					CT QUICK REFERENCE - 15.Aug.11
                             BRIGHTSPEED
                                SELECT
                          EDGE, EXCEL, ELITE
                                   &
                            ELITE ENANCED
gantry/table flr.
anchored?                          YES
gantry cable
access                            REAR
conduit sizes
-OC to Gantry                      3 1/2"
-OC to PDU            -            2 1/2"
PDU to Gantry                2 1/2" and 3 1/2"

                          39" to gantry, OC, AW,
                          injector electronics and
PDU & UPS                    varian & laserlight
interference limits                controls
gauss field                      OC = 10G
proximity limit                 Gantry = 1G
                             OC - Gtry = 60'
standard cable               OC - PDU = 60'
lengths                      PDU - Gtry = 20'
                             OC - Gtry = 75'
alternate cable              OC - PDU = 75'
lengths                      PDU - Gtry = 55'

voltage/amps/MDC                480V/90A
cat NO.                         E4502AB

UPS                       B7999ZA or E4502KY
recommended
ceiling height                     9'-0"
absolute min.
ceiling height                     8'-0"
absolute min. room
size
(LIMITED ACCESS)             10'-10" X 19'-5"


typical room size            12'-10" X 19'-3"
CT QUICK REFERENCE - 15.Aug.11

                          BRIGHTSPEED ELITE,
                             EDGE & EXCEL
gantry/table flr.
anchored?                          YES
gantry cable
access                            REAR
conduit sizes
-OC to Gantry                      3 1/2"
-OC to PDU            -            2 1/2"
PDU to Gantry                2 1/2" and 3 1/2"

                          39" to gantry, OC, AW,
                          injector electronics and
PDU & UPS                    varian & laserlight
interference limits                controls
gauss field                      OC = 10G
proximity limit                 Gantry = 1G
                             OC - Gtry = 60'
standard cable               OC - PDU = 60'
lengths                      PDU - Gtry = 20'
                             OC - Gtry = 75'
alternate cable              OC - PDU = 75'
lengths                      PDU - Gtry = 55'

voltage/amps/MDC                480V/90A
cat NO.                         E4502AB

UPS                       B7999ZA or E4502KY
recommended
ceiling height                     9'-0"
absolute min.
ceiling height                     8'-0"
absolute min. room
size
(LIMITED ACCESS)              11'-0" X 20'-0"


typical room size            12'-10" X 19'-3"
CT QUICK REFERENCE - 15.Aug.11
                     BRIGHTSPEED
                    ELITE ENHANCED
gantry/table flr.
anchored?                   YES
gantry cable
access                    REAR
conduit sizes
-OC to Gantry              3 1/2"
-OC to PDU        -        2 1/2"
PDU to Gantry        2 1/2" and 3 1/2"

                      39" to gantry, OC, AW,
                      injector electronics and
PDU & UPS                varian & laserlight
interference limits            controls
gauss field                  OC = 10G
proximity limit             Gantry = 1G
                         OC - Gtry = 60'
standard cable           OC - PDU = 60'
lengths                  PDU - Gtry = 20'
                         OC - Gtry = 75'
alternate cable          OC - PDU = 75'
lengths                  PDU - Gtry = 55'

voltage/amps/MDC            480V/90A
cat NO.                     E4502AB

UPS                   B7999ZA or E4502KY
recommended
ceiling height                 9'-0"
absolute min.
ceiling height                 8'-0"
absolute min. room
size
(LIMITED ACCESS)         12'-10" X 19'-4"


typical room size         13'-0" X 20'-8"
CT QUICK REFERENCE - 15.Aug.11

                           DISCOVERY CT750
                          HD LIGHTSPEED VCT
gantry/table flr.
anchored?                            YES
gantry cable
access                              REAR
conduit sizes
-OC to Gantry                       3 1/2"
-OC to PDU            -             2 1/2"
PDU to Gantry                 2 1/2" and 3 1/2"

PDU & UPS
interference limits        39" to gantry and OC
gauss field                      OC = 10G
proximity limit                 Gantry = 1G
                              OC - Gtry = 60'
standard cable                OC - PDU = 60'
lengths                       PDU - Gtry = 20'
                              OC - Gtry = 75'
alternate cable               OC - PDU = 75'
lengths                       PDU - Gtry = 55'



voltage/amps/MDC
cat NO.                          480V/125A


                           B7864PZ, P5064PS &
                                 E4502F
                            ((OLD) B7864PP or
UPS                             E4502KT)
recommended
ceiling height                       9'-0"
absolute min.
ceiling height                       7'-6"


                               11'-8" X 19'-2"
absolute min. room               (GT1700)
size                           11'-8" X 21'-2"
(LIMITED ACCESS)                 (GT2000)



                               13'-6" X 20'-0"
                                 (GT1700)
                               13'-6" X 22'-0"
typical room size                (GT2000)
 Delivery: min.           5ft. corr. = 78"
 opening, incl.           6 ft. corr. = 66"
Dollies w/o side          7 ft. corr. = 56"
      rails               8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11
                            DISCOVERY CT590
                            RT, OPTIMA CT580
                               LIGHTSPEED
                                RT16/XTRA
gantry/table flr.
anchored?                            YES
gantry cable
access                              REAR
conduit sizes
-OC to Gantry                       3 1/2"
-OC to PDU            -             2 1/2"
PDU to Gantry                 2 1/2" and 3 1/2"

PDU & UPS
interference limits        39" to gantry and OC
gauss field                      OC = 10G
proximity limit                 Gantry = 1G
                              OC - Gtry = 60'
standard cable                OC - PDU = 60'
lengths                       PDU - Gtry = 20'
                              OC - Gtry = 75'
alternate cable               OC - PDU = 75'
lengths                       PDU - Gtry = 55'



voltage/amps/MDC                 480V/125A
cat NO.                           E4502AE




UPS                        B7999ZA or E4502KY
recommended
ceiling height                       9'-0"
absolute min.
ceiling height                       8'-0"


                               12'-2" X 20'-0"
absolute min. room               (GT1700)
size                           12'-2" X 22'-0"
(LIMITED ACCESS)                 (GT2000)



                               14'-0" X 20'-0"
                                 (GT1700)
                               14'-0" X 24'-0"
typical room size                (GT2000)
 Delivery: min.           5ft. corr. = 78"
 opening, incl.           6 ft. corr. = 66"
Dollies w/o side          7 ft. corr. = 56"
      rails               8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11
                             LIGHTSPEED RT
gantry/table flr.
anchored?                            YES
gantry cable
access                              REAR
conduit sizes
-OC to Gantry                       3 1/2"
-OC to PDU            -             2 1/2"
PDU to Gantry                 2 1/2" and 3 1/2"

PDU & UPS
interference limits        39" to gantry and OC
gauss field                      OC = 10G
proximity limit                 Gantry = 1G
                              OC - Gtry = 60'
standard cable                OC - PDU = 60'
lengths                       PDU - Gtry = 20'
                              OC - Gtry = 75'
alternate cable               OC - PDU = 75'
lengths                       PDU - Gtry = 55'



voltage/amps/MDC                 480V/125A
cat NO.                           E4502AE




UPS                        B7999ZA or E4502KY
recommended
ceiling height                       9'-0"
absolute min.
ceiling height                       8'-0"



absolute min. room
size
(LIMITED ACCESS)               12'-2" X 19'-3"




typical room size              14'-0" X 19'-3"
 Delivery: min.           5ft. corr. = 78"
 opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
      rails      8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11
                      LIGHTSPEED 5x
                    SERIES LIGHTSPEED
                        16 & 16PRO
gantry/table flr.
anchored?                    YES
gantry cable
access                     REAR
conduit sizes
-OC to Gantry               3 1/2"
-OC to PDU        -         2 1/2"
PDU to Gantry         2 1/2" and 3 1/2"

PDU & UPS
interference limits    39" to gantry and OC
gauss field                  OC = 10G
proximity limit             Gantry = 1G
                          OC - Gtry = 60'
standard cable            OC - PDU = 60'
lengths                   PDU - Gtry = 20'
                          OC - Gtry = 75'
alternate cable           OC - PDU = 75'
lengths                   PDU - Gtry = 55'

                             480V/90A
                         E4502AB (LS 16),
voltage/amps/MDC            480V/125A
cat NO.                  E4502AE (16PRO)




UPS                    B7999ZA or E4502KY
recommended
ceiling height                   9'-0"
absolute min.
ceiling height                   8'-0"




absolute min. room
size                       13'-6" X 20'-0"




typical room size          13'-6" X 20'-0"
 Delivery: min.       5ft. corr. = 78"
 opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
      rails      8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11
                         LIGHTSPEED
                       SERIES Qxi, PLUS,
                            ULTRA
gantry/table flr.
anchored?                       YES

gantry cable access            REAR
conduit sizes
-OC to Gantry                  3 1/2"
-OC to PDU         -           2 1/2"
PDU to Gantry            2 1/2" and 3 1/2"

PDU & UPS
interference limits    39" to gantry and OC
gauss field                  OC = 10G
proximity limit             Gantry = 1G
                          OC - Gtry = 60'
standard cable            OC - PDU = 60'
lengths                   PDU - Gtry =20'
                         OC - Gtry = 75'
alternate cable          OC - PDU = 75'
lengths                  PDU - Gtry = 55'
voltage/amps/MDC            480V/90A
cat NO.                     E4502AB
                          B7999ZA or
UPS                         E4502KY
recommended
ceiling height                  9'-0"
absolute min.
ceiling height                  8'-0"
absolute min. room
size                       13'-6" X 19'-4"
 Delivery: min.        5ft. corr. = 78"
 opening, incl.        6 ft. corr. = 66"
Dollies w/o side       7 ft. corr. = 56"
      rails            8 ft. corr = 48"
CT QUICK REFERENCE -15.Aug.11
                      HISPEED Qxi
gantry/table flr.
anchored?                  YES
gantry cable
access                    REAR
conduit sizes
-OC to Gantry             3 1/2"
-OC to PDU                2 1/2"
-PDU to Gantry      2 1/2" and 3 1/2"

PDU & UPS           39" to gantry and
interference limits        OC
gauss field             OC = 1G
proximity limit       Gantry = 1G
                    OC - Gtry = 75'
standard cable      OC - PDU = 75'
lengths             PDU - Gtry = 55'
alternate cable
lengths                    NO
voltage/amps/MD         480V/90A
C cat NO.               E4502AD
                        B7999ZA
UPS                     E4502KY
recommended
ceiling height             9'-0"
absolute min.
ceiling height             7'-6"
absolute min.
room size             13'-0" X 19'-4"
Delivery: min.      5ft. corr. = 78"
opening, incl.      6 ft. corr. = 66"
 Dollies w/o        7 ft. corr. = 56"
  side rails        8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11

                      YMS HISPEED
                      SERIES Xi, Xi
                    PLUS, Xi PRO, Nxi,
                    Nxi PLUS, Nxi PRO
gantry/table flr.
anchored?                  YES
gantry cable
access                    REAR

conduit sizes
-OC to Gantry                2"
-OC to PDU                   2"
-PDU to Gantry           (2) 2 1/2"
PDU & UPS
interference           20" to gantry
limits                  39" to OC

gauss field             OC = 10G
proximity limit        Gantry = 1G
                    OC - Gantry = 59'
standard cable       OC - PDU = 59'
lengths             PDU - Gantry = 59'
alternate cable
lengths                    NO
voltage/amps/MD         480V/90A
C cat NO.               E4502AD

                    E4504AL     (req's
UPS                     E4502AD)
recommended
ceiling height             8'-6"

absolute min.
ceiling height             7'-6"

absolute min.
room size             11'-6" X 16'-6"

Delivery: min.
               5ft. corr. = 61"
opening, incl.
               6 ft. corr. = 52"
 Dollies w/o
  side rails
Delivery: min.
opening, incl.
 Dollies w/o
               7 ft. corr. = 46"
  side rails
                  8 ft. corr = 43"

				