Embed
Email

SOP_New_Brunswick_Scientific_Bioreactor

Document Sample

Shared by: qingyunliuliu
Categories
Tags
Stats
views:
5
posted:
12/5/2011
language:
English
pages:
7
Modular Benchtop Fermentor Operation

New Brunswick Scientific Company, BioFlo 110,

by Edward M. Trujillo

9/4/2003

Revised 10/14/09







This is an abbreviated operating procedure for the bioreactor in our lab. Refer to the BioFlo 110

manual for more specific information.



1. Open the main cold water valves and air valves show in Figure 1. Be sure the drain to the

bioreactor jacket leads into the sink.









Figure 1: Air and Water Valves for the Bioreactor.



2. Turn on PC computer on the west side of the Biolab and sign in (user name: biolab, and

password: biolab). The Opto program is only usable if you log in, on the local machine,

using the biolab account user name: BIO_LAB COMPUTE.

3. Assemble the bioreactor, if it is not already assembled for the type of experiment you

plan to do. Make sure both the dissolved oxygen electrode and pH electrode are

immersed in a distilled water solution or the

aqueous solution you plan to use for fermentation.

I suggest you fill the reactor vessel up to the 1.5

liter mark. Fill the vessel using an open port in top

of the reactor and a funnel and empty the vessel

using the peristaltic pump provided.

4. Switch on the main power to the bioreactor located

on the Power Controller module (see Figure 2).

This turns on everything else (unless someone has

changed the switches on the different units),

including the primary control unit located on top of

the power control unit (see Figure 2).

5. The orange display on the primary control unit

should default to the setup screen on startup (Figure

3). If not, push the Startup Key. The bioreactor can

be setup to run in either a fermentation mode or a

cell culture mode. Make sure the unit you are using Figure 2: Bioreactor Modules.

(the control unit can control up to 4 reactors,

we only have one – Unit 1) is setup for

fermentation mode by pressing the USE key

using the Select Buttons on the setup screen.

Then set the unit to fermentation mode by

pressing the Select Button again on the next

screen. Also, you have the option to select

the language used for the controller. I

suggest you use English.

6. The Main Menu (showing the status of the

bioreactor) is obtained by pushing the Main

Menu Key depicted in Figure 3.

7. Note that on startup (default) the O2 loop on Figure 3: Bioreactor Primary Control Unit

the Main Menu is set to Manual mode, while Control Panel.

all the other loops are set to OFF. With our

setup this activates the gas coming from the gas mixer. Set the O2 loop to OFF to turn

off the gas coming from the gas mixer. You can set this back to MANUAL when you are

ready to use the gas mixer and have connected the line from the mixer to the rotameter

(see #10 below).

8. Fill the water jacket with tap water using the temperature control setting (Main Menu) on

the primary control unit by

(a) Setting the temperature control to AUTO

(b) Entering a temperature setpoint that is at least 5 C below the current bioreactor

temperature. The controller will respond to the call for cooling by opening the solenoid

valve, filling the water jacket. The water overflow will go into an overflow tube located

in the water jacket (in the back) and drain to the sink. Flush with at least 2 jacket volumes

of water. You can stop the water from flowing to the drain by setting the setpoint to the

desired bioreactor temperature (typically above the current temperature).

9. Test the Agitator Control by setting the agitator control to AUTO and setting a setpoint

between 0 and 1200 rpm. This is how you control the agitation speed if you are not using

the Gas Mix Controller or the dO2 controller.

10. Gas flow is controlled for fermentation two ways:

a. Manifold Control (Figure 1e) – make sure the connection from the outlet of the

manifold is connected to the inlet of the rotameter (using the quick disconnects).

Each gas is controlled by a separate needle valve located on the manifold. The

flow rate to the bioreactor is controlled by the rotameter located downstream from

the manifold. Typically, the needle valves will be in an ON/OFF situation, adjust

them only to give the maximum flow rate of 1200 CCM (cubic centimeters per

minute) on the rotameter. Use the rotameter to adjust the total volumetric flow to

the bioreactor. Make sure the pressure regulators on the cylinders that you are

using are set to approximately 10 psig.

b. Gas Mix Control – when using this control, the dO2 setting, make sure the output

connection from the gas mixer is connected to the inlet of the rotameter (using the

quick disconnects) and the valves on the individual gas cylinders are open (open

only those you will use). Make sure the pressure regulators on the cylinders are

set to approximately 10 psig.

11. You have four control choices to choose from in this mode: Agitation, Oxygen,

Agitation/Oxygen or None. Each of those is described in the manual, however the

Agitation/Oxygen control is the one used most frequently, since it controls the dissolved

oxygen level by both agitation and the mix of air and oxygen. However, if you want to

keep a constant agitation rate throughout your fermentation, you may want to use the

Oxygen control mode. In the dO2 details screen, press the setup button. From here you

can set the high and low limits for agitation and the gains for process control. Change the

P-Gain and I-Gain ONLY if you understand what you are doing. Now press the Cascade

selector button. From here you can choose one of the four options. Usually, you will

choose the Agitation/Oxygen option but the Oxygen option is also possible.

12. Dissolved Oxygen Probe Calibration – The dO2 sensor is a polarographic probe

manufactured by Ingold (Mettler-Toledo, InPro 6000 series O2 sensor). If the sensor has

been disconnected from the transmitter for longer than 5 minutes it has to be polarized

prior to calibration by connecting it to the transmitter or a polarization module for 6

hours. You should calibrate the dissolved oxygen probe before using it for your

fermentation. Normally, a single point calibration is sufficient. This is done by placing

the probe into a medium of known dissolved oxygen concentration, typically distilled

water at a constant temperature and pressure that is in equilibrium with a sparging gas

(typically air). Note the probe only measures % saturation. The actual dissolved oxygen

concentration has to be obtained from reference tables. Keep in mind that most reference

tables for dissolved oxygen assume atmospheric pressure at sea level. A correction must

be made to those values for pressure at our altitude.

13. To calibrate the dO2 probe choose the dO2 screen from the Main Menu. From here you

can choose the setpoint, choose the upper and lower limits for the display screen (Disp Hi

– Disp Lo), calibrate the probe (Calibrate) and set the process control parameters

(SETUP) such as the deadband, output multiplier, and the P and I gain. Set the control to

AUTO. Next, push the CALIBRATE button. Normally you do not have to set the zero,

but if you want to (dual point calibration), the easiest method is to momentarily

disconnect the O2 probe from the transmitter (no more than 5 minutes) and once the

reading is stable at zero, push the set zero selector button and enter zero. Reconnect the

O2 probe to the transmitter. The Set Span selector is used for the 100% calibration.

Once the probe reaches a steady value, indicating equilibrium with the medium and the

temperature is constant, press the Set Span button and enter 100, indicating the probe

(and medium) is 100% saturated with the sparging gas. Keep in mind that it usually takes

a long time for a solution to get completely 100% saturated.

14. pH Control – Choosing the pH selector button from the Main Menu, brings up the pH

control screen. From here you can choose the setpoint, choose the upper and lower limits

for the display screen (Disp Hi – Disp Lo), calibrate the probe (Calibrate) and set the

process control parameters (SETUP) such as the deadband, output multiplier, and the P

and I gain. Typically, you will use the default parameters for SETUP. You should,

however, calibrate your pH probe before you use it. This is explained in the bioreactor

manual, starting on page 63 (actually page 65 starts on calibration). Note that you have to

physically remove the probe from the bioreactor in order to do the calibration. Be sure to

rinse the probe with distilled water before you place it back into the bioreactor.

pH is controlled by two concentrated solutions of acid and base. Typically, you will use

a sulfuric acid solution and a sodium hydroxide solution, approximately 10% by weight.

DO NOT USE HYDROCLORIC ACID IN THE BIOREACTOR. HCl corrodes

stainless steel. The acid and base solutions are placed in their respective sterile bottles

and connected to the appropriate pumps using the existing plastic tubing. Make sure all

the lines to and from the pumps are filled before connecting the lines to the bioreactor.

Use the triport on top of the reactor for the acid, base and antifoam lines.

15. Once everything is setup, you can choose the setpoint for the pH and the controller will

control the pH in the bioreactor at that setpoint by turning on and off the acid and base

pumps. I suggest this is the last controller you setup.

16. Antifoam Control – the bioreactor is setup for antifoam control. Pumping a concentrated

antifoam solution into the vessel using one of the peristaltic pumps does this. There is a

foam sensor mounted in the headplate of the vessel that senses the presence of foam and

responds by turning on the antifoam pump. Make sure there is ample antifoam in the

addition bottle and that the tubing going to and coming from the pump is completely

filled and connected to the triport on the headplate.

17. Go back to the Main Menu and make sure the readings are reasonable for the variables

you selected. You are now ready to monitor the readings using the computer software.

COMPUTER SOFTWARE AND DATA ACQUISITION



The New Brunswick Scientific software, BIOCOMMAND LITE, is used for data acquisition

and for on-line plots of the various parameters. See the manual for complete details of the

software.



1. Open BCLite from the Desktop.

2. In the first window select “Open an Existing Recipe”.

3. Select File name “Recipe Bio Bob.rcp”.

4. Go back to desktop

5. Open the shortcut to OptoDisR.

a. You should see the dO2, Agitation, Temperature, and pH logging.

6. Go back to BioCommand Lite.

7. Click “Start Batch”.

8. Select no, “do not save” changes to recipe.

9. Enter a novel Batch Display Name and then a Destination File Name.

10. After you click OK for the Destination File Name data will being recording.

11. To monitor/change set points go to Views, and double click “Batch Summary.1”. To

change a set point, double click on the line for that controlled (e.g. “Agit”)

12. To monitor trends and history go to Views. Double click on “Trend” under “New

Views” and click “Loops”. Then click “Set up Loop” and select the loop you wish to

monitor (e.g. dO2, pH, or so on). Use the arrow keys to add additional loops to the plot.

13. To add an alarm, go to the Alarm Tab, and choose the controller and set alarm bounds.

14. When finished gathering data, click “End Batch” on the BioCommand software, and click

“Exit Opto” in the OptoDisR software.

15. You can open OptoDisplay results in Excel.

a. Open Excel.

b. Go to file => open.

c. My Computer => C: Drive.

d. Open “Opto Data” folder.

e. Change Files of Type to “All Files”.

f. The logs are saved by date, so find the correct file.

g. When the text import wizard opens, click Next.

h. In Step 2, put a check in comma and hit next.

i. Now select “Finish”.

j. When it is open, columns G & H do not pertain to our reactor.

16. You can later read these files using separate software by New Brunswick, DBViewer32,

and then transfer the file to MS Excel.

a. Open the DBViewer32 program.

b. Select “Open” and find the file name you gave to your experiential run in Step 9.

MEASURING GLUCOSE CONCENTRATIONS



There are several ways to measure glucose concentrations during the fermentation. Two that

are available in our lab are the following:



(1) Sigma Chemical Company’s GLUCOSE TRINDER, [Procedure No. 315]. This kit is

used primary for the quantitative, enzymatic determination of glucose in serum or plasma

using a spectrophotometer at 505 nm and is linear up to 750 mg/dL. See the procedure

written by Sigma for detailed information.

(2) Molecular Probes’ AMPLEX RED GLUCOSE/GLUCOSE OXIDASE ASSAY KIT (A-

22189). This kit uses a fluorescence microplate reader and is also an enzymatic

determination, but uses excitation I the range of 530-560 nm and fluorescence emission

detection at approximately 590 nm. The reagents in this kit are very sensitive to light and

somewhat air sensitive. Thus, they must be use promptly. See the procedure and product

information sheets from the company for detailed information. This kit is used primarily



(3) HPLC. See the Projects Lab HPLC Page.







MEASURING YEAST CONCENTRATIONS



There are two ways to measure yeast concentrations during fermentation:



(1) Turbidimetric (Optical Density) Method – This is a very simple method that is based

strictly on how turbid the solution is, assuming that the yeast cells are the only

contribution to the turbidity of the solution. Turbidity is measured by absorbance in a

spectrophotometer set at a suggested wavelength of 660 nm. You must develop a

calibration curve (linear works best) for the yeast then compare the unknown sample with

this curve. You may use the Projects Lab’s UV-Vis or Plate Reader.

(2) ATP Assay Kit by BioTrace International, - This kit is based on the measurement of total

adenosine triphosphate as an indicator of the amount of yeast cells. Again, an enzymatic

reaction of d-Luciferin with ATP in the presence of Luciferase is used which produces

light and the light intensity is measured with a luminometer. There are several

information sheets on this method with procedures. Consult those documents.

(3) The Chemometec Cell Counter. See the Projects Lab Cell Counter Page.

SHUTDOWN AND CLEANUP PROCEDURES



1. Turn off the agitator, gas flow, and temperature control.

2. Empty the vessel using the peristaltic pump and then flush/agitate the reactor at least

twice with hot distilled water. Use the agitator to clean the vessel during each flush, but

make sure to turn off the agitator during filling and emptying. You may have to use a

dilute acid solution (sulfuric acid) or a mild detergent solution to get the vessel

completely clean.

3. MAKE SURE THE VESSEL IS FILLED WITH DISTILLED WATER (1.5 LITERS)

AT THE END OF FLUSHING AND CLEANING. The dO2 probe and the pH probe

need to be immersed in water when left overnight. For long-term storage, see the

bioreactor manual.

4. Turn off all the controls on the primary display unit

5. Turn off the main power switch.

6. Turn off the main water valves and the main air valves show, as well as all the valves on

the gas cylinders.

7. Sign off and shut down the computer.



Other docs by qingyunliuliu
Work Resume - Grauwald Creative
Views: 11  |  Downloads: 0
Gulf_Impacts_v2
Views: 0  |  Downloads: 0
1615_RPI_091208
Views: 0  |  Downloads: 0
Gudelines_for_applicants
Views: 0  |  Downloads: 0
Suzuki Violin Lessons
Views: 14  |  Downloads: 0
court_2005_028_fr
Views: 0  |  Downloads: 0
082607homily
Views: 0  |  Downloads: 0
Role of the Chartered Man Acct_i_
Views: 8  |  Downloads: 0
By registering with docstoc.com you agree to our
privacy policy

You are almost ready to download!

You are almost ready to download!