Start Labbook and choose job
ISIS uses the Labbook application to manage data files and to launch applications. Each
registered user will have a job assigned to them according to their last use. Therefore, it is good
to have multiple users registered so that both user and job do not have to be changed at the
beginning of each session.
Start Labbook (if not already running) with the icon on the QuickLaunch toolbar or on the
desktop.
Choose user from the list and click OK.
Change Job, if necessary.
Start the “standard” programs. AutoColumn, AutoUpdate, and Tetra should be started as a matter
of course.
AutoColumn is used to display (and set) the SEM parameters. It is helpful to have this running to
ensure that the conditions which the ISIS system has read are indeed the same as the those to
which the SEM is set. There is a delay (sometimes considerable) in the update of values from the
SEM. It is best to ensure that the correct values are on record before collecting data, especially
images.
AutoUpdate is used to toggle polling of SEM conditions on and off. This will normally be set on
when Labbook is first started. However, it may be off if Labbook was already running. It is best
to run AutoUpdate to be certain. Polling should be turned off when taking multiple photographs.
It may also be turned off once the important conditions (kV and WD) have been read.
Magnification is important for images and especially for line scans, but it may be ignored for x-
ray spectra and maps.
Tetra is used to control the Tetra BSE detector. Since that detector is used for most composition
imaging, the control program should be started.
Click on the appropriate icon to start the desired task, usually X-ray analysis. Imaging and
Automation are also common tasks.
Set SEM conditions for the task. The following conditions are used for more than 95% of x-ray
work. Other conditions may be chosen or recommended in conjunction with lab staff. Because
these conditions are used frequently, they have been stored under parameter set #5 under Shift-
F10.
25 mm working distance. (the detector is aimed at this height)
X-ray detector in.
20 kV (needed for good excitation of all x-ray lines)
195 beam current (for good count rates)
40 Pa atmosphere (needed only for insulating specimens)
X-ray setup for Selectable, Light element mode, Process constant 4 (these are not the default
settings, but are normally setup for new users). For higher throughput, light element mode
may be turned off (i.e., select Fast Counting mode).
Check count rate using X-ray application.
Locate an area of the sample with a good intensity. Ensure the raster is completely on the
area of interest and not overhanging the edge of the sample. Metals are usually more
intense than oxides which are more intense than inorganic materials or polymers
Start acquiring a spectrum.
Adjust beam current (F10) and or EDS detector position to 30-40% deadtime. Note: some
samples may not tolerate high currents without charging or thermal damage.
Note: there is a tradeoff between signal quality and spatial resolution. BSE signal quality
benefits from high current. Current may be set high for a good quality image and the EDS
detector may be withdrawn to reduce the count rate. For highest count rate with the least
current, insert the EDS detector as far as it can go and reduce the beam current with F10
to a maximum of 40% deadtime. Even lower count rates may be used but will prolong the
counting time for the same precision in the results.
Set the signal contrast and brightness for the beam current chosen.
Collecting X-ray maps
Recall or paint x-ray windows. Many are already defined and may be used as a starting point.
Consider a background window using an element not present, e.g., Argon.
Edit names. Rename background window to “bkgd” and rename windows for overlapping
lines, e.g., P-Zr or Na-Zn. You may also shorten long window designations (e.g., Na Ka1 to
Na or NaKa).
Start Speed-Map (icon on X-ray analysis program)
Define acquisition setups
Number of frames. We normally find 40 frames to be quite adequate at moderate count
rates. More frames will improve the appearance of the maps. I discourage collecting less
then 20 frames.
Pick map size. Maps smaller than 256 pixels are suitable only for very gross structures.
Maps of 512 pixels are limited to only 6 elements along with the image.
256 pixel maps require about 15 seconds per frame scan. Thus, a 40-frame scan requires
about 10 minutes.
512 pixel maps require about 60 seconds per frame scan. Thus, a 50-frame scan requires
more than 45 minutes. However, fewer frames might give quite suitable detail in the
maps.
Start the scan.
Wait.
Save the results.
Recall the file. This brings up the sample ID and allows for faster processing. It also allows
for point analysis at areas of interest in the map.
Size the window to fit the maps and “tidy” the display. Fine-tune the window size to just
encompass the maps of interest. Save the layout for future use.
Adjust map brightness and contrast (Ctrl-E).
Copy bitmap (Edit, Copy Results) and paste into PCI or similar program. Convert the image
to grayscale and save the file in GIF format for portability.
Print the map set.
Collecting images
Working distance for whatever is suitable for imaging.
X-ray detector out for working distances of less than 25 mm.
Beam voltage and current set for “good” image signal and resolution. Higher resolution
generally requires lower current while good quality requires more current and/or shorter
working distance.
Start Speed-Map
Define acquisition setups
Pick map size. Maps smaller than 256 pixels are suitable only for very gross structures. Maps
of 512 pixels are limited to only 6 elements along with the image.
256 pixel maps require about 15 seconds per frame scan. Thus, a 50-frame scan requires
about 12 minutes.
512 pixel maps require about 60 seconds per frame scan. Thus, a 50-frame scan requires
more than 45 minutes. However, fewer frames might give quite suitable detail in the maps.
Start the scan.
Wait.
Save the results.
Recall the file if you want the label to appear.
Size the window to fit the maps and “tidy” the display. Fine tune the window size to just
encompass the maps of interest. Save the layout for future use.
Adjust map brightness and contrast.
Save bitmap to a file for portable, digital format (File, Print, Save Results). Limit file name to
8 characters. The name may be lengthened later, and the file may be converted from DIB
(~BMP) format to GIF for portability.
Print the map s