					NIRSpec Test Specification – IMA-PSF


Test ID: IMA-PSF

Author: T. Böker

Reviewers: P. Ferruit (29/09/2010), TBD

Purpose:
This test obtains "clean" and well-sampled PSF images at a large
number of focus positions. We will use the RCSS point source
observed through the S1600 slit aperture.

Calibration sequences covered:

None.

Description:

The following OGSE/NIRSpec configuration parameters do not change
throughout the test:

CMO_abs: "RCSS"
CLS:      "CLOSE"
RCSS:     "MOS1_N"
CAA:      "OFF"
FWA:      "F110W"
GWA:      "MIRROR"
-----------------------------------------------------------------

The test will obtain exposures of a 32x32 pixel detector sub-array
at the position the S1600 slit image. The RCSS will be positioned to
illuminate the center of the slit, and the MOS1_N filament will be
turned on. A suitable set of exposures will be obtained at each
position of a 5-point dither pattern (+/- one third of a detector
pixel in x and y), in order to improve the spatial sampling. Because
of the high flux of the RCSS, the detectors have to be operated in
window mode.

This sequence will be repeated for a full focus sweep as described
in IMA-FOCUS.

Comments:

None.
Specific dependencies for this procedure:

Pre-requisites                              Status Comments

None.




Step-by-step procedure:

     Test conductor:                              Date and time:



#    Steps                                        Status Comments

01   Set the DHAS ingestion profile:                     Write down
        The REP path should be set to IMA-              the REP path
          PSF-xx where xx is a two-digit                 name.
          running number starting at 01.
        Check that the SRC file pattern
          includes files with the extension
          .csv and .msa

02   Start the ingestion (Auto Ingest mode).             Write down
                                                         date and
                                                         time.

03   Run script TBD with the following input
     CSV file:

     IMA-PSF.csv

04   Check that all files have been ingested by          Write down
     the DHAS and stop the ingestion.                    date and
                                                         time.
Data Acceptance Criteria (before moving to another test sequence):

ID           Criteria                            Status Comments

SUCCESS-01   Procedure    must    have    been
             successfully executed and all the
             exposures successfully obtained.

SUCCESS-02   The headers of the exposures
             shall be consistent with the
             contents of the input CSV files.

SUCCESS-03   Each raw exposure FITS file must
             have   been   visually inspected
             using the IQLAC and no obvious
             anomaly detected.


The signal-to-noise ratio in each exposure must be at least 20
(TBC), i.e. the count level in the peak pixels of the PSF must be
~20 times (TBC) the RMS variation of the “dark” areas of the
exposure.


Expected test duration: 2h45
Appendix A – Inputs to calibration readiness review checklist:

ID                Checklist items                   Comments

CHK-IMA-PSF-01    The CSV file IMA-PSF.csv:
                      Shall have been updated to
                       take into account the
                       exposure number of groups
                       derived during the
                       preparation period.

CHK-IMA-PSF-02    The CSV file IMA-PSF.csv:
                      Shall have been updated to
                       make sure that, for the
                       MOS-mode exposures, the
                       CMO_X and CMO_Y positions
                       do position the RCSS
                       source at the center of
                       A1600.

CHK-IMA-PSF-03    The CSV file IMA-PSF.csv:
                      Shall have been updated to
                       make sure that the readout
                       window corners are
                       consistent with the
                       position of the RCSS
                       source on the detector.

CHK-IMA-PSF-04    Once updated, the CSV file IMA-   Write down
                  PSF.csv:                          baseline CSV
                      Shall have been verified     file repository
                       and put under                version.
                       configuration control.
                      Shall be available in the
                       baseline CSV file
                       repository.
Appendix B – Data analysis products:

ID                Products                       Comments

ANA-IMA-PSF-01    A library of PSF for a wide
                  range of RMA positions (i.e.
                  ranging from in-focus to
                  completely out-of-focus).

				