Scanner Chart
The following chart list the factor used by must Insta Trak, ENTrak & ConneCTstat Users:
GE Light Siemens
GE Siemens Picker Maroni Elscint Toshiba Phillips
Speed Volume Zoom
Scanning Protocal:
Gantry Angle Zero Zero Zero Zero Zero Zero Zero Zero Zero
Matrix 512 513 514 515 516 517 518 519 520
Scanning Plane Axial Axial Axial Axial Axial Axial Axial Axial Axial
Table Swivel --- --- --- --- --- --- Zero --- ---
Scanning Guidelines:
Set Up Image Scout Scout Topogram Topogram Pilol Pilot Surview Scanogram Scanogram
Scan Mode Helical Helical Spiral Spiral Helical Helical Helical Helical Helical
Slice Thinkness 3mm 2.5 / 1.25mm 3mm 3mm 3mm 3.2mm 3.2 / 2.5mm 3mm 3mm
Reconstructed Table Increment 1mm 1 / 1.25mm 1mm 1mm 1mm 1mm 1mm 1mm 1mm
Algorithm / Filter Bone Bone Ulta-high Ulta-high Bone Facil Bone Filter D FC30 Filter B
Cranal Algorithm Standard Standard Brian Brian Brian Brian Brian Brian Brian
Resolution --- --- --- --- --- Standard Standard --- ---
Brief Compress --- --- No NO --- NO --- --- ---
Pitch 1;1 3 / HQ 1;1 1;1 1;1 1 1;1 1;1 1;1
Headholder ok? No No Yes Yes No No Yes No No
Field of View 25cm 25cm 25cm 250mm 250mm 250mm 250mm 250mm 25cm
SFOV Large Body Body --- --- --- --- --- --- ---
Be sure to keep all parametes consistent throughout scan!