        ATTENTION SHIPPER:


              PLEASE READ!!
              PLEASE READ!!
Important Unpacking Information
           Included.
                      Technical Bulletin TB-08-005-06
Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   1
                                   DRAFT
        Immucor Galileo Echo
If you have questions about the information contained in this
     document, please contact Immucor Technical Support
                     at 800.492.2583.

 If you think you may have run into an issue that would
        prevent you from completing the delivery....
          DO NOT LEAVE THE DELIVERY SITE
           call Matt Gaines at TECH TRANS at
              1-800-952-3411-IMMEDIATELY




                         Technical Bulletin TB-08-005-06
   Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                      DRAFT                                         2
 Immucor Galileo Echo

Tech Trans System Installation



                      Technical Bulletin TB-08-005-06
Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                   DRAFT                                         3
           Immucor Galileo Echo
Tools required
• Flathead screwdriver, 5/32”
  Allen wrench, and a box cutter

Or

For Instrument Serial numbers
   M00252-M00291 and M00309
   and above.
• #2 Phillips screwdriver and a
   box cutter



                              Technical Bulletin TB-08-005-06
        Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   4
                                           DRAFT
              Immucor Galileo Echo
There are two (2) cartons that are labeled
   Main Unit and Fluidics
•   Tech Trans is responsible for
    unpacking the “Main Unit” and
    “Fluidics” cartons only.
•   There are additional cartons for the
    computer, monitor, printer, UPS, and
    accessories. These cartons will be
    unpacked by an Immucor
    Representative. These boxes
    should NOT be unpacked by
    Tech Trans personnel.




                              Technical Bulletin TB-08-005-06
        Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   5
                                           DRAFT
           Immucor Galileo Echo
Unpacking the Main
 Unit carton
• Open the large box and
  remove the four foam
  pads.
• Open the inner small box.




                           Technical Bulletin TB-08-005-06
     Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                        DRAFT                                         6
          Immucor Galileo Echo
Unpacking the Main
 Unit carton
• Remove the cardboard
  insert and the plastic bag




                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   7
                                         DRAFT
          Immucor Galileo Echo
Unpacking the Main
 Unit carton
• Grasp the instrument by
  the handles near the rear
  of the instrument
• Grasp the back panel of
  the instrument for stability
  while lifting
• Lift the instrument out of
  the box

                              Technical Bulletin TB-08-005-06                            8
        Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                           DRAFT
         Immucor Galileo Echo
Unpacking the Main
 Unit carton
• Place the instrument on
  the counter with the
  bottom plywood exposed
• Using the screwdriver,
  unscrew the seven
  screws and remove the
  plywood from the bottom
  of the instrument


                           Technical Bulletin TB-08-005-06
     Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   9
                                        DRAFT
           Immucor Galileo Echo
Unpacking the Main
 Unit carton
• Place the instrument
  upright on the right side
  of the counter




                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   10
                                         DRAFT
          Immucor Galileo Echo
Unpacking the Main
 Unit carton
• Place the plywood, plastic
  bag and cardboard insert
  back into the box for
  removal
• Place the foam padding
  on the four corners of the
  inner box for removal


                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   11
                                         DRAFT
           Immucor Galileo Echo
Unpacking the Fluidics
 carton
• The Fluidics carton consists of
  the instrument cover for the
  Main unit and the Fluidics
  module
   – Open the large box and
      remove the four foam pads.
   – Open the inner box and
      remove the cardboard
      divider. The instrument
      cover is located on top.

                             Technical Bulletin TB-08-005-06
       Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   12
                                          DRAFT
         Immucor Galileo Echo
Unpacking the Fluidics
 carton
• Remove the instrument
  cover from the box and
  remove the bubble wrap
  from the cover




                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   13
                                         DRAFT
          Immucor Galileo Echo
Unpacking the Fluidics
 carton
• Place the instrument
  cover on the Main unit by
  sliding it onto the front of
  the instrument.




                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   14
                                         DRAFT
           Immucor Galileo Echo
Unpacking the Fluidics
 carton
• Remove the cardboard
  divider. The Fluidics
  Module is in the bottom of
  the carton.
• Lift the cardboard insert
  out of the box.



                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                         DRAFT
           Immucor Galileo Echo
Unpacking the Fluidics
 carton
• Lift the Fluidics module
  out of the box.
• Place on the counter with
  the bottom plywood
  exposed.




                           Technical Bulletin TB-08-005-06
     Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                        DRAFT                                         16
            Immucor Galileo Echo
Unpacking the Fluidics
 carton
• Using the 5/32” Allen wrench
  or the Phillips screwdriver for
  fluidics modules F00252-
  F00291 and F00309 and
  above, remove the five screws
  and remove the plywood
• Place the Fluidics module
  upright to the left of the main
  instrument


                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   17
                                         DRAFT
            Immucor Galileo Echo
Unpacking the Fluidics
 carton
• Return the plywood, cardboard
  insert, cardboard dividers and
  bubble wrap inside the box for
  removal
• Replace the foam padding
  onto the corners of the inner
  box for removal




                          Technical Bulletin TB-08-005-06
    Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA   18
                                       DRAFT
          Immucor Galileo Echo
• Remove the empty Main unit and Fluidics cartons from
  the facility
• The G3-Access and all additional cartons will be
  unpacked by Immucor
• If requested by the customer to unpack all additional
  accessory boxes, call Matt Gaines at TECH TRANS
  at 1-800-952-3411-IMMEDIATELY



                            Technical Bulletin TB-08-005-06
      Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                         DRAFT
                                                                                       19
            Immucor Galileo Echo
Minimum space requirements for instrument placement:
• Width – 43 1/8 inches
• Height – 26 inches
• Depth – 22 1/2 inches
• 12 inches of permanent clearance in back of instrument for service
   and maintenance access
• Verify level surface for instrument placement. Instrument must be
   placed on a stationary bench, not on an equipment cart.
• Additional space required for printer: 14.6" W x 9.6" D x 9.5” H




                             Technical Bulletin TB-08-005-06
       Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA
                                          DRAFT
                                                                                        20

				