x.x to 6.x.x)
        Software : Symantec Ghost ™ 8.0 or later (English version, Enterprise Product) is needed. In case
         you need to install, send a message to KID,SVC-SUPPORT (A-Japan,exgen1).
        Specific execution file for the ENA without Video Card (PANEL.exe) : Available (exe, 1KB,
         Mar.'03)

Performance Test Program <Agilent Service Center Use Only>
        Agilent Part Number : N/A
        Revision Number : A.02.40
        Release Date: 7 Dec., 2006
        OS : Windows 2000 / XP
        Program Language : VEE 6.2 or VEE 7.xx
        Program Download : Available (exe, 364KB)
        Operation Manual : Available (pdf, 948KB, Oct.'02)
      NOTE: It is necessary to use the Edit Cal

Adjustment Program <Agilent Service Center Use Only>
      Agilent Part Number : N/A
      Revision Number : A.02.60
      Release Date: 9 Jul., 2007
      OS : Windows 2000 / XP
      Program Language : VEE 6.2 or VEE 7.xx
      Program Download : Available (exe, 374KB)
      Operation Manual : Available (pdf, 702KB, Oct.'02)
      NOTE: It is necessary to use the Edit Cal.

				