Docstoc

BONE MILL

Document Sample
BONE MILL Powered By Docstoc
					           Equipment Specifications for TUBULAR EXTERNAL FIXATOR

           UNSPSC Code:
           ECRICode:
1 Description of Function
SI         Name                                                         Technical   Bidders
                                                                        Specs       Deviation
                                                                        quoted      if any
                                                                        by bidder

     1.1     These instruments are used in orthopedic surgery for
             fixation of open fractures/comminuted fractures or
             fracture pelvis externally by pins and rods.
2           Operational Requirements
SI         Name                                                         Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted    if any
                                                                        by bidder

     2.1     As specified

     3 Technical specifications
SI         Name                                                         Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted    if any
                                                                        by bidder

     3.1   TUBULAR EXTERNAL FIXATOR SYSTEM
     3.2 The contents described below are the minimum essential
           for the set. The quantity and size of the implant may vary
           with the demand of the user.

     3.3 ITEM DESCRIPTION                       QTY

            External Fixator

       Universal Clamp
       Tubular rods with Caps 6"-18" with 5 cm increment
       Sigma Clamps (Twin Clamps)
       Tube to Tube Clamp
       Drill Bit 3.2/3.5mm , 6-9” length
       Schanz Pin 4.5 mm Cortical of different lengths 6,9inch.
       Schanz Pin 5 mm Cancellous
       Schanz Pin 6.5 mm Cancellous
       Transverse Clamp
       Inter Connecting Rod Assorted Sizes
       Spanner(10-11)
       Hinged Box Spanner
       Triple Trocar with Sleeves 6.5 mm
       T-Handled with Chuck
Dr. Deepak Mittal                                         Dr. V.K. Sharma
        Forearm/ Small Bone External Fixator
        Instrument Set for above:
        Drill Bits 2.5 mm
        Schanz pin 3.5 mm Cortical
        Schanz pin 2.5 mm Cortical with 3.5mm shaft.
       Universal Clamp 3.5mm
        Spanner No. 8,9
        Assorted sized interconnecting rods

        External fixator for Metacarpal and Phalanges, small tubular bones- radius etc

        Instrument for above should consist of following
        Small, Medium, large distractors 1.5/2mm/2.5/3.5mm of assorted length
        Drill Bits 1.5 mm
        Allen Key of suitable size
        Threaded/ plain K wires/schanz pins 1.5/2mm/2.5mm/3mm.
        Link Joints of various configurations for above wires and pins
        Interconnecting rods

3      System Configuration Accessories, spare and consumables

Sl                       Name                      Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.
3.1    As specified

4      Environmental factors
Sl                      Name                       Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.

4.1    1. The unit shall be capable of being stored continuously in ambient temperature of 0-50
       deg C and relative humidity of 15 -90%.

5      Power Supply
Sl                       Name                      Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.
None

6      Standards, safety and training
Sl                       Name                      Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.

6.1    Should be manufactured by company etc. which are licensed by Drug Controller of India
       and conform to the standards laid down by Drug Controller of India.
6.2    Material should be of implant grade steel, carbon fibre, titanium as required.

6.3    Shelf Life of more than 2 years.


7      Documentation
Sl                       Name                     Technical Specs           Bidders deviation if
                                                  Quoted by the bidder. any.
7.1    1. User/Technical / Maintenance manuals to be supplied in English.
       2. List of important spare parts and accessories with their part number and costing.

         -2-


Dr. Deepak Mittal                                            Dr. V.K. Sharma
Equipment Specifications for ELECTRO-HYDRAULIC OPERATING
                     THEATRE TABLE FOR ORTHOPAEDIC

UNSPSC Code:
ECRI Code:
1 Description of Function

   Sl    Name                                                           Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted    if any
                                                                        by bidder

   1.1   Operating tables provide an elevated surface that
         supports the patient’s body during surgical procedures,
         stabilizing the patient’s position and providing optimal
         exposure of the surgical field.

2 Operational Requirements

   Sl    Name                                                           Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted    if any
                                                                        by bidder

   2.1   C-arm compatiblity with electro-hydroulic operation table

3 Technical Specifications

   Sl    Name                                                           Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted    if any
                                                                        by bidder

   3.1   Following are the minimum essentials, quantity&
         size,minor details to be specified by the user:
         It should be removable, seven or more sections of
         tabletop:
         * Independently removable section
         * Full Length radio transparency along entire length with
         ability to move C arm under and around table throughout
         its length.
         * Besides, normal elevation of column, the table top
         should be capable of lift by control unit
         * There should be provision for lifting the trunk and pelvis
         sections either separately or together, anatomically
         elevated at the same height
         * Head section with adjustment for optimum positioning
         * The table top should allow the use X-ray cassette all
         over its length
         *The pads should be removable for hygiene and
         cleaning,bacteriostatic, should be made of long life
         material.
         All types of Orthopaedic position like beach chair, for

Dr. Deepak Mittal                                           Dr. V.K. Sharma
         discectomy surgery, femoral& tibial interlocking etc
         should be feasible.
   3.2   Motorized Movements:
         * Height Adjustment Range : 200 mm or more
         * Table Top Lift Range : 100 mm or more
         * Lateral Tilt : 15 deg. or more
         * Trendelenberg : +30deg / -15deg.
         * Lengthwise manual movement of trunk sections for
         easy access to image intensifier : 300 mm or more


   3.3   Column: Column for above table top, complete with built-
         in maintenance free battery. Column should be movable
         on trolley.
   3.4   Trolley: For moving the table top and column with facility
         for height adjustment and trendlenberg tilting

4 System Configuration Accessories, spares and consumables

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   4.1   Standard Accessories:
         * Double remote control One Cordless Infra Red and with
         Cables
         * Motor Driven emergency movement module
         * Water proof foot pedal for table top height adjustment
         and indicators
         * Thigh Holder: Pair of thigh holder, swiveling, adjustable
         via ball & socket joint, lengthwise and laterally
         adjustable. Complete with pads, fixing straps and clamps
         * Instrument tray, sliding or rail, Dimension: App x 320x
         350 mm
         * Mobile Frame: Mobile frame, with stainless steel bucket,
         complete with 4 castors
         * Cassette Holder: Holding arm for positioning x-ray
         cassettes under radio transparent table top. Mounted on
         side rails by means of clamp.
         * Insertion Bar: Bar for insertion of x-ray cassette, under
         the table top.
   4.2   Attachments:
         * Lower Limbs Traction Device
         * Femoral Nailing Device
         * Knee Arthroscopy Device
         * Meniscus Positioning Device
         * Hip Prosthesis Device
         * Humerus Positioning Device
         * Shoulder Arthroscopy Device
         * Spine Cord Positioning Device
         * Hand Surgery Table

5 Environmental factors

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation


Dr. Deepak Mittal                                          Dr. V.K. Sharma
                                                                      quoted    if any
                                                                      by bidder

   5.1   The unit shall be capable of being stored continuously in
         ambient temperature of 0 -50deg C and relative humidity
         of 15-90%
   5.2   The unit shall be capable of operating in ambient
         temperature of 20-30 deg C and relative humidity of less
         than 70%

6 Power Supply

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   6.1   Power input to be 220-240VAC, 50Hz fitted with Indian
         plug
   6.2   Energy requirement and efficiency should conform to
         Indian Bureau of energy efficiency etc.

7 Standards, Safety and Training

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   7.1   Manufactures/Supplier should have ISO certificate to
         Quality Standard.
   7.2   Should be compliant with IEC 61010-1:(or any
         international equivalent eg EN/UL 61010) covering safety
         requirements for electrical equipment for measurement
         control and laboratory use
   7.3   Should be CE, FDA approved .
   7.4   Comprehensive training for lab staff and support services
         till familiarity with the system.

8 Documentation

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   8.1   User/Technical/Maintenance manuals to be supplied in
         English.
   8.2   Certificate of calibration and inspection.
   8.3   List of Equipments available for providing calibration and
         routine maintenance support as per manufacturer
         documentation in service / technical manual.
   8.4   List of important spare parts and accessories with their
         part number and costing


Dr. Deepak Mittal                                          Dr. V.K. Sharma
   8.5   Log book with instructions for daily, weekly, monthly and
         quarterly maintenance checklist. The job description of
         the hospital technician and company service engineer
         should be clearly spelt out.
   8.6   Compliance Report to be submitted in a tabulated and
         point wise manner clearly mentioning the page/para
         number of original catalogue/data sheet.Any point, if not
         substantiated with authenticated catalogue/manual, will
         not be considered.




Dr. Deepak Mittal                                         Dr. V.K. Sharma
       Equipment Specifications for TOTAL HIP PROSTHESIS

       UNSPSC Code:
       ECRI Code:
1 Description of Function
SI         Name                                                               Technical Bidders
                                                                              Specs     Deviation
                                                                              quoted by if any
                                                                              bidder
     1.1     Hip replacement, also hip arthroplasty, is a surgical
            Procedure in which the hip joint is replaced by a prosthetic
            implant. Such joint replacement orthopedic surgery generally is
            conducted to relieve arthritis pain or fix severe physical joint
            damage as part of the hip fracture treatment.


2 Operational Requirements
SI         Name                                                               Technical Bidders
                                                                              Specs     Deviation
                                                                              quoted    if any
                                                                              by bidder
     2.1 Cemented stem should be highly polished collarless and smooth.
     2.2    Cementless stem should have good bone ingrowth properties with metal / hydroxyappati
            Te coat which may be variable in length.
2.3        Revision stem with proximal tag holes , variable length of coating, and stem length fixed/
           modular.
2.4        Modularity in neck / stem length with option of varying horizontal/ vertical/ anterior offset.
2.5 Good quality locking mechanism for acetabular shell and liner. And good filling femoral stme
in case of cementless total hip prosthesis.
2.6 Head size of 22,28,36mm or as per user demand for longevity of implant.


     3 Technical Specifications
SI         Name                                                               Technical Bidders
                                                                              Specs     Deviation
                                                                              quoted by if any
                                                                              bidder

     3.1    01 Cemented Total Hip Prosthesis with / without
            disposables for cement gun and low viscosity cement


            Sizes of implant and quantity will vary with the user demand & following are the
            minimum essentials.

            A Cemented THR Should consist of:
            Femoral Stem of various sizes
            Femoral Head of 22.28,36mm as applicable
            Acetabular Cup polyethylene/ ULTRA high molecular weight polyethylene with/without
            pods of different sizes
Dr. Deepak Mittal                                                   Dr. V.K. Sharma
       Cement Restrictor 8mm to 14mm polyethylene
       Bone Cement of the required viscosity Normal/low/high viscosity as applicable.
       Cement Injector Kit
       Stem centralizer

       B Bipolar Hemiarthroplasty ( cemented) Should consist of:
       Femoral Stem of different sizes
       Femoral Head of appropriate size
       Bipolar Shell with or without seperate Polyethylene Liner for Bipolar Shell of different
       sizes

       Cement Restrictor 8mm to 14mm
       Bone Cement of the required viscosity Normal/low/high viscosity as applicable.
       Cement injector Kit Disposable
       Stem centralizer


       02 Cementless Total Hip Prosthesis

        A Should consist of:

        Cementless Stem of different sizes with metal/ hydroxyapatite
       coat etc.

       Femoral head of different sizes:
       22,28,36mm

       Acetabular shell with metal/ hydroxyapatite coat etc.

       Polyethylene/ UHMWPE liner

       Acetabular screws

       B Ceramic on ceramic should consist of:

       Cementless stem

       Ceramic of high quality and fatigue strength with ability to
       bear cyclical loads femoral head

       Acetabular shell

       Ceramic liner

       Acetabular screws

       C should consist of:

       Cementless stem

       Femoral head

       Acetabular shell

       Polyethylene liner with metal surface

       Acetabular screws

Dr. Deepak Mittal                                              Dr. V.K. Sharma
       D Revision THR should consist of :

       Cementless femoral stem with holes for suture
       tying AND MODULARITY IN STEM/ NECK
       LENGTH AND OFFSET

       Femoral head

       Acetabular shell

       Polyethylene liner with metal surface

       Acetabular screws

       E should consist of:

       Cementless femoral stem

       Femoral head

       Bipolar shell

       Polyethylene liner for bipolar shell

       03 hybrid total hip prosthesis with
       disposables for cement gun and
       low/ HIGH/ NORMAL viscosity bone
       cement
       A Should consist of :
       Femoral Stem
       Femoral Head
       Acetabular shell
       Polyethylene liner
       Acetabular screws
       Cement Restrictor 8mm to 14mm
       Bone Cement low viscosity
       Cement Injector Kit disposable
       B Should consist of:
       Femoral Stem
       Femoral Head
       Acetabular shell
       Polyethylene liner with metal surface
       Acetabular screws
       Cement Restrictor 8mm to 14mm
       Bone Cement low viscosity
       Cement Injector Kit disposable


       C eramicshould consist of :

       Femoral stem

       Ceramic femoral head

       Acetabular shell

       Ceramic liner


Dr. Deepak Mittal                                      Dr. V.K. Sharma
         Acetabular screws
         Cement Restrictor 8mm to 14mm
         Bone Cement low viscosity 2 Packet
         Cement Injector Kit disposable 1 Set

      4. System Configuration Accessories, spare and consumables

Sl                        Name                     Technical Specs          Bidders deviation if
                                                   Quoted by the bidder. any.
             a. Should be of high quality metal/alloy/ceramic to sustain maximum number of
                cyclical load. Like chromium-cobalt; Titanium alloy/ zirconia etc.
             b. Polyethylene may be normal or ultra high molecular weight as specified
             c. Cement of powder and liquid mixing parts – polymethylmethacrylate – antibiotic/
                not and of different viscosity. Should be conforming to high international
                standards/ least damaging.

      5. Environmental factors
Sl                       Name                      Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.

             5.1. The unit shall be capable of being stored continuously in ambient temperature
             of 0-50 deg C and relative humidity of 15 -90%.
       5.2. The unit shall be capable of operating continuously in ambient temperature of 10-40
       deg C and relative humidity of 15-90%.


      6. Power Supply
Sly                       Name                     Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.
None
   7. Standards, safety and training
Sl                    Name                         Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.

7.1 Company/ supplier Should have a CE, FDA approved ceritfication.
7.2 Material should be of implant grade with long life cobalt chromium etc. as required.
7.3 Comprehensive training for lab staff and support services till familiarity with the system on
    site.
7.4 Long shelf life.

      8. Documentation
Sl                        Name                       Technical Specs           Bidders deviation if
                                                     Quoted by the bidder. any.
8.1    1. User/Technical / Maintenance manuals to be supplied in English.
       2. Certificate of calibration and inspection.
       3. list of equipments available for providing calibration and routine preventive maintenance
       support as manufacturer service/maintenance manual.
       4. List of important spare parts and accessories with their part number and costing.
       5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
       checklist. The job description is clearly spelt out.
       6. Compliance report to be submitted in a tabulated and point wise manner clearly
       mentioning the page/para/number with authenticated catalogue/manual without which it
       will not be considered.




Dr. Deepak Mittal                                            Dr. V.K. Sharma
     Equipment Specifications for SMALL FRAGMENT INSTRUMENT
                      SET and distal radius locking plate

UNSPSC Code:
ECRI Code:
 1 Description of Function
SI     Name                                                                          Technical Bidders
                                                                                     Specs     Deviation
                                                                                     quoted    if any
                                                                                     by bidder

     1.1 These instruments are used in orthopedic surgery for fixing
 small bone fractures as of forearm and wrist etc .
     2 Operational Requirements
SI     Name                                                                          Technical Bidders
                                                                                     Specs     Deviation
                                                                                     quoted by if any
                                                                                     bidder
2.1 LCP DISTAL RADIUS 2.4 PLATE : SHOULD HAVE The LCP combi-hole

In the shaft portion of plate WITH the choice of using either standard screws or locking screws for fracture fixation,

or a combination of the two screw types. It should consist of:

DC hole for standard screws The DC hole corresponds to the DCU (Dynamic Compression

Unit) of the LC-DCP plate 2.0 or 2.4 and is intended for standard cortex screws.

Tapered, threaded hole for locking screws

Elongated LCP combi-hole. At junction

Round hole Threaded round holes accept LCP Locking Screws 2.4 mm(or LCP Locking Screws

2.7 mm, head LCP 2.4) without protrusion of the screw head. AT T OR RIGHT ARM OF PLATES.




     As specified
3 Technical Specifications
SI     Name                                                                          Technical Bidders
                                                                                     Specs     Deviation
                                                                                     quoted by if any
                                                                                     bidder



Dr. Deepak Mittal                                                        Dr. V.K. Sharma
3.1 Following are the minimum essentials and configuration, size, quantity to vary with the user
demand.

Dorsal radius plates:

SEPERATE FOR right and left (where applicable) and in standard and long design

– LCP L Distal Radius Plate 2.4, 2 holes, left angled

– LCP L Distal Radius Plate 2.4, 2 holes, right angled

– LCP L Distal Radius Plate 2.4, 3 holes, right angled

– LCP L Distal Radius Plate 2.4, 3 holes, left angled

– LCP L Distal Radius Plate 2.4, 3 holes, oblique right angled

– LCP L Distal Radius Plate 2.4, 3 holes, oblique left angled

– LCP T Distal Radius Plate 2.4

Volar distal radius plates

2 types of volar plates, available in right and left and in standard and long design

– LCP Distal Radius Plate 2.4, volar

– LCP T Buttress Plate 2.4, volar

LCP Distal Radius Plates 2.4

Implants

1 type of volar plate in DIFFERENT lengths, shaft with 6-8HOLES

Locking screws

– LCP Locking Screw 2.4 mm, self-tapping

– LCP Locking Screw 2.7 mm (head LCP 2.4), self-tapping

Standard screws
– Cortex Screw 2.4 mm,
– Cortex Screw 2.7 mm,


Special instruments
    , Description Qty.
Drill bit 2.5mm dia for quick coupling

Drill bit 3.5mm dia for quick coupling

Drill Bit 1.8 mm with marking, length ,

Dr. Deepak Mittal                                                           Dr. V.K. Sharma
Drill Bit 2.0 mm with marking, length

          Tap for 3.5mm Cortex Screws
          Tap for 4.0 mm Cancellous Bone Screws
          Counter sink
          Insert Drill Sleeve with Stop 3.5/ 2.5mm
          Double drill Sleeve 3.5/2.5mm

Double drill Sleeve 3.5/ 2.5mm

LCP Drill Sleeve 2.4, with scale up to 30 mm, for Drill Bit 1.8 mm –2nos

LCP Drill Sleeve for LCP screws 2.7 mm (head LCP 2.4), with scale up to 30 mm, for Drill Bit 2.0 mm



          Screwdrivers Shaft, hexagonal/etc, for 3.5/ 2.7/2.4.Omm screw
          drivers
          Holding sleeve
          Depth Gauge for 3.5 & 4mm Screws
          Bending iron for small plates
          Bending Press
          Small Bone holding Forceps, Self Centering, Handle 190mm

          Reduction Forceps, Serrated Jaws, Handle , 140mm
          Hohmann retractor 150 X 7mm
          Hohmannn retractor 160 X6 X 3mm
          Hohmann retractor 160 X 8X3 2
          Periosteal Elevator curved straight cutting edge 3mm X 185
          mm
          Periosteal Elevator curved straight cutting edge 6mm X
          185mm
          T handle with Quick coupling
          Screw driver for 3.5 & 4.0 mm hexagonal screws
          Neutral and compression position drill guide for 2.5 mm drill bits and 3.5 mm screws
          Reduction forceps with points length 125 mm
          Triple drill guide 2mm
          Screw forceps self holding
          Universal chuck cannulated with T handle
          Sterilization case with trays

Sterilization and storage screw box for 3.5 screws

Available in stainless steel (SSt) or titanium alloy (TAN).



     4. System Configuration Accessories, spare and consumables

Sl                           Name                          Technical Specs             Bidders deviation if
                                                           Quoted by the bidder.       any.
4.1 As specified

     5. Environmental factors
Sl                      Name                               Technical Specs             Bidders deviation if
                                                           Quoted by the bidder.       any.

Dr. Deepak Mittal                                                      Dr. V.K. Sharma
5.1.1     . The unit shall be capable of being stored continuously in ambient temperature of 0-50
          deg C and relative humidity of 15 -90%.
        2. The unit shall be capable of operating continuously in ambient temperature of 10-40
        deg C and relative humidity of 15-90%.

        Parameters:
      6. Power Supply
Sl                         Name                     Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.
None
   7. Standards, safety and training
Sl                    Name                          Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.

 7.1. Company/ supplier Should have a license for manufacture by Drug Controller India and
conform to the laid down standards by them in this regard .
       2. Material should be of implant grade steel, titanium as required.
       3. Comprehensive training for lab staff and support services till familiarity with the system
       on site.
       4. shelf life should be long.
       5. Comprehensive warranty for 2 years and 5 years AMC after warranty.


      8. Documentation
Sl                         Name                       Technical Specs           Bidders deviation if
                                                      Quoted by the bidder. any.
5.1     1. User/Technical / Maintenance manuals to be supplied in English.
        2. Certificate of calibration and inspection.
        3. list of equipments available for providing calibration and routine preventive maintenance
        support as manufacturer service/maintenance manual.
        4. List of important spare parts and accessories with their part number and costing.
        5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
        checklist. The job description is clearly spelt out.
        6. Compliance report to be submitted in a tabulated and point wise manner clearly
        mentioning the page/para/number with authenticated catalogue/manual without which it
        will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
  Equipment Specifications for SET FOR FIXATION OF PROXIMAL
                                           FEMUR

UNSPSCCode:
ECRI Code:
1 Description of Function
SI     Name                                                       Technical Bidders
                                                                  Specs     Deviation
                                                                  quoted    if any
                                                                  by bidder

     1.1 These instruments are used in orthopedic surgery for

fixation of proximal thigh fractures.

     2 Operational Requirements

SI     Name                                                       Technical Bidders
                                                                  Specs     Deviation
                                                                  quoted by if any
                                                                  bidder

     2.1 As specified

     3 Technical

specifications

SI     Name                                                       Technical Bidders
                                                                  Specs     Deviation
                                                                  quoted by if any
                                                                  bidder

     3.1 DESCRIPTION              QTY
     Following are the minimum essentials ; quantity
          &size where appropriate may vary with the
          user demand.
     Dynamic Hip Screw INSTRUMENT
     ADJUSTABLE PIN GUDE 1
GUIDE WIRE PINS PROXIMALLY THREADED 15
 PIN RE LOCATOR 1
4.5 MM TAP SLEEVE 1
 6.5 MM TAP SLEEVE 11
T-HANDLE FOR DHS TAP 1
 NEUTRAL AND LOAD DRILL GUIDE 1
130 TO 150 DEG. (ADJUSTABLE) PIN TEMPLATE
 SHARP HOOK 1
T-HANDLE FOR DHS TAP 1
 IMPACTOR GUIDE 1
 TUBE/PLATE TRIALS 1
LARGE SCREW DRIVER BIT WITH QUICK COUPLING 2
EXTRACTOR RETAINING SHAFT AND KNOB 1
SCREW FORCEPS 1
6.5 MM TAP FOR 6.5 MM CANCELLOUR SCREWS 1

Dr. Deepak Mittal                                         Dr. V.K. Sharma
 EXTRACTOR FOR LONG SCREWS 1
DEPTH GAUGE 1
SCREW DRIVER 1
3.2 MM DRIL BIT WITH QUICK COUPLING 1
 4.5 MM TAP SLEEVE 1
DEPTH GAUGE 1
LAG SCREW INSERTER 1
 CENTERING SLEEVE 1
6.4 MM TWIST DRILL 1
STANDARD TRIPLE REAMER one
DHS reamer shaft one
Large cannulated bone tap for DHS one
Sterilization and storage box for implants and
instruments one.

Dynamic Condylar Screw INSTRUMENTS

ADJUSTABLE PIN GUDE 1
GUIDE WIRE PINS PROXIMALLY THREADED 15
 PIN RE LOCATOR 1
4.5 MM TAP SLEEVE 1
 6.5 MM TAP SLEEVE 11
T-HANDLE FOR DCS TAP 1
 NEUTRAL AND LOAD DRILL GUIDE 1
95 DEGREE PIN TENPLATE1
90 DEGREE PIN TEMPLATE 1
 SHARP HOOK 1
T-HANDLE FOR DCS TAP 1
T-HANDLE WITH QUICK COUPLING FOR SCREW DRIVER
BIT DRILL BITS AND TAPS 1
 IMPACTOR GUIDE 1
 TUBE/PLATE TRIALS 1
LARGE SCREW DRIVER BIT 2
EXTRACTOR RETAINING SHAFT AND KNOB 1
SCREW FORCEPS 1
6.5 MM TAP FOR 6.5 MM CANCELLOUR SCREWS 1
DEPTH GAUGE 1
SCREW DRIVER 1
3.2 MM DRIL BIT WITH QUICK COUPLING 1
 4.5 MM TAP SLEEVE 1
LAG SCREW INSERTER 1
 CENTERING SLEEVE 1
6.4 MM TWIST DRILL 1
DCS reamer shaft
Large cannulated bone tap for DCS one
Sterilization and storage box for implants and
instruments one.

Implants for DHS
DHS PLATES 135 deg. 4,5,6,7,8 holed etc
DHS Screws 70,75,80,85,90mm etc.
Compression screws of different sizes
4.5mm and 6.5mm screw box with commonly used
screw sizes.

Implants for DCS
DCS PLATES 95 deg. 8,10,12 holed etc.
DHS Screws 50,55,60,65,70mm etc.
Dr. Deepak Mittal                                Dr. V.K. Sharma
Compression screws different sizes
4.5mm and 6.5mm screw box with commonly used
screw sizes.

IMPLANTS FOR CANNULATED SCREW WITH
GUIDE WIRE DRILL BITS, SCREW DRIVER AND
TAPS FOR INSERTION.

INSTRUMENT SET FOR FIXATION FOR 7 MM
CANNULATED SCREWS
Threaded guide pins
Cannulated reamers
Cannulated tap
Cannula for reamers
Parallel pin guide
Depth guage (direct measuring device)
Quick coupling handle
Cannulated screw driver
Screw extractor with handle
Sterilization and storage box for implants and
instruments 50,55,60,65,70,75,80,85mm etc

      4. System Configuration Accessories, spare and consumables

Sl                         Name                     Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.
              a. As specified

      5. Environmental factors
Sl                      Name                        Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.

              5.1 1. The unit shall be capable of being stored continuously in ambient
                  temperature of 0-50 deg C and relative humidity of 15 -90%.
        2. The unit shall be capable of operating continuously in ambient temperature of 10-40
        deg C and relative humidity of 15-90%.

        Parameters:
      6. Power Supply
Sl                         Name                     Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.
None
   7. Standards, safety and training
Sl                    Name                          Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.

7.5      Company/ supplier Should have a license for manufacture by Drug Controller India and
      conform to the laid down standards by them in this regard
7.6    Material should be of implant grade steel, carbon fibre, titanium as required
7.7    Comprehensive training for lab staff and support services till familiarity with the system on
      site
7.8    Long shelf life.
7.9   Comprehensive warranty for 2 years and 5 years AMC after warranty.



        .
Dr. Deepak Mittal                                              Dr. V.K. Sharma
      8. Documentation
Sl                        Name                       Technical Specs          Bidders deviation if
                                                     Quoted by the bidder. any.
8.2    1. User/Technical / Maintenance manuals to be supplied in English.
       2. Certificate of calibration and inspection.
       3. list of equipments available for providing calibration and routine preventive
       maintenance support as manufacturer service/maintenance manual.
       4. List of important spare parts and accessories with their part number and costing.
       5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
       checklist. The job description is clearly spelt out.
       6. Compliance report to be submitted in a tabulated and point wise manner clearly
       mentioning the page/para/number with authenticated catalogue/manual without which it
       will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
                   Specifications for RING EXTERNAL FIXATOR SYSTEM

                   UNSPSCCode:
                   ECRI Code:
1 Description of Function
SI         Name                                                        Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted by if any
                                                                       bidder

     1.1 These instruments are used in orthopedic surgery for fixing
comminuted fractures, treating gap fracture nonunions, limb length
discrepancy etc.
     2 Operational Requirements
3 SI        Name                                                        Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted by if any
                                                                        bidder
     2.1    As specified

     Technical Specifications
     SI     Name                                                        Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted by if any
                                                                        bidder

     3.1 SHOULD BE MADE FOR IMPLANT GRADE STAINLESS STEEL/carbon fibre as specified by
     user. Some sizes and quantity to vary with user demand.


            DESCRIPTION                                  QTY

            STAINLESS STELL 1/2 RINGS FOR 20 NUTS
            80
            100
            120
            140
            160
            180
            200
            220
            240

            CARBON ½ RINGS 60-180 MM WITH 10MM INCREMENT


            THREADED ROD

            4"

Dr. Deepak Mittal                                           Dr. V.K. Sharma
        6"
        8"
        10"
        12"
        14"
        16"
        18"
        NUT
OLIVE WIRE


        1.5MM

        1.8MM

KWIRE BAYONET TIPS

        1.5MM

        1.8MM

WIRE FIXATION BOLTS

CANNULATED

SLOTTED

RING FIXATION BOLT

15MM

20MM

25MM

ITALIAN ARCH ASSORTED

1400

1600

1800

2000

OBLIQUE SUPPORT

FEMALE POSTS

2HOLE

Dr. Deepak Mittal       Dr. V.K. Sharma
3HOLE

4HOLE

MALE POSTS

2HOLE

3HOLE

4HOLE

MALE HINGES

FEMALE HINGES

SHORT CONNECTION PLATE

5   HOLE

6   HOLE

7   HOLE

8   HOLE

9   HOLE

10 HOLE

11 HOLE

12 HOLE

TWISTED PLATES

1 HOLE

2 HOLE

3 HOLE

4 HOLE

THREADED SOCKETS

WASHERS

SINGLE PIN FIXATOR CENTRAL SLOT

SCHANZ PIN

SPANNER 10-11 MM 6

BOX SPANNER FIX 4

BOX SPANNER MOBILE 4

WIRE CUTTING PLIER 2

Dr. Deepak Mittal                 Dr. V.K. Sharma
DYNAMOMETRIC TENSIONER 2

DYNAMEWIRE TENSIONER 2




      4. System Configuration Accessories, spare and consumables

Sl                        Name                     Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.
             a. As specified

      5. Environmental factors
Sl                      Name                       Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

             5.1 1. The unit shall be capable of being stored continuously in ambient
             temperature of 0-50 deg C and relative humidity of 15 -90%.
       2. The unit shall be capable of operating continuously in ambient temperature of 10-40
       deg C and relative humidity of 15-90%.

        Parameters:
      6. Power Supply
Sl                        Name                     Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.
NONE
   7. Standards, safety and training
Sl                    Name                         Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

7.1 Company/ supplier Should have a license for manufacture by Drug Controller India and
conform to the laid down standards by them in this regard .
7.2 Material should be of implant grade steel, carbon fibre, titanium as required.
7.3 Comprehensive training for lab staff and support services till familiarity with the system on
site.
7.4 Comprehensive warranty for 2 years and 5 years AMC after warranty.


      8. Documentation
Sl                        Name                       Technical Specs          Bidders deviation if
                                                     Quoted by the bidder. any.
8.3    1. User/Technical / Maintenance manuals to be supplied in English.
       2. Certificate of calibration and inspection.
       3. list of equipments available for providing calibration and routine preventive
       maintenance support as manufacturer service/maintenance manual.
       4. List of important spare parts and accessories with their part number and costing.
       5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
       checklist. The job description is clearly spelt out.
       6. Compliance report to be submitted in a tabulated and point wise manner clearly
       mentioning the page/oara/number with authenticated catalogue/manual without which it
       will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
                Equipment Specifications for Plaster Cutting Saw

                UNSPSC Code:
                ECRI Code:
1 Description of Function
SI         Name                                                                    Technical Bidders
                                                                                   Specs     Deviation
                                                                                   quoted    if any
                                                                                   by bidder

     1.1 These instruments are used in orthopedic surgery for cutting of
plaster of paris/synthetic cast applied for fractured bones.
     2 Operational Requirements
SI         Name                                                                    Technical Bidders
                                                                                   Specs     Deviation
                                                                                   quoted    if any
                                                                                   by bidder

     2.1 Should be of high quality metal for cutting Plaster of paris or
synthetic cast
     2.2    Should have an oscillating mode which is not damaging to
human tissues of appropriate voltage and oscillations.
     3 Technical Specifications
SI         Name                                                                    Technical Bidders
                                                                                   Specs     Deviation
                                                                                   quoted by if any
                                                                                   bidder
     3.1    SPECIFICATION FOR PLASTER CUTTING SAW with/without cast vacuum system.

            1. LIGHT WEIGHT & HEAVY DUTY; pistol type/handy grip
            2. SHOCK PROOF FIBER GLASS /metal BODY
            3. w o rk i n I n di a n e l ec t ric a l c o n d i ti o n s 2 2 0 V e tc .
            4. CAPABLE OF CUTTING POP & FIBRE GLASS (
            SYNETHETIC)
            5. LONG LIFE TEFLON/OTHER MATERIAL COATED BLADE/others
            6. Blade Size 63 mm Dia etc Qty. 10 Nos.

4 System Configuration Accessories, spares and consumables
SI         Name                                                                    Technical Bidders
                                                                                   Specs     Deviation
                                                                                   quoted    if any
                                                                                   by bidder
4.1        As specified necessary one plaster spreader, shear etc.

5 Environmental factors


Dr. Deepak Mittal                                                       Dr. V.K. Sharma
SI    Name                                                             Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted by if any
                                                                       bidder


5.11. The unit shall be capable of being stored continuously in ambient temperature of 0-50 deg
C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

   Parameters:
6 Power Supply
Sl                       Name                   Technical Specs             Bidders deviation if
                                                Quoted by the bidder.       any.
6.1 Power input to be 220-240 V AC, 50 Hz fitted with Indian plug.

7 Standards, safety and training
Sl                    Name                        Technical Specs           Bidders deviation if
                                                  Quoted by the bidder.     any.

7.1 1. Company/ supplier Should have a CE, FDA certification .
     2. Material should be of implant grade steel, , titanium as required.
     3. Comprehensive training for lab staff and support services till familiarity with the system
     on site.
     4. Comprehensive warranty for 2 years and 5 years AMC after warranty.

7.2 Certified to be compliant with IEC 60601 for Medical Electrical Equipments or BEE.

8 Documentation
Sl                       Name                      Technical Specs          Bidders deviation if
                                                   Quoted by the bidder. any.
8.1 1. User/Technical / Maintenance manuals to be supplied in English.
     2. Certificate of calibration and inspection.
     3. list of equipments available for providing calibration and routine preventive
     maintenance support as manufacturer service/maintenance manual.
     4. List of important spare parts and accessories with their part number and costing.
     5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
     checklist. The job description is clearly spelt out.
     6. Compliance report to be submitted in a tabulated and point wise manner clearly
     mentioning the page/number with authenticated catalogue/manual without which it will not
     be considered.




Dr. Deepak Mittal                                            Dr. V.K. Sharma
      Equipment Specifications for Pedicle Fixation and Reduction
      &cervical plating system


UNSPSCCode:
ECRI Code:
1 Description of Function
SI     Name                                                          Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder

     1.1 These instruments are used in orthopedic surgery for fixation of unstable spine due
to infection, trauma, tumour,degenerative/inflammatory disease .
     2 Operational Requirements
SI     Name                                                          Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted by if any
                                                                     bidder

     2.1 As specified

 3 Technical Specifications
SI     Name                                                         Technical Bidders
                                                                    Specs     Deviation
                                                                    quoted by if any
                                                                    bidder

     3.1 quantity to vary with user demand and size variation may occur where
     appropriate To human anatomy.
     Broad Specifications
          Titanium Mono and Multiaxial Pedicle top loading, top
          tightening 4.5/5.5/6.5mmscrew fixation system for
          stabilization / fusion, dislocation, fracture, degenerative &
          Deformity of Anterior/posterior thoracic, lumbar and/or sacral
          spine.

        Specifications
        • System should be of Titanium Alloy and MRI Compatible.
        • Should be universal system for anterior and posterior
        fusion.
        • System should have top loading Multiaxial and monoaxial
        screws
        • Long length 5.5mm rod of atleast 50cm.
        • Should have single closure top tightening Titanium Break
        off plugs .
        • Should have Adult, Pediatric,laminar , Transverse Process
        ,Extended Body,Left and right offset hooks
        • Titanium Multiaxial Reduction pedicle screws
        (5.5mm/6.5mm).
        • Should have low profile monoaxial and multiaxial Screws
        in 4.5/5.5,6.5mm Diameter and from 25mm-60mm in
Dr. Deepak Mittal                                          Dr. V.K. Sharma
         length.
         • Should have anterior staples for 13/15/17mm
         • Should have titanium mesh cages for the cervical
         ,thoracic , lumbar and interbody fusion.

         A) Cervical cages 10X30, 13X70,10X100,10X8MM
         B) Thoracic cages 13X30, 16X30,16X60mm

         C) Lumbar cages 19X40, 19X90,13X10mm.

         D) Cervical plating system plate should be of
            various shapes with variable angle/locking
            mechanism screws MRI compatible . complete
            with of various hole length as required by
            user.guarded oscillating drill bits, self tapping
            screws, screw drivers etc.




     4. System Configuration Accessories, spare and consumables

Sl                       Name                      Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.
            b. As specified

     5. Environmental factors
Sl                      Name                       Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

            5.1 1. The unit shall be capable of being stored continuously in ambient
                temperature of 0-50 deg C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

       Parameters:
     6. Power Supply
Sl                       Name                      Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

     7. Standards, safety and training
Sl                      Name                       Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

7.1 Company/ supplier Should have CE, FDA approved certification.
7.2 Material should be of implant grade steel, carbon fibre, titanium as required.
7.3 Comprehensive training for lab staff and support services till familiarity with the system on
site.
7.4 long shelf life.


     8. Documentation
Sl                       Name                       Technical Specs           Bidders deviation if
                                                    Quoted by the bidder. any.
8.4 1. User/Technical / Maintenance manuals to be supplied in English.
      2. Certificate of calibration and inspection.
      3. list of equipments available for providing calibration and routine preventive maintenance
      support as manufacturer service/maintenance manual.
      4. List of important spare parts and accessories with their part number and costing.
Dr. Deepak Mittal                                              Dr. V.K. Sharma
     5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
     checklist. The job description is clearly spelt out.
     6. Compliance report to be submitted in a tabulated and point wise manner clearly
     mentioning the page/para/number with authenticated catalogue/manual without which it
     will not be considered.




Dr. Deepak Mittal                                       Dr. V.K. Sharma
           Equipment Specifications for MODULAR TOTAL KNEE AND
           HIP REPLACEMENT

UNSPSC Code:
ECRI Code:
1 Description of Function
SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

     1.1 These instruments are used in orthopedic surgery for providing patients with physical

destruction of joints due to some pathology with an artificial painless,mobile, stable joint.

     2 Operational Requirements

SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted by if any
                                                                       bidder

     2.1 As specified

     3 Technical Specifications
SI     Name                                                           Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted by if any
                                                                      bidder

     3.1 Following are the minimum essentials sizes and configuration,quantity may vary
         With the user demand.& with the aim of convenient and productive knee
     replacement.
     Description                                   Qty.
          Intramedullary Drill w/Step 1
          IM Femoral A/P Sizing Guide 1
          Universal Handle Peg Driver 1
          Intramedullary Alignment Guide 9 & 4 Inches 2
          Anterior Femoral Cutting Guide 1x2 2
          Rotational Alignment Guide 1
          IM Distal Femoral Cutting Guide/single multicut zig etc. 1
          IM Femoral A/P Measuring Guide 1
          IM Femoral Finishing Guide of various sizes 5
          Trochlear Recess Finishing Guide of various sizes 5
          Notch/Chamfer Guide of various sizes 5
          Allignment Rod, Allignment rod With Coupler 2
          Extramedullary Alignment Arch 1
          Femoral Trial Left and Right of various sizes 10
          Universal Femoral File 1
          IM Femoral Impactor 1
          Spacer/Align Guide of various sizes 7
          Femoral Provisional Extractor 1
          T. Handle with Chuk 1
          Extramedullary Tibial Cutting Guide 2
          Universal Ligament Retract Spring 1
Dr. Deepak Mittal                                            Dr. V.K. Sharma
        Tibial Depth Resection Gauge 1
        Femoral Recutter 1
        Stemmed Tibial Sizing Plates of various sizes 4
        Tibial Provisional/ Holding Clamp 1

        Cemented stemmed drill guide one
        Cemented stemmed tibial drill
        Stemmed tibial broaches of barious sizes
        Tibial provisional extractor
        Tibial impactor
        Stemmed tibial provisional impactor
        Short spring screws pins
        Grooved shout head holding pins four
        Holding pins five
        Headless holding pin three
        Hexhead holding pins three
        Tibial resection guide one
        Articular surface insertion instrument one
        Articular surface removal instrument one
        Female hex driver/ extractor one
        Patella tendon retractor one
        Hex head screwdriver one
        Holding pin plier one
        Patellar clamp two
        Patellar drill guides one
        Patellar saw guide one
         Patellar/femoral drill one
        Patella trials of various sizes three
        Female hexa screw driver one
        Varus/valgus tibial recutter one
        LPS Art. Surface provisionsl locking screw one
        Stemmed tibial plate provisional of various sizes four
        Townley femur caliper two
        Tibial retractor one
         slap hammer one
        bone screw drill 3.2mm one
        recutter 2mm one
        A/Surface trial of various sizes ten


        CONSTRAINED CONDYLAR KNEE

        Hex head holding pin eight
        Torque wrench one
        Torque wrench screw driver one
        Augment holding clamp head one
        Augment holding clamps one
        Rotation alignment guide one
        Standard cut with screw one
        Revision plus 3mm cut and screw one
        Revision standard cut and screw one
        Torque lining femoral set screw driver one
        Placement guide one
        0deg. Distal resection one
        Distal cutting guide one
        Grooved short head holding pin four
        Headless pin four
Dr. Deepak Mittal                                          Dr. V.K. Sharma
        Revision IM Guide one
        Femoral sizing temp. of various sizes four
        Notch champher guide of various sizes five
        Femoral alignment screw driver one
        Offset stem lock nut one
        Offset stem wrench one
        Posterior saw blade attachment one
        Femoral stem drill 18mm one
        Stem prov. Adapter one
        Femoral base guide flange one
        Offset bushing one
         Drill guide bushing five
        Tibial plates extn. Wrenches three
        Bushing one
        Femoral tibial lt. and right of various sizes eight
        Posterior femoral retractor two
        I.M Hole locator one
        Modular box prov. Five
        (LPS/) Art. Surf prov.locking screw one
        Femoral alignment prov screw short four
        Art. Sur. Trial of various sizes eight

        IMPLANTS:
        Femoral component various sizes to fit various anatomy
        Tibial component various sizes to fit various anatomy
        Articulating surface component various sizes to fit various anatomy
        Flexion extension gap

        Patella component component various sizes to fit various anatomy

        TOTAL HIP : CONFIGURATION, SIZE AND QUANTITY MAY VARY BUT SHOULD
        ACHIEVE THE ULTIMATE AIM OF CONVINIENT AND PRODUCTIVE TOTAL HIP
        REPLACEMENT
        CEMENTED CUP INSTRUMENTS

        Osteotomy template one
        Box osteotome one
        Taper reamer one
        Trochanteric reamer one
        straight reamer9mm one
        rasp of various sizes four
        rasp handle one
        cone provisional one
        allen medullary plug of various sizes four
        allen plug introducer one
        cement restrictor plate and seal two
        femoral pressuriser plate and seal two
        stem impactor one
        stem extractor one
        head trials 22/28/36 mmX S2
        head trials 22/28/36 mmX M2
        head trials 22/28/36 mmX L2
        HEAD impactor one
        Acetabular reamer
        Shaft for acetabular reamer one
        Coupling T handle one
        Reamer adapter on
Dr. Deepak Mittal                                             Dr. V.K. Sharma
        Acetabular cup trials 43,45,47,49,51,53,55,57,59
        Cup pusher cum positioner one
        Alignment rod one
        Pusher rod with plastic head 22 and 28/36 mm two
        Acetabualar preparation drill with stop 9,11,13 mm

         UNCEMENTED TOTAL HIP INSTRUMENTS
         Vacuum liner inserter 28 or 22 mm
         Mechanical liner inserter 22mm or 28mm
         Liner/ shell disassembly instrument
         Liner extractor
         Gun sight alignment guide
         Alignment rod
         Drill guide
         Screw driver universal
         Screw driver straight
         Modular universal handle
         Screw forceps 150
         Flexible depth guage
         Screw for left forceps 450
Frame alignment guide
Cup impactor –positioner
Liner provisional 46X (22or 28/36mm)
Liner provisional 48X (22or 28/36mm)
Liner provisional 50,52,54,56X (22or 28mm)
Shell provisionals 46mm
Shell provisionals 48mm
Shell provisionals 50mm
Shell provisionals 52mm
Shell provisionals 54mm
Shell provisionals 56mm
Flexible drill bit 15mm, 30mm and 45mm
Impacting plate for shell for various sizes
Acetabular cup pusher
Impactor lock nut
Acetabular cup remover
Acetabular reamer 40mm or 42mm
Acetabular reamer 44mm
Acetabular reamer 46mm
Acetabular reamer 48mm,50,52,54and 56mm
Shaft for acetabular reamer
Coupling handle for shaft
Nylon face impactor

HYBRID :CEMENTED STEM AND BIPOLAR CUP
Osteotomy template one
Box osteotome one
Taper reamer one
Trochanteric reamer one
straight reamer 9mm one
coupling t handle one
rasp of various sizes four
rasp handle one
cone provisional one
allen medullary plug of various sizes four
allen plug introducer one
cement restrictor plate and seal two
Dr. Deepak Mittal                                      Dr. V.K. Sharma
femoral pressuriser plate and seal two
stem impactor one
stem extractor one
head trials 22/28/36 mmX S2
head trials 22/28/36 mmX M2
head trials 22/28/36 mmX L2
HEAD impactor one
Liner trial cup 39to 43 mm one
Liner trial cup 45 to 47 mm one
Liner trial cup 51to 45 mm one
Shell trial cups 39,41,43,45,47,49,51,53,55,57
Shell trial holding rod one
Forceps one
Reamer adapter
Judet head extractor one
Shell, liner head may be separate or in various configurations/combination for bipolar
hemiarthroplasty.

UNCEMENTED THR
Osteotomy guide 9-10 mm, 11,12-13,14-15 four
Box osteotome small one
Tapered awl one
T handle with chuck one
Trochanteric reamer one
IM Reamer 9 to 15mm –half mm difference     thirteen
rasp 9,10 mm, 11,12,13,14,15mm seven
rasp alignment tip 9,10,11,12,13,14,15  seven
rasp tip wrench one
calcar planer one
cone provisional 09/10,11,12/13,14/15 four
femoral HEAD provisional 22mm 3
femoral head provisional 28mm /36mm 5
rasp handle one
alignment rod one
stem impactor one
stem driver/inserter adaptor one
extractor hammer one
head impactor one
tommy bar one
various configurations/combination of above to satisfy the ultimate aim of good
uncemented THR though above are the general specefications.

     4. System Configuration Accessories, spare and consumables

Sl                      Name                     Technical Specs         Bidders deviation if
                                                 Quoted by the bidder.   any.
            a. As specified

     5. Environmental factors
Sl                      Name                     Technical Specs         Bidders deviation if
                                                 Quoted by the bidder.   any.

            5.1 1. The unit shall be capable of being stored continuously in ambient
                temperature of 0-50 deg C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

Dr. Deepak Mittal                                          Dr. V.K. Sharma
       Parameters:
     6. Power Supply
Sl                        Name                      Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.
None
   7. Standards, safety and training
Sl                    Name                          Technical Specs           Bidders deviation if
                                                    Quoted by the bidder.     any.

7.10       1. Company/ supplier Should have a CE, FDA approved ceritfication.
       2. Material should be of implant grade of long life like stainless steel,etc. as appropriate.
       3. Comprehensive training for lab staff and support services till familiarity with the system
       on site.
       4. Comprehensive warranty for 2 years and 5 years AMC after warranty.


     8. Documentation
Sl                        Name                       Technical Specs           Bidders deviation if
                                                     Quoted by the bidder. any.
               8.1 1. User/Technical / Maintenance manuals to be supplied in English.
       2. Certificate of calibration and inspection.
       3. list of equipments available for providing calibration and routine preventive maintenance
       support as manufacturer service/maintenance manual.
       4. List of important spare parts and accessories with their part number and costing.
       5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
       checklist. The job description is clearly spelt out.
       6. Compliance report to be submitted in a tabulated and point wise manner clearly
       mentioning the page/para/number with authenticated catalogue/manual without which it
       will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
             Equipment Specifications for MINI INSTRUMENT SET

             UNSPSC Code:
             ECRI Code:
1 Description of Function
SI         Name                                                       Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted by if any
                                                                      bidder

  1.1 These instruments are used in orthopedic/general surgery for

fixation of fractures of small bones of hand and feet.

     2 Operational Requirements
SI         Name                                                       Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

  2.1 As specified

3 Technical Specifications
SI         Name                                                       Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder
     3.1 S..NO. FOLLOWING ARE THE MINIMUM ESSENTIAL, QUANTITY AND
         CONFIGURATION MAY BE AS PER USER DEMAND
     3.2 Description                               Qty.
         1. Drill bit 1.5mm dia for quick coupling 3
         2. Drill bit 2.0mm dia for quick coupling 3

            3. Countersink Shaft 2.7 1
            4. Countersink Shaft 1.5/2.0 mm Screws 1
            5. Tap for 1.5 mm Cortex Screws
            6. Tap for 2.00mm Cortex Screws
            7. Tap for 2.7mm to Cortex Screws
            8. Depth Gauge for 1.5mm to 2.0 mm Screws
            9. Depth Gauge for 2.7 mm to 4.00 mm Screws
            10. Double Drill guide 1.5/1.1
            11. Double Drill guide 2.0/1.5
            12 Double Drill guide 2.7/2.0
            13. Self compression drill guide for 1.5mm drill & 2 mm
            screws
            14. Drill guide for 2 mm drill, 2.7 mm screws & self
            compression plates
            15. Bending iron for 1.5 & 2mm plates
            16. Bone holding forceps for Mini Fragment length 155mm
            17. Holding forceps for mini plates with ratchet lock
            18. Reduction forceps with points 125mm
            19. Reduction forceps with serrated jaws length 140 mm
            20. Hohmann retractor 150 X 7mm
Dr. Deepak Mittal                                           Dr. V.K. Sharma
       21.   Hohmann retractor 160 X 6 X 3mm
       22.   Hohmann retractor 160 X 8 X 3mm
       23.   Periosteal elevator 3 mm
       24.   Periosteal elevator 6mm '




Dr. Deepak Mittal                              Dr. V.K. Sharma
           Equipment Specifications for MINI FRAGMENT IMPLANTS

           UNSPSCCode:
           ECRI Code:
1 Description of Function
SI         Name                                                            Technical Bidders
                                                                           Specs     Deviation
                                                                           quoted by if any
                                                                           bidder
     1.1 A implant may be defined as an artificial organic material that can be surgically
         implanted into a person's body to replace damaged tissue. This implant is for fixation
         of small bone s of hand and feet.
2 Operational Requirements
SI         Name                                                            Technical Bidders
                                                                           Specs     Deviation
                                                                           quoted by if any
                                                                           bidder

     2.1 As specified
     3 Technical Specifications
SI         Name                                                            Technical Bidders
                                                                           Specs     Deviation
                                                                           quoted by if any
                                                                           bidder
     3.1    Mini plates 2Mm for the fixation of fracture of hand foot
            Straight place 3,4,5,6, holed - 2 Each

            Straight place 20 hold ( may be cut to the desired length)3.
            nos.
            T- Plates 2/2 Holes 3 Nos.
            L-Plates 2/2 Holes ( RT & Left) 2 each
            L-plates 2/2 Holes ( 110 degree RT & LT ) 2 each
            2 MM cortical screws 10 mm to 20mm with 2mm
            increments 20 each


            1.5 MM screws & plates for 1.5 mm screws for the fixation
            of articular fractures of phalanx

            Screws 8 mm to 16mm with 1 mm increment plates - 12
            each.
            Straight plate 20 holed may be cut to the desired ienth -
            2nos.
            T-plates 3 holes in head length 50 mm ( may be cut to
            desired Length )- 3 nos.
            Cercelage wire with eye length 280mm DIA 08.mm- - 10

4. System                     Accessories, spares and consumables
Configuration

Dr. Deepak Mittal                                               Dr. V.K. Sharma
SI    Name                                                              Technical Bidders

            As specified

     5.Environmental factors
Sl                       Name                      Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

            c.  1. The unit shall be capable of being stored continuously in ambient
                temperature of 0-50 deg C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

     Parameters:
6.Power Supply
Sl                         Name                    Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.
None
7. Standards, safety and training
Sl                      Name                       Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

7.11      1. Should be manufactured by company which has license of Drug Controller of India
    and conforms to standards laid down by the Drug Controller of India vide notification of
    Ministry of Health & family welfare .
      2. Material should be of implant grade steel, titanium as required.
      3. Comprehensive training for lab staff and support services till familiarity with the system
      on site.
      4. Comprehensive warranty for 2 years and 5 years AMC after warranty. AND LONG
      SHELF LIFE.


     8. Documentation
Sl                         Name                    Technical Specs           Bidders deviation if
                                                   Quoted by the bidder. any.
8.1 1. User/Technical / Maintenance manuals to be supplied in English.
     2. Certificate of calibration and inspection.
     3. list of equipments available for providing calibration and routine preventive
     maintenance support as manufacturer service/maintenance manual.
     4. List of important spare parts and accessories with their part number and costing.
     5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
     checklist. The job description is clearly spelt out.
     6. Compliance report to be submitted in a tabulated and point wise manner clearly
     mentioning the page/para/number with authenticated catalogue/manual without which it
     will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
Equipment Specifications for MANUAL and battery CANNULATED
HAND DRILL

UNSPSCCode:
ECRICode:
1 Description of Function
SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder
     1.1 Drilling machines are used in a number of orthopaedic
         surgical procedures, for example, in making holes in bones
         for bone screws,in drilling out the medulla or marrow
         areas of bones and sawing of bones.
2 Operational Requirements
SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted by if any
                                                                       bidder

     2.1 As specified
     3 Technical
Specifications
SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted by if any
                                                                       bidder


     3.1 Ergonomic pistol designed light weight of durable and long lasting material.

     3.2 Fully Cannulated at least to 3.2mm

     3.3 Should have versatile attachments for reaming, drilling, ,sawing, and inserting K wires
         ETC AND ANY ORTHOPAEDIC PROCEDURE AS APPROPRIATE.

     3.4 Should have power, speed range of at least of 0-4000rpm and torque range of atleast 0-
         6N adequate for all orthopaedic procedures of drilling , reaming and sawing (both
         oscillating and reverse).

     3.5 Battery of Ni-Cd or any good material of long life/DC voltage 12-15/ .5-.6ampere hours
         and good short charging time. Should have long running time post charging. Provided
         with battery charger to BIS standards

     3.6 Should be chuckless, and have instant start/ stop trigger option, variable speed/torque ,
         oscillating/forward –reverse drilling switch mechanism on the hand piece.

     3.7 Should be bio compatible and not harm tissues.
Dr. Deepak Mittal                                             Dr. V.K. Sharma
     3.2 FOR HAND DRILL : Chuck in hand drill should be such
         which can be tightened without the
         Key.
     3.3 Tiltable Aluminum Handle.for hand drill
     3.4 Drill (and battery) should be autoclavable.
4
     System Configuration Accessories, spares and consumables
     SI     Name                                                              Technical Bidders
                                                                              Specs     Deviation
                                                                              quoted by if any
                                                                              bidder

     4.1 All standard accessories that is 4 batteries with charger,
         flexible high quality, low pressure deep fluted sharp
         CUTTING intramedullary reamers of 6-14mm, acetabular
         reamers of 39-65mm sizes , and reciprocating/ oscillating
         saw blade of sizes desired by user.quantity may vary.
     4.2 All versatile attachment of drills, reamers –flexible/ dynamic
         hip screw-triple reamer, K-wire and saw desired for proper
         functioning of the machine
           .

5 Environmental factors
SI        Name                                                              Technical Bidders
                                                                            Specs     Deviation
                                                                            quoted    if any
                                                                            by bidder
5.1     1.The unit shall be capable of being stored continuously in in ambient temperature of
0-50 deg C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

        Parameters:
      6. Power Supply
Sl                           Name                Technical Specs        Bidders deviation if
                                                 Quoted by the bidder. any.
6.1Should conform to national Electrical standards Should be compliant with IEC 61010-1:(or
any international equivalent eg EN/UL 61010) covering safety requirements for electrical
equipment for measurement control and laboratory use.
6.2 Should be chargeable on Indian condition power supply.
    7. Standards, safety and training
Sl                       Name                  Technical Specs                   Bidders deviation if
                                               Quoted by the bidder.             any.

7.1       1 . Should have a CE, FDA approved ceritfication.
          2. Material should be of stainless steel in case of hand drill and of high quality material
          for battery drill as appropriate and fully sterlizable.
          3. Comprehensive training for lab staff and support services till familiarity with the system
          on site.
          4. Comprehensive warranty for 2 years with spares and 5 years AMC after warranty.


      8.Documentation
Sl                           Name                      Technical Specs           Bidders deviation if
Dr. Deepak Mittal                                                 Dr. V.K. Sharma
                                                   Quoted by the bidder. any.
8.1 1. User/Technical / Maintenance manuals to be supplied in English.
     2. Certificate of calibration and inspection.
     3. list of equipments available for providing calibration and routine preventive
     maintenance support as manufacturer service/maintenance manual.
     4. List of important spare parts and accessories with their part number and costing.
     5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
     checklist. The job description is clearly spelt out.
     6. Compliance report to be submitted in a tabulated and point wise manner clearly
     mentioning the page/para/number with authenticated catalogue/manual without which it
     will not be considered.




Dr. Deepak Mittal                                       Dr. V.K. Sharma
       Equipment Specifications for LARGE FRAGMENT FIXATION
      IMPLANTS 4.5 MM and locking compression plate 5mm/4.5mm.

UNSPSCCode:
ECRI Code:
1 Description of Function
SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder
     1.1 A implant may be defined as an artificial organic material that can be surgically
         implanted into a person's body to replace damaged tissue.This is used for fixation of
         large bones of arm , thigh etc.

2 Operational Requirements
SI     Name                                                            Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted by if any
                                                                       bidder
2.1    The surgical instruments should not get rust after repetitive
       routine autoclaving & should last long.
2.2    Thread of screw and locking thread mechanism should be of
       high quality. Locking screw mechanism should be angularly
       stable.




3 Technical Specifications
SI     Name                                                            Technical Bidders




Dr. Deepak Mittal                                           Dr. V.K. Sharma
.1     4.5/5.0 Broad/NARROW LCP FOR LONG BONES should be of low               Specs     Deviation
       profile. To support and facilitate minimal invasive surgery the LCP    quoted by if any
       4.5/5.0 Broad Curved may be designed with TAPERED nose on              bidder
       either side. TheRE SHOULD BE small holes in the nose allow for
       primary fixation with K-Wires.
3.2    Locking PLATE SHOULD HAVE a COMBI HOLE OF LOCKING
       AND DYNAMIC COMPRESSION UNIT AND UNDERSURFACE
       SHOULD BE UNDERCUT.
       The LOCKING Hole should accept a 5MM locking threaded head
       SCREW OF AND DYNAMIC HOLE SHOULD ACCEPT 4.5MM
       SCREW.
       Screws in locking plate should be fully recessed in the plate.
3.3
       Locking screw should be self tapping.
3.4
3.5    Distal third tibial narrow limited compression plate should be
       anatomically precontoured.
3.6     Distal femoral should have a expanded distal end anatomically
       contoured /designed with varied arrangement of multiple locking only
       round holes.
        Proximal tibial plate should be hockey shaped; proximal end
       expanded & anatomically contoured /designed with varied
       arrangement of multiple locking only round holes.
       The shaft of these plates should have combi 4.5/5mm
       locking/dynamic compression holes.


     3,1   LARGE FRAGMENT FIXATION IMPLANTS 4.5MM; 5/4.5mm LCP

           LC-D.C/ LCP. NARROW PLATES 4.5MM , 5 HOLE .
           LC-D.C/ LCP. . NARROW PLATES 4.5MM , HOLE .
           LC-D.C/ LCP. . NARROW PLATES 4.5MM 7 HOLE,8,9,,10,12
           HOLE .
           Distal and proximal tibial ;and femoral LCP from 8 to 14 holes.
           4.5 MM CORTICAL SCREWS 20MM TO 48MM
           WITH INCREMENT OF 2MM - 15 each
           4.5MM CORTCAL SCREWS 50 TO 70 MM WITH
           5MM INCREMENT - 10 each
           5mm locking cortical and 6.5mm/ cancellous screw from 24 to 70mm.

4 System Configuration Accessories, spares and consumables
SI     Name                                                                   Technical Bidders
                                                                              Specs     Deviation
                                                                              quoted by if any
                                                                              bidder

     4.1 As specified

     5 Environmental factors
SI     Name                                                                   Technical Bidders
                                                                              Specs     Deviation

5.1     1.The unit shall be capable of being stored continuously in in ambient temperature of
0-50 deg C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

       Parameters:

Dr. Deepak Mittal                                                Dr. V.K. Sharma
     7. Power Supply
Sl                       Name                      Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.
None
   7. Standards, safety and training
Sl                      Name                       Technical Specs          Bidders deviation if
                                                   Quoted by the bidder.    any.

7.1 1. Should be manufactured by company which has license of Drug Controller of India and
conforms to standards laid down by the Drug Controller of India vide notification of Ministry of
Health & family welfare.
      2. Material should be of implant grade steel, titanium as required.
      3. Comprehensive training for lab staff and support services till familiarity with the system
      on site.
      4. Comprehensive warranty for 5 years AMC after warranty of instruments and long shelf
      life of implants.


     8.Documentation
Sl                       Name                      Technical Specs           Bidders deviation if
                                                   Quoted by the bidder. any.
8.1 1. User/Technical / Maintenance manuals to be supplied in English.
     2. Certificate of calibration and inspection.
     3. list of equipments available for providing calibration and routine preventive
     maintenance support as manufacturer service/maintenance manual.
     4. List of important spare parts and accessories with their part number and costing.
     5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
     checklist. The job description is clearly spelt out.
     6. Compliance report to be submitted in a tabulated and point wise manner clearly
     mentioning the page/para/number with authenticated catalogue/manual without which it
     will not be considered.




Dr. Deepak Mittal                                            Dr. V.K. Sharma
     Equipment Specifications for K NAIL INSTRUMENT SET WITH
                                          IMPLANTS

UNSPSCCode:
ECRICode:
1 Description of Function
SI     Name                                                           Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

     1.1 These instruments are used in orthopedic surgery for fixation of fracture of thigh/long
bones.
     2 Operational Requirements
SI    Name                                                            Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted by if any
                                                                      bidder
      None
Technical Specifications

SI     Name                                                           Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted by if any
                                                                      bidder

     3.1 Instrument Set should consist of the following,
         quantity may vary with the user demand
         following are the general guidelines:
         . Kuntscher's diamond pointed awl
       Metal Scale
              o 'K' nail extractor with 3 hooks (Different Sizes)- Three
                 Nos. Bone hammer
              o Kocher's bone hook- Two No 'K' nail impactor nail
                 setter- Four Nos.

             o   'K' nail reamer for femur (different sizes)
                 cannulated/noncannulated 6-13 with I mm increment

             o   Guide wire for 'K' nail Femur - Two Nos.

             o   Bone holding forceps - Four Nos. ( For
                 reduction)

             o   Periosteal Elevator Medium (Two)

             o   Large (Two) Box for instruments and implants
                 for sterilization and Storage
             o   Sizes ( 36,37, 38 cm) X 8,9,10,11 - 10 Each (120)
                 Sizes ( 39, 40, 41, 42 Cm) x 9,10, 11,12 - 10 Each
                 (160)
4 System Configuration Accessories, spares and consumables
Dr. Deepak Mittal                                            Dr. V.K. Sharma
SI     Name                                                             Technical Bidders
                                                                        Specs     Deviation
                                                                        quoted by if any
                                                                        bidder
4.1 As specified


     5 Environmental factors
SI     Name                                                             Technical Bidders
                                                                        Specs     Deviation

5.1     1.The unit shall be capable of being stored continuously in in ambient temperature of
0-50 deg C and relative humidity of 15 -90%.
      2. The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 15-90%.

        Parameters:
      8. Power Supply
Sl                       Name                      Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.
None
   7. Standards, safety and training
Sl                      Name                       Technical Specs           Bidders deviation if
                                                   Quoted by the bidder.     any.

7.1 1. Should be manufactured by company which has license of Drug Controller of India and
conforms to standards laid down by the Drug Controller of India vide notification of Ministry of
Health & family welfare.
      2. Material should be of implant grade steel, titanium as required.
      3. Comprehensive training for lab staff and support services till familiarity with the system
      on site.
      4. Comprehensive warranty for 5 years warranty of instruments and long shelf life of
      implants .


      8.Documentation
Sl                       Name                      Technical Specs           Bidders deviation if
                                                   Quoted by the bidder. any.
8.1 1. User/Technical / Maintenance manuals to be supplied in English.
     2. Certificate of calibration and inspection.
     3. list of equipments available for providing calibration and routine preventive maintenance
     support as manufacturer service/maintenance manual.
     4. List of important spare parts and accessories with their part number and costing.
     5. Log book with instructions for daily, weekly, monthly and quarterly maintenance
     checklist. The job description is clearly spelt out.
     6. Compliance report to be submitted in a tabulated and point wise manner clearly
     mentioning the page/para/number with authenticated catalogue/manual without which it
     will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
      Equipment Specifications for INTRAMEDULLARY NAILING SET
                              WITH FLEXIBLE REAMERS

UNSPSC Code:
ECRICode:
1 Description of Function
SI     Name                                                           Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

     1.1 These instruments are used in orthopedic surgery for reaming

of bone before fixation of long bones with nails like in thigh, leg

,armetc..

     2 Operational Requirements

 SI      Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted by if any
                                                                      bidder

     2,1 Following are the general guidelines and

configuration, quantity may vary with the user with the

ultimate aim of femoral intramedullary locked nailing;

reconstruction locked nailing; proximal femoral and distal

femoral locked nailing.

     2.2 Locking screw appropriated for the nail and site to be

provided.As specified

     2.3 Anatomic precurvature required as appropriate for

different nails as anterior bow, coronal angulation for

trochanteric entry nail for femur etc and same apply for

tibia.




Dr. Deepak Mittal                                            Dr. V.K. Sharma
     3 Technical Specifications

SI      Name                                                    Technical Bidders
                                                                Specs     Deviation
                                                                quoted    if any
                                                                by bidder
     3.1 One System should be capable of providing instrumentation
         for Femoral, Tibial & Humeral reconstruction and retrograde
         nailing preferably
     3.2 General Instruments
         1 ST-PIN 9X1/8IN STYLES 5 1
         2 SLAP HAMMER 1
         3 RULER RADIOLUCENT 1
         4 7MM DIA ANGLED FEMORAL AWL 1
         5 FLARED EXCHANGE TUBE 1
         6 SKIN PROTECTOR 1
         7 TIBIAL AWL 1
         8 MALLET 1
         9 THREADED DRIVER 1
         10 TROCHANTERIC REAMER 1
         11 WIRE GRIP T-HANDLE 1
         12 THREADED EXTRACTOR 1
         13 ANGLED TIP FEMORAL AWL 1
         Femoral Instruments
         1 9/16IN PIN WRENCH 1
         2 PROX TRAGETING GUIDE RADIOLUCENT 1
         3 FEM GUIDE BARREL LT 1
         4 FEM GUIDE BARREL RT 1
         5 OFFSET DRIVER ADAPTOR 1
         6 LOCKING BOLT 1
         7 FEM PIN/DRILL BUSHING 1
         8 FEM DRILL BUSHING 5.0MM 1




         9 FEMORAL DRILL BUSHING 1
         10 FEM SCREW BUSHING 8MM 1
         11 BUSHING AWL FEMORAL 1
         12 FEMORAL DRILL LARGE 5MM 1
         13 RECON SCREW COUNTER 1
         14 GUIDE PIN 3.2MM DIA X 304MM 1

         PROXIMAL RETROGRADE (RECONSTRUCTION) FEMORAL
         INSTRUMENT SET
         1 PROXIMAL BUSHING 3.2MM 1
         2 PROXIMAL BUSHING 4.5MM 1
         3 PROXIMAL BUSHING 8.0MM 1
         4 PROX DRILL W/STOP MED, 3.2MM 1
         5 PROX DRILL W/STOP, LRG 4.5MM 1
         6 DRILL, MEDIUM 3.2MM 1
         7 DRILL LARGE 4.5MM 1
         8 TROCAR MEDIUM 3.2MM 1
         9 TROCAR LARGE, 4.5MM 1
         10 SCREW DRIVER SHORT HEX, 3.5MM 1
         11 TARGELING SET of various sizes 1

Dr. Deepak Mittal                                       Dr. V.K. Sharma
       12 SCREW DEPTH GAUGE 1

       RETROGRADE FEMORAL INSTRUMENT SET
       1 SCREW DEPTH GAUGE 1
       2 DISTAL TARGETING GUIDE 1
       3 CORTICAL NUT COUNTERBORE 1
       4 CORTICAL NUT SCREW DRIVER 1
       5 FEM PIN/DRILL BUSHING 3.2MM 1
       6 FEM DRILL BUSHING 4.5MM 1
       7 STENMANN PIN 12 INCH 1
       8 LOCKING BOLT 1
       9 FEM SCREW BUSHING 8.0MM 1
       10 FEMORAL BUSHING AWL 1
       11 FEMORAL DRILL, 4.5MM 1
       12 T-HANDLE HEXHEAD SCREW DRIVER 3.5MM 1
       13 PIN WRENCH, 9/16 INCH 1
       14 RETROGRADE INSTRUMENT CASE 1

       Tibial Instruments
       1 9/16IN PIN WRENCH 1
       2 LOCKING BOLT 1
       3 PROX TRAGETING GUIDE RADIOLUCENT 1
       4 TIBIAL OBILQUE HOLE 1
       5 TIBIAL DRILL BUSHING 3.7 1
       6 TIBIAL SCREW BUSHING 1
       7 BUSHING AWL TIBIAL/HUM 1
       8 TIBIAL DRILL 1

       HUMERAL INSTRUMENT SET

       1 PIN WRENCH 1
       2 LOCKING BOLT 1
       3 PROX TRAGETING GUIDE preferably RADIOLUCENT 1
       4 OBLIQUE HOLE ADAPTOR 1
       5 HUMERAL PROX TARGETING GUIDE 1
       6 HUMERAL OBLIQUE HOLE ADAPTOR 1
       7 HUMERAL DRILL BUSHING MEDIUM 3.2MM 1
       8 HUMERAL SCREW BUSHING, 8.0MM 1
       9 HUMERAL BUSHING AWL 1
       10 HUMERAL DRILL, MEDIUM 3.2MM 1

       Distal Set Instruments
       1 SCREW DEPTH GUAGE 1
       2 DISTAL DRILL 3.2MM 1
       3 DISTAL DRILL 3.7MM 1
       4 DISTAL TROCAR 3.2MM 1
       5 DISTAL TROCAR 3.7MM 1
       6 T- HANDLE HEX HEAD SCREW DRIVER 1
       7 DISTAL SCREW DRIVER 3.5 1
       8 DISTAL INSERT 1
       9 SET SCREW 1
       10 HANDLE 1
       11 BUSHING INSERT 1
       12 DISTAL SCREW DEPTH G 1
       13 DISTAL TROCAR 5MM 1
       14 DISTAL DRILL 5MM 1




Dr. Deepak Mittal                                 Dr. V.K. Sharma
       1 8.0 MM FLEXIBLE REAMERS 1
       2 8.5 MM FLEXIBLE REAMERS 1
       3 9.0 MM FLEXIBLE REAMERS 1
       4 9.5 MM FLEXIBLE REAMERS 1
       5 10.0 MM FLEXIBLE REAMERS 1
       6 10.5 MM FLEXIBLE REAMERS 1
       7 11.0 MM FLEXIBLE REAMERS 1
       8 11.5 MM FLEXIBLE REAMERS 1
       9 12.0 MM FLEXIBLE REAMERS 1
       10 12.5 MM FLEXIBLE REAMERS 1
       11 13.0 MM FLEXIBLE REAMERS 1
       12 13.5 MM FLEXIBLE REAMERS 1
       13 14.0 MM FLEXIBLE REAMERS 1

       IMPLANTS configuration and quantity may vary with the user
       demand with the ultimate aim of stable , locked nailing.



       1 INTERLOCKING NAILS (S.S.)
       A Femoral Nails - Stainless Steel/ titanium
       Suitable for interlocking and reconstruction purposes
       With facility for dynamization.
       Nail should be Cannulated but designed for unreamed insertion if
       required.

         Proximal Drill Guide should be preferably radiolucent especially
       so in proximal and recon femaoral nailingt.
       Sizes:
       Dia 8mm to 14mm Length 24 cm to 46cm
       Recon Screws Dia 5.5mm 60mm to 130mm
       Low Profile Cortical Locking Screws Dia 4.2mm Length 20 to
       90mm
       B Tibial Nail - Stainless Steel /titanium
       with Proximal Transverse Holes so that fractures close to the
       joint space can be treated by inserting oblique screws that
       should not compromise the Tibial plateau posteriorly/ other
       stable locked configuration.
       Further distal hole to treat distal metaphyseal fractures.
       Proximal Dynamic slot to allow compression.
       Should be supplied if possible Sterile Packed
       Proximal Drill guide should be preferably radiolucent.
       Sizes:
       Dia 7mm to 12mm Length 28cm to 42cm
       C Retrograde Femoral Nails Stainless Steel/ltitanium
       With Dynamic slot
       Three closely located distal holes/other stable locking
       configuration for Supracondylar fracture


       Drill guide should be radiolucent.
       Sizes:
       Dia 10mm to 14mm Length 16cm to 42cm
       D Humeral Nail - Stainless Steel
       Suture holes at proximal end
       Two holes proximally
       distal locking.
       Proximal drill guide should be radiolucent.
Dr. Deepak Mittal                                         Dr. V.K. Sharma
4 System Configuration Accessories, spares and consumables

   Sl     Name                                                         Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   4.1    As specified

5 Environmental factors

   Sl     Name                                                         Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   5.1    The unit shall be capable of being stored continuously in
          ambient temperature of 0 -50 deg C and relative
          humidity of 15-90%

6 Power Supply

   Sl Name                                                             Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

         None

7 Standards, Safety and Training

   Sl     Name                                                         Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   7.1    1. Should be manufactured by company which has license
          of Drug Controller of India and conforms to standards laid
          down by the Drug Controller of India vide notification of
          Ministry of Health & family welfare
          2. Manufacturer should have ISO certification for quality
          standards.
          3. Comprehensive training for lab staff and support
          services till familiarity with the system on site.
          4. Comprehensive warranty for 5 years and long shelf life
          for implants .
          5. Material should be of high quality used for medical
          equipments.

8 Documentation

   Sl     Name                                                         Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   8.1    1. User/Technical/Maintenance manuals to be supplied in
          English.
          2. Certificate of calibration and inspection.
          3. List of Equipments available for providing calibration

Dr. Deepak Mittal                                            Dr. V.K. Sharma
        and routine Preventive Maintenance Support. as per
        manufacturer service/ maintenance manual.
        4. List of important spare parts and accessories with their
        part number and costing.
        5. Log book with instructions for daily, weekly, monthly
        and quarterly maintenance checklist. The job description
        of clearly spelt out.
        6. Compliance Report to be submitted in a tabulated and
        point wise manner clearly mentioning the page/para
        number with authenticated catalogue/manual,without
        which it will not be considered.




Dr. Deepak Mittal                                         Dr. V.K. Sharma
  Equipment Specifications for Instruments for removal of
                               damaged screw

  UNSPSC Code:
  ECRI Code:
  1 Description of Function

   Sl      Name                                                                              Technical B
                                                                                             Specs     D
                                                                                             quoted    if
                                                                                             by bidder

   1.1 T These instruments are used in orthopedic surgery for removal of damaged screws in
       P previously operated/fixed fractures.

  2 Operational Requirements

     Sl     Name                                        Technical Bidders
                                                        Specs     Deviation
                                                        quoted    if any
                                                        by bidder

     2.1   As specified; following are the general
           requirements and any configuration/system
           which achieves above purpose of removal of
           damaged screw efficiently and conveniently
           are acceptable.

  3 Technical Specifications

     Sl     Name                                        Technical Bidders
                                                        Specs     Deviation
                                                        quoted    if any
                                                        by bidder

     3.1   Sr. No. Description Qty. user demand
           specefic
           1. HOLLOW REAMER FOR 3.5/4.0 MM SCREWS
           1 no.
           2. HOLLOW REAMER FOR 4.5 MM SCREWS 1
           no.
           3. HOLLOW REAMER FOR 5.0/6. 0.6. 5.7.0 MM
           SCREWS 1 no.
           4. SPARE CENTERING PIN FOR HOLLOW
           REMDER FOR
           3.5/4.0 MM SCREWS 1 no.
           5. SPARE CENTERING PIN FOR HOLOW
           REAMER FOR 4.5MM
           SCREWS 1 no.
           6. SPARE CENTERINGPIN FOR HOLLOW
           REAMER FOR
           5.0.6.5/7.0MM SCREWS 1 no.
           7. SPARE REAMER TUBE FOR HOLLOW
           REAMDER FOR
           3.5/4.0 MM SCREWS 1
           8. SPARE REAMER TUBE FOR HOLLOW

Dr. Deepak Mittal                                       Dr. V.K. Sharma
        REAMER FOR
        4.5MM SCREWS 1
        9. SPARE REAMER TUBE FOR HOLLOW
        REAMER FOR
        5.0/6.5/7.0MM SCREWS 1
        10. EXTRACTION BOLT FOR 3.5/4.MM
        SCREWS 1
        11. EXTRACTION BOLT FOR 4.5 MM SCREWS
        1
        12. EXTRACTION BOLT FOR 5.0/6.5/7.0MM
        SCREWS 1
        13. EXTRACTION SCREW CONICAL FOR 2.7,
        3.5 & 4MM
        SCREWS 1
        14. EXTRACTION SCREW CONICAL FOR 4.5 &
        6.5 Mm
        SCREWS 1
        15. FORCEPS FOR SCREWS REMOVAL L205MM
        1
        16. GOUGE 10MM WIDE L 205MM 1
        17. ASSEMBLY PIN 1
        18. ASSEMBLY PIN 1
        19. CONNECTING END TO HOLLOW REAMER
        1.5, 2,
        2.7,3.5/4.0, 4.5 & 6.5/7.0 ( 1 each) 6
        20. CONNECTING END WITH CENTERING PIN
        1.5, 2,2.7,
        3.5,/4.0, 4.5 & 6.5/7.0 ( I each) 6
        21. HOLLOW REAMER FOR SCREWS 1.5, 2,
        2.7,3.5/4.0, 4.5
        & 6.5/7.0 ( 1 each) 6
        22. EXTRACTION BOLT 1.5, 2, 2.7, 3.5/4.0,
        4.5 & 6.5/7.0
        ( 1 each) 6
        23. CONICAL EXTRACTION SCREW 1.5/2,
        2.7/3.5/5.4.0,
        4.5/6.5/7.0 ( 1 each) 3
        24. HOLDING FORCEPS FOR SCREW 3
        25. HOLLOW GOUGE 10MM 3
        26. T HANDLE WITH QC 1
        27. T HANDLE WITH CHUCK & KEY 1
        28. HAMMER 1
        29. STERILIZATON & STORAGE CASE 1

4 System Configuration Accessories, spares and consumables

   Sl   Name                                        Technical Bidders
                                                    Specs     Deviation
                                                    quoted    if any
                                                    by bidder

  4.1   As specified

5 Environmental factors

   Sl   Name                                        Technical Bidders
                                                    Specs     Deviation
                                                    quoted    if any
                                                    by bidder

Dr. Deepak Mittal                                    Dr. V.K. Sharma
  5.1    The unit shall be capable of being stored
         continuously in ambient temperature of 0 -50
         deg C and relative humidity of 15-90%

6 Power Supply

   Sl Name                                                      Technical Bidders
                                                                Specs     Deviation
                                                                quoted    if any
                                                                by bidder

        None

7 Standards, Safety and Training

   Sl    Name                                                   Technical Bidders
                                                                Specs     Deviation
                                                                quoted    if any
                                                                by bidder

  7.1    1. Should be manufactured by company which
         has license of Drug Controller of India and
         conforms to standards laid down by the Drug
         Controller of India vide notification of Ministry of
         Health & family welfare..
         2 Comprehensive training for lab staff and
         support services till familiarity with the system
         on site.
         3 Comprehensive warranty for 5 years.

8 Documentation

   Sl    Name                                                   Technical Bidders
                                                                Specs     Deviation
                                                                quoted    if any
                                                                by bidder

  8.1    1. User/Technical/Maintenance manuals to be
         supplied in English.
         2. Certificate of calibration and inspection.
         3. List of Equipments available for providing
         calibration and routine Preventive
         Maintenance Support. as per manufacturer
         service/ maintenance manual.
         4. List of important spare parts and
         accessories with their part number and
         costing.
         5. Log book with instructions for daily, weekly,
         monthly and quarterly maintenance checklist.
         The job description of clerarlyt spelt out.
         6. Compliance Report to be submitted in a
         tabulated and point wise manner clearly
         mentioning the page/para number with
         authenticated catalogue/manual,without which
         it will not be considered.




Dr. Deepak Mittal                                                Dr. V.K. Sharma
Equipment Specifications for Implant for 3.5 mm D.C. D.C. Small
                                 Plates and Cortical Screw

UNSPSC Code:
ECRI Code:
1 Description of Function

   Sl    Name                                                                                            Technical
                                                                                                         Specs
                                                                                                         quoted
                                                                                                         by bidder

  1.1   An implant may be defined as an artificial organic material that can be surgically implanted
         into a person's body to replace damaged tissue. These small plates are used for fixation of
        small bones as of forearm,ankle etc.
  1.2   Locking plates are useful for performing mimimally invasive fracture fixation of bones and fixation of
        osteoporotic and periprosthetic/implant failure fractures

2 Operational Requirements

   Sl   Name                                                                   Technical Bidders
                                                                               Specs     Deviation
                                                                               quoted    if any
                                                                               by bidder

  2.1 The surgical instruments should not get rust after repetitive
      routine autoclaving & should last long. LC-DCP SHOULD
  2.2 HAVE HOLES OF DYNAMIC COMPRESSION TYPE WITH
      POSSIBLE 40 DEG. SCREW ANGULATION ; THE HOLES
      SHOULD BE UNDERCUT AND CONFORM TO APPROPRIATE
  2.3 STANDARDS.
      Thread of screw and locking thread mechanism should be of
      high quality. Locking screw mechanism should be angularly
      stable.

  2.4 3.5/4.0 Small fragment/ reconstruction LCP FOR tubular BONES
      should be of low profile. To support and facilitate minimal invasive
      surgery the LCP 3.5/4.0 Broad Curved may be designed with
  2.5 TAPERED nose on either side. TheRE SHOULD BE small holes in
      the nose allow for primary fixation with K-Wires.

  2.6 Locking PLATE SHOULD HAVE a COMBI HOLE OF LOCKING
      AND DYNAMIC COMPRESSION UNIT AND UNDERSURFACE
  2.7 SHOULD BE UNDERCUT.
      The LOCKING Hole should accept a 4MM locking threaded head
  2.8 SCREW OF AND DYNAMIC HOLE SHOULD ACCEPT 3.5MM
      SCREW.
      Screws in locking plate should be fully recessed in the plate.

        Locking screw should be self tapping.

        Distal tibial pilon limited compression plate should be anatomically
        precontoured. It should have a expanded distal end anatomically
        contoured /designed with varied arrangement of multiple locking
        only round holes 2.7mm.


Dr. Deepak Mittal                                                  Dr. V.K. Sharma
        Proximal humeral locking compression plate should be raquet
        shaped /expanded & anatomically contoured /designed at proximal
        end expanded; with varied arrangement of divergent/ convergent /
        parallel multiple locking only round holes 2.7mm to catch the
        humerus head . Proximal end should also have small holes at
        periphery to catch the different fracture fragment with threads. The
        shaft of these plates should have combi 3.5/4mm locking/dynamic
        compression holes and an elongated one at junction.




3 Technical Specifications

   Sl    Name                                                                  Technical Bidders
                                                                               Specs     Deviation
                                                                               quoted    if any
                                                                               by bidder

  3.1   1. QUANTITY AS REQUIRED BY USER
        Screws box for a) Dia 3.5mm cortical length 14mm to
        34mm with 2mm
        increment – 10 Each
        b) 2.7mm/ 3.5mm locking threaded head screws of 12 to
        44mm
        c) 4mm cancellous screws of 16 to 60mm.

        2. LC-D.C. Small Plates 4 holes - 10 Nos.

        3.D.C. Small Plates 5 holes - 10 Nos

        4.D.C. Small Plates 6 holes - 15 Nos.

        5. D.C. Small Plates 7 holes - 15 Nos.

        6. Locking compression 3.5 mm plates, reconstruction
        plates, distal tibial pilon and proximal humerus locking
        plates of 4 to 9 hole length.

4 System Configuration Accessories, spares and consumables

   Sl    Name                                                                  Technical Bidders
                                                                               Specs     Deviation
                                                                               quoted    if any
                                                                               by bidder

  4.1   As specified

5 Environmental factors

   Sl    Name                                                                  Technical Bidders
                                                                               Specs     Deviation
                                                                               quoted    if any
                                                                               by bidder

  5.1   The unit shall be capable of being stored continuously in
        ambient temperature of 0 -50 deg C and relative humidity
        of 15-90%


Dr. Deepak Mittal                                                 Dr. V.K. Sharma
6 Power Supply

   Sl Name                                                                  Technical Bidders
                                                                            Specs     Deviation
                                                                            quoted    if any
                                                                            by bidder

        None

7 Standards, Safety and Training

   Sl    Name                                                               Technical Bidders
                                                                            Specs     Deviation
                                                                            quoted    if any
                                                                            by bidder

  7.1    1. Should be manufactured by company which has license of
         Drug Controller of India and conforms to standards laid down
         by the Drug Controller of India vide notification of Ministry of
         Health & family welfare.
         2. Manufacturer should USE medical orthopaedic implant
         grade stainless steel//titanium.
         3. Comprehensive training for lab staff and support
         services till familiarity with the system on site.
         4. long shelf life.

8 Documentation

   Sl    Name                                                               Technical Bidders
                                                                            Specs     Deviation
                                                                            quoted    if any
                                                                            by bidder

  8.1    1. User/Technical/Maintenance manuals to be supplied in
         English.
         2. Certificate of calibration and inspection.
         3. List of Equipments available for providing calibration and
         routine Preventive Maintenance Support. as per
         manufacturer service/ maintenance manual.
         4. List of important spare parts and accessories with their
         part number and costing.
         5. Log book with instructions for daily, weekly, monthly
         and quarterly maintenance checklist. The job description of
         clerarly spelt out.
         6. Compliance Report to be submitted in a tabulated and
         point wise manner clearly mentioning the page/para
         number with authenticated catalogue/manual,without
         which it will not be considered.




Dr. Deepak Mittal                                             Dr. V.K. Sharma
  Equipment Specifications for HEMIREPLACEMENT SET-austine
                       moore prosthesis WITH IMPLANTS

UNSPSC Code:
ECRI Code:
1 Description of Function

   Sl   Name                                                                                                 Technic
                                                                                                             Specs
                                                                                                             quoted
                                                                                                             by bidd

  1.1   A implant may be defined as an artificial organic material that can be surgically implanted into a
        person's body to replace damaged tissue. Hemireplacement for hip is replacement of femoral
        head that is half of the hip joint in case of old patient with fracture neck of femur etc.

2 Operational Requirements

   Sl   Name                                                          Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

  2.1   As specified

3 Technical Specifications

   Sl   Name                                                          Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

  3.1   Instrument Set should consist of the following: quantity
        as per user demand

        Austin Moore Extractor - 1NO.
        Head Extractor - 1 NO.
        Head Gauge - 1 No.
        Aluminum impactor with nylon face - 1 No.
        Lanes small bone lever - 1 Pair
        Lanes serrated bone lever - 1 Pair
        Lanes large bone lever - 1 Pair
        Moore type rasp – Three all sizes.
        Lagenbeck retractor Small - 1 Pair
        Lagenbeck retractor Medium - 1 Pair
        Lagenbeck retractor Large - 1 Pair
        Murphy skid - 1 No.
        Cobra Retractor - 1 No.

        IMPLANTS AUSTIN MOORE- STERILE PACKED

        Size   37- five
        Size   39- Five
        Size   41,43,45,47,49, Ten Each
        Size   51, 53 – Five Each

4 System Configuration Accessories, spares and consumables

Dr. Deepak Mittal                                         Dr. V.K. Sharma
   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

  4.1    As specified

5 Environmental factors

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

  5.1    The unit shall be capable of being stored continuously in
         ambient temperature of 0 -50 deg C and relative humidity
         of 15-90%

6 Power Supply

   Sl Name                                                             Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

        None

7 Standards, Safety and Training

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

  7.1    1. Should be manufactured by company which has license
         of Drug Controller of India and conforms to standards laid
         down by the Drug Controller of India vide notification of
         Ministry of Health & family welfare.
         3. Comprehensive training for lab staff and support
         services till familiarity with lthe system on site.
         4. Comprehensive warranty for 5 years for instruments
         and long shelf life of implants.

8 Documentation

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

  8.1    1. User/Technical/Maintenance manuals to be supplied in
         English.
         2. Certificate of calibration and inspection.
         3. List of Equipments available for providing calibration
         and routine Preventive Maintenance Support. as per
         manufacturer service/ maintenance manual.
         4. List of important spare parts and accessories with their
         part number and costing.
         5. Log book with instructions for daily, weekly, monthly
         and quarterly maintenance checklist. The job description of

Dr. Deepak Mittal                                           Dr. V.K. Sharma
        clearly spelt out.
        6. Compliance Report to be submitted in a tabulated and
        point wise manner clearly mentioning the page/para
        number with authenticated catalogue/manual,without
        which it will not be considered.




Dr. Deepak Mittal                                       Dr. V.K. Sharma
   Equipment Specifications for ELECTRIC OPERATED DRILL &
                                            SAW

UNSPSC Code:
ECRI Code:
1 Description of Function

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   1.1   Drilling machines are used in a number of orthopaedic
         surgical procedures, for example, in making holes in
         bones for bone screws and in drilling out the medulla or
         marrow areas of bones.

2 Operational Requirements

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   2.1   Electric driven, autoclavable, versatile, forward & reverse
         mode with oscillating saw handpieces
   2.2   Following are the general guidelines – it should be able to
         perform orthopaedic procedures of drilling ,
         intramedullary reaming, sawing etc.

3 Technical Specifications

   Sl    Name                                                          Technical Bidders
                                                                       Specs     Deviation
                                                                       quoted    if any
                                                                       by bidder

   3.1   Driving unit:- Includes motor, sturdy stand with wheels
   3.2   Flexible shaft:- Minimum length-2 meters Autoclavable
         quick connection
   3.3   Hand Piece for Drill:- Cannulated Autoclavable Pistol
         Type
         - Speed-1200 to 1500 RPM.
         - Jacob chuck.
         - Quick coupling chuck (Synthesis type).
         - Hudsons chuck.
         - Chuck for K-wire.
         - Forward & Reverse options.
   3.4   Hand Piece for Reamers :- Cannulated Autoclavable Pistol
         Type
         - Speed-400 RPM minimum.
         - Chuck for cannulated Reamers.
         - Forward & Reverse options.
   3.5   Hand Piece for Saggital Saw:- Autoclavable Pistal Type

Dr. Deepak Mittal                                          Dr. V.K. Sharma
         - Easy Attachments of blades (without Instrument)
         - 2 Blades each of different size used routinely in
         Orthopaedic surgery (Total No, 10).
         - ACL Blades optional
         Quantity as per user demand


4 System Configuration Accessories, spares and consumables

   Sl    Name                                                        Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder

   4.1   As specified

5 Environmental factors

   Sl    Name                                                        Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder

   5.1   The unit shall be capable of being stored continuously in
         ambient temperature of 0 -50 deg C and relative humidity
         of 15-90%
   5.2   The unit shall be capable of operating in ambient
         temperature of 20-30 deg C and relative humidity of less
         than 70%

6 Power Supply

   Sl    Name                                                        Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder

   6.1   Power input to be 220-240VAC, 50Hz fitted with Indian
         plug
   6.2   Suitable UPS with maintenance free batteries for
         minimum one-hour back-up should be supplied with the
         system.
   6.3   Resettable overcurrent breaker/ any other protection
         device shall be fitted for protection

7 Standards, Safety and Training

   Sl    Name                                                        Technical Bidders
                                                                     Specs     Deviation
                                                                     quoted    if any
                                                                     by bidder

   7.1   Should be CE, FDA approved preferably.

         Manufactures/Supplier should have ISO certificate to
         Quality Standard.
   7.2   Should be compliant with IEC 61010-1:(or any

Dr. Deepak Mittal                                         Dr. V.K. Sharma
         international equivalent eg EN/UL 61010) covering safety
         requirements for electrical equipment for measurement
         control and laboratory use
   7.3   Should be approved by national government agency for
         medical electrical equipment –BEE, BIS approved product
   7.4   Comprehensive training for lab staff and support services
         till familiarity with the system.

8 Documentation

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   8.1   User/Technical/Maintenance manuals to be supplied in
         English.
   8.2   Certificate of calibration and inspection.
   8.3   List of important spare parts and accessories with their
         part number and costing.
   8.4   List of Equipments available for providing calibration and
         routine Preventive Maintenance Support. as per
         manufacturer documentation in service/technical manual.
   8.5   Compliance Report to be submitted in a tabulated and
         point wise manner clearly mentioning the page/para
         number of original catalogue/data sheet.Any point ,if not
         substantiated with authenticated catalogue/manual, will
         not be considered.
   8.6   Log book with instructions for daily, weekly, monthly and
         quarterly maintenance checklist. The job description of
         the hospital technician and company service engineer
         should be clearly spelt out.




Dr. Deepak Mittal                                          Dr. V.K. Sharma
    Equipment Specifications for -C-Arm Image Intensifier for
                                      orthopaedics

UNSPSC Code:
ECRI Code:
1 Description of Function

   Sl Name                                                            Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

        This equipment is used in orthopaedic fractures for
        imaging of bone pathology or fractures on display monitor
        during operation/ reduction of fractures.

2 Operational Requirements

   Sl Name                                                            Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

        UNIPLANAR OR BIPLANAR WHICH IS CONVINIENT , LEAST
        BIO/RADIATION-HAZARDOUS, EFFICIENT, FOR ALL TYPE
        OF ORTHOPAEDIC IMAGING OF BONES
        PEROPERATIVELY/OTHERWISE.

3 Technical Specifications

   Sl    Name                                                                                                   Tec
                                                                                                                Spe
                                                                                                                quo
                                                                                                                by b

  3.1    A Versatile, Compact C-arm unit should allow unobstructed positioning and enhanced case of operation in
         interventions. C-arm to have following or better mechanics for 9” Image Intensifier.

         a) Motorized Vertical travel : MINIMUM 500 mm or more
         b) Privotal rotation : =/- 12.5 deg. Or more
         c) Orbital rotation : = 90 deg. – minimum 30 deg. Or better
         d) Depth/Radius of C-arm : 660 mm or better
         e) SID : 950 mm or more
         f) Horizontal travel : 220 mm or better
         g) Free space between
         II & X-ray tube : MINIMUM 740 mm or more
         h) Rotation of C-arm : +/- 270 deg. Or more
         i) Total WIDTH of C-arm : MAXIMUM 800 mm or less
         K) option of biplanar imaging as per user demand if required.

         Remote controlled collimator with rotable parallel shutter,. The C-arm should have facility of locking the C
         movements with easy to turn handle on control unit. Rear wheels must be freely movable for easy to turn
         control unit. Rear wheels must be freely movable for easy positioning of the complete C-arm around the

         Image Intensifier

Dr. Deepak Mittal                                          Dr. V.K. Sharma
        Atleast triple field 9”/6”/4” input dia offering resolution ( Minimum 64lp/cm or better for 4” input ) & cont
        or better)


        TV Camera

        Ultra Compact CCD camera with high No of pixels ( > 450000) and video band width ( atleast 20 MHz of
        with 2 Nos. 17” 625 lines 100 Hz flicker free TV monitors with facility to rotate the image continuously.

        Direct Radiography

        Radiography should be possible on a cassette to be fitted in a holder for 10X 12 inches cassette. The unit
        complete with one such holder and 1 No. cassettes including high speed intensifying screens.

        X- ray generator

        High frequency ( 25- 40 KHz or more) 3.5 to 5.0 KW or even better X-ray generator with high capacity ro
        ray tube of dual foci of 0.3 and o.6 mm ( 200 KHU) or better.

        Fluoroscopy output : 40-120 KV in IKV steps
        mA output : MINIMUM Up to 8. mA or better
        Snapshort : MINIMUM Up to 12.0 mA or better

        Pul fluoroscopy rate selectable:- 1 image per second to 1 image per 5 second or better

        Automatic dose rate regulation with KV & mV control
        Time totalizer for fluoroscopy with facility to alarm after every 5 minutes of fluoroscopy

        Radiography output : 40-120 KV in 1 KV steps
        -mA range : Up to 250 mAs or better
        -mA max : Up to 90 mA or better


        Image Memory

        At least 1 (LIH) + minimum 20,000 frames dynamijkc digital memory on Hard Disk with 576 X 576 matri
        There should be facility to insert patient name through alphanumeric key board. They system must be up
        functions of performing REAL TIME digital subtraction angiograpny wiyth acquisition up to 6 frames/sec. o
        Road –mapping functions etc. at any later date for peripheral angiography.

        Essential Accessories

        The complete functional system must be quoted with DUAL CHANNEL Laser LIGHT SOURCE ON, X-RAY TU
        making a cross to reduce the X-ray dose, Built in DODE AREA PRODUCT meter for disply of X-ray dose, li
        aprons ( 6) required CVT and thermal imaging film printer with 12 film rolls.
        CD/DVD Writer.
        ABOVE ARE GENERAL APPROXIMATE GUIDELINES/PARAMETER VALUES ; ANY SYSTEM
        THAT DOES THIS IMAGING EFFECIENTLY CONVINIENTLYAND SAFELY IS WELCOME
         4 System Configuration Accessories, spares and consumables

            Sl      Name                                                            Technical Bidders
                                                                                    Specs     Deviation
                                                                                    quoted    if any
                                                                                    by bidder

           4.1      As specified



Dr. Deepak Mittal                                          Dr. V.K. Sharma
         5 Environmental factors

            Sl      Name                                                             Technical Bidders
                                                                                     Specs     Deviation
                                                                                     quoted    if any
                                                                                     by bidder

           5.1      The unit shall be capable of being stored continuously in
                    ambient temperature of 0 -50 deg C and relative humidity of
                    15-90%
           5.2      The unit shall be capable of operating in ambient temperature
                    of 20-30 deg C and relative humidity of less than 70%

         6 Power Supply

            Sl      Name                                                             Technical Bidders
                                                                                     Specs     Deviation
                                                                                     quoted    if any
                                                                                     by bidder

           6.1      Power input to be 220-240VAC, 50Hz fitted with Indian plug
           6.2      Suitable UPS with maintenance free batteries for minimum
                    one-hour back-up should be supplied with the system.
           6.3      Resettable overcurrent breaker/ANY OTHER PROTECTION
                    DEVICE shall be fitted for protection
                          6.4. Comprehensive warranty for 2 years with spares and
                          5 years AMC after warranty without spares.

                    6.5 Certified to be compliant with IEC 60601 for Medical
                    Electrical Equipments.



         7 Standards, Safety and Training

            Sl      Name                                                             Technical Bidders
                                                                                     Specs     Deviation
                                                                                     quoted    if any
                                                                                     by bidder

           7.1      Company/ supplier Should have a CE, FDA approved
                    certification preferably.
                    Manufactures/Supplier should have ISO certificate to Quality
                    Standard.
           7.2      Should be compliant with IEC 61010-1:(or any international
                    equivalent eg EN/UL 61010) covering safety requirements for
                    electrical equipment for measurement control and laboratory
                    use
           7.3      Should be BEE/ BIS and AERB approved product OR ANY
                    NATIONAL AGENCY APPROVED FOR MEDICAL ELECTRICAL
                    DEVICES
           7.4      Comprehensive training for lab staff and support services till
                    familiarity with the system.

         8 Documentation

Dr. Deepak Mittal                                             Dr. V.K. Sharma
            Sl      Name                                                            Technical Bidders
                                                                                    Specs     Deviation
                                                                                    quoted    if any
                                                                                    by bidder

           8.1      User/Technical/Maintenance manuals to be supplied in English.
           8.2      Certificate of calibration and inspection.
           8.3      List of important spare parts and accessories with their part
                    number and costing.
           8.4      List of Equipments available for providing calibration and
                    routine Preventive Maintenance Support. as per manufacturer
                    documentation in service/technical manual.
           8.5      Compliance Report to be submitted in a tabulated and point
                    wise manner clearly mentioning the page/para number of
                    original catalogue/data sheet.Any point ,if not substantiated
                    with authenticated catalogue/manual, will not be considered.
           8.6      Log book with instructions for daily, weekly, monthly and
                    quarterly maintenance checklist. The job description of the
                    hospital technician and company service engineer should be
                    clearly spelt out.




Dr. Deepak Mittal                                                Dr. V.K. Sharma
                    Equipment Specifications for BONE MILL

                    UNSPSC Code:
                    ECRI Code:
1 Description of Function

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   1.1   A bone mill for use in a surgical or otherwise sterile
         environment includes a particle reducer such as a
         pulverizing blade, a motor such as an electrical/pneumatic
         motor, and a coupling such as a rotary shaft for
         connecting the particle reducer to the motor.This bone is
         used for filling bone cavities/defects.

2 Operational Requirements

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   2.1   As specified

3 Technical Specifications

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   3.1   The unit housing and working part should be of stainless
         steel and should be entirely autoclavabLe.
   3.2   Should be able to produce fine (2mm to 4mm) or course
         (4 mm to 6 mm) bone particles.
   3.3   Should be with 2 interchangeable blades/MORE for fine
         and coarse cutting
   3.4   Should be able to operate pneumatically and should have
         10ft/APPROPRIATE air hose
   3.5   The unit should have pushing rod to enable bone in the
         bone chamber to be guided into the cutting chamber.

4 System Configuration Accessories, spares and consumables

   Sl    Name                                                         Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   4.1   All standard accessories desired for proper functioning of
         the machine.



Dr. Deepak Mittal                                          Dr. V.K. Sharma
5 Environmental factors

   Sl     Name                                                        Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   5.1    The unit shall be capable of being stored continuously in
          ambient temperature of 0 -50 deg C and relative humidity
          of 15-90%
   5.2    The unit shall be capable of operating in ambient
          temperature of 20-30 deg C and relative humidity of less
          than 70%

6 Power Supply

   Sl Name                                                            Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

         None

7 Standards, Safety and Training

   Sl     Name                                                        Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   7.1    Company/ supplier Should have a CE, FDA approved
          ceritfication
          Manufactures/Supplier should have ISO certificate to
          Quality Standard.
   7.2    Should be BEE/BIS / NATIONAL GOVERNMENT AGENCY
          FOR MEDICAL ELECTRICAL DEVICES approved product
   7.3    Comprehensive training for lab staff and support services
          till familiarity with the system.

8 Documentation

   Sl     Name                                                        Technical Bidders
                                                                      Specs     Deviation
                                                                      quoted    if any
                                                                      by bidder

   8.1    User/Technical/Maintenance manuals to be supplied in
          English.
   8.2    List of important spare parts and accessories with their
          part number and costing.
   8.3    Log book with instruction for daily , weekly, monthly and
          quarterly maintenance checklist.
          The job description of the hospital technician and
          company service engineer should be clearly spelt out




Dr. Deepak Mittal                                           Dr. V.K. Sharma

				
DOCUMENT INFO
Shared By:
Categories:
Tags:
Stats:
views:24
posted:10/27/2011
language:English
pages:69