Respironics

BiPAP-Vision









1

Classification

 Electrically powered – internal battery powers

Vent Inoperative & audible alarms if AC power lost. Error code

205 if internal low battery condition.

 2 modes

 CPAP



 S/T



 Set Rate



 IT in seconds



 IPAP



 EPAP



2

Circuit

 BiPAP circuit (with

exhalation port or

Whisper Swivel)

 Humidifier or HME

 With heated humidifier

need circuit with water

traps. Cannot cut into regular

BiPAP circuit to add traps.









3

Start Up

 Plug into AC mains & Connect to 50 psi gas

source

 Connect

BiPAP circuit Dispos-

able

- don’t cap O2 inlet

 Power switch module filter



RU corner

on back

panel – Fuses

ON







4

5

Power-on-self-test (POST)

initiates









6

After POST, Start “Test

Exhalation Port”









7

Follow steps on screen, block

circuit, “Start Test”.









Exhalation (castle) port

8

Exhalation port test in

progress









9

Exhalation Port Test

 Analyzes leak rate of exhalation port

 If successful – the value of “Pt. Leak” will

be displayed

 If unsuccessful – “Tot. Leak” will be

displayed

 EPT must be performed when using the

Plateau Exhalation Valve because the leak

is significantly different than the port or

swivel.

10

Test Complete, press

Monitoring to exit.









Monitoring





11

Trigger Criteria in S/T

 Time Trigger (Rate)

 Flow Trigger

 6 ml inspired volume detected

 Patient flow exceeds the shape signal (a

shadow of the flow signal delayed by 300 ms)

Pressure





Flow







12

In S/T mode



 Only patient (flow) triggered breaths are

flow cycled (PS).

 Time triggered breaths are time cycled

according to the set Inspiratory Time.









13

Cycle Criteria in S/T

 Time cycled

 Set Inspiratory Time

 A 3 sec maximum IPAP time has occurred

 Flow cycled

 Flow reversal detected

 Patient flow is less than signal flow

 Spontaneous Expiratory Threshold achieved –

another way to flow cycle

Pressure

SET



Flow



14

Alarm

Silence

IPAP

8

EPAP

3 Alarm

Rate Reset

12





Pt Data

Display

Area



Monitoring







Parameters Alarms

Mode

15

To change settings, press Parameters –

then soft key next to parameter and

adjust with knob.

Soft keys









Knob 16

Can set Parameters while in

Stand-by









17

Stand-by Mode



 Can set parameters and alarms while in

SB

 Bumping the circuit may simulate pt

inspiration, in which case Vision exits SB

automatically and goes into ventilation

mode.







18

To change Mode, press Mode, then

parameter soft key to adjust, and then

“Activate New Mode” soft key



2









3









1 19

Alarms displayed in

Mode/Message Area









20

 Lo Min

 Hi Press Alarms Vent

 Lo Press  Hi Rate

 Lo Press  Lo Rate

Delay

 Apnea

 Learn

Base

Flow









21

Cannot adjust Parameters

when Monitoring displayed









22

IPAP Rise Time









Timed

Inspiration









23

Can Change Bar Graphs to Waveforms

with “Options” soft key









24

Mini-heart Nebulizer Tx









25



Related docs
Other docs by liwenting
第04章 类的重用
Views: 89  |  Downloads: 0
摘要
Views: 81  |  Downloads: 0
摘要
Views: 85  |  Downloads: 0
摘要_2_
Views: 68  |  Downloads: 0
國泰醫院2012年紙本期刊到刊總表
Views: 134  |  Downloads: 0
”Lyme_disease”_-_the_European_history
Views: 66  |  Downloads: 0
تعریف و تاریخچهPRP
Views: 77  |  Downloads: 0
_C6C28D15-9903-407A-8FEE-77A0422212B0_
Views: 113  |  Downloads: 0
__________
Views: 96  |  Downloads: 0
By registering with docstoc.com you agree to our
privacy policy

You are almost ready to download!

You are almost ready to download!